Vascular and Interventional Radiology |
1 From the Department of Radiology, Division of Vascular and Interventional Radiology, Los Angeles CountyUniversity of Southern California Medical Center, Los Angeles, Calif. Received July 16, 1999; revision requested August 25; revision received November 22; accepted December 7. Address correspondence to M.D.K., Department of Radiology, University of Southern California Medical CenterUniversity Hospital, 1500 San Pablo St, Los Angeles, CA 90033 (e-mail: mkatz@hsc.usc.edu).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: One hundred nineteen patients were prospectively enrolled in this study. CO2 cavograms were obtained and evaluated for IVC diameter, location of renal veins, and presence of thrombus and venous anomalies. If CO2 cavography was judged to be adequate, an IVC filter was deployed. After filter placement, cavography was performed with iodinated contrast material; these images were compared with the CO2 cavograms.
RESULTS: Two patients experienced mild side effects related to venous CO2 injection. Comparison of cavograms obtained with CO2 and iodinated contrastenhanced material showed the caval size to be within 3 mm in all 119 patients. In 116 patients (97.5%), CO2 cavography was judged to be adequate, and in 115 patients, filters were placed. In three (2.5%) patients, it was necessary to perform iodinated contrastenhanced cavography before filter deployment. All six cases of venous anomaly and 11 (78.6%) of 14 cases of thrombosis were clearly identified with CO2 cavography. One filter was maldeployed owing to misinterpretation of the CO2 cavogram.
CONCLUSION: CO2 cavography is well tolerated, safe, and adequate for identification of the parameters necessary for filter deployment. It is especially valuable in patients with a history of reaction to iodinated contrast material or renal insufficiency.
Index terms: Angiography, contrast media, 982.1249 Carbon dioxide Contrast media, comparative studies, 982.1249 Venae cavae, filters, 982.1267
| INTRODUCTION |
|---|
|
|
|---|
Although vena cavography has traditionally been performed with use of iodinated contrast material, carbon dioxide (CO2) has been substituted as an intravascular contrast material in patients who are allergic to iodinated contrast material and in those with renal insufficiency (47). The use of CO2 has been described in both the arterial and the venous systems. Venographic examples include performing CO2 wedged hepatic portography to guide and follow up transjugular intrahepatic portosystemic shunt placement (8), venography of subclavian and brachiocephalic veins for venous occlusion (4) or for venous catheter placement (9), and fistulography of dialysis grafts (10). With specific reference to vena cavography, Sullivan et al (4) reported that IVC filters were successfully deployed by using CO2 alone in seven of nine patients; however, it was unclear whether CO2 cavography enabled an accurate assessment of IVC diameter.
The purpose of this study was to determine the safety and adequacy of CO2 as a contrast material for inferior vena cavography before IVC filter deployment. We also performed vena cavography with iodinated contrast material to provide a standard of reference for caval size and the presence of anomalies or thrombus.
| MATERIALS AND METHODS |
|---|
|
|
|---|
1.6 mg/dL [141.4 µmol/L]), and (e) had a history of major reaction to intravascular iodinated contrast material. We obtained approval for the study from our institutional review board and written informed consent from all patients. A total of 119 patients (65 male, 54 female; mean age, 48.7 years; age range, 1789 years) were prospectively included in this study. One hundred nineteen (38%) of 316 IVC filters placed at our institution during the 33-month study period were placed in these patients. The two main indications for filter placement were contraindication to anticoagulation in patients with deep venous thrombosis or pulmonary emboli (n = 57 [47.9%]) and prophylactic treatment of patients deemed to be at high risk for the development of deep venous thrombosis (n = 38 [31.9%]). Other indications for filter placement included the failure of conventional anticoagulation agentsthat is, the progression of deep venous thrombosis or recurrent pulmonary emboli (n = 10 [8.4%]), complications from anticoagulatives (n = 9 [7.6%]), and patient noncompliance while receiving anticoagulant regimens (n = 5 [4.2%]).
Standard venous access was obtained through either a common femoral vein or internal jugular vein by using the Seldinger technique. After a 5-F pigtail or similar catheter was positioned in the IVC just superior to the iliac veins, posteroanterior vena cavography was performed after hand injection of approximately 60 mL of CO2 per angiographic run. Medical-grade (99.99% pure) CO2 (Altair Medical Gases, San Ramon, Calif) was used. A 60-mL syringe was purged of room air three times to prevent air contamination and was then allowed to fill with CO2 at low pressure. Rapid injection was then performed after the angiographic catheter was primed with CO2.
Images were obtained with digital subtraction angiography (DSA) at a rate of three to six images per second. If needed, multiple runs were performed to obtain a cavogram that was adequate for IVC filter placement. Occasionally, the patient was placed in the ipsilateral anterior oblique position to help identify renal venous inflow. When needed, additional tube angulation was used to displace bowel gas with respect to the IVC. When CO2 cavography was deemed adequate, radiographic landmarks were noted and the caval filter was deployed (Fig 1). If CO2 cavography was considered to be inadequate for safe filter deployment, iodinated contrast materialenhanced cavography was performed. The filter was then placed, and the study was recorded as a CO2 cavography failure.
|
|
Routine noninvasive monitoring of the patients vital signs included electrocardiographic tracing, intermittent blood pressure monitoring, and pulse oximetry. Sedation and/or anesthesia was achieved with a local anesthetic at the entry site, with the addition of conscious sedation in most instances. The patient signs and symptoms related specifically to CO2 cavography were recorded. Those sought included nausea, vomiting, pain, dyspnea, light-headedness, hypotension or hypertension, and dysrhythmias.
The images were evaluated for the location and type of venous anomaly, presence of IVC thrombus, maximum IVC diameter with both contrast media, and complications related to filter deployment. The images were reviewed by two interventional radiology attending staff members (M.D.K., C.C.J.) together by means of consensus.
Statistical analysis was performed by using the paired, two-tailed Student t test and an array correlation coefficient.
| RESULTS |
|---|
|
|
|---|
One hundred eighteen caval filters were placed in the 119 patients. A Vena Tech or LGM filter (B. Braun Medical, Bethlehem, Pa) was placed in 58 patients (49.2%), and a titanium Greenfield filter (Boston Scientific/Meditech, Natick, Mass) was placed in 56 (47.5%). In two patients (1.7%), stainless steel "over-the-wire" Greenfield filters (Boston Scientific/Meditech), which became available toward the end of our study period, were placed. In two patients (1.7%), Birds Nest filters (Cook) were placed because the CO2 IVC diameter was greater than or equal to 28 mm. In one patient, an 89-year-old man, CO2 and iodinated contrast cavographic studies revealed IVC clot extending to near the right atrium, and an IVC filter was not placed.
The IVC diameter measurements obtained with CO2 and iodinated contrastenhanced cavography are detailed in Table 1; a histogram of the differences is presented in Figure 2. Overall, the size of the IVC was underestimated with CO2 cavography compared with iodinated contrastenhanced cavography, with an average difference of 0.27 mm. When analyzed by using a paired, two-tailed Student t test, this difference in diameter measurement was statistically significant (P
.05). In 43 patients (36.1%), there was no difference in IVC diameters obtained with CO2 and iodinated contrastenhanced cavography. In 86 patients (72.3%), the diameters obtained with CO2 and iodinated contrastenhanced cavography were within 1 mm of each other. In three patients, CO2 cavography demonstrated an IVC diameter of 28 mm or less, whereas iodinated contrastenhanced cavography demonstrated an IVC diameter of more than 28 mm. In two of these three patients, an LGM or titanium Greenfield filter was placed and no clinical sequelae were noted.
|
|
|
|
Cavography with CO2 was prospectively judged to be a failure in three patients (2.5%): In two patients (1.7%), CO2 cavography inadequately demonstrated the renal vein anatomy, and in one (0.8%), iodinated contrastenhanced cavography was necessary to further define a small suprarenal IVC. In all three patients, an IVC filter was safely placed after iodinated contrastenhanced cavography was performed.
Overall, the average number of CO2 cavography DSA runs performed was 2.3 per patient (mean, two; range, one to eight). The maximum amount of CO2 administered to any one individual in this study group was approximately 500 mL.
When the two cavographic examinations were compared for overall quality and information, the CO2 cavograms were judged to be equal to the iodinated contrastenhanced cavograms in 84 patients (70.6%). In 24 patients (20.2%), the iodinated contrastenhanced cavogram was judged to be superior, and in 11 patients (9.2%), the CO2 study was judged to superior.
A complication of filter deployment occurred in one patient (0.8%). A titanium Greenfield filter originally placed in the immediate infrarenal IVC after the CO2 study was found to have one leg protruding into the right renal vein at iodinated contrastenhanced cavography. Because this renal vein was not clearly identified with CO2 cavography alone, the filter positioning was believed to represent a failure of the CO2 study.
Finally, there was a surprising lack of side effects related to CO2 injection among our study patients. One patient had nausea with both CO2 and iodinated contrast material, but it resolved spontaneously. One patient vomited several minutes after the administration of CO2. This also resolved without further treatment, and it is unclear whether this episode was truly related to the CO2 or whether it was due to the recent intravenous administration of meperidine hydrochloride. No episodes of dyspnea, light-headedness, induced hypotension or hypertension, or cardiac dysrhythmias were observed.
| DISCUSSION |
|---|
|
|
|---|
Because CO2 is a normal constituent of the human body, there is no possibility of an allergic reaction (7). Although the use of CO2 in the arterial system has been deemed safe (4,1924), more strict attention has been recommended for venous CO2 injection (7). Concerns regarding CO2 flooding of the right ventricle necessitate careful administration with respect to the injected volume of CO2 and the interval between CO2 injections. Kerns et al (7) performed vena cavography in humans using up to 70 mL of CO2 per injection and reported that it was well tolerated, with dissolution or elimination of CO2 in a matter of seconds. A more recent animal study revealed that although there was an immediate and reversible increase in the pulmonary arterial pressure after CO2 cavography, diagnostic doses of CO2 had no substantial effect on cardiopulmonary function regardless of patient positioning (25). In addition, Bendib et al (26) reported 1,600 cases without major complication in which 200 mL of CO2 was injected intravenously for the detection of pericardial effusion. We agree with the recommendation of Kerns et al (7) that no more than 200 mL should be administered in any one injection and suggest the use of substantially less in the case of vena cavography. Under usual conditions, diagnostic images are obtained with 60 mL or less of CO2.
Our experience suggests that CO2 cavography is a safe procedure when performed appropriately. Episodes of transient tingling, abdominal pain, and bowel ischemia and infarction have been reported with CO2 use in the arterial system (7,27), and two fatalities have been reported (28,29), although causality was unproved, though suggestive, in both cases (27). As to venous injections, few adverse reports have been described in the literature. We noted two mild reactions (nausea and/or vomiting) in our series, one of which occurred with iodinated contrast material administration as well. The other was mild and self-limited and may have been caused by concomitantly administered narcotic medications.
In this series, CO2 cavography adequately depicted all anatomic parameters necessary for IVC filter placement in 94% of patients. All of the anatomic caval anomalies were identified with CO2 imaging.
IVC thrombus was clearly delineated with CO2 cavography in 11 of 14 patients. The three instances of incomplete thrombus delineation with CO2 cavography occurred early in our study. Although two of these cases were clearly related to an incomplete IVC CO2 study, the filter positioning and approach was correct. The third was not clearly a failure of the CO2 study, because the initial thrombus was found and filter positioning appeared to be appropriate.
Overall, there were seven (5.9%) CO2 cavography failures (Table 2). This number could likely have been minimized with stricter attention to careful positioning of the catheter and patient. Specific attention to the IVC just superior to the iliac veins is recommended with CO2 cavography because incomplete evaluation of this region led to nearly one-third of our cavography failures. Patients with incomplete IVC evaluation at CO2 cavography should undergo iodinated contrastenhanced cavography or selective venous imaging.
|
Finally, the overall quality of the CO2 cavograms in our study was rated as good to excellent, with nearly 80% of CO2 cavograms having a diagnostic quality equal or superior to that of iodinated contrastenhanced cavograms. Essentially all of the remaining CO2 cavograms, even though judged to be of lesser quality, were believed to contain sufficient diagnostic information to allow safe caval filter placement.
Overall, we believe that CO2 is a safe contrast material for radiography of the vena cava, is well tolerated, and provides sufficient diagnostic information for vena cavography and for guiding IVC filter placement. Although appropriate for use in almost any patient, CO2 appears to be clearly advantageous in those with a history of iodinated contrast material reaction or renal insufficiency. Because IVC diameter tends to be slightly underestimated with CO2 cavography, caution is recommended in filter selection when the diameter of the IVC is greater than 25 mm at CO2 cavography.
| FOOTNOTES |
|---|
Author contributions: Guarantor of integrity of entire study, C.L.D.; study concepts, M.D.K.; study design, G.L.P., S.E.H.; definition of intellectual content, M.D.K.; literature research, C.L.D.; clinical studies, M.D.K.; experimental studies, D.S.H.; data acquisition, Y.H.P.; data and statistical analyses, C.L.D., C.C.J.; manuscript preparation, C.L.D.; manuscript editing, C.C.J.; manuscript review, M.D.K.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S K Morcos Acute serious and fatal reactions to contrast media: our current understanding Br. J. Radiol., August 1, 2005; 78(932): 686 - 693. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Streiff Vena Caval Filters: A Review for Intensive Care Specialists J Intensive Care Med, March 1, 2003; 18(2): 59 - 79. [Abstract] [PDF] |
||||
![]() |
P. Girard, J.-B. Stern, and F. Parent Medical Literature and Vena Cava Filters* : So Far So Weak Chest, September 1, 2002; 122(3): 963 - 967. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||