(Radiology. 1999;213:23-37.)
© RSNA, 1999
Glandular Breast Dose for Monoenergetic and High-Energy X-ray Beams: Monte Carlo Assessment1
John M. Boone, PhD
1 From the Department of Radiology, University of California, Davis, Medical Center, 4701 X St, Radiology Research Laboratories, Sacramento, CA 95817. Received August 26, 1998; revision requested October 23; final revision received January 14, 1999; accepted March 26. Supported in part by grants from the United States Army Breast Cancer Research Program (DAMD17-94-J-4424 and DAMD17-98-1-8176), the California Breast Cancer Research Program (0192), and the National Cancer Institute (R21 CA 82077). Address reprint requests to the author (e-mail: jmboone@ucdavis.edu).
 |
Abstract
|
|---|
PURPOSE: To extend the utility of normalized glandular dose (DgN) calculations to higher x-ray energies (up to 120 keV) and to provide the tools for investigators to calculate DgN values for arbitrary mammographic and x-ray spectra.
MATERIALS AND METHODS: Validated Monte Carlo methods were used to assess DgN values. One million x-ray photons (1120 keV, in 1-keV increments) were input to a semicircular breast geometry of thicknesses from 2 to 12 cm and breast compositions from 0% to 100% glandular. DgN values for monoenergetic (1120 keV) x-ray beams, polyenergetic (40120 kV, tungsten anode) x-ray spectra, and polyenergetic mammographic spectra were computed. Skin thicknesses of 45 mm were used.
RESULTS: The calculated DgN values were in agreement within approximately 1%6% with previously published data, depending on breast composition. DgN tables were constructed for a variety of x-ray tube anode-filter combinations, including molybdenum anodemolybdenum filter, molybdenum anoderhodium filter, rhodium anoderhodium filter, tungsten anoderhodium filter, tungsten anodepalladium filter, and tungsten anodesilver filter. DgN values also were graphed for monoenergetic beams to 120 keV and for general diagnostic x-ray beams to 120 kV.
CONCLUSION: The tables and graphs may be useful for optimizing mammographic procedures. The higher energy data may be useful for investigations of the potential of dual-energy mammography or for calculation of dose in general diagnostic or computed tomographic procedures.
Index terms: Breast radiography, radiation dose, 00.47, 0.99 Breast radiography, technology, 00.12 Breast radiography, utilization, 00.99 Physics
 |
Introduction
|
|---|
The assessment of radiation dose to the breast during mammography has been of interest to many investigators (119). Over the years, the normalized glandular dose (DgN) has come to serve as the benchmark parameter, useful for calculating the glandular dose in clinical mammography. The DgN values are essentially the roentgen-to-rad conversion values, calculated for the "at-risk" glandular component of the breast. Recent efforts to calculate DgN tables for the mammography community have primarily been focused on clinically relevant spectra (4,5,7) with molybdenum anodemolybdenum filter (Mo-Mo), molybdenum anoderhodium filter (Mo-Rh), or rhodium anoderhodium filter (Rh-Rh) combinations in the 2035-kV range.
In this work, DgN tables were computed for much thicker breasts than for those in previous reports, with values reported here for breast thicknesses from 2 to 12 cm in 1-cm increments. While the typical compressed breast thickness in the United States is approximately 4.2 cm, there are many women with a compressed breast thickness that ranges to 12 cm or thicker. The tables provided in this article may be useful for these patients.
The motivation to extend DgN tables to encompass higher energy levels was based on an interest in dual-energy mammography, where the optimal high-energy beam is likely to be very high (>100 keV), well beyond current clinical mammographic x-ray beam energies. In addition, with the recent introduction of full-field digital mammography systems into the clinical environment, it is likely that slightly higher energy x-ray beams may become useful in some instances. This study was intended to extend the utility of DgN calculations to higher x-ray energies (up to 120 keV) and to provide the tools for investigators to calculate DgN values for arbitrary x-ray spectra, including monoenergetic x-ray beams (for example, produced by synchrotron sources [20], free-electron lasers [21], or other exotic x-ray sources). To this end, tables of DgN values have been provided for the x-ray tube anode-filter combinations of Mo-Mo, Mo-Rh, Rh-Rh, tungsten anoderhodium filter (W-Rh), tungsten anodepalladium filter (W-Pd), and tungsten anodesilver filter (W-Ag). Graphical data also are provided to demonstrate DgN values for monoenergetic and polyenergetic x-ray beams.
 |
MATERIALS AND METHODS
|
|---|
Details of the Monte Carlo Study
Monte Carlo procedures were used to compute the glandular dose to the breast. Although I have developed independent computer code for Monte Carlo studies (22,23), the TART97 Monte Carlo code was purchased from the Radiation Safety Information Computational Center, Oak Ridge National Laboratory (Oak Ridge, Tenn) for use in this study. The TART97 code was developed primarily at Lawrence Livermore National Laboratory (24) in Livermore, Calif, and is a thoroughly verified and mature Monte Carlo program. A full description of the TART97 Monte Carlo program is available in the literature (24); however, a brief description is appropriate here.
In a Monte Carlo simulation, each of the millions of photons traced in computations undergoes absorption or scattering, depending on the outcome of a random number generator. The probabilities of the multiple scattering calculations are weighted by the probability of that event at each x-ray energy studied. The TART97 Monte Carlo routine uses multiple scattering calculations, follows the history of all photons, and includes the photoelectric, Raleigh, and Compton scatter interaction mechanisms in the energy region reported. All photons were followed until they either left the volume of interest, were completely absorbed, or reached an arbitrarily small energy level (0.10 keV).
Monoenergetic x-ray photons at 1-keV intervals were input into a mathematic phantom in each of the simulation runs. Each photon run made use of 1 million photons at each monoenergetic energy level, and these data were used to construct monoenergetic DgN tables in a procedure described later in this article. The lowest energy simulated was 1 keV, and the highest was 120 keV. For the polyenergetic spectra reported, weighted sums of the monoenergetic DgN data were computed. The x-ray spectra used for this study were generated by using mathematic spectral models described previously (25,26). The x-ray attenuation coefficients for the filters also were reported previously (27).
Geometry and Composition Issues
The geometry simulated in this study is shown in Figure 1. Instead of a D-shaped semicircular breast shape, as others (4,5) have used, a cylindric breast shape was simulated (Fig 1a). The cone-shaped radiation field emitted from the source was collimated to irradiate half of the breast (a semicircle). The semicircular field geometry was particularly simple to simulate with the TART97 code and was efficient to run. The semicircle of breast tissue that was not in the radiation field was intended to simulate the presence of the torso of the patient (the chest wall). For their geometries, Wu et al (4) and Dance (5) assumed a D-shaped breast (no chest wall). The presence of tissue outside of the radiation field may have a minor influence in terms of backscatter, and this is of particular concern in this study due to the much higher x-ray energies studied here. While the nonirradiated semicircle is not the exact geometry of the chest wall, it was thought that the presence of some tissue behind the breast was slightly more representative of the geometry encountered in mammography, rather than no tissue outside of the radiation field.

View larger version (26K):
[in this window]
[in a new window]
|
Figure 1. Diagram shows the geometry used for the Monte Carlo simulations. R1 = radius of breast (including skin layer) in millimeters, R2 = radius of breast (excluding skin layer) in millimeters, SID = source-to-image distance, T = breast thickness, Tskin = skin thickness in millimeters.
|
|
Various breast compositions were studied, from 0% glandular100% adipose to 100% glandular0% adipose, by mass. DgN values were computed for the proportion of glandular tissue mass to total breast tissue mass. For concise reference henceforward, the breast composition is referred to in terms of the glandular percentage alone. The compositional data from Hammerstein et al (28) were used. X-ray coefficients for compound (multielemental) substances such as breast tissue were prorated on the basis of the weight fraction of the element; however, in substances where the density changed with the composition, the calculation of proportions is a little more complicated, and the techniques described in the following paragraphs were used.
For a tissue containing a weight fraction fg of glandular tissue (and, correspondingly, a weight fraction of 1 - fg for adipose tissue), it can be shown that the glandular volume fraction vg is given by
where
g is the density of 100% glandular tissue (
g = 1.04 g/cm3, from Hammerstein et al [28]) and
a is the density of 100% adipose tissue (
a = 0.93 g/cm3). Let the total volume be set to unit volume (1 cm3) for simplicity, such that vg + va = 1 cm3, and the compound density is
The mass m of each component in the unit volume is simply mg =
gvg and ma =
ava, where the "g" subscripts refer to glandular tissue and the "a" subscripts refer to adipose tissue. For completeness, the elemental compositions and densities for a variety of glandular fractions are given in the Appendix. By using the above procedure, the linear attenuation coefficients for 0%, 50%, and 100% glandular tissues were compared with those reported by Hammerstein et al (28). These data are shown in Figure 2.

View larger version (25K):
[in this window]
[in a new window]
|
Figure 2. Graph shows comparison of the linear attenuation coefficients computed in the present study with those reported by Hammerstein et al (28) for 0%, 50%, and 100% glandular tissue. The data from Hammerstein et al are shown as the symbols, and the coefficients used in the present study are shown as lines. Excellent agreement in terms of attenuation coefficients was observed over the energy range compared.
|
|
The breast tissue (glandular and adipose compound) is enclosed in a layer of skin, as illustrated in Figure 1. The skin thickness was varied in this study. For comparison with the results of Dance (5), a skin thickness of 5 mm was used. For comparison with the results of Wu et al (4), a skin thickness of 4 mm was used. For a single geometry and breast composition, the influence of skin thickness from 2 to 6 mm was studied. The density and elemental composition of skin, taken from Hammerstein et al (28), are reported in the Appendix.
Conversion of Monte Carlo Results to DgN Values
For a given breast composition, photon energy, and geometry, the output produced by the TART97 Monte Carlo package that was of interest in this study was the energy deposited (normalized per input photon) in the breast tissue compartment (Fig 1). The average energy to the breast tissue compartment, per incident x-ray photon to the breast, was normalized by means of the energy of the incident photons (all Monte Carlo runs used monoenergetic spectra), such that the fractional energy absorption, f(E), was calculated as follows:
The value of E is expressed in kiloelectron volts, and the x-ray photon spectra
(E) is normalized to the number of photons corresponding to 1 R (0.258 mC/kg) (for the entire spectrum). This type of normalization is typical (26) for investigators working with x-ray spectra. DgN values were calculated by using
where the value of f(E) was defined in Equation (3), the constant corrects for various unit conversions, G is defined later in Equation (6), area is the surface area at the top of the breast (in the entrance plane) exposed to x-rays, and mass is that of the purely glandular portion of the breast tissue. Let fg be the glandular fraction, by weight, of the breast tissue. For example, fg = 1.0 for a 100% glandular breast, and fg = 0.5 for a 50% glandular breast. For a semicircular breast tissue compartment of radius R2 (Fig 1), a breast density
, a compressed breast thickness T, and a skin layer thickness Tskin, the mass term in Equation 4 is given by
The G term in Equation (4) corrects the normalized dose calculation specifically to the glandular component of the breast tissue (DgN) in a heterogeneous tissue matrix. Values for breasts with a 0% glandular fraction are computed by extrapolation from DgN calculations of glandular fractions in the 2%5% range:
where the mass energy absorption coefficients (µen/
) are specified with an "a" subscript for adipose tissue and with a "g" subscript for glandular tissue. DgN and G were derived from first principles and by consulting previous publications (4,5). Units for DgN (Eq [4]) were derived as follows: DgN is expressed in millirad per roentgen, f(E) has no unit, E is expressed in kiloelectron volts per photon,
(E) is expressed as photons per square millimeter per roentgen, G has no unit, area is in square millimeters, and mass is in grams. The constant 1.6021 x 10-8 was derived as follows:
 |
RESULTS
|
|---|
Monte Carlo Results for f(E) Values
The Monte Carlo results for 0% and 100% glandular breasts are illustrated in Figure 3. If the breast tissue were not encapsulated in a layer of skin, these curves would be pseudoexponential, f(E) being unity at very low energy levels and decreasing almost exponentially with increasing energy level. However, x-ray photons of the lowest energy are unable to penetrate the relatively thin skin layer to contribute a fraction of their energy to the breast tissues; rather, the energy is deposited in the skin layer. The value of f(E) is, therefore, substantially dampened at the low energy levels because of the effect of skin filtration. As the glandular fraction increases, the f(E) value of the incident x-rays increases slightly, as would be expected due to the changing composition and increasing density.

View larger version (35K):
[in this window]
[in a new window]
|
Figure 3a. Graphs show f(E) as a function of incident x-ray energy for (a) 0% glandular tissue and (b) 100% glandular tissue. The curves are shown for 11 tissue thicknesses ranging from 2 to 12 cm in 1-cm increments. The bottom curve represents the data for the 2-cm breast thickness, and the top curve represents the data for the 12-cm-thick breast; intermediate curves are not marked for clarity, but are in order from 3 to 11 cm. Dotted lines = curves for odd-numbered tissue thicknesses (3, 5, 7, 9, and 11 cm), solid lines = curves for even-numbered thicknesses (2, 4, 6, 8, 10, and 12 cm).
|
|

View larger version (37K):
[in this window]
[in a new window]
|
Figure 3b. Graphs show f(E) as a function of incident x-ray energy for (a) 0% glandular tissue and (b) 100% glandular tissue. The curves are shown for 11 tissue thicknesses ranging from 2 to 12 cm in 1-cm increments. The bottom curve represents the data for the 2-cm breast thickness, and the top curve represents the data for the 12-cm-thick breast; intermediate curves are not marked for clarity, but are in order from 3 to 11 cm. Dotted lines = curves for odd-numbered tissue thicknesses (3, 5, 7, 9, and 11 cm), solid lines = curves for even-numbered thicknesses (2, 4, 6, 8, 10, and 12 cm).
|
|
It is suspected that subject contrast will be near the maximum when imaging at the kiloelectron voltage corresponding to the peak f(E) value for each breast thickness and composition. If one plots the data illustrated in Figure 3 differently, one can see the nonlinear effects of breast thickness on f(E) (Fig 4). In Figure 4, the absorption of 10-keV photons was constant across breast thicknesses, which suggested that the absorption dynamics of these low-energy photons occurred in the first 2 cm of tissue. At higher incident photon energy levels, the f(E) value increased with increasing breast thickness, as would be expected.
Figure 5 demonstrates yet a different perspective on these data: The f(E) value is shown as a function of glandular fraction for different incident photon energy levels in a 4-cm-thick breast. Monte Carlo runs were performed with several intermediate glandular fractions (0%, 20%, 40%, 50%, 60%, 80%, and 100%) for the data shown in Figure 5. The curves shown in Figure 4f(E) versus breast thicknessare nonlinear, whereas the curves in Figure 5f(E) versus glandular fractionare linear. This observation suggests that linear interpolation between glandular fraction data, for the f(E) values, was reasonable and preferable in comparison with interpolation between breast thicknesses. As a result, only 0% glandular (100% adipose) and 100% glandular results need to be reported because other proportions can easily be calculated.
Comparison with DgN Values in the Literature
The DgN values computed in the present study were compared with the results of Dance (5), as illustrated in Figure 6. Dance presented conversion factors in different units (mean glandular dose per incident air kerma, mGy · mGy-1), which were recomputed to the units of millirad per roentgen for this comparison. The skin thickness was set to 5 mm, and a 50% glandular breast was modeled, consistent with Dance's method. Four breast thicknesses, 2, 4, 6, and 8 cm, were studied for this comparison. When the different Monte Carlo routines, and particularly the different x-ray spectra, used to produce these data are considered, the qualitative agreement seen in Figure 6 between Dance's data and the data from this study is good. The 4-cm-thickness data were subjected to quantitative comparison. Because the half-value layers (HVLs) corresponding to both data sets differed on a point-by-point basis, direct comparisons between the data sets were not possible. Therefore, both data sets were computer fit by using commercially available software (TABLECURVE 3.0; Jandel Scientific, Corta Madera, Calif), with excellent precision (4-cm-thickness data from Dance: r2 = 0.9999; 4-cm-thickness data from present study: r2 = 0.9975), and comparison was then made between the computer-fit DgN values (between data sets) over the HVL range of 0.25-mm aluminum to 1.3-mm aluminum. The 4-cm-thickness data from the present study were found to differ from those of Dance, on average, by -1.12% (SD, 2.66). In terms of absolute DgN values, the difference was -3.34 mrad/R (SD, 10.6).
The data of Wu et al (4) are the most commonly used DgN values in the United States. The DgN values reported in the present study were calculated at exactly the same HVLs as those of Wu et al; thus, a direct comparison was possible. These comparative data are shown in Figure 7. DgN values calculated at three breast thicknesses (4, 6, and 8 cm) and at seven kilovoltages (2335 kV in increments of 2 kV) are shown for each breast composition. The DgN values observed in this study were consistent with, but slightly lower than, those of Wu et al for a 0% glandular breast but were seen to be in excellent agreement with the values of Wu et al for 50% and 100% glandular breast compositions. The slight qualitative differences (5%7%) in the data for a 0% glandular breast may be a consequence of different extrapolation techniques for a 0% glandular breast. The spectra computed for this study were hardened by adding an acrylic plastic sheet, such that the HVLs exactly matched those reported by Wu et al. While the kilovoltages and HVLs were identical, the DgN values calculated in the present study did make use of a different spectral model (25) than that (29) used by Wu et al, and this may explain the slight differences between the DgN values derived in the present study and those derived in the study by Wu et al.

View larger version (21K):
[in this window]
[in a new window]
|
Figure 7. Graph shows DgN values reported by Wu et al (4) along the y axis in comparison with DgN values from the present study (along the x axis). Individual points represent the data obtained with an energy range of 23-35 kV (in 2-kV increments). The DgN values reported here, averaged over all energy levels and breast thicknesses, differed from those of Wu et al by -6.2% for the 0% glandular breast, -1.9% for the 50% glandular breast, and +0.9% for the 100% glandular breast.
|
|
For the 0% glandular data, the mean differences (and SDs) averaged over the seven spectra (23, 25, 27, 29, 31, 33, and 35 kV) were -5.5% (SD, 1.3), -6.3% (SD, 1.7), and -6.8% (SD, 1.5) for the 4-, 6-, and 8-cm breast thicknesses, respectively. For the 50% glandular breast, the mean differences were -1.5% (SD, 1.4), -1.9% (SD, 1.5), and -2.3% (SD, 1.6) for the 4-, 6-, and 8-cm breast thicknesses, respectively. For the 100% glandular breast, the mean differences were 0.6% (SD, 1.5), 1.0% (SD, 1.4), and 1.1% (SD, 1.7) for the 4-, 6-, and 8-cm breast thicknesses, respectively.
Influence of Skin Thickness
The calculation of glandular dose for a patient has many uncertainties associated with it, and estimation of cancer risk on the basis of the glandular dose has even more uncertainties. The uncertainties in calculating glandular dose include uncertainties not only in the tabulated DgN values but also practical uncertainties in assessing the thickness of the breast, the breast composition, the precise milliampere-second value used, the differences between the actual mammographic geometry and that used in Monte Carlo simulations, and so on.
Figure 3 illustrates that the highest f(E) of incident x-ray photons occurred in the energy region from about 15 to 25 keV, where the f(E) curves peaked. Not coincidentally, this is the energy region where the vast majority of the x-ray photons in conventional x-ray spectra (eg, Mo-Mo combination at 26 kVp) exist. The fact that there was high absorption in the breast in this energy region also suggests that photons in this energy range are useful for the production of high-contrast images. As mentioned earlier, the left edges of the peaks of the f(E) curves seen in Figure 3 are a consequence of the absorption of incident x-rays by the skin layer. The steep slope of the left edges of the f(E) peaks suggests that a small difference in the assumption of skin thickness may have a large influence on the overall accuracy of the dose calculation.
To examine this in the case of the typical breast, Monte Carlo simulations were performed by using different skin thicknesses. Figure 8 illustrates the calculated DgN values for a 50% glandular, 4-cm-thick breast. Averaged over the different x-ray spectra (2335 kV), the change in DgN values (relative to a 4-mm skin thickness) that resulted from different skin thicknesses was 15.2% (SD, 2.1) for a 2-mm-thick skin layer, 7.1% (SD, 0.9) for a 3-mm-thick skin layer, -6.4% (SD, 0.7) for a 5-mm-thick skin layer, and -11.8% (SD, 1.3) for a 6-mm-thick skin layer.

View larger version (32K):
[in this window]
[in a new window]
|
Figure 8. Graph illustrates the influence of skin thickness on the DgN value. In comparison with the 4-mm skin thickness data from the study by Wu et al (4) (), the DgN values increased, on average, by 7% for a 3-mm skin thickness and by 15% for a 2-mm skin thickness. The DgN values decreased by 6% for a 5-mm skin thickness and by 12% for a 6-mm skin thickness. These simulations used a Mo-Mo spectrum.
|
|
The observation that skin thickness has a large influence on the DgN value is not surprising given the shape of the absorption curves (Fig 3). The purpose of presenting these data is to demonstrate that, among the uncertainties involved in dose calculations, it is likely that the slight differences (<6%) in tabulated DgN values produced by different investigators are small, as compared with the large errors that can occur in making the wrong assumptions or generalizations about an individual patient's breast characteristics.
Monoenergetic Beams
Figure 9 illustrates the monoenergetic DgN values expressed in millirad per 106 photons per energy interval. Although the general shapes of these curves are similar to those of the f(E) curves in Figure 3, the influence of tissue thickness is inverted. DgN values for the same x-ray energy and breast composition increased with decreasing breast thickness, because there is less self shielding in the thinner breast.

View larger version (37K):
[in this window]
[in a new window]
|
Figure 9a. Graphs show DgN values normalized per entrant photon instead of per roentgen (as is more typical) for (a) 0% glandular breasts and (b) 100% glandular breasts. Curves are shown for breast thicknesses ranging from 2 to 12 cm in 1-cm increments. The DgN values shown here are higher for thin breasts and lower for thicker breasts, which is the reverse of the trend seen for f(E) in This graph illustrates that, on a per photon basis, photons in the energy region between approximately 12 and 30 keV contribute the most to the DgN value. This is the energy region where most conventional mammographic x-ray spectra are centered.
|
|

View larger version (37K):
[in this window]
[in a new window]
|
Figure 9b. Graphs show DgN values normalized per entrant photon instead of per roentgen (as is more typical) for (a) 0% glandular breasts and (b) 100% glandular breasts. Curves are shown for breast thicknesses ranging from 2 to 12 cm in 1-cm increments. The DgN values shown here are higher for thin breasts and lower for thicker breasts, which is the reverse of the trend seen for f(E) in This graph illustrates that, on a per photon basis, photons in the energy region between approximately 12 and 30 keV contribute the most to the DgN value. This is the energy region where most conventional mammographic x-ray spectra are centered.
|
|
Although the DgN values are high for thinner breasts, the entrance exposure during a mammogram is markedly lower in thin breasts; therefore, thin breasts typically receive substantially lower glandular doses than do thicker breasts. An alternate label for the y axis in Figure 9 would be DgN x 10-6 per photon, or DgN(E). Therefore, by multiplying the data in Figure 9 by 10-6 and then integrating the product of an incident x-ray spectrum (
[E]) and the appropriate curve in Figure 9 (DgN[E]), the DgN value for an arbitrary spectrum can be calculated as shown below:
where it is understood that the total number of photons in
(E) in Equation (8) is normalized to 1 R (0.258 mC/kg). The curves shown in Figure 9 coupled with the formula given in Equation (8) are most useful when one is dealing with an arbitrary x-ray spectrum, which is typically computed in units of photons per square millimeter per energy interval.
Photons at different energy levels contribute differently to exposure in air, owing to the energy dependence of the mass energy attenuation coefficient of air. As a consequence, DgN values expressed in the traditional units of millirad per roentgen (vs x-ray energy) (Fig 10) have a different shape than those of the DgN per photon curves in Figure 9. Figure 10 illustrates the millirad per roentgen DgN values for 0% (Fig 10a) and 100% (Fig 10b) glandular tissue. Figure 10 is directly useful if one is interested in the DgN value for a monoenergetic beam of 1-R (0.258-mC/kg) incident exposure to the breast.

View larger version (37K):
[in this window]
[in a new window]
|
Figure 10a. Graphs show DgN values for monoenergetic x-ray energies for (a) 0% glandular breasts and (b) 100% glandular breasts. Breast thickness ranged from 2 to 12 cm in 2-cm increments. The DgN values shown in these graphs are plotted in the conventional unit of millirad per roentgen, as opposed to millirad per photon as in The curves in this graph can be read directly when assessing DgN for monoenergetic beams.
|
|

View larger version (36K):
[in this window]
[in a new window]
|
Figure 10b. Graphs show DgN values for monoenergetic x-ray energies for (a) 0% glandular breasts and (b) 100% glandular breasts. Breast thickness ranged from 2 to 12 cm in 2-cm increments. The DgN values shown in these graphs are plotted in the conventional unit of millirad per roentgen, as opposed to millirad per photon as in The curves in this graph can be read directly when assessing DgN for monoenergetic beams.
|
|
Mammographic Spectra
Tables 112 show the DgN values for several conventional and unconventional mammographic spectra. Tables 1, 3, 5, 7, 9, and 11 give DgN values for the 0% glandular breast; Tables 2, 4, 6, 8, 10, and 12 give these values for the 100% glandular breast. For other glandular fractions, DgN values may be linearly interpolated from the 0% and 100% glandular tables. The data in Tables 112 were computed from simulations in which a 4-mm skin thickness was used. X-ray spectra were computed by using previously reported techniques (25). Tables 16 present DgN values for conventional mammographic spectra, including Mo-Mo, Mo-Rh, and Rh-Rh combinations. Tables 7-12 present DgN values for unconventional mammographic spectra that are being considered as substitutes for use in women with thicker breasts and for digital mammography systems; these include W-Rh, W-Pd, and W-Ag combinations. Because of the strong L-characteristic x-ray emission of the tungsten anode around 12 keV, a 50-µm thickness of filter material is needed to effectively eliminate these x-rays from the entrance beam.
Tables 112 allow interpolation across energy level (by using the HVL), thickness, and glandular fraction of the breast composition. For example, consider the case of a 4.2-cm-thick breast composed of 30% glandular and 70% adipose tissue imaged at 26 kV with a Mo-Rh x-ray spectrum and an HVL of 0.37 mm aluminum: From Table 3 (0% glandular, Mo-Rh), the DgN value at 4.2 cm and an HVL of 0.37 is interpolated from the four data points corresponding to 4- and 5-cm-thick breasts at HVLs of 0.363 mm aluminum (at 26 kV) and 0.375 mm aluminum (at 27 kV). In this case, DgN = [236(0.8) + 195(0.2)] x (0.417) + [244(0.8) + 202(0.2)] x (0.583) = 232.35 mrad/R. The values in parentheses are the interpolation weights, and the values in square brackets are the thickness-interpolation values. The corresponding calculation was used to compute the DgN value for a 100% glandular breast by using the same interpolation coefficients and the DgN values in Table 4, which yielded DgN = [162(0.8) + 128(0.2)] x (0.417) + [168(0.8) + 133(0.2)] x (0.583) = 158.43 mrad/R. By interpolating these two values to the 30% glandular fraction of the example, the result is DgN-30 = DgN-0(0.7) + DgN-100(0.3) = 232.35(0.7) + 158.43(0.3) = 210 mrad/R, where DgN-30, DgN-0, and DgN-100 are the DgN values for 30%, 0%, and 100% glandular breasts, respectively.
High-Energy Polyenergetic Beams
Figure 11 illustrates the DgN values for polyenergetic x-ray beams in the general diagnostic energy region. Figure 11a shows results for the 0% glandular breast, and Figure 11b shows results for the 100% glandular breast. The x-ray spectra used for these calculations were generated by using a spectral model developed by the author (26). General radiographic (tungsten anode) x-ray spectra were computed from 40 to 120 kV, with the assumption of a 5% kilovoltage ripple (approximating an inverter generator), and with 2.5 mm of added aluminum filtration. The HVL ranged from 1.6 mm aluminum at 40 kV to 5.0 mm aluminum at 120 kV, with an approximately linear relationship (r2 = 0.998) between HVL and kilovolt level, where kV = 23.318 x HVL - 0.237. This relationship can be used to convert HVL values to the kilovolt values in Figure 11.

View larger version (35K):
[in this window]
[in a new window]
|
Figure 11a. Graphs show DgN values for conventional polyenergetic x-ray beams in which a tungsten anode and 2.5 mm of added aluminum filtration are used. DgN values are shown for (a) 0% glandular breasts and (b) 100% glandular breasts.
|
|

View larger version (37K):
[in this window]
[in a new window]
|
Figure 11b. Graphs show DgN values for conventional polyenergetic x-ray beams in which a tungsten anode and 2.5 mm of added aluminum filtration are used. DgN values are shown for (a) 0% glandular breasts and (b) 100% glandular breasts.
|
|
The DgN values for these high-energy x-ray beams may be useful for calculating glandular breast dose in some general diagnostic radiographic studies (lateral views) or in computed tomographic studies, if certain assumptions are made. One of the higher energy beams may be useful as the high-energy component of a dual-energy mammography system.
 |
DISCUSSION
|
|---|
In this study, Monte Carlo techniques were used to calculate DgN values. As validation of the procedures used here, DgN values for conventional mammographic spectra were compared with the results of the seminal work of Dance (5) and Wu et al (4). The comparison with Dance's results showed agreement within about 1%, and comparisons with the data of Wu et al showed agreement within a range of about 1%6%, depending upon breast glandularity. The influence of skin thickness was evaluated (Fig 8), and it was seen that a difference of 1 mm in skin thickness (eg, 3-mm instead of 4-mm skin thickness) had an influence of approximately 7% on the DgN values and that a difference of 2 mm (eg, 2-mm instead of 4-mm skin thickness) had an influence of 15%. As a consequence, the difference between the DgN values of Wu et al and those in the present study is smaller than typical differences in skin thickness.
Once the results of the present study had been verified against existing results for conventional x-ray spectra, the methods were used to extend DgN Monte Carlo calculations to 120 keV. A series of tables for possible mammographic spectral candidates has been provided to allow direct calculations of breast dose. Monoenergetic results also were computed and are presented in Figure 10. These data may be useful for computing the DgN values for arbitrary x-ray spectra, including those that may be useful for dual-energy mammography in the high-energy region.
Mammography as a modality continues to mature, and, at present, there are several digital mammography systems nearing the marketplace. Digital images allow the ability to retrospectively manipulate the displayed contrast. While it is impossible to recover subject contrast that is not recorded by the detector, it is thought that the ability to enhance displayed contrast retrospectively will be useful in improving image contrast in the clinical setting. If this assumption proves to be true after experience, it is likely that some compromises in subject contrast may be appropriate under some circumstances. For example, in women with larger breasts, where dose levels are much higher, a shift to harder x-ray spectra may be effective, and this is one reason why tungsten anodes with higher-atomic-number filtration are under investigation (and why the relevant DgN values are reported here). At least one design under commercial investigation involves a scanning slot beam of x rays; such a design places high heat-loading demands on the x-ray tube. Tungsten is a remarkable anode material because of its high melting point, and this is another reason why tungsten may become more common in some digital mammography systems. To be fully evaluated, new spectra with unconventional anode and filter materials will be studied for their influence on both image quality and patient dose. The DgN values reported here may be useful toward that end.
It is likely that alternate spectra will continue to be studied to determine whether further optimization of the mammographic examination can be achieved, given various new technologic developments. Furthermore, there is a small group of women who have a compressed breast thickness exceeding 8 cm; in these cases, DgN tables were not available, and, for such cases, x-ray spectra have not been optimized.
This study was intended to provide clinical medical physicists, as well as researchers, with the tools needed to calculate glandular breast dose for any arbitrary x-ray spectra in a simple but accurate manner. Efforts to computer fit these curves with adequate precision proved to be unsuccessful; therefore, the raw data in Figures 9 and 10 and in Tables 112 will be made available to all interested parties via e-mail request.
 |
Appendix 1
|
|---|
The elemental compositions and densities of breasts with different glandular proportions and of skin are given in Table A1.
 |
Acknowledgments
|
|---|
Dermott Cullen, PhD, of the Lawrence Livermore National Laboratory (Livermore, Calif) was most helpful in discussions concerning the use of the TART97 code, and his contribution to this study is gratefully acknowledged.
 |
Footnotes
|
|---|
Abbreviations: DgN = normalized glandular dose
HVL = half-value layer
f(E) = fractional energy absorption
Mo-Mo = molybdenum anodemolybdenum filter
Mo-Rh = molybdenum anoderhodium filter
Rh-Rh = rhodium anoderhodium filter
W-Ag = tungsten anodesilver filter
W-Pd = tungsten anodepalladium filter
W-Rh = tungsten anoderhodium filter
See also the editorial by Kimme-Smith (pp 710
) in this issue.
Author contribution: Guarantor of integrity of entire study, J.M.B
 |
References
|
|---|
-
Kulkarni RN, Supe SJ. Radiation dose to the breast during mammography: a comprehensive, realistic Monte Carlo calculation. Phys Med Biol 1984; 29:1257-1264.[Medline]
-
Doi K, Chan HP. Evaluation of absorbed dose in mammography: Monte Carlo simulation studies. Radiology 1980; 135:199-208.[Abstract/Free Full Text]
-
Dance DR. The Monte Carlo calculation of integral radiation dose in xeromammography. Phys Med Biol 1980; 25:25-37.[Medline]
-
Wu X, Barnes GT, Tucker DM. Spectral dependence of glandular tissue dose in screen-film mammography. Radiology 1991; 179:143-148.[Abstract/Free Full Text]
-
Dance DR. Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose. Phys Med Biol 1990; 35:1211-1219.[Medline]
-
Gingold EL, Wu X, Barnes GT. Contrast and dose with Mo-Mo, Mo-Rh, and Rh-Rh target-filter combinations in mammography. Radiology 1995; 195:639-644.[Abstract/Free Full Text]
-
Wu X, Gingold EL, Barnes GT, Tucker DM. Normalized average glandular dose in molybdenum target-rhodium filter and rhodium target-rhodium filter mammography. Radiology 1994; 193:83-89.[Abstract/Free Full Text]
-
Dance DR, Persliden J, Carlsson GA. Calculation of dose and contrast for two mammographic grids. Phys Med Biol 1992; 37:235-248.[Medline]
-
Ng KH, Aus RJ, DeWerd LA, Vetter JR. Entrance skin exposure and mean glandular dose: effect of scatter and field gradient at mammography. Radiology 1997; 205:395-398.[Abstract/Free Full Text]
-
Arfelli F, Bonvicini V, Bravin A, et al. A linear array silicon pixel detector: images of a mammographic test object and evaluation of delivered doses. Phys Med Biol 1997; 42:1565-1573.[Medline]
-
Thilander-Klang AC, Ackerholm PH, Berlin IC, et al. Influence of anode-filter combinations on image quality and radiation dose in 965 women undergoing mammography. Radiology 1997; 203:348-354.[Abstract/Free Full Text]
-
Calicchia A, Gambaccini M, Indovina PL, Mazzei F, Pugliani L. Niobium/molybdenum K-edge filtration in mammography: contrast and dose evaluation. Phys Med Biol 1996; 41:1717-1726.[Medline]
-
Thiele DL, Craig AR. Mean glandular dose and the standard breast. Australas Phys Eng Sci Med 1996; 19:94-96.[Medline]
-
Gentry JR, DeWerd LA. TLD measurements of in vivo mammographic exposures and the calculated mean glandular dose across the United States. Med Phys 1996; 23:899-903.[Medline]
-
Monticciolo DL, Sprawls P, Kruse BD, Peterson JE. Optimization of radiation dose and image quality in mammography: a clinical evaluation of rhodium versus molybdenum. South Med J 1996; 89:391-394.[Medline]
-
Lee KH, Kandaiya S. Estimation of breast radiation dose in a mammographic system. Appl Radiat Isot 1996; 47:361-363.[Medline]
-
Faulkner K, Law J, Robson KJ. Assessment of mean glandular dose in mammography. Br J Radiol 1995; 68:877-881.[Abstract]
-
Helvie MA, Chan HP, Adler DD, Boyd PG. Breast thickness in routine mammograms: effect on image quality and radiation dose. AJR 1994; 163:1371-1374.[Abstract/Free Full Text]
-
Boone JM, Shaber GS, Tecotzky M. Dual-energy mammography: a detector analysis. Med Phys 1990; 17:665-675.[Medline]
-
Burattini E, Cossu E, Di Maggio C, et al. Mammography with synchrotron radiation. Radiology 1995; 195:239-244.[Abstract/Free Full Text]
-
Tompkins PA, Abreu CC, Carroll FE, Xiao QF, MacDonald CA. Use of capillary optics as a beam intensifier for a Compton x-ray source. Med Phys 1994; 21:1777-1784.[Medline]
-
Boone JM, Seibert JA. Monte Carlo simulation of the scattered radiation distribution in diagnostic radiology. Med Phys 1988; 15:713-720.[Medline]
-
Boone JM, Seibert JA. Scattered energy deposition under shielding. Invest Radiol 1988; 23:627-631.[Medline]
-
Cullen DE. TART97 a coupled neutron-photon 3-D, combinatorial geometry Monte Carlo transport code Livermore, Calif: Lawrence Livermore National Laboratory, 1997.
-
Boone JM, Fewell TR, Jennings RJ. Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography. Med Phys 1997; 24:1863-1874.[Medline]
-
Boone JM, Seibert JA. An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV. Med Phys 1997; 24:1661-1670.[Medline]
-
Boone JM, Chavez AE. Comparison of x-ray cross sections for diagnostic and therapeutic medical physics. Med Phys 1996; 23:1997-2005.[Medline]
-
Hammerstein GR, Miller DW, White DR, Masterson ME, Woodard HQ, Laughlin JS. Absorbed radiation dose in mammography. Radiology 1979; 130:485-491.[Abstract]
-
Tucker DM, Barnes GT, Wu XZ. Molybdenum target x-ray spectra: a semiempirical model. Med Phys 1991; 18:402-407.[Medline]
Related Article
-
New Digital Mammography Systems May Require Different X-ray Spectra and, Therefore, More General Normalized Glandular Dose Values
- Carolyn Kimme-Smith
Radiology 1999 213: 7-10.
[Full Text]
[PDF]
This article has been cited by other articles:

|
 |

|
 |
 
J. Keyrilainen, M. Fernandez, M.-L. Karjalainen-Lindsberg, P. Virkkunen, M. Leidenius, K. von Smitten, P. Sipila, S. Fiedler, H. Suhonen, P. Suortti, et al.
Toward High-Contrast Breast CT at Low Radiation Dose
Radiology,
October 1, 2008;
249(1):
321 - 327.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. Sechopoulos, S. Vedantham, S. Suryanarayanan, C. J. D'Orsi, and A. Karellas
Monte Carlo and Phantom Study of the Radiation Dose to the Body from Dedicated CT of the Breast
Radiology,
April 1, 2008;
247(1):
98 - 105.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. J. Martin
RADIATION DOSIMETRY FOR DIAGNOSTIC MEDICAL EXPOSURES
Radiat Prot Dosimetry,
March 28, 2008;
(2008)
ncm495v1.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. Sechopoulos, S. Suryanarayanan, S. Vedantham, C. J. D'Orsi, and A. Karellas
Radiation Dose to Organs and Tissues from Mammography: Monte Carlo and Phantom Study
Radiology,
December 4, 2007;
(2007)
2462070256.
[Abstract]
[Full Text]
|
 |
|