1 From the Dept of Medical Imaging, Mount Sinai Hospital and the University Health Network, University of Toronto, 600 University Ave, Toronto, Ontario, Canada M5G 1X5 (L.M.W., J.K.K., M.M., N.M.); Dept of Orthopedic Surgery, Mount Sinai Hospital, Toronto (C.R.H., A.E.G.); and Dept of Radiology, Thomas Jefferson University Hospital, Philadelphia, Pa (M.E.S., W.B.M.). From the 1998 RSNA scientific assembly. Received Feb 12, 1999; revision requested Mar 30; final revision received Jul 28; accepted Aug 30. Supported in part by RSNA Research and Education Foundation and a grant from Berlex, Canada. Address reprint requests to L.M.W. (e-mail: lwhite@mtsinai.on.ca).
| Abstract |
|---|
|
|
|---|
MATERIALS AND METHODS: An initial phantom study was performed with cobalt-chrome and titanium prostheses to establish the imaging parameters for a subsequent clinical study. In the clinical study, coronal and transverse MR imaging of 14 THA prostheses was performed before and after intravenous contrast material administration in 12 patients who were being considered for revision arthroplasty. The images were reviewed for evidence of juxtaarticular or periprosthetic abnormalities, patterns of contrast enhancement, and quality of periprosthetic tissue depiction.
RESULTS: Phantom study results showed improved periprosthetic tissue depiction with use of thin sections, increased frequency-encoding gradient strength, and fast spin-echo sequences. The clinical study results demonstrated periprosthetic abnormalities in 11 cases: mechanical loosening in two cases (including one case with an associated periprosthetic fracture); granulomatosis, eight; and infection, one. In 100% of cases, tissue depiction around the femoral component was judged to be of "diagnostic quality." Tissue depiction around the acetabular component was of diagnostic quality in five (36%) cases. In all seven surgically confirmed cases, a correct diagnosis was made preoperatively with MR imaging.
CONCLUSION: By using simple modifications to standard MR imaging sequences, diagnostic-quality MR imaging of THA complications can be performed, particularly around the femoral prosthetic stem.
Index terms: Hip, abnormalities, 44.20, 44.454 Hip, MR, 44.121411, 44.121413, 44.121415, 44.121416, 44.12143 Hip, prostheses, 44.454 Magnetic resonance (MR), pulse sequences, 44.121411, 44.121413, 44.121415, 44.121416
| Introduction |
|---|
|
|
|---|
The appropriate examination of a symptomatic patient who has undergone THA includes diagnostic imaging evaluation. This can involve radiography (3), nuclear medicine studies with bone and gallium scanning (4), joint aspiration (4,5), and arthrography (6,7). Cross-sectional imaging, however, typically is not used because of imaging artifacts caused by metal hip prostheses. These imaging artifacts include beam-hardening artifacts at computed tomography (CT) and metal susceptibility artifacts at magnetic resonance (MR) imaging.
Investigators have described the use of simple strategies to decrease the MR imaging artifacts caused by a variety of metal orthopedic devices used throughout the body (813). These strategies have included selective orientation of the frequency-encoding direction along the long axis of the device (8,12,13), reducing the voxel size (9,13), increasing the readout gradient strength (9,12), using fast spin-echo imaging (812), and using lower static magnetic field strengths.
Despite the prevalence of THA procedures, however, few MR studies have focused on imaging around the metal orthopedic hip prostheses (10,12,14). Tormanen et al (12) performed an in vitro investigation that focused on optimizing standard MR sequences for imaging around metal orthopedic hip implants. Potter et al (10) performed an in vivo investigation that focused on the use of MR angiography for the detection of deep venous thrombosis and use of MR imaging for the detection of hip dislocation in patients who had undergone THA previously.
Lemmens et al (14) performed an earlier clinical MR imaging study of THA prostheses with a 0.5-T magnet. The results of this investigation indicated good depiction of loosening around the distal part of the femoral component of a THA prosthesis. Elsewhere, however, poorer depiction of periprosthetic soft tissue was noted.
Higher magnetic field strengths, however, are associated with proportionately larger metal artifacts. This has contributed to the continuing and prevailing notion that MR imaging is not appropriate for the assessment of possible THA complications. To our knowledge, no clinical study to date has focused on MR imaging of the numerous periprosthetic complications of metal hip implants at 1.5 T, nor have there been further studies performed with lower magnetic field strengths.
In this study, we performed an initial clinical investigation aimed at assessing the feasibility of using standard MR imaging sequences at 1.5-T magnetic field strength and with simple imaging parameter modifications for the assessment of various complications of primary THA.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Initial Gelatin Phantom Study
An initial gelatin phantom study was performed by using cobalt-chrome and titanium alloy femoral prostheses to help establish the section thickness and frequency-encoding gradient strength criteria for the patient study. A brief summary of the effects of magnetic static field inhomogeneities on MR images and simple methods to reduce these effects is provided in the Appendix.
Both types of prostheses were placed into separate plastic tubs filled with 2.6% commercially available gelatin. The prostheses were positioned longitudinally at the isocenter of the magnet. MR imaging was performed by using a coronal intermediate-weighted fast spin-echo sequence. The frequency-encoding direction was set along the longitudinal axis of the prosthesis. Coronal sections were obtained by using the following imaging parameters: 2,000/1325 (repetition time msec/echo time msec), echo train length of four, 30-cm field of view, 256 x 256 matrix, and one signal acquired. The section thicknesses varied between 5, 10, and 20 mm. The readout bandwidths (commensurate with frequency-encoding gradient strength for the same field of view) varied between 8, 16, and 32 kHz. In addition, standard coronal intermediate-weighted spin-echo imaging with similar parameters was performed to qualitatively assess the differences between spin-echo and fast spin-echo imaging.
Patient Study
All patients were referred to our institution for orthopedic consultation and consideration for revision arthroplastic surgery. Fourteen metal hip implants in 12 patients (seven women, five men; mean age, 65 years; age range, 4183 years) were studied. The mean time between the primary THA procedure and imaging was 15 years (range, 6 months to 30 years). All patients presented with complaints of pain and/or had abnormalities at radiography. The symptomatic complaints included local hip-groin pain in seven patients, hip and thigh pain in two patients, and isolated thigh pain and radiating pain from the hip to the knee, respectively, in one patient each. One patient was asymptomatic but had a focal area of periprosthetic hyperlucency in the proximal femur. In the two patients who had undergone bilateral arthroplasty, both hip implants were imaged. Thirteen of the hip prostheses were composed of a cobalt-chrome alloy, and one was composed of stainless steel.
Four THA prostheses were cemented, seven were not cemented, and three were hybrid devices with cemented acetabular components and noncemented femoral prostheses. Cement fixation, however, was not expected to have a major effect on imaging (15).
MR Imaging Parameters
Transverse T2-weighted images were acquired by using a fast spin-echo sequence. The imaging parameters were 4,5005,500/100, echo train length of 16, 5-mm section thickness with no intersection gap, 2230-cm field of view, 32-kHz readout bandwidth, 256 x 192 matrix, and three signals acquired.
Coronal fast inversion-recovery (ie, fast multiplanar inversion-recovery) images were obtained with a relatively long echo time to provide T2 weighting. The imaging parameters were 5,000/40/150 (inversion time msec), echo train length of eight, 5-mm section thickness with no intersection gap, 22-cm field of view, 256 x 192 matrix, and two signals acquired.
Transverse and coronal T1-weighted images were acquired with a T1-weighted fast spin-echo sequence. The imaging parameters were 450600/1015, echo train length of four, section thicknesses of 10 mm for transverse images and 5 mm for coronal images with no intersection gap, 2230-cm field of view, 32-kHz readout bandwidth, 256 x 256 matrix, and two signals acquired.
T1-weighted sequences were used before and after the intravenous administration of 2.0 mL/kg gadopentetate dimeglumine (Magnevist; Berlex Laboratories, Wayne, NJ) into the antecubital fossa. To avoid problems caused by the inhomogeneity of spectral fat suppression in the vicinity of the metal prostheses, no fat suppression was used at postcontrast T1-weighted imaging.
Image Evaluation
The MR images were reviewed independently by two radiologists (L.M.W., N.M.) who were blinded to the radiographic findings and clinical symptoms. The images were assessed for evidence of juxtaarticular or periprosthetic fluid collections, periprosthetic osseous defects, periprosthetic soft-tissue masses, and patterns of periprosthetic enhancement.
In cases in which revision arthroplasty was subsequently performed, the surgical pathologic findings of periprosthetic tissue specimens were correlated with the prior MR imaging findings. In addition, each observer made a qualitative assessment of the bone and soft-tissue resolution surrounding the prosthetic components. The periprosthetic structures assessed included juxtaprosthetic marrow and cortical bone, as well as the normal soft-tissue structures surrounding the acetabulum and proximal femur, including the sciatic nerve and obturator, iliopsoas, and adductor musculature. Periprosthetic image quality was considered to be poor or nondiagnostic when there was no depiction of periprosthetic structures, good when there was diagnostic but suboptimal resolution of periprosthetic bone and soft tissue, or excellent when there was clear depiction of periprosthetic structures.
Interobserver variability in the assessment of periprosthetic image quality around the acetabular and femoral components was recorded, and a weighted
statistic was calculated (16).
| RESULTS |
|---|
|
|
|---|
|
|
|
|
|
With our MR imaging system, as with most clinical MR imaging systems, a 32-kHz readout bandwidth is the highest currently available bandwidth option on the operator console. A 32-kHz readout bandwidth was therefore chosen for our clinical study (9,12).
As expected, a mild improvement in the appearance but not in the size of the signal voids was noted with the use of fast spin-echo imaging versus spin-echo imaging (11). Only fast spin-echo imaging sequences were therefore used in our study.
Patient Study
By using the imaging techniques described, periprosthetic abnormalities were detected at MR imaging of 11 of the 14 metal hip implants (in 11 patients) assessed by both observers. Three arthroplastic prostheses demonstrated no focal periprosthetic abnormalities at MR imaging.
Seven of the 11 patients with periprosthetic abnormalities at MR imaging subsequently underwent THA revision. In each of these cases, the surgical and pathologic findings correctly correlated with the prior MR imaging findings. Two cases showed evidence of mechanical loosening, one of which had an associated periprosthetic fracture. Four cases demonstrated foreign-body granulomatosis. One case of infection also was demonstrated. The remaining four patients with periprosthetic abnormalities at MR imaging, who had not undergone surgery at the time this article was written, all had MR imaging findings that were consistent with foreign-body granulomatosis. Three of these patients are currently awaiting THA revision surgery. Revision surgery has been deferred for the last patient owing to a superimposed medical condition.
In all eight suspected or proved cases of foreign-body granulomatosis, a focal periprosthetic intraosseous mass with low T1 signal intensity and heterogeneous, predominately low to intermediate T2 signal intensity was seen (Figs 3, 4). With the intravenous administration of contrast material, all cases of suspected foreign-body granulomatosis demonstrated peripheral enhancement and some irregular internal enhancement. Five of the cases involved granulomatosis situated around both the femoral and acetabular components. One case involved the acetabular component only, and two involved the femoral component only. Of the four cases of granulomatosis in which the patient underwent surgery (all proved to be positive), two involved both the femoral and the acetabular periprosthetic tissue, and two involved only the periprosthetic aspect of the femoral component.
|
|
|
|
|
|
|
|
|
|
value of 0.90. Depiction of the periprosthetic bone and soft tissue surrounding the femoral prosthesis component on the acquired MR images was considered by the first reviewer to be excellent in six of 14 cases. The second reviewer judged seven cases as having excellent periprosthetic depiction. Depiction was considered by both reviewers to be good in all the remaining cases. No acquired MR image was judged to have poor or nondiagnostic depiction of tissue around the femoral prosthesis component. Thus, the images of tissue surrounding all the femoral prosthesis components were of diagnostic quality in 100% of cases. Depiction of the acetabular periprosthetic structures was judged to be excellent in two of 14 cases by the first reviewer and in one case by the second reviewer. Depiction was deemed good in three cases and four cases, respectively, by the two reviewers. Both reviewers judged the depiction of periacetabular structures to be poor in nine cases. Thus, depiction of the acetabular periprosthetic structures was considered by both reviewers to be of diagnostic quality in five (36%) of 14 cases.
The depiction of periprosthetic structures in the one case with a stainless steel prosthesis was judged by both reviewers to be good surrounding the femoral prosthesis and poor surrounding the acetabular component. No substantial differences in periprosthetic imaging artifacts between the cemented and noncemented prosthetic components were noted.
| DISCUSSION |
|---|
|
|
|---|
Mechanical loosening is usually diagnosed radiographically on the basis of periprosthetic areas of radiolucency, temporal changes in prosthetic positioning, and the filling of areas of periprosthetic hyperlucency with radiographic contrast material at joint arthrography. In this study, the two cases of mechanical loosening (both surgically proved) manifested on the preoperative images as nonenhancing fluid collections with low T1 signal intensity and high T2 signal intensity surrounding the femoral stem (Fig 5).
Foreign-body granulomatosis, or microparticle disease, is a process of local foreign-body reactions to various substrates of arthroplasty, including poly-methyl-methacrylate, metal, and polyethylene; this condition results in periprosthetic bone loss. On radiographs, foreign-body granulomatosis typically manifests as lobulated regions of periprosthetic osteolysis. In this study, the cases of suspected foreign-body granulomatosis manifested as focal periprosthetic intraosseous masses with low T1 signal intensity and heterogeneous, predominately low to intermediate T2 signal intensity. With the intravenous administration of contrast material, foreign-body granulomatosis demonstrated peripheral enhancement and some irregular internal enhancement.
MR imaging has been shown to be highly sensitive and specific for the detection of osseous fractures (1921). In the single case of periprosthetic fracture in this study, we observed cortical penetration of the proximal femur by the stem of the femoral prosthesis in the setting of underlying mechanical loosening (Fig 6). On the basis of the conventional radiographic findings, the patient acquired this fracture during the interval between initial clinical consultation and MR imaging. It is plausible that, with appropriate MR imaging, other types of fractures in addition to occult longitudinal periprosthetic fractures may be detected.
Infection is a relatively uncommon THA complication that is seen with less than 1% of all THA procedures performed (22). Radiographs in the setting of infection are often nonspecific, and the diagnostic evaluation for possible infection can include nuclear medicine imaging and/or joint aspiration. In the single case in this study, an irregular, peripherally enhancing fluid collection with low signal intensity at T1-weighted imaging and high signal intensity at T2-weighted imaging was demonstrated, and this high or low signal intensity communicated with the underlying prosthetic hip joint (Fig 7). Although all infections do not result in periarticular abscess or fluid collections, the presence of juxtaosseous fluid at MR imaging has been recognized as an important sign in the diagnosis of osteomyelitis (2325). The specificity of the finding of periprosthetic fluid collections in the setting of pain following THA and the features that distinguish juxtaarticular abscess collections from fluid secondary to mechanical loosening are not yet known.
In this investigation, there was diagnostic-quality depiction of the periprosthetic structures surrounding the femoral prostheses in 100% of cases according to the two observers. Reduced (36%) diagnostic depiction of the periprosthetic soft tissue and bone surrounding the acetabular components was noted. This latter result is reasonable, considering the complicated section thickness and orientation variation that occurs in the proximity of a structure with spherical geometry (17) and the complex influence of the metallic acetabular component.
The potential clinical use of MR imaging in the investigation of suspected THA complications was highlighted in three cases in which the MR imaging findings directly affected the subsequent clinical management. In one case, infection was diagnosed at MR imaging 6 months after the initial joint replacement. Clinical management was altered in this case, with joint aspiration, resection arthroplasty, and long-term antibiotic therapy initiated before revision arthroplastic surgery.
In the second case, the patient's radiographic findings suggested possible foreign-body granuloma formation within the proximal femur, but the MR images showed the fatty bone marrow to have normal signal intensity without a focal periprosthetic abnormality; these findings were suggestive of stress shielding. On the basis of the MR imaging findings in this patient, the initial plans for revision surgery were deferred.
In the third case, the MR images showed a large periprosthetic abnormality that was consistent with foreign-body granuloma involving the entire proximal femur (Fig 4). The preoperative surgical plan, which was initially based on the radiographic findings (Fig 4a), was altered. MR imaging delineated the size and extent of granuloma formation and the lack of residual bone stock within the proximal femur (Fig 4b, 4c). The initial treatment plans were changed from local curettage and impaction bone grafting to resection and allograft reconstruction of the proximal femur. MR imaging, therefore, aided in specifying the type and amount of bone grafting required for successful revision surgery; these factors may be particularly critical for surgical planning at centers that do not have an on-site bone bank.
One limitation of our study was the relatively small number of THA prostheses imaged; this was mainly because of our patient selection criterion. Only those patients who were being considered for revision surgery on the basis of radiographic and/or clinical abnormalities were selected. This criterion, however, allowed us to assess the imaging of advanced stages of disease and THA complications, which is ideal for an initial assessment of MR imaging of THA complications. Eventually, the capability of MR imaging to help assess the early changes of disease and THA complications will need to be investigated. It should be noted, however, that a large number of cases is not necessarily required to demonstrate a technique's ability to depict periprosthetic soft tissue and bone regions in the setting of THA.
Another limitation of our study was the fact that the three patients who did not have focal periprosthetic abnormalities at MR imaging did not undergo revision surgery. Thus, the negative findings in the study could not be confirmed. In addition, although MR imaging had a clinical effect in some of the cases described, no testing of the accuracy of the technique was performed as a component of this initial investigation. Thus, although MR imaging appears to be promising, the true clinical value of the technique in the setting of suspected THA complication has yet to be assessed.
In this initial investigation, we did not perform a direct comparison between MR imaging findings and radiographic findings. Although comparison of the relative sensitivity, specificity, and accuracy of MR imaging versus those of radiography in patients with suspected complications of THA may be an important future investigation, such a comparison was not attempted in this investigation. The purpose of this initial study was to assess the feasibility of using MR imaging, with simple modifications in technique, in the examination of patients with possible complications of primary THA.
Although it is clear that diagnostic-quality MR imaging around the femoral component can be performed, the difficulty in imaging around the acetabular component is a problem intrinsic to the geometry of the acetabular component and its effect on two-dimensional MR imaging. This remains a difficult problem to address efficiently.
Decreasing the primary magnetic field strength might be beneficial to MR imaging around metal hip implants. This would commensurately reduce geometric distortion artifact with the same frequency-encoding gradient strengths. The use of lower primary magnetic field strength should be considered, particularly with regard to its effect of reduced signal-to-noise resolution.
Using alternative MR imaging sequences or adaptations of those currently used may result in further minimization of metal artifacts. For instance, the use of three-dimensional techniques will help to reduce the voxel volume and eliminate the section thickness variation caused by static field distortions. These sequences, however, tend to be time-consuming. In addition, a two-dimensional, view-angle tilting adaptation, as described by Cho et al (26), may be implemented as a means of eliminating misregistration.
In conclusion, the use of MR imaging in examining patients with possible THA-related complications is an underinvestigated clinical area. The results of this study of 14 THA prostheses in symptomatic patients demonstrate that practical and consistent diagnostic-quality 1.5-T MR imaging of THA complications can be performed, especially around the femoral stem.
| APPENDIX |
|---|
|
|
|---|
Misregistration in the frequency-encoding direction is well described and creates characteristic signal intensity variations, depending on the implant's shape and orientation (8,9,12,13,27). There is no misregistration in the phase-encoding direction because phase encoding performed with variable amplitudes is immune to misregistration.
Section thickness variation is a result of altered excitation profiles in a distorted primary magnetic field (9). In the presence of a mildly distorted primary magnetic field, an excited section takes on a warped appearance, as it follows isomagnetic lines within the imaging volume. In the presence of a ferromagnetic materialfor example, a metal spheresection thickness variation increases markedly and the profile of the section takes on a complex structure (17). Much of the signal void proximal to a spherical metal object is related to section thickness variation (17).
Diffusion-related signal intensity loss increases in the proximity of a metal implant. In spin-echo imaging, a refocusing pulse recovers the transverse signal intensity loss caused by static influences such as primary magnetic field inhomogeneities and bulk susceptibility differences. Transverse signal intensity that is not recoverable with a refocusing pulse contributes to T2 relaxation. In the presence of very large magnetic field gradients, such as those imposed by a metal hip implant in its vicinity, randomly moving water molecules undergo increased spin dephasing that is not completely recoverable with a refocusing pulse. Therefore, near a metal implant, a long echo time, such as that in T2-weighted spin-echo imaging, causes increased signal intensity loss from random microscopic processes such as diffusion.
Unlike in spin-echo imaging, in gradient-echo imaging of metal implants there is no refocusing pulse and, thus, there is additional intravoxel dephasing (T2*) that creates marked signal voids. Although decreasing the voxel size (28) or shortening the echo time (29) can improve gradient-echo imaging to some degree, the extent of intravoxel dephasing with this sequence usually precludes its regular use in the evaluation of tissue around metal implants.
If spectral fat saturation is used during an MR imaging sequence, static field or applied radio-frequency inhomogeneities can cause nonuniform signal intensities (30).
Other MR imaging artifacts, such as those related to static field (4) or applied radio-frequency (31) eddy currents, exist. These, however, are relatively nonsubstantial in magnitude.
MR Imaging Metal Artifact Reduction Techniques
Simple modifications to imaging parameters can be made to reduce the MR imaging artifacts caused by metal implants. First, misregistration artifact is proportional to static field inhomogeneity and inversely proportional to applied frequency-encoding gradient strength. Therefore, increasing the frequency-encoding gradient strength decreases the misregistration artifact proportionately (12,32).
Second, selective orientation of the frequency- and phase-encoding gradients improves the resolution of tissues surrounding metal implants by changing the misregistration artifact direction (8,12,13). In addition, the degree of such a resultant misregistration signal void is smallest with metal implants that are oriented longitudinally in the static magnetic field (8,9,12,13,27).
Third, to our knowledge, the effect of variable section thickness has been largely ignored in previously published literature. With the femoral stem placed longitudinally in the primary magnetic field, minimal effects on section curvature and thickness in the transverse or coronal orientation would be expected because of the symmetric geometry of the stem (33). In the region of the prosthesis head, however, which is roughly spherical in geometry, extreme variation in section thickness can occur (17). Although increases or decreases in the selected section thickness may aid in visualization, the section thickness variation can be completely controlled only by using three-dimensional acquisition with phase encoding of the section-select direction. This option, however, is not always practical because of examination time constraints.
Fourth, fast spin-echo imaging techniques can produce a modest increase in MR signal intensity near metal implants owing to several complex reasons. Fast spin-echo imaging refocuses spins at a faster interval than does conventional spin-echo imaging and thus causes a small increase in signal intensity by reducing diffusion-related signal intensity loss. This effect increases with smaller interecho spacing (8). Furthermore, fast spin-echo sequences are robust to malrotations of signal. Malrotation of signal is a potential problem with spin-echo multiecho sequences, wherein malrotated signal interferes with the primary signal pathway and thus creates artifacts; this effect is markedly reduced with crusher gradients around each 180° pulse. The fast spin-echo k-space pathway, however, is special in that it causes most malrotated signal to run in phase with the primary signal pathway; no crusher gradients are required (34). Because fast spin-echo sequences use rather than eliminate malrotated signal, there is a modest increase in signal intensity relative to that with multiecho spin-echo sequences per effective echo time. Accordingly, fast spin-echo sequences can use 180° refocusing pulses that are slightly more spatially thick than the 90° excitation pulse, to result in better refocusing of excited signal without concern for increased regions of malrotation.
It should be noted that, contrary to common misconception, fast spin-echo imaging does not reduce signal voids created as a result of misregistration artifacts by reducing misregistration artifact, but rather it reduces these voids by increasing the signal intensity inherent in the described mechanisms.
Fifth, reducing the voxel size increases the spatial definition of implant-induced signal void and thus reduces the apparent size of the void (9). A simple technique to reduce the voxel size without increasing the examination time is to increase the number of pixels in the frequency-encoding direction (9). Furthermore, smaller voxels help to reduce diffusion-related signal intensity loss.
Sixth, the signal intensity variation associated with spectral fat saturation in a nonuniform magnetic field may be reduced by using a fast multiplanar inversion-recovery sequence, which is an effective fat saturation technique that is less vulnerable to magnetic field variation (30).
| Footnotes |
|---|
Author contributions: Guarantors of integrity of entire study, L.M.W., J.K.K., M.M.; study concepts and design, L.M.W., J.K.K., M.M., M.E.S.; definition of intellectual content, L.M.W., J.K.K., M.M.; literature research, L.M.W., J.K.K., M.M.; clinical studies, L.M.W., M.M., W.B.M.; experimental studies, L.M.W., J.K.K.; data acquisition, L.M.W., J.K.K., M.M., W.B.M., N.M.; data analysis, L.M.W., J.K.K., M.M., N.M., A.E.G., C.R.H.; statistical analysis, L.M.W.; manuscript preparation, L.M.W., J.K.K.; manuscript editing, L.M.W., J.K.K., M.E.S.; manuscript review, L.M.W., J.K.K., C.R.H., A.E.G., M.E.S.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H Marmery and S Ostlere Imaging of prosthetic joints Imaging, September 1, 2007; 19(3): 299 - 309. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.-J. Lee, S. Kim, S.-A. Lee, H.-T. Song, Y.-M. Huh, D.-H. Kim, S. H. Han, and J.-S. Suh Overcoming Artifacts from Metallic Orthopedic Implants at High-Field-Strength MR Imaging and Multi-detector CT RadioGraphics, May 1, 2007; 27(3): 791 - 803. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. C. Genson, C. E. Blane, M. A. Helvie, S. A. Waits, and T. L. Chenevert Effects on Breast MRI of Artifacts Caused by Metallic Tissue Marker Clips Am. J. Roentgenol., February 1, 2007; 188(2): 372 - 376. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Raphael, A. H. Haims, J. S. Wu, L. D. Katz, L. M. White, and K. Lynch MRI comparison of periprosthetic structures around zirconium knee prostheses and cobalt chrome prostheses. Am. J. Roentgenol., June 1, 2006; 186(6): 1771 - 1777. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. D. Toms, D. Davidson, B. A. Masri, and C. P. Duncan The management of peri-prosthetic infection in total joint arthroplasty J Bone Joint Surg Br, February 1, 2006; 88-B(2): 149 - 155. [Full Text] [PDF] |
||||
![]() |
P. J. Duffy, B. A. Masri, D. S. Garbuz, and C. P. Duncan Evaluation of Patients with Pain Following Total Hip Replacement J. Bone Joint Surg. Am., November 1, 2005; 87(11): 2566 - 2575. [Full Text] [PDF] |
||||
![]() |
K. Ohashi, G. Y. El-Khoury, D. L. Bennett, J. M. Restrepo, and K. S. Berbaum Orthopedic Hardware Complications Diagnosed with Multi-Detector Row CT Radiology, November 1, 2005; 237(2): 570 - 577. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. W. A. Pfirrmann, H. P. Notzli, C. Dora, J. Hodler, and M. Zanetti Abductor Tendons and Muscles Assessed at MR Imaging after Total Hip Arthroplasty in Asymptomatic and Symptomatic Patients Radiology, June 1, 2005; 235(3): 969 - 976. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. G. Potter, B. J. Nestor, C. M. Sofka, S. T. Ho, L. E. Peters, and E. A. Salvati Magnetic Resonance Imaging After Total Hip Arthroplasty: Evaluation of Periprosthetic Soft Tissue J. Bone Joint Surg. Am., September 1, 2004; 86(9): 1947 - 1954. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. D. M. Stumpe, H. P. Notzli, M. Zanetti, E. M. Kamel, T. F. Hany, G. W. Gorres, G. K. von Schulthess, and J. Hodler FDG PET for Differentiation of Infection and Aseptic Loosening in Total Hip Replacements: Comparison with Conventional Radiography and Three-Phase Bone Scintigraphy Radiology, May 1, 2004; 231(2): 333 - 341. [Abstract] [Full Text] [PDF] |
||||