Technical Developments |
1 From the Departments of Radiology and Biomedical Engineering, (J.M.S., E.A., P.K., H.H.Q., X.Y.), Biomedical Engineering (J.D.), and Division of Cardiology (K.A.S., A.W.H.), Johns Hopkins University School of Medicine, Outpatient Center, Room 4243, 601 North Caroline St, Baltimore, MD 21287-0845. Received May 26, 1999; revision requested July 19; final revision received December 15; accepted January 17, 2000. Supported in part by a Lavoisier Program grant (J.M.S.), a Société Francaise de Radiologie grant (J.M.S.), National Institutes of Health RO1HL61672 grant (E.A.), and National Institutes of Health R29HL57483 grant (E.A.). Address correspondence to X.Y. (e-mail: xyang@mri.jhu.edu).
| ABSTRACT |
|---|
|
|
|---|
Index terms: Angiography, contrast media, 50.121412, 50.12142, 54.121412, 54.12142 Gadolinium, 50.121412, 50.12142, 54.121412, 54.12142 Magnetic resonance (MR), contrast enhancement, 50.121412, 50.12142, 54.121412, 54.12142 Magnetic resonance (MR), guidance Magnetic resonance (MR), vascular studies, 50.121412, 50.12142, 54.121412, 54.12142
| INTRODUCTION |
|---|
|
|
|---|
The potential to perform accurate two-dimensional and three-dimensional angiography (4), obtain high-spatial-resolution MR images with excellent soft-tissue contrast (5,6), and track interventional devices (79) make magnetic resonance (MR) imaging a potential alternative. By allowing the imaging of plaque morphology and composition, as well as the assessment of related organ function during a single vascular intervention, MR imaging may modify the management of vascular interventions and improve the clinical outcome for patients with atherosclerosis.
However, conventional MR angiography sequences (time-of-flight angiography, phase-contrast angiography, and contrast mediumenhanced angiography), when used for the guidance of vascular intervention devices through the arterial tree (1012), have general limitations. First, acquisition times are longer than 1 second, which precludes obtaining information about the local hemodynamic condition. Second, long postprocessing analysis is required to generate road maps; and third, contrast-enhanced angiography necessitates the use of large amounts of contrast medium, which precludes the possibility of multiple injections because of increased background signal intensity. Recently, some authors have reported that two-dimensional projection techniques coupled with intravenous injections of gadopentetate dimeglumine can depict arterial flow in large arteries (13,14).
The goal of this study was to evaluate whether the combination of selective intraarterial injections of small doses of gadopentetate dimeglumine with a fast two-dimensional gradient-echo sequence could generate sufficient information to guide vascular interventions and allow analysis of local hemodynamic conditions. A two-dimensional projection technique with no section selection was used to avoid problems relative to arterial localization and to enable imaging of a tortuous vessel in a single image.
| MATERIALS AND METHODS |
|---|
|
|
|---|
|
For all these experiments, imaging parameters were used that were identical to those described in the preliminary study. However, a projection technique, disabling the section-selective gradients, was used. All animals were treated according to the principles of laboratory animal care of the National Society for Medical Research and the National Institutes of Health (15). The experimental protocol was approved by the animal care and use committee and by the institutional review board at Johns Hopkins University. Oral informed consent was obtained from the volunteer.
Human volunteer experiment.A 2-cm-diameter tube was taped on the chest of a 30-year-old male volunteer (weight, 80 kg; height, 1 m 80 cm). A roller flow pump provided a pulsatile flow compatible with the human pulse rate (70 beats per minute) and aortic flow (3 L/min) (Fig 2). A 60% stenosis was created by tightening a string around the tube. Since the obtained images were essentially projection images, the tube on the chest of the body was expected to generate a signal intensity similar to that of a human aorta. To simulate an intraarterial injection, a 4-F introducer was connected to the descending flow tube, upstream from the stenosis. The body coil was used for imaging. Injection of 10 mL of gadopentetate dimeglumine through the introducer was performed simultaneously to obtain images at variable concentrations of gadopentetate dimeglumine: 31.25, 62.50, 125.00, 250.00, and 500.00 mmol/L. The signal intensity of the vessel was measured by a radiologist (J.M.S.) for each injection, and the concentration that provided the best vascular signal intensity compared with that for adjacent tissues was recorded.
|
The rabbit was then placed in a supine position in the imager and aligned with the main magnetic field. An extremity coil was used to obtain the images. Coronal projection images were obtained first to determine the anatomic position of the rabbit. Transverse and sagittal scout images were obtained to prescribe the angulation of the projection. The next step was to inject contrast medium intravenously to obtain a road-map image and allow positioning of a catheter in the aorta; 1.5 mL of gadopentetate dimeglumine (500.00 mmol/L) was injected into the ear vein of the rabbit. There were 60 images obtained in 17 seconds.
An MR-compatible 3-F angiographic catheter constructed in-house was then sent through the introducer up to the ascending aorta to allow intraarterial injections of gadopentetate dimeglumine. Injections of 1 mL of gadopentetate dimeglumine with variable concentrations were administered through the catheter, followed by 3 mL of an iso-osmolar saline solution flush. We tested concentrations of 31.25, 62.50, 125.00, 250.00, and 500.00 mmol/L. The signal intensity of the vessel was measured for each injection by a radiologist (J.M.S.), and the concentration that provided the best vascular signal intensity compared with that for adjacent tissues was recorded. By using the optimal gadopentetate dimeglumine concentration, images with a 24- and a 12-cm field of view were obtained to determine whether higher spatial resolution angiography could be achieved.
Canine coronary experiment.Two healthy 30-kg mongrel dogs (Biomedical, Friedensburg, Pa) were studied for this experiment and killed at the end of the experiment. Anesthesia was induced by administering 25 mg/kg of pentobarbital sodium (Sodium Nembutal, Veterinary Laboratories) intravenously and was maintained by means of general inhalational anesthesia (1%2% isoflurane; Isoflurane; Abbott Laboratories, North Chicago, Ill). A 9-F introducer sheath (Boston Scientific) was placed through a right carotid arterial cutdown. The left main coronary artery was engaged with a 7-F magnetically compatible catheter (Cordis Endovascular, a Johnson & Johnson Company, Miami Lakes, Fla) with radiographic guidance.
After each dog was transported to the magnet, a cardiac phased-array coil (GE Medical Systems) was placed around its chest. The previously determined rabbit protocol was used, but with a smaller field of view (20 and 16 cm), with no intravenous injection, and with 4 mL rather than 1 mL of gadopentetate dimeglumine injected for each acquisition because of the larger diameter of the angiographic catheter. The signal intensity of the vessel was measured for each injection by a radiologist (J.M.S.). The concentration that provided the best vascular signal intensity compared with that for adjacent tissues was recorded. Multiple projection images were obtained, which included standard right anterior oblique images and left anterior oblique images with or without cranial or caudal angulation to assess the left coronary artery.
| RESULTS |
|---|
|
|
|---|
Rabbit Experiment
The intravenous injection allowed the acquisition of dynamic angiographic images of the aorta and its branches (Fig 3). Enhancement was seen first in the pulmonary artery, then in the heart chambers on the left side and the aorta. Body signal intensity, including that of fat and periarterial tissues, was suppressed completely on all images. The rate of 3.5 frames per second showed arterial filling and washout, as well as the venous contrast phase, visible in the portal vein. Intraarterial injections in the ascending aorta, with use of gadopentetate dimeglumine concentrations between 31.25 and 125.00 mmol/L, displayed arterial phases in the aorta. Carotid and renal vessels, which are as small as human coronary vessels, were depicted clearly. Concentrations of 250.00 and 500.00 mmol/L did not enhance the aorta. The best concentration was 62.50 mmol/L because it allowed higher spatial resolution angiography with 24- and 16-cm fields of view; however, arterial signal intensity decreased. Overlapped tissues in acquisitions with a small field of view were not a problem, as their signal intensity was below the noise level. No substantial enhancement due to gadopentetate dimeglumine accumulation in background tissues was noticed during the experiment, which lasted approximately 1 hour.
|
|
|
| DISCUSSION |
|---|
|
|
|---|
The technique we present in this article was inspired by contrast-enhanced MR angiography; however, the existing technology is applied in a different manner. First, rather than obtaining three-dimensional high-spatial-resolution images in a few seconds, our technique allows the acquisition of a single-projection image in 300 msec. This method is similar to that of x-ray fluoroscopy. Second, instead of injecting large doses and concentrations of gadopentetate dimeglumine into veins, the optimal concentration of gadopentetate dimeglumine is attained within the artery by injecting low doses of low-concentration contrast medium directly into the arteries.
Figures 25 show the potential of such an approach. Arterial filling and washout are clearly demonstrated. The anatomy of the left anterior descending coronary artery and the circumflex coronary artery of the dog is well displayed, with a spatial resolution per pixel of 0.8 x 1.5 mm. No postprocessing data viewing technique is needed. In addition, the technique provides easy acquisition of different projection images. Figure 2b and 2e illustrates this advantage by showing a lateral projection image that allows easy depiction of a 60% stenosis, which was not easy to quantify on the coronal image.
Another advantage of this method is the use of a low dose of gadopentetate dimeglumine injected per angiographic examination, which avoids background tissue enhancement, even if it is used multiple times during an interventional procedure. Figure 1 shows that when gadopentetate dimeglumine concentration is lower than 1 mmol/L, the signal intensity is low. Thus, as long as gadopentetate dimeglumine concentration in the background tissues is lower than 1 mmol/L, a vessel filled with a gadopentetate dimeglumine concentration of 15.6062.50 mmol/L will be enhanced clearly (Fig 1). If we approximate the distribution volume of gadopentetate dimeglumine to 12 L in vascular and interstitial compartments (22) to maintain a concentration lower than 1 mmol/L, the radiologist is limited to a maximum injection of 10 mmol of gadopentetate dimeglumine in the body. In our experiment, angiography of a coronary artery necessitated the injection of 0.1 mmol of gadopentetate dimeglumine (3 mL of gadopentetate dimeglumine with a concentration of 31.25 mmol/L). It would have been possible, therefore, without even considering gadopentetate dimeglumine elimination by the kidneys, to perform 120 injections of gadopentetate dimeglumine without substantial background enhancement.
After dilution in the body, the dose is equivalent to one 500th of the standard gadopentetate dimeglumine concentration (0.125 mmol/L compared with 500.00 mmol/L) which, as shown in Figure 1, gives a negligible signal intensity. This small dose allows (a) multiple successive angiographic examinations and (b) assessment of small fields of view because overlapped tissues are always suppressed efficiently (Fig 3).
For x-ray angiography, the more concentrated the contrast medium injected, the better the arterial enhancement will be, whatever the contrast medium dilution after injection into the vessel. For MR angiography, high concentrations lead to susceptibility artifacts if dilution of contrast medium in the vessel is insufficient, which would preclude arterial enhancement (Fig 1). Thus, to calculate the appropriate gadopentetate dimeglumine concentration needed in the syringe to enhance the vessel, the radiologist needs to estimate the dilution level of gadopentetate dimeglumine after injection into the bloodstream. In theory, the dilution level depends on the ratio between blood flow in the vessel and the injection rate of gadopentetate dimeglumine in milliliters per second. As vessel flow is difficult to evaluate, such calculations were not considered in this study.
To achieve optimal vessel enhancement, a gadopentetate dimeglumine concentration of 31.25 mmol/L was always tested first, followed by more concentrated injections of gadopentetate dimeglumine, until the maximum vessel enhancement was reached. This method was efficient and rapid and could be applied easily in all animals. However, optimal concentration levels500.00 mmol/L for the tube, 62.50 mmol/L for the rabbit aorta, and 31.25 mmol/L for the canine coronary vesselscorrelated with vessel and catheter diameter ratios in all these experiments, which were 1:20, 1:5, and 2:3, respectively.
For future clinical applications, this technique involves several issues. (a) An MR imager with a high-field-strength magnet and fast and high gradients is required, and not all imagers are so equipped. (b) For interventional procedures, an open magnet or short magnet is required to obtain easy access to the patient. Despite dramatic progress in this direction, these open or short magnets currently are not available with high magnetic fields, which limits their actual use in clinics. (c) The spatial and temporal resolution are still lower compared with those attained with x-ray fluoroscopy, which may be insufficient for stenosis quantification in small vessels, such as coronary arteries. (d) For the depiction of coronary arteries, the acquisition time per image (300 msec) is too long and causes motion blurring of the vessels during the systolic phase (Fig 4). Acquisition time will have to be shortened to minimize this blurring. (e) For selective injection of contrast medium, it is necessary to determine the relationship of the catheter to the arterial structures, an issue that was not addressed in this experiment. Although the results of some studies (8,23) have shown the possibility of depicting this relationship with MR imaging guide wires, guiding catheters, or both, this issue will need to be assessed.
Assessing the anatomic and hemodynamic conditions of a vessel is the first step before placing or advancing any intravascular device during an interventional procedure. To date, this has been accomplished with x-ray fluoroscopy, which allows high spatial and temporal resolution, with frequent updating of the data. The technique we present here shows that placement and monitoring of interventional devices also may be possible by using MR imaging. Arterial filling and washout, as well as venous and perfusion phases, were clearly displayed, demonstrating that real-time angiography with use of an MR technique is feasible. Even if 3.5 frames per second represents a low fluoroscopic rate, technologic evolution of gradient hardware and new pulse sequences will provide faster acquisition times in the near future (24,25), which will allow clinicians to take full advantage of the benefits of MR imaging in the field of interventional radiology and cardiology.
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. Arepally, P. V. Karmarkar, C. Weiss, and E. Atalar Percutaneous MR Imaging-guided Transvascular Access of Mesenteric Venous System: Study in Swine Model Radiology, January 1, 2006; 238(1): 113 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Zorger, M. Volk, O. W. Hamer, M. Lenhart, J. Seitz, B. Butz, and C. Paetzel Intraarterial Gadolinium-Enhanced MR Angiography in Humans for the Detection of Infrainguinal Arterial Stenoses Before and After Percutaneous Angioplasty Am. J. Roentgenol., October 1, 2005; 185(4): 867 - 872. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.-C. Schulte, G. Bongartz, R. Huegli, M. Aschwanden, K. A. Jaeger, W. Ostheim-Dzerowycz, A. L. Jacob, and D. Bilecen Intraarterial Versus IV Gadolinium Injections for MR Angiography: Quantitative and Qualitative Assessment of the Infrainguinal Arteries Am. J. Roentgenol., September 1, 2005; 185(3): 735 - 740. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Bilecen, A.-C. Schulte, H. G. Heidecker, M. Aschwanden, R. Huegli, K. A. Jaeger, W. Ostheim-Dzerowycz, and G. Bongartz Lower Extremity: Low-Dose Contrast Agent Intraarterial MR Angiography in Patients--Initial Results Radiology, January 1, 2005; 234(1): 250 - 255. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Hager, H. Kaemmerer, A. Leppert, M. Prokop, S. Blucher, H. Stern, and J. Hess Follow-up of Adults With Coarctation of the Aorta: Comparison of Helical CT and MRI, and Impact on Assessing Diameter Changes Chest, October 1, 2004; 126(4): 1169 - 1176. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. K. Wacker, R. M. Maes, J. A. Jesberger, S. G. Nour, J. L. Duerk, and J. S. Lewin MR Imaging-Guided Vascular Procedures Using CO2 as a Contrast Agent Am. J. Roentgenol., August 1, 2003; 181(2): 485 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Omary, J. D. Green, B. E. Schirf, Y. Li, J. P. Finn, and D. Li Real-Time Magnetic Resonance Imaging-Guided Coronary Catheterization in Swine Circulation, June 3, 2003; 107(21): 2656 - 2659. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Green, R. A. Omary, J. P. Finn, Y.-C. Chung, R. Tang, Y. Li, S. Vasireddy, and D. Li Two- and Three-dimensional MR Coronary Angiography with Intraarterial Injections of Contrast Agent in Dogs: A Feasibility Study Radiology, January 1, 2003; 226(1): 272 - 277. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Lederman, M. A. Guttman, D. C. Peters, R. B. Thompson, J. M. Sorger, A. J. Dick, V. K. Raman, and E. R. McVeigh Catheter-Based Endomyocardial Injection With Real-Time Magnetic Resonance Imaging Circulation, March 19, 2002; 105(11): 1282 - 1284. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Omary, K. P. Henseler, O. Unal, L. J. Maciolek, J. P. Finn, D. Li, A. A. Nemcek Jr., R. L. Vogelzang, and T. M. Grist Comparison of Intraarterial and IV Gadolinium-Enhanced MR Angiography with Digital Subtraction Angiography for the Detection of Renal Artery Stenosis in Pigs Am. J. Roentgenol., January 1, 2002; 178(1): 119 - 123. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. B. Reeder, Y. P. Du, J. A. C. Lima, and D. A. Bluemke Advanced Cardiac MR Imaging of Ischemic Heart Disease RadioGraphics, July 1, 2001; 21(4): 1047 - 1074. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||