Technical Developments |
1 From the Department of Biomedical Engineering, Technion, Israel Institute of Technology, Haifa 32000, Israel. Received September 15, 1997; revision requested November 24; final revision received November 20, 1998; accepted January 15, 1999. Supported in part by the Israel Science Foundation (grant Z-953/93), the Technion Vice Provost for Research Fund (J. Tal Equipment and Research Fund 130-304), the Irving and Adele Rosenberg Foundation, and the Israel Cancer Association (grants 960023-B and 972025-B). Address reprint requests to H.A. (e-mail: haim@biomed.technion.ac.il).
| Abstract |
|---|
|
|
|---|
Index terms: Breast, US, 00.12989 Computed tomography (CT), ultrasound (US) Ultrasound, experimental studies, 00.12989 Ultrasound computed tomography (US/CT), 00.12989
| Introduction |
|---|
|
|
|---|
Ultrasonic imaging, in comparison, is considered safe; is suitable for women in any age group; enables tissue characterization on the basis of a variety of parameters such as backscatter, attenuation, and speed of sound (36); depicts tissue morphology; and is a relatively cost-effective modality. Currently, pulse-echo imaging (B scan) is the most widely applied ultrasonic technique, but its ability to discriminate between solid masses is limited (7). Ultrasonic computed tomography (CT) has been suggested as an alternative imaging technique that provides better tissue characterization (8,9), but the large number of projections required to obtain a high-spatial-resolution image requires use of large (and costly) transducer arrays or results in a long acquisition time.
Spiral CT is a relatively new approach implemented in modern x-ray imaging (1012). With spiral CT, a continuous scan is obtained of a volume of interest, which provides the data for a three-dimensional (3D) reconstruction of the studied object. Spiral CT provides information for a much more detailed 3D reconstruction of the object and is more efficient than is conventional CT performed with the same scanning parameters. On the basis of its 3D nature, spiral CT can provide longitudinal and arbitrarily oriented cross sections in addition to the axial images commonly obtained with conventional CT.
The objective of this study was to examine the feasibility of implementing spiral CT in ultrasonic imaging as a potential method for 3D breast imaging.
| Materials and Methods |
|---|
|
|
|---|
Refraction index projections can be estimated by measuring the time of flight of the ultrasonic wave through the object by using the following line integral (along each ray) (8):
Attenuation coefficient projections can be estimated by measuring the amplitude of the through transmitted ultrasonic wave. In this case, the following line integral is used (8):
Spiral CT
Spiral CT provides volumetric information from the object. Unlike conventional CT, which scans the object section by section (Fig 1a), spiral CT scans the entire object along a spiral path (Fig 1b). Accordingly, if the helical pitch (defined as the axial displacement of the detectors per revolution) is properly selected, spiral CT contains information from all regions of the object. Consequently, small abnormal regions located between two consecutive planes, which may disappear in the 3D reconstruction obtained with conventional CT, are likely to be detected with spiral CT. Naturally, this also depends on other scanning parameters (ie, section thickness [which is fixed by the collimator], the number of sampled rays, the number of projections acquired, and the quality of the reconstruction algorithm used).
|
t) multiplied by the number of projections (NP) required to reconstruct the object. Given a total scanning time (T) as a constraint, the number of available projections is determined by NP = T/(
t). If the number of projections used to reconstruct a single section is N1, conventional CT can yield only NP/N1 sections, whereas spiral CT can yield NP - (2 · N1) sections. For example, if the total number of projections acquired is 640 and N1 = 64, conventional CT can produce only 10 sections whereas spiral CT can produce as many as 512 sections.
With ultrasonic CT, no hazardous radiation is involved. Therefore, total acquisition time is not limited by dose factors but merely by the patient's comfort. The time required to acquire a single ultrasonic projection (
t) is determined mainly by engineering factors (data storage and mechanical movements). The physical limitation here is almost negligible (ultrasonic travel time through 200-mm-thick tissue is about 0.13 msec). Thus, a projection of 128 rays, for example, will require about 17 msec plus the time required for the hardware to sample and store the data. (With modern hardware and array transducers, an acquisition rate of more than 10 projections per second can be achieved.)
Object reconstruction from spiral CT data can be achieved by using alternative approaches. In this study, linear interpolation of the projection data into a set of parallel planes (17,18) was used. The interpolation method is described in detail in the Appendix. Basically with this method, the spiral CT data set is first resampled into a set of many planar Radon transforms by using the interpolation algorithm. Then, by applying a conventional backprojection algorithm to each Radon transform, a set of parallel cross sections is generated. Consequently, the number of planes reconstructed is substantially larger than the number obtained by conventional CT with the same scanning parameters. This leads to a much more detailed 3D reconstruction of the object. However, the penalty for that procedure is an increase in the section sensitivity profile (ie, the effective section width is increased [1922]). It should be noted that the section sensitivity profile depends strongly on the helical pitch used (ie, a smaller pitch will provide a narrower effective section width and vice versa). Nevertheless, with a smaller pitch, a larger number of windings will be required to cover the object, which leads to an increase in the total scanning time.
Spiral Ultrasonic CT Experimental System
The experimental system used to implement spiral CT to ultrasonic waves comprises a water bath with a specially built computer-controlled mechanism that can produce spatial motion with 3 df [x, z,
] for a pair of 5-MHz focused transducers (Panametrics; Waltham, Mass). Section thickness is determined on the basis of the diameter of the acoustic beam generated by the transducers. We estimated it to be about 5 mm in this case. The system can scan a cylindric volume defined by the user (up to 20 cm in diameter and 30 cm high) located in the center of the water bath. The system can currently acquire a 15-cm-long projection with 600 rays in 10 seconds. The helical pitch, defined as the axial displacement of the ultrasonic transducers per revolution, can be set arbitrarily by the user (0300 mm per revolution).
A pulser receiver (model 5800; Panametrics) was used to transmit and receive the ultrasonic waves. A two-channel 20-MHz analog-to-digital converter (model R2000; Rapid Systems, Seattle, Wash) was used to digitally store the detected waves. For image reconstruction, the peak-to-peak amplitude of each wave was registered. The system was controlled with a 66-MHz computer (PC/486; Fox Computers, Taiwan). Codes for data acquisition, image reconstruction, and image display were written in our laboratory by using a C++ compiler (Borland International, Scotts Valley, Calif).
Objects Tested and Protocols Used
Three objects were used to evaluate spiral ultrasonic CT performance.
Plasticine spherical target.A 7-mm-diameter plasticine sphere (Cope 4516; PanArt, Jerusalem, Palestine) was used to simulate a single tissue target and allow comparison of conventional ultrasonic CT with spiral ultrasonic CT. The sphere was positioned within the scanner water tank. It was positioned in the water tank by using a very thin wire (about 0.2-mm diameter), which was too thin to be detected by the system (the corresponding ultrasonic wavelength in water is about 0.3 mm). Scout acoustic attenuation projections based on Equation (2) were acquired for general orientation. With use of these scout projections, the coordinates of the desired imaging windows (ie, minimum and maximum spatial coordinates of the scanned volume) were set manually.
Tomographic images of the sphere were initially obtained by using reconstructions of four conventional ultrasonic CT sections with a resolution of 128 x 128 pixels. Each pixel was 1 x 1 mm. A 10-mm distance between sections was used. (This distance was used to ensure the sphere was detected in only one section of the conventional ultrasonic CT scan.) Next, the same sphere was reconstructed with use of a spiral ultrasonic CT scan with four spiral windings and a 10-mm pitch. The corresponding cross sections were also reconstructed with a resolution of 128 x 128 pixels. In-plane resolution was again 1 x 1 mm.
On the ultrasonic attenuation tomograms, brighter gray levels indicated higher attenuation. Thus, because the ultrasonic attenuation coefficient of plasticine is higher than that of the background (ie, water), the sphere appeared as a bright spot on a relatively dark background. The sphere was detected within the tomographic reconstructions by two observers (H.A., D.S.) on the basis of consensus. The sphere was easily detected (Fig 2). Because we merely compared the two methods for detection, no tests of reproducibility were performed.
|
|
x
y
z) was 1.00 x 1.00 x 0.62 mm. Again the images were analyzed by consensus. The purpose of this study was merely to validate that the reconstructions produced with spiral ultrasonic CT matched the expected cross sections of the hard-boiled egg. Hence, no tests of reproducibility were performed.
Breast phantom.The breast biopsy phantom (model BB-1; ATS Laboratories, Bridgeport, Conn) mimics the appearance, size, and acoustic properties of the human breast. The phantom contains eight target structures embedded randomly throughout. The target structures are similar to abnormal tissue and are 510 mm in diameter. The phantom was imaged with spiral ultrasonic CT with 16 spiral windings and 5-mm pitch. A 128 x 128 x 64-voxel reconstruction of the phantom was obtained. Voxel size (
x
y
z) was 1.40 x 1.40 x 1.25 mm.
Abnormal tissuemimicking targets were characterized in the tomograms with relatively higher attenuation coefficients (ie, high gray level) and a dark corona (presumably a diffraction effect). The same two observers detected the targets by consensus. For validation, the phantom was also scanned manually by a third observer with a conventional ultrasonic scanner (Synergy-B; Diasonics, Israel) with a 3.5-MHz transducer in the B-scan mode.
Reproducibility of Data Acquisition
The reproducibility of the system in acquiring the data required for image reconstruction was also evaluated. The system was given a reset command and the transducers moved to their home position. Then the system was ordered to acquire a single acoustic projection of the breast phantom with coordinates arbitrarily set by the user. The acquired data were stored on the computer disk. The procedure (ie, a reset command followed by an acquisition command) was repeated several times. Finally, the correlation coefficients between 10 pairs of these acoustic projections (acquired from the same location of the breast phantom) were calculated by using a library function (MATLAB; Math Works, Natick, Mass). A high correlation coefficient indicates high reproducibility.
| Results |
|---|
|
|
|---|
Hard-boiled egg phantom.Axial, longitudinal, and oblique (35° relative to the axial plane) cross sections of the hard- boiled egg obtained from the reconstructed spiral ultrasonic CT data are depicted in Figure 3. As can be noted, the structure of the egg is clearly visible in all the cross-sectional images. However, the longitudinal cross section is blurred and inferior in quality relative to the axial cross section. This may be attributed to the large helical pitch chosen.
|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Spiral ultrasonic CT offers several advantages. The first stems from the inherent properties of the ultrasonic modality, which is considered hazardless and relatively cost-effective. This implies applicability to mass screening and potential use in clinics and hospitals.
The second advantage is derived from the information provided by the ultrasonic CT images. Although the commonly used pulse-echo techniques have been very useful in depicting breast cysts (1,2), their ability to differentiate solid masses is limited (2). On the other hand, analyses based on speed of sound (23,24) and acoustic attenuation have demonstrated clinical potential (eg, Scherzinger et al [25] achieved approximately 90% sensitivity and specificity in depicting breast cancer with ultrasonic CT).
In addition, spiral ultrasonic CT offers the main advantage provided by x-ray spiral CT, which is a volumetric coverage of the object with an efficient scan. Volumetric imaging also offers better diagnostic potential over conventional two-dimensional imaging. This property was demonstrated in our study for the spherical target and the egg and breast phantoms. As shown in Figure 2, spiral ultrasonic CT depicted the spherical target that conventional ultrasonic CT may have missed. Furthermore, the 3D reconstructions obtained for the egg and breast phantoms enabled the display of arbitrarily oriented cross sections, as depicted in Figures 3 and 4, respectively. In addition, by using a dynamic display, the diagnostic physician could virtually move through the object and locate the lesions. Finally, with the 3D display of the breast (Fig 5), the surgeon could obtain a more realistic orientation and possibly design a better surgical intervention.
However, spiral ultrasonic CT also has several disadvantages that should be noted. Owing to the facts that through transmission is required and ultrasonic waves can barely penetrate the bones and the lungs, spiral ultrasonic CT is limited almost exclusively to the breast. Moreover, axillary regions of the breast or regions very proximal to the ribs cannot be imaged with this technique. Hence, spiral ultrasonic CT must be coupled with the standard pulse-echo technique. In addition, the major disadvantage of x-ray spiral CT also applies here. As a result of the interpolation between data acquired at different heights (relative to the spiral axis), the effective section profile is substantially increased (twice the spiral pitch [1921]). Consequently, if the spiral pitch is large, notable blurring may occur along the longitudinal direction. This was demonstrated on the longitudinal cross-sectional images obtained in the egg (Fig 3). Optimization between the desired section sensitivity profile and the width of the pitch might give better results for a given imaged object.
In addition, several limitations of the present study should be pointed out. For computational simplicity, we ignored diffraction effects. This assumption is not entirely valid. A more accurate model should also account for diffraction effects. An algorithm that accounts for diffraction effects, such as the filtered backpropagation algorithm suggested by Devaney (15), might produce more accurate reconstructions. However, this algorithm is computationally much more intensive. Furthermore, Schreiman et al (26) have already demonstrated that clinically valuable images of the breast can be obtained with ultrasonic CT by means of a straight line model. As the objective of this study was merely to demonstrate feasibility, we chose to implement the simpler model.
A second simplification was to ignore the dependency of the attenuation coefficient on the frequency. A more accurate approach would be to apply spectral analysis to the signals. Alternatively, one can assume a linear relation between the attenuation coefficients and the frequency and can implement one of the methods reviewed by Kak and Slaney (14).
Finally, it should be pointed out that although attenuation tomograms were obtained in this study, with a slight modification (ie, measuring time of flight instead of wave amplitude), tomograms depicting speed of sound distribution may be produced. This acoustic parameter (ie, speed of sound) has been shown to provide valuable clinical information (23,24).
In conclusion, findings in this study demonstrate that spiral ultrasonic CT is feasible. This imaging method may find clinical application as a breast screening tool.
| Appendix |
|---|
|
|
|---|
In spiral CT, the position, z, along the spiral axis and the angle,
P, at which the projection is acquired, are linearly related by
To resample the spiral projections into a set of parallel Radon planes, the z value of each desired plane is first set. By using this relation, three consecutive projections (previous [prev], current [cur], and next) are selected that were acquired at different heights (z values) but correspond to the same angle. Then, the weighted average of these projections is calculated by
P are given by
P
2
are given by
The resultant vector P(
) is the new resampled projection at angle
. The procedure is repeated to cover all
values from 0 to
, and the filtered backprojection is applied to reconstruct the corresponding section (at the desired height z).
| Footnotes |
|---|
Author contributions: Guarantor of integrity of entire study, H.A.; study concepts and design, H.A.; definition of intellectual content, H.A.; literature research, D.S.; experimental studies, H.A., D.S.; data acquisition and analysis, D.S.; manuscript preparation and editing, H.A., D.S.; manuscript review, H.A.
| References |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||