|
|
||||||||
Breast Imaging |
1 From the Iris Cantor Center for Breast Imaging, University of California Los Angeles School of Medicine (L.W.B.); the Breast Imaging Center, Thomas Jefferson University, Philadelphia, Pa (D.M.F.); the American College of Radiology, Reston, Va (S.B., M.A.F., P.A.W.); and the Department of Radiology, Mount Sinai School of Medicine, New York, NY (S.A.F.). Received June 18, 1999; revision requested July 14; revision received September 1; accepted September 14. Address correspondence to D.M.F., Mallinckrodt Institute of Radiology, Washington University Medical Center, 510 S Kingshighway Blvd, St Louis, MO 63110 (e-mail: farriad@mir.wustl.edu).
| Abstract |
|---|
|
|
|---|
MATERIALS AND METHODS: In 1997, the American College of Radiology Mammography Accreditation Program reviewed clinical images for 2,341 mammography units. For each mammography unit, the facility submitted bilateral mediolateral oblique and craniocaudal mammograms obtained in a woman with fatty breasts and a woman with dense breasts. Images were reviewed independently by two experienced radiologists. Reviewers listed the general categories and specific deficiencies that led to a decision to fail the unit that produced the clinical images.
RESULTS: Of the 2,341 mammography units, 1,034 (44%) failed the clinical image evaluation process. Of 6,128 categories cited by reviewers as deficient, 1,250 (20%) involved problems in positioning; 944 (15%), exposure; 887 (14%), compression; 806 (13%), sharpness; 785 (13%), contrast; 703 (11%), labeling; 465 (8%), artifacts; and 288 (5%), noise. A significantly higher proportion of failures was attributed to positioning deficiencies for fatty breasts than for dense breasts (P = .028). Higher proportions of failures in dense breasts were related to compression (P < .001) and exposure (P < .001) deficiencies.
CONCLUSION: Common problems in clinical image quality have been identified. This information should be useful for educators and facilities striving to improve the quality of mammography.
Index terms: Breast radiography, quality assurance, 00.11, 00.93, 00.99 Quality assurance
| Introduction |
|---|
|
|
|---|
In ACR MAP, there are more than 110 radiologists who conduct peer reviews of clinical images that are submitted for accreditation. Two independent clinical image reviewers assess each set of mammograms. In the event of disagreement, the outcome is determined by a third reviewer. All ACR MAP clinical image reviewers meet the following standards: (a) an active practice in which 50% or more of the practice is dedicated to breast imaging, (b) member of the ACR, (c) MQSA-qualified interpreting physician, (d) American Board of Radiology certification, (e) practice in an accredited facility, and (f) recommendation for the position from a current clinical image reviewer. The performance of all reviewers is monitored on a quarterly basis. These reports evaluate consistency among reviewers and provide each reviewer with an assessment of his or her reviews for the most recent quarter. Reviewers are informed of their individual rates and whether their individual rates are within 1 SD of the mean rate for all reviewers. Corrective action is initiated for reviewers who score outside of the acceptable range.
Clinical image evaluation has been a component of the ACR MAP since the inception of ACR MAP in 1987 (4). In the 3 years before this article was written, ACR MAP had reviewed clinical images from nearly 11,300 facilities, which represents approximately 94% of mammography facilities nationwide (5). We conducted a retrospective analysis of all ACR MAP clinical image evaluations in 1997 to identify the types and numbers of deficiencies that cause a mammography unit to fail the clinical imageevaluation process.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Image quality was reviewed for the following eight image quality categories: positioning, compression, exposure, contrast, sharpness, noise, artifacts, and labeling. For each of these categories, specific deficiencies were noted and recorded by ACR MAP reviewers for all submitted sets of clinical images. Thus, there could be multiple deficiencies in any of the categories (Table 1).
|
To identify the most common categories with problems that led to a failed clinical image review, we tabulated the frequencies with which each of the eight general image quality categories contributed to a decision of a failure grade. We further identified and tabulated the specific types and frequencies of deficiencies within each of these eight categories. To identify image quality problems specific to fatty versus dense breasts, we compared image quality categories identified as contributing to failure for these two breast compositions.
Using the test of significance of proportions (7), we identified the categories where the percentages of failure were significantly different (P < .05) in fatty versus dense breasts. We also used the Spearman rank correlation coefficient to determine whether failure in one category significantly correlated with failure in other categories (7). Since clinical image reviewers do not fail a mammography unit in a facility for isolated artifact or image-labeling deficiencies, we also tabulated all serious deficiencies in these two categories, regardless of whether a mammography unit in a facility failed the review process. This additional information was a more accurate reflection of the frequency of serious labeling and artifact problems in the sample. A "serious" problem was defined as an image quality category that received a "1" or "2" rating on a scale of 1 to 5 (1 = worst, 5 = best).
| RESULTS |
|---|
|
|
|---|
Table 2 reports the frequency of problems in the eight image quality categories that were judged to be serious enough to lead to failure. Several categories were usually cited as responsible for a failed clinical image evaluation. Overall, individual categories were cited 6,128 times as reasons for failure. The most common image quality category leading to failure was positioning, which affected 1,250 sets of submitted images, followed by exposure, compression, sharpness, and contrast. In the 1,250 sets of images with problems in positioning, there were 3,400 specific positioning deficiencies cited by reviewers (Table 3).
|
|
There were 806 sets of submitted images with sharpness problems. In the 806 sets of images, there were 1,514 specific deficiencies related to image sharpness, including 889 (59%) for poor delineation of linear structures, 489 (32%) for unsharpness of feature margins, and 136 (9%) for blurring of microcalcifications. In 785 sets of images with contrast problems, there were 964 total contrast deficiencies: Inadequate contrast was responsible for 862 (89%) citations; excessive contrast, for 96 (10%) citations; and other reasons, for six (1%) citations. In 288 sets of images, there were 458 noise-related deficiencies, including 339 (74%) for a visually striking mottle pattern and 119 (26%) for limited visualization of detail because of noise.
The data showed significant correlation among compression, sharpness, contrast, and exposure deficiencies. The greatest correlations were noted between compression and sharpness (r = 0.4046, P < .001) and between exposure and contrast (r = 0.4766, P < .001). Significant correlations were also noted between compression and exposure (r = 0.1537, P < .001), sharpness and exposure (r = 0.1083, P < .001), and contrast and sharpness (r = 0.1413, P < .001). Table 2 compares the deficiencies in the submitted images of the primarily fatty versus the primarily dense breasts.
There were 703 sets of images with artifact problems that resulted in failure in the clinical imageevaluation process. In the total sample of submitted images, there were 1,241 serious deficiencies in the artifact category, including 465 (37%) for lint, 362 (29%) for scratches or pickoff (small defects in film emulsion due to processing), 148 (12%) for grid-related artifacts, 115 (9%) for roller marks, and 151 (12%) for other artifacts.
There were 465 sets of images with labeling problems among cases that resulted in failure in the clinical imagereview process. In the total sample of submitted images, there were 831 serious deficiencies in the labeling category, including 419 (50%) for inadequate identification of the facility, 101 (12%) for failure to identify the radiologic technologist, 94 (11%) for failure to place an identifying number on the intensifying screen for each cassette, 87 (10%) for improperly labeled mammographic view, 78 (9%) for inadequate identification of the patient, and 52 (6%) for other labeling problems.
We calculated the frequency of serious artifact and labeling problems in the total sample of submitted images. Of the 831 serious labeling problems, 542 (65%) were identified in the sample of images that passed the clinical image review.
| DISCUSSION |
|---|
|
|
|---|
Findings of our review demonstrated that common problems were positioning and compression. Problems in the positioning category were most frequently due to an inadequate amount of pectoralis major muscle on the mediolateral oblique view, poor visualization of posterior tissue on the craniocaudal or mediolateral oblique view, sagging breasts, or skin folds. Inadequate compression was manifested by overlapping breast structures, nonuniform tissue exposure, and motion unsharpness. Instructions for achieving proper positioning and compression are explained and illustrated in a number of publications (810). Thus, it appears that many practicing radiologic technologists still have a need for actual hands-on training for the standard views, the mediolateral oblique and the craniocaudal views.
Image clarity, or the ability of the mammogram to portray diagnostic information clearly, depends on a number of other factors, which include exposure, sharpness, contrast, and noise (1118). In the image quality category of exposure, generalized underexposure was a far more frequent deficiency than overexposure. This is of concern because lesions can be obscured in dense underexposed breast tissue, which leads to false-negative mammograms. Generalized underexposure can result from inadequate compression, technical factors, equipment malfunction, or a combination of these factors. Nonuniform exposure of the breast parenchyma with inadequate penetration of denser tissues was another commonly cited deficiency.
Sharpness is the ability to define the edges of breast structures. Unsharpness, also referred to as blur, is manifested by poor delineation of the edges of linear structures, tissue margins, and microcalcifications. Common causes of blur include patient motion and poor screen-film contact. Contrast, which is the degree of variation in optical density between different areas of the film, is affected by film processing, exposure, film type, and scatter reduction (with grids, compression). Inadequate contrast was a more frequent deficiency than excessive contrast.
Noise, manifested as a visually striking mottle pattern, was the least common category that led to failure of the image-review process. However, when noise was substantial enough to limit the visualization of detail, this deficiency sometimes led to failure of the clinical imagereview process. Quantum mottle is the major source of noise in mammography. The fewer photons used to make the image, the greater the observed quantum mottle. Other causes of noise include vigorous or prolonged processing, fast image-recording systems, and underexposure.
Although failure of the clinical imagereview process was not based solely on the presence of artifacts, egregious artifacts sometimes led to failure. The most common artifacts cited by reviewers were dirt or lint, scratches or pickoff, grid lines, and processor roller marks. Careful film handling and diligent quality control measures should eliminate many of these artifacts.
While deficiencies in labeling alone usually do not lead to a failed clinical image evaluation, inappropriate labeling was a common problem on submitted images (19). The most common deficiency in this category was inadequate identification of the mammography facility. The final regulations of the MQSA, which went into effect on April 28, 1999, list specific requirements for proper image identification (Figure). According to the final regulations, "each image shall have the following information indicated on it in a permanent, legible, and unambiguous manner and placed so as not to obscure anatomic structures": (a) name of the patient and additional patient identifier, (b) date of examination, (c) view and laterality, (d) facility name and location (at a minimum, the location should include city, state, and zip code of the facility), (e) technologist identification, (f) cassette or screen identification, and (g) mammography unit identification if there is more than one unit in the facility (2).
|
According to our data, deficiencies in exposure, contrast, sharpness, and compression are frequently interrelated. For example, there is a correlation between compression and exposure deficiencies, because adequate exposure depends on sufficient compression of dense fibroglandular tissue. Compression and sharpness deficiencies also often coexist on the same images. This correlation is likely due to the motion unsharpness that results from inadequate compression.
In summary, the high proportion of mammography units in facilities that fail the clinical imagereview process is not surprising because of the stringent criteria used for image evaluation. The high expectations are in keeping with the intent of the MQSA, the educational emphasis of the MAP, and the fact that facilities select their best representative images for review. Although many facilities have deficiencies in their clinical image evaluation when applying for accreditation, a majority of facilities successfully complete the clinical imageevaluation process upon reapplication. In fact, it is generally acknowledged that the quality of mammography has markedly improved since the inception of ACR MAP (20).
| Acknowledgments |
|---|
| Footnotes |
|---|
Author contributions: Guarantor of integrity of entire study, L.W.B.; study concepts, D.M.F., L.W.B.; study design, M.A.F., S.B., L.W.B., D.M.F.; definition of intellectual content, D.M.F., L.W.B.; literature research, L.W.B.; data acquisition, P.A.W., M.A.F.; data analysis, L.W.B., S.B., D.M.F.; statistical analysis, S.B.; manuscript preparation, D.M.F., L.W.B.; manuscript editing, D.M.F., L.W.B., S.A.F.; manuscript review, S.A.F., D.M.F., L.W.B., P.A.W.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. MINIGH Mammographic Film Artifacts. Radiol. Technol., May 1, 2006; 77(5): 389M - 402M. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |