|
|
||||||||
Medical Physics |
1 From the U.S. Food and Drug Administration, Center for Devices and Radiological Health, 1350 Piccard Dr, Rockville, MD 20850. Received December 15, 1997; revision requested March 11, 1998; final revision received October 4, 1999; accepted October 26. Address correspondence to R.V.K. (e-mail: rvk@cdrh.fda.gov).
| Abstract |
|---|
|
|
|---|
MATERIALS AND METHODS: Surveyed facilities were randomly selected from each state. Patient exposure was evaluated along with film processing, half-value layer, and image quality. Additional information obtained concerned type of equipment, facility work load, radiographic technique, screen-film system, and grid type.
RESULTS: Mean entrance air kerma in all facilities was 141 µGy (16.1 mR). Mean kilovoltage in all facilities was 101 kV. In 1994, 140 (90%) of 156 hospitals (vs 71% in 1984) and 92 (58%; nearly double the percentage in 1984) of 159 nonhospital sites were using grids. Scoring with the imaging test tool resulted in a mean spatial resolution of 2.3 cycles per millimeter, and a mean low-contrast sensitivity of about 3%. Two hundred fifty-three (80%) of 315 facilities surveyed were processing film at minimum acceptable performance levels.
CONCLUSION: Mean entrance air kerma for all facilities did not substantially change. Although increased grid usage would lead to the expectation of higher measured exposures, this was offset by an increase in the use of faster screen-film combinations.
Index terms: Images, quality, 60.11 Quality assurance Radiations, exposure to patients and personnel, 60.11 Radiations, measurement, 60.11 Screens and films Thorax, radiography, 60.11
| Introduction |
|---|
|
|
|---|
The Conference of Radiation Control Program Directors (CRCPD), state radiologic health programs, and the U.S. Food and Drug Administration (FDA) Center for Devices and Radiological Health (CDRH) jointly cooperate to carry out the Nationwide Evaluation of X-ray Trends (NEXT) program. This program was begun in the 1970s when the CDRH encouraged diagnostic facilities to quantify radiation exposures to patients during typical diagnostic radiographic examinations. This information was also believed to be useful to state agencies in the evaluation of patterns of usage and in the improvement of specific areas of practice in radiology.
In 1984, attention was turned to chest radiography, and a nationwide survey was performed in hospitals (10). In 1986, a survey of private practice facilities was performed. Although patient exposure levels at chest radiographic examination are considered to be relatively low, consideration was given to the fact that a large proportion of the population undergoes this examination, often repeatedly.
The purpose of this study was to report the findings of the survey conducted by the CDRH, CRCPD, and state agencies in 1994 to determine what changes may have taken place in the practice of chest radiography and to compare the findings with those of a previous survey.
| MATERIALS AND METHODS |
|---|
|
|
|---|
A written test protocol was developed by the CRCPD-NEXT committee. Surveyors attend a training session to ensure that all followed the same standard method of data collection. All information collected was entered into a computerized database maintained at the FDA.
Each surveyor used equipment that included an electrometer, a 6-mL ionization chamber (Model 1015; Radcal, Monrovia, Calif), several aluminum sheets (type 1100) for the determination of the half-value layer, the FDA LuCal chest phantom (11), and a 5 x 10-cm imaging test tool. This tool was attached with Velcro to the chest phantom and was imaged when the exposure measurements were obtained. The phantom simulated the attenuation of a 22.5-cm-thick chest radiographed in the posteroanterior projection. The x-ray exit spectra at diagnostic energy levels matched those obtained with an anthropomorphic chest-lung phantom. This phantom was originally developed as part of the pilot testing of the 1984 NEXT survey to address the fact that automatic exposure control was becoming more widely used (12).
Scoring of image spatial resolution and low-contrast sensitivity was performed in 1994. An image test tool, assembled at the FDA, was intended to provide a simple measure of spatial resolution and low-contrast sensitivity (Fig 1). The tool contained a series of eight mesh screens made of copper wire (which had 0.8, 1.2, 1.6, 2.0, 2.4, 3.2, 3.9, and 4.7 wires per millimeter) and a 6.1-mm-thick aluminum disk that had eight 7.94-mm-diameter pockets bored to the following depths: 0.08, 0.11, 0.16, 0.23, 0.32, 0.46, 0.64, and 0.89 mm. The image test tool was positioned in the x-ray beam away from the ionization chamber and its holding bracket to ensure that it was imaged completely. The surveyors recorded the number of spatial resolution mesh patterns and the number of low-contrast holes depicted on the radiographic image.
|
The chest phantom was placed in the beam from the overhead tube of a conventional radiographic-fluoroscopic unit. Each sheet was attached to the phantom, in turn, and an ionization chamber was held in place behind the phantom (with an air gap) to record the relative exposure as the beam traversed the phantom and aluminum sheet. Exit exposure measurements were made at 70, 80, 90, 100, 110, and 120 kV. The range of half-value layers was 2.56 (70 kV) to 4.38 mm (120 kV) of aluminum. A monitor chamber was used to compensate for variations in tube output. The relative transmission through both the phantom and each sheet was recorded. These data were used to plot the relative subject contrast against the hole score (linear fit with residual values of 0.8620.976) obtained at each energy level (Fig 2).
|
A processing speed of 100 represented adherence to the recommendations of the manufacturer, a speed of less than 80 indicated underprocessing, and a speed of greater than 120 indicated overprocessing. This ±20% tolerance was used to account for variability in the film emulsion over the useful life of the film, variations attributable to calibration of the densitometer and sensitometer, and normal fluctuations in the day-to-day range of operation with automatic processors (14). Darkroom fog was evaluated by removing an exposed film from its cassette holder and by allowing it to remain in the darkroom work area for 2 minutes with part of the film covered before it was processed. Normally, films are processed immediately after they are removed from the cassette holder.
At each surveyed facility, the type of generator, brand of film and intensifying screen, grid ratio, and names of the equipment and processor manufacturers were recorded. The approximate number of chest examinations performed per week and per unit at each facility were determined. The ionization chamber used to measure x-ray output was positioned 24 cm from the front surface of the phantom so that its measurements would be essentially free of backscatter. X-ray output was measured with the phantom positioned against the chest stand. The phantom was imaged by using clinical techniques, including the use of automatic exposure control, for density setting and film size. Entrance air kerma (free of backscatter) was calculated for the posteroanterior view in a patient with a 22.5-cm-thick chest by using an inverse-square correction from the position of the ionization chamber to the entrance surface of the patient. Beam quality was determined in millimeters of aluminum.
| RESULTS |
|---|
|
|
|---|
The use of older single-phase generators declined. Use of three-phase equipment, including high-frequency generators, increased from 55% of hospitals surveyed in 1984 to 77% (121 of 156) in 1994. More than 96% of nonhospital facilities reported the use of single-phase generators in 1986; this proportion was reduced to 80% (127 of 159) in 1994.
The use of automatic exposure control (phototiming) increased, although more so at hospitals. In 1984, only 54% of hospitals surveyed used automatic exposure control; this percentage increased in 1994 to 82% (128 of 156). Twenty (13%) of the 159 nonhospital facilities surveyed in 1994 used automatic exposure control; this percentage increased from 1986, when only 6% of nonhospital facilities surveyed reported the use of automatic exposure control.
The mean kilovoltage in hospitals was 110 kV (median, 110 kV). The mean kilovoltage in nonhospital facilities was 93 kV (median, 90 kV). The mean kilovoltage in all facilities was 101 kV. The SD was about 15 kV. The distribution is depicted in Figure 3. In both hospitals and nonhospital facilities, the mean values increased by 6 kV, since the mean was 104 kV in hospitals in 1984 and 87 kV in nonhospital facilities in 1986.
|
The use of grids to improve image quality has increased. In the recent combined survey, 140 (90%) of 156 hospitals were using grids (median grid ratio, 12:1), representing an increase from 71% in 1984. The percentage of nonhospital facilities using grids (median grid ratio, 10:1) had almost doubled from 33% (1986) to 58% (92 of 159). The mean and median grid ratios in use at hospitals were the same, and these values were the same as those found in 1984. The mean ratio at nonhospital facilities increased from 8:1 to 10:1. The mean entrance air kerma observed was 131 µGy (14.7 mR; median, 105 µGy) at hospitals and 157 µGy (17.6 mR; median, 114 µGy) at nonhospital facilities (Fig 4). Mean entrance air kerma in all facilities was 141 µGy (16.1 mR).
|
When processing was evaluated, 78 (50%) of 156 hospitals had processing speeds of 80120, which we considered to be in accordance with the specifications of the film manufacturers. The mean processing speed at hospitals was 115 (median, 115). Fifty (32%) hospitals were overprocessing (speeds greater than 120). Eighty-six (54%) of 159 nonhospital facilities had speeds of 80120 (mean, 107; median, 107). Thirty-nine (25%) nonhospital facilities were overprocessing (Fig 5). The mean film density (measured in the region of the lungs) at hospitals was 1.61 (median, 1.59), whereas the mean film density measured at nonhospital facilities was 1.74 (median, 1.67).
|
In 19841986, the mean film optical density was virtually the same at both hospitals and nonhospital facilities (1.43 and 1.42). In 1994, the mean film optical density increased by 0.18 at hospitals and 0.32 at nonhospital facilities.
The surveyors counted the number of wire mesh patterns and holes visible on radiographs obtained with the image-quality test tool positioned on the phantom. The median number of mesh screens seen was five of eight (Fig 6). This number corresponded to 2.4 wires per millimeter (60 wires per inch). The median number of holes seen was five of eight (Fig 7).
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Overall, the mean entrance skin exposure was essentially unchanged from the exposure from the last NEXT chest survey. The mean slightly decreased by about 9 µGy (1.03 mR) at hospitals and increased by about 18 µGy (2.06 mR) at nonhospital facilities. The mean entrance air kerma (entrance skin exposure) for posteroanterior examinations at both hospitals and nonhospital facilities in 19841986 was 140 µGy (16.04 mR). A more complete tabular and graphical summary of all of the survey results will be published by the CRCPD (16).
There were two important technical changes between 1984 and 1986 that affect entrance air kerma, and the effects of the changes tend to offset each other. The increased use of faster screen-film combinations allowed shorter exposure times (lower mAs) and lower patient exposures. However, the use of a grid results in more x-ray beam absorption and will increase patient exposure if no other factor is changed. Therefore, the initial expectation that the mean entrance exposure in patients would decrease was balanced by the finding that more facilities were using antiscatter grids.
When we evaluated film processing, a shift toward faster processing speeds resulted in an overall improvement compared with the findings of prior surveys. As mentioned previously, when processing speed was determined, slightly more than half of both hospitals and nonhospital facilities were within the normal range. The mean STEP value reported by hospitals in 1984 was 97, and 72% of the facilities were in the range of 80120. The increase in the mean STEP value in hospitals to 107 in the 1994 survey indicated an overall increase in processing speed. This finding is further evidenced by the large increase in the percentage of hospitals that were overprocessing, that is, 32% (50 of 156) in 1994 versus 10% in 1984. The percentage of hospitals that developed film in the 80120 range decreased apparently because they were now overprocessing; however, this finding is not necessarily undesirable. Underprocessing would be more cause for concern because it might lead facilities to raise patient exposures unnecessarily to compensate for unsatisfactory radiographs. Underprocessing declined from 17% of hospitals surveyed in 1984 to just 4% (six of 156) in 1994.
The film processing results were similar at nonhospital facilities. The mean processing speed increased while still remaining within the target range (speed was 85 in 1986). In contrast to hospitals, however, the number of facilities processing within the normal range increased from 45% to 54% (86 of 159). The percentage of facilities found to be underprocessing decreased substantially from 41% in 1986 to only 14% (23 of 159) in 1994. At the same time, the percentage of nonhospital facilities found to be overprocessing increased from 14% in 1986 to 25% (39 of 159). While the nonhospital facilities were processing at speeds lower that those of hospitals, more nonhospital facilities used speeds in the desired range.
With respect to the scores of the image-quality test tool, use of this aid may prove to be a useful and quick way to evaluate the performance of an imaging facility. The geometry of the tool duplicated that which is clinically used, and imaging was performed under conditions in which scattered radiation was produced by the phantom. Individual facilities that perform conventional screen-film chest radiography may consider the data in Figure 8 as indicators of the low-contrast sensitivity they can detect. The comparative results in the different ranges of the kilovolt peak indicated a slight improvement in the detectable subject contrast as the tube voltage increased. This finding may be the result of a relationship between technique selection and grid use. The data show that the mean kilovolt peat at facilities that use grids was 107 kV, compared with 82 kV for the remainder. Hospitals reported slightly higher scores compared with those of nonhospital facilities, and, for those facilities that did not use grids, their mean of 4.3 was lower than the overall mean.
After the completion of the initial NEXT study, the Swedish National Institute of Radiation Protection used a Lucite and aluminum phantom to perform a NEXT-type survey in that country (17). Despite considerable differences in technique factors, the entrance air kerma values from our study were almost identical in facilities that used scatter suppression. The tube voltage used in Sweden ranged from 100 to 150 kV, and grid usage was almost universal. The 1994 NEXT results showed mean film processing speeds that were slightly lower that those reported in Sweden, although this finding was somewhat offset by slightly higher mean screen-film speeds.
| Acknowledgments |
|---|
| Footnotes |
|---|
Author contributions: Guarantors of integrity of entire study, R.V.K., O.H.S.; study concepts, B.J.C., O.H.S.; study design, B.J.C., O.H.S., R.O.S.; definition of intellectual content, B.J.C., O.H.S., R.V.K.; literature research, R.V.K.; data analysis, R.V.K.; statistical analysis, R.V.K.; manuscript preparation, R.V.K.; manuscript editing and review, R.V.K., O.H.S.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. S. Hambali, K.-H. Ng, B. J. J. Abdullah, H.-B. Wang, N. Jamal, D. C. Spelic, and O. H. Suleiman ENTRANCE SURFACE DOSE AND IMAGE QUALITY: COMPARISON OF ADULT CHEST AND ABDOMINAL X-RAY EXAMINATIONS IN GENERAL PRACTITIONER CLINICS, PUBLIC AND PRIVATE HOSPITALS IN MALAYSIA Radiat Prot Dosimetry, February 16, 2009; (2009) ncp007v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. E. Muhogora, N. A. Ahmed, A. Almosabihi, J. S. Alsuwaidi, A. Beganovic, O. Ciraj-Bjelac, F. K. Kabuya, A. Krisanachinda, M. Milakovic, G. Mukwada, et al. Patient Doses in Radiographic Examinations in 12 Countries in Asia, Africa, and Eastern Europe: Initial Results from IAEA Projects Am. J. Roentgenol., June 1, 2008; 190(6): 1453 - 1461. [Abstract] [Full Text] [PDF] |
||||
![]() |
C S Moore, G P Liney, A W Beavis, and J R Saunderson A method to optimize the processing algorithm of a computed radiography system for chest radiography Br. J. Radiol., September 1, 2007; 80(957): 724 - 730. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. O'Daniel, D. M. Stevens, and D. D. Cody Reducing Radiation Exposure from Survey CT Scans Am. J. Roentgenol., August 1, 2005; 185(2): 509 - 515. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Spelic, R. V. Kaczmarek, and O. H. Suleiman Nationwide Evaluation of X-ray Trends Survey of Abdomen and Lumbosacral Spine Radiography Radiology, July 1, 2004; 232(1): 115 - 125. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Vassileva A phantom for dose-image quality optimization in chest radiography Br. J. Radiol., October 1, 2002; 75(898): 837 - 842. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Vano, J. M. Fernandez, J. I. Ten, E. Guibelalde, L. Gonzalez, and C. S. A. Pedrosa Real-Time Measurement and Audit of Radiation Dose to Patients Undergoing Computed Radiography Radiology, October 1, 2002; 225(1): 283 - 288. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |