|
|
||||||||
Medical Physics |
1 From the Department of Radiology, Brigham and Women's Hospital, 75 Francis St, Boston, MA 02115. From the 1998 RSNA scientific assembly. Received July 22; revision requested September 2; revision received October 21; accepted November 22. Address correspondence to R.D.N. (e-mail: nawfel@bwh.harvard.edu).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Surface dose was estimated on a CT dosimetric phantom by using thermoluminescent dosimetric (TLD) and CT pencil chamber measurements. Scatter exposure was estimated from scattered radiation measured at distances of 10 cm to 1 m from the phantom. Scatter exposures measured with and without placement of a lead drape on the phantom surface adjacent to the scanning plane were compared.
RESULTS: Phantom surface dose rates ranged from 2.3 to 10.4 mGy/sec. Scattered exposure rates for a commonly used CT fluoroscopic technique (120 kVp, 50 mA, 10-mm section thickness) were 27 and 1.2 µGy/sec at 10 cm and 1 m, respectively, from the phantom. Lead drapes reduced the scattered exposure by approximately 71% and 14% at distances of 10 and 60 cm from the scanning plane, respectively.
CONCLUSION: High exposures to patients and personnel may occur during CT fluoroscopyguided interventions. Radiation exposure to patients and personnel may be reduced by modifying CT scanning techniques and by limiting fluoroscopic time. In addition, scatter exposure to personnel may be substantially reduced by placing a lead drape adjacent to the scanning plane.
Index terms: Computed tomography (CT), guidance, 70.12119 Computed tomography (CT), radiation exposure, 70.12119 Fluoroscopy Phantoms Radiations, exposure to patients and personnel, 70.12119 Radiations, measurement, 70.12119 Radiations, protective and therapeutic agents and devices, 70.12119
| INTRODUCTION |
|---|
|
|
|---|
During CT fluoroscopy, the patient is exposed to radiation at or near the needle puncture site. The duration of exposure may range from seconds to a few minutes (36). Since the fluoroscopic exposure to the patient at the needle entry site is cumulative, deterministic effects can be substantial, as they are during conventional fluoroscopy (7,8). Hence, there is a potential radiation hazard to the patient during CT fluoroscopyguided interventional procedures. Furthermore, the level of scatter exposure to personnel could be high at locations close to the scanning plane with use of CT fluoroscopic techniques during these procedures. Personnel exposure, especially exposure to the hands of the physician, is of concern, since the hands are closest to the scanning plane during the intervention.
Since there is a possibility for high exposures to occur during CT fluoroscopy, the purpose of our study was to assess the potential for high radiation exposure by using data from phantom measurements. The exposure rate during CT fluoroscopy was measured at the surface of an acrylic CT dosimetric phantom and at locations adjacent to the phantom. In addition, we investigated the effect of lead drape shielding for the reduction of radiation exposure to personnel.
Results were used to develop guidelines for exposure time during CT fluoroscopy with the goal of minimizing patient dose and personnel exposure. These guidelines are specific for the particular CT scanner and scanning conditions used during CT fluoroscopic interventional procedures (6).
| MATERIALS AND METHODS |
|---|
|
|
|---|
At acceptance testing, dose measurements of the computed tomography dose index (CTDI) (917) obtained by using the head (16-cm-diameter) and body (32-cm-diameter) dosimetric phantoms (Nuclear Associates; Carle Place, NY) were 13.0 mGy per 100 mAs and 8.8 mGy per 100 mAs, respectively, at 120 kVp with a 10-mm section thickness.
Surface Dose and Scatter Exposure Rate Measurements
Surface dose and scatter exposure were assessed during CT fluoroscopy by using a 20-cm-diameter acrylic CT dosimetric phantom. This phantom was used because it better characterizes the attenuation caused by a human abdomen, which has an intermediate size between those of the 16-cmdiameter head and 32-cmdiameter body phantoms.
Dose rates were estimated at the surface and at a 2-cm depth in the phantom from both thermoluminescent dosimetric (TLD) and pencil ionization chamber (model PC 4P; Capintec, Ramsey, NJ) measurements. These data were collected with the table at a fixed position during CT fluoroscopic scanning so that the entire exposure occurred at the same location (section) in the phantom. The TLD chips (TLD100; Harshaw Bircon, Solon, Ohio) were 3 x 3 x 1-mm and were arranged in a linear pattern with no spacing between them. The TLDs were sent to be read out (Landauer; Glenwood, Ill).
Calibration was performed by exposing the TLDs to a diagnostic x-ray beam that was calibrated with an ionization chamber traceable to the National Institute of Standards and Technology (NIST). The tube potential of the calibration beam was 120 kVp, with a half-value layer of 8 mm of aluminum.
Scatter exposures were estimated from measurements of phantom scatter radiation. A digital dosimeter (model 192X; Capintec) and diagnostic chamber (model PM-30 30-mL diagnostic chamber with NIST traceable calibration; Capintec) were used to measure scatter radiation. Scatter exposures were specified as air kerma. Tube potentials were 80 and 120 kVp, and tube currents were 50135 mA. Section thicknesses were 210 mm. Dose rates (both surface and scatter) were calculated by dividing the total cumulative exposure by the exposure time (10 seconds). The reproducibility determined from six measurements of the CTDI was 3.3%.
Scatter Radiation Exposure
Secondary exposure was measured at points relative to the scanning plane and phantom. Figure 1 illustrates the geometry and experimental setup for the measurement of the scatter exposure rate, with measurement points chosen at distances from the scanning plane toward the foot end of the table.
|
Scatter Exposure Reduction
The potential for high exposures to personnel during CT fluoroscopy persuaded us to investigate a practical method for controlling scatter exposure. Several of the previous scatter measurements were repeated with a lead drape placed on top of the phantom adjacent to the scanning plane (Fig 2). Reduction in scatter was estimated at various distances from the scanning plane and with different section thicknesses. It was assumed that the measured exposure rate at distances along the z axis was due to scattered radiation from the phantom.
|
| RESULTS |
|---|
|
|
|---|
|
Scatter Exposure Rate
Scatter exposure rates for all techniques ranged from 51.7 to 1.22 µGy/sec (5.90.13 mR/sec) at distances of 10100 cm from the scanning plane. The scatter exposure rate at 10 cm from the scanning plane was substantially lower with 80 kVp and 75 mA, as opposed to 120 kVp and 50 mA (Fig 3). With the same CT fluoroscopic scanning time, use of the lower-exposure technique would correspond to a 60% reduction in personnel hand exposure.
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
During 80 seconds of CT fluoroscopic scanning, the dose is substantially reduced with the use of lower-exposure CT fluoroscopic techniques. For example, surface doses at 120 kVp and 50 mA and at 80 kVp and 135 mA are, respectively, 52% and 44% of the surface dose at 120 kVp and 90 mA. Eighty seconds was used, since this is the maximum duration of CT fluoroscopic exposure before the timer on the system resets. It was also approximately the mean scanning time of the CT fluoroscopyguided abdominal biopsies that were performed in patients over a 6-month period (6). The estimated skin dose was a maximum, since we assumed that continuous CT fluoroscopic scanning was used in the same section location. In clinical procedures, this may not always occur.
The radiation dose from CT scanning is high when compared with the dose from other radiographic exposures (10,17). The typical CTDI estimated for an abdominal CT examination at our institution is 31.4 mGy. This is comparable to the range of doses reported in the literature (10, 1719,20). Actual patient doses may vary considerably because of variations in patient size, scanner geometry (source collimation, scanning field diameter), and scanning parameters (tube potential, tube current, scanning time).
With the highest-exposure technique considered in our study (120 kVp, 90 mA, 10-mm section), the dose would be approximately 830 mGy. This skin dose is comparable to doses at approximately 2030 minutes of fluoroscopy during cardiac catheterization or angiographic procedures at our institution and to skin doses at fluoroscopy, as others report (21,22).
Radiation-induced deterministic effects in the patient's skin, such as erythema, may occur at certain dose thresholds during CT fluoroscopy. We hope that dose rate estimates provide information to physicians for the modification of the scanning technique to reduce exposure. In Table 2, the CT fluoroscopic scanning times required to reach dose thresholds of 1,000 and 2,000 mGy with various techniques are given. Transient erythema may occur at a threshold skin dose of approximately 2,000 mGy, but it has been observed at 1,000 mGy (7,8). Thus, these thresholds were used in the development of a dose management plan to reduce patient and personnel exposure during CT fluoroscopic procedures.
|
Our dose management plan requires technologists to notify the physician performing the procedure and other staff when the cumulative CT fluoroscopic scanning time reaches level 1. Subsequently, technologists are required to inform the physician when the scanning time is approaching level 2. If level 2 is exceeded, the rationale for the long exposure time is investigated, and the medical physicist is consulted to provide an estimate of patient skin dose. In addition, we recommend that the patient be examined for radiation-induced skin changes after completion of any procedure when the scanning time exceeds level 2. The ideal intervals for follow-up are at 24 hours and at 12 weeks after the procedure. The intent of this plan is to provide the physician with additional information that may assist him or her in the reduction of unnecessary exposure.
In addition to high patient exposures, the exposure to personnel can be substantial and comparable to personnel exposures at fluoroscopy during cardiac catheterization and interventional radiology procedures (2326). Although the physician's hands will most likely receive the highest levels of exposure, the thyroid and lens of the eye are also at risk.
Scatter exposure to personnel during CT fluoroscopic procedures was estimated from the exposure rate given by the equation, Er = (E/t)f, and the scanning time, tf, for a particular CT fluoroscopic scanning technique (6). With a procedure consisting of an 80-second scanning time with 120 kVp, 50 mA, and a 10-mm section thickness, our results predict a maximum dose of 2.2 mGy to the physician's hand (10 cm from the scanning plane) with a potential head and/or neck dose of 0.1 mGy (1 m from the scanning plane). If the physician's workload were 10 procedures per month, the corresponding monthly hand dose to the physician would be 22 mGy. This assumes that the physician's hands are located approximately 10 cm from the scanning plane for the entire scanning time. The hand doses estimated from a single ring badge exposure for one of our radiologists was 17.7 mGy for 1 month and 16.7 mGy for a second consecutive month. The radiologist performed approximately 10 procedures during each month. Each of these ring badge exposures is consistent with the hand dose predicted from our measurements.
Kato et al (4) demonstrated that the use of needle holders would allow physicians to perform substantially more procedures without exceeding the maximum permissible dose, as compared with not using needle holders. Our surface dose rate at 80 kVp (3.1 mGy per 100 mAs) compares reasonably well with that of Kato et al (3.8 mGy per 100 mAs). Thus, we would expect similar patient doses and personnel exposures with use of the same clinical procedures and fluoroscopic scanning times.
A decrease in the patient dose during CT fluoroscopy will decrease scatter radiation exposure, as is true with other diagnostic radiologic procedures. The patient dose can be reduced by decreasing the CT technique factors and/or CT fluoroscopic scanning time. The choice of technique will depend to large degree on patient habitus and type of procedure. The use of a lower tube potential and tube current could yield reductions of 60% or greater in personnel hand exposure, while a reduction in the section thickness from 10 to 5 or 2 mm can result in personnel exposure reductions of 50%80%. A decrease in the tube potential, tube current, or section thickness, however, may compromise image quality. The physician should choose the technique with the lowest exposure that provides sufficient visualization of the target lesion.
Our results also indicate that positioning a lead drape adjacent to the scanning plane is an effective method for the reduction of scattered radiation during CT fluoroscopy. It would allow sufficient room for needle placement and provide considerable protection to personnel. The greatest effect of the lead drape on the reduction of scatter exposure is close to the patient surface. At greater distances from the patient, the reduction due to inverse square law most likely dominates, and the lead drape becomes less advantageous.
Substantial reduction in personnel hand exposure is possible when a lead drape is placed over the patient. If the drape provides an overall reduction in scattered radiation of 71%, the physician's hand exposure (at a distance of 10 cm from the scanning plane) would be reduced from 2.2 to 0.64 mGy, with the assumption of 80 seconds of scanning time with the use of procedures performed with 120 kVp, 50 mA, and a 10-mm section thickness. The physician could then be allowed to perform substantially more procedures per year without exceeding the maximum permissible dose.
In conclusion, we determined that there is a potential for high exposures to occur during CT fluoroscopy; personnel hand exposure may be reduced by positioning a lead drape caudal to the scanning plane at the needle entry site. In addition, patient dose and personnel exposure can be controlled by using the lowest CT fluoroscopic scanning techniques that are adequate to image the lesion and surrounding tissue structures and by limiting CT fluoroscopy scanning time to only that which is necessary.
The reduction in personnel exposure may be substantial for physicians who perform multiple procedures. Physicians and other users of this technology must develop an awareness of the factors that influence these exposures; they should use standard radiation protection principles regarding time, distance, and shielding, when appropriate. Assessment of radiation dose will provide physicians with useful information for the development of standard imaging protocols as the use of this technology becomes more prevalent.
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
Abbreviations: CTDI = computed tomography dose index, NIST = National Institute of Standards and Technology, TLD = thermoluminescent dosimeter
Author contributions: Guarantor of integrity of entire study, R.D.N.; study concepts and design, all authors; definition of intellectual content, all authors; literature research, R.D.N., P.F.J., S.G.S., K.T.; experimental studies, R.D.N., S.H.; data acquisition, R.D.N., S.H.; data analysis, R.D.N., P.F.J., S.G.S., K.T.; manuscript preparation, R.D.N., P.F.J., S.G.S., K.T.; manuscript editing, R.D.N., P.F.J., S.G.S.; manuscript review, all authors.
| REFERENCES |
|---|
|
|
|---|
Related Article
This article has been cited by other articles:
![]() |
J. Fritz, C. Thomas, S. Clasen, C. D. Claussen, J. S. Lewin, and P. L. Pereira Freehand Real-Time MRI-Guided Lumbar Spinal Injection Procedures at 1.5 T: Feasibility, Accuracy, and Safety Am. J. Roentgenol., April 1, 2009; 192(4): W161 - W167. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. A. Tsalafoutas, V. Tsapaki, C. Triantopoulou, C. Pouli, V. Kouridou, I. Fagadaki, and J. Papailiou Comparison of Measured and Calculated Skin Doses in CT-Guided Interventional Procedures Am. J. Roentgenol., November 1, 2008; 191(5): 1601 - 1607. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hohl, C. Suess, J. E. Wildberger, D. Honnef, M. Das, G. Muhlenbruch, A. Schaller, R. W. Gunther, and A. H. Mahnken Dose Reduction during CT Fluoroscopy: Phantom Study of Angular Beam Modulation Radiology, February 1, 2008; 246(2): 519 - 525. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Das, F. Sauer, U. J. Schoepf, A. Khamene, S. K. Vogt, S. Schaller, R. Kikinis, E. vanSonnenberg, and S. G. Silverman Augmented Reality Visualization for CT-guided Interventions: System Description, Feasibility, and Initial Evaluation in an Abdominal Phantom Radiology, July 1, 2006; 240(1): 230 - 235. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Carlson, J. P. Felmlee, C. E. Bender, R. L. Ehman, K. L. Classic, T. L. Hoskin, W. S. Harmsen, and H. H. Hu CT Fluoroscopy-guided Biopsy of the Lung or Upper Abdomen with a Breath-hold Monitoring and Feedback System: A Prospective Randomized Controlled Clinical Trial Radiology, November 1, 2005; 237(2): 701 - 708. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Aviles Lucas, D. R. Dance, I. A. Castellano, and E. Vano Estimation of the peak entrance surface air kerma for patients undergoing computed tomography-guided procedures Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 317 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Meleka, A. Patra, E. Minkoff, and K. Murphy Value of CT Fluoroscopy for Lumbar Facet Blocks AJNR Am. J. Neuroradiol., May 1, 2005; 26(5): 1001 - 1003. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Wagner Selective Lumbar Nerve Root Blocks with CT Fluoroscopic Guidance: Technique, Results, Procedure Time, and Radiation Dose AJNR Am. J. Neuroradiol., October 1, 2004; 25(9): 1592 - 1594. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Nawfel, P. F. Judy, A. R. Schleipman, and S. G. Silverman Patient Radiation Dose at CT Urography and Conventional Urography Radiology, July 1, 2004; 232(1): 126 - 132. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Maher, D. A. Gervais, M. K. Kalra, B. Lucey, D. V. Sahani, R. Arellano, P. F. Hahn, and P. R. Mueller The Inaccessible or Undrainable Abscess: How to Drain It RadioGraphics, May 1, 2004; 24(3): 717 - 735. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Carlson, J. P. Felmlee, C. E. Bender, R. L. Ehman, K. L. Classic, H. H. Hu, and T. L. Hoskin Intermittent-Mode CT Fluoroscopy-guided Biopsy of the Lung or Upper Abdomen with Breath-hold Monitoring and Feedback: System Development and Feasibility Radiology, December 1, 2003; 229(3): 906 - 912. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Binkert, F. R. Verdun, M. Zanetti, C. W. Pfirrmann, and J. Hodler CT Arthrography of the Glenohumeral Joint: CT Fluoroscopy Versus Conventional CT and Fluoroscopy--Comparison of Image-Guidance Techniques Radiology, October 1, 2003; 229(1): 153 - 158. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Sakarya, O. Unal, B. Ozbay, K. Uzun, I. Kati, S. Ozen, and O. Etlik MR Fluoroscopy-guided Transthoracic Fine-Needle Aspiration Biopsy: Feasibility Radiology, August 1, 2003; 228(2): 589 - 592. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. B. Solomon, A. Patriciu, M. E. Bohlman, L. R. Kavoussi, and D. Stoianovici Robotically Driven Interventions: A Method of Using CT Fluoroscopy without Radiation Exposure to the Physician Radiology, October 1, 2002; 225(1): 277 - 282. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. K. Paulson, D. H. Sheafor, D. S. Enterline, H. P. McAdams, and T. T. Yoshizumi CT Fluoroscopy-guided Interventional Procedures: Techniques and Radiation Dose to Radiologists Radiology, July 1, 2001; 220(1): 161 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. Wagner CT Fluoroscopy: Another Advancement with Additional Challenges in Radiation Management Radiology, July 1, 2000; 216(1): 9 - 10. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |