|
|
||||||||
Medical Physics |
1 From the Departments of Radiology (H.G., T.A.) and Medical Physics (K.W.B., B.J., J.P.), Örebro Medical Centre Hospital, 701 85 Örebro, Sweden. From the 1999 RSNA scientific assembly. Received February 8, 2000; revision requested April 7; revision received May 25; accepted July 31. Supported by the research committee of Örebro County Council and the Swedish Radiation Protection Institute, project SSI P1141.99. Address correspondence to H.G. (e-mail:hakan.geijer@orebroll.se).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Multiple images were reconstructed into one image at a workstation. A low-dose alternative was to use digital pulsed fluoroscopy. Dose measurements were performed with thermoluminescent dosimeters in an Alderson phantom. At the same time, kerma area-product values were recorded. A Monte Carlo dose calculation also was performed. Image quality was evaluated with a contrast-detail phantom and visual grading system. Angle measurements were evaluated with an angle phantom and measurements obtained on patient images.
RESULTS: The effective radiation dose was 0.087 mSv for screen-film imaging, 0.16 mSv for digital exposure imaging, and 0.017 mSv for digital fluoroscopy; the corresponding kerma area-product values were 0.43, 0.87, and 0.097 Gy · cm2, respectively. The image quality of the digital exposure and screen-film images was about equal at visual grading, whereas fluoroscopy had lower image quality. The angle phantom had lower angle values with digital fluoroscopy, although the difference in measured angles was less than 0.5°. The patient images showed no difference in angles.
CONCLUSION: The described digital scanning method has acceptable image quality and adequate accuracy in angle measurements. The radiation dose required for digital exposure imaging is higher than that required for screen-film imaging, but that required for digital fluoroscopy is much lower.
Index terms: Dosimetry Images, quality, 30.11, 30.1215 Radiations, exposure to patients and personnel Radiography, digital, 30.1215 Spine, curvature, 30.861 Spine, radiography, 30.11, 30.1215
| INTRODUCTION |
|---|
|
|
|---|
In diagnostic radiology, a low radiation dose based on the ALARA (as low as reasonably achievable) principle (2) is important. In scoliosis radiography, this principle is especially important, because patients with scoliosis are young and often subjected to repeated studies owing to long treatment periods. Most evaluations of scoliosis radiography in the literature have included vague estimates of the radiation doses used.
The digital method described by van Eeuwijk et al (3) helps overcome the problem of detector size. Briefly, this examination consists of a multiple-image scan of the entire spine. The images are then reconstructed into one image at a workstation. A variant of this method is grid-controlled pulsed fluoroscopy.
The aim of our study was to investigate the radiation dose, image quality, and accuracy of Cobb angle measurements obtained with the digital scanning method compared with those obtained with the conventional screen-film method.
| MATERIALS AND METHODS |
|---|
|
|
|---|
This study was designed to compare three methods of scoliosis radiographytwo digital techniques and one screen-film method. The two digital methodsdigital exposure imaging and digital fluoroscopyare similar in many ways and were performed with the equipment described in the next section. The screen-film method was evaluated as a reference standard for comparison with the digital methods.
Radiographic Techniques
All the digital imaging (ie, digital exposure and digital fluoroscopy) examinations were performed with a Multi Diagnost 4 (Philips Medical Systems, Eindhoven, the Netherlands) digital image intensifierbased unit with capabilities for digital exposure and grid-controlled pulsed fluoroscopy. Equipment data are shown in Table 1. The equipment was used in the factory setting, and all methods were evaluated the way they were set during installation.
|
|
Lateral views are not routinely used at our institution and thus were not included in this study. However, these views can be obtained in the same way posteroanterior projections are obtained.
All images are exported to a workstation with commercially available software (EasyVision; Philips Medical Systems). The images are merged and reconstructed into one overview image in less than 30 seconds with the aid of pattern recognition (Fig 2). The borders of the individual images cannot be seen on overview images. An overview image has a size of, for example, 512 x 1,3002,000 pixels. With the same software, measurements of Cobb angles, femur height difference, and vertical alignment can be performed easily. Finally, annotated images can be printed out or, as at our institution, exported to a picture archiving and communicating system (PACS) (Sectra and Philips Medical Systems, Linköping, Sweden). All images are viewed on a monitor, and image manipulations, such as gray-scale correction, magnification, and measurement, can be performed at any workstation in the hospital that is connected to the PACS.
|
|
The TLD-loaded phantom was irradiated at three separate sessions for each method; reading and reloading were performed between each session to assess absorbed dose variability. With screen-film and digital exposure imaging, the phantom was examined five times at each irradiation event to increase the radiation exposure levels enough to obtain reliable TLD readings. With digital fluoroscopy, we chose to examine the phantom 20 times at each irradiation event for the same reason. With digital fluoroscopy, a higher pulse frequency (three pulses per second) was used because this was the lowest frequency available.
Kerma area-product values.Kerma area-product values were measured by using a transmission ion chamber connected to an electrometer (Doseguard 100; RTI Electronics, Mölndal, Sweden).
Monte Carlo calculations.With the Monte Carlo method, a computer simulation of the experimental situation was performed. For all three methods, the organ doses and effective dose were calculated with the PCXMC computer program (Finnish Centre for Radiation and Nuclear Safety, Helsinki, Finland) (7). For the two digital methods, a separate calculation had to be performed for each image; this resulted in 37 calculations for digital exposure imaging and 58 calculations for digital fluoroscopy. The results of these separate calculations were added to obtain the results for each method. All Monte Carlo calculations were tailored to be applicable for a 15-year-old patient of the same size as the phantom.
Image Quality
Contrast-detail phantom.The CDRAD 2.0 contrast-detail phantom (Instrumentale Dienst, Nijmegen, the Netherlands) (8) was used to perform a semiobjective evaluation of image quality. This phantom is a 1-cm-thick slab of acrylic with drilled holes of different sizes and depths. The number of holes visible in each column can be drawn as a graph, which gives a depiction of image quality. A numeric value, the image quality figure, also can be calculated. A low image quality figure means better image quality. The phantom was placed between two 7.5-cm-thick acrylic sheets to simulate the scattering conditions of the human body. The phantom was imaged once with each method. Each reconstructed image was read once by three observers (including H.G. and T.A.) independently.
Visual grading.The image quality of the patient images was evaluated by using visual grading (9) in which each image was compared with a reference screen-film image. An image database was formed from 30 screen-film examinations, performed in 30 patients (six male, 24 female; mean age, 13.8 years; age range, 719 years), from 1995 that were sequentially retrieved from the archives; 30 digital exposure images, obtained in 30 patients (six male, 24 female; mean age, 13.9 years; age range, 818 years), from 1998 to 1999; and 10 digital fluoroscopy images, obtained in 10 female patients (mean age, 15.0 years; age range, 1118 years), from 1998 to 1999. Ten of these patients, after giving informed consent, underwent a double examination with both digital exposure imaging and digital fluoroscopy. Because it was impossible to perform imaging in the same patient with all three methods, the study was performed in two steps.
Thirty screen-film and 30 digital exposure images were compared with one reference image, which was chosen as a representative screen-film scoliosis radiograph of good quality. Each image was evaluated separately and independently by the three observers.
The following levels were evaluated: third thoracic vertebra (T3), 10th thoracic vertebra (T10), and third lumbar vertebra (L3). These levels were chosen to focus the observers on smaller parts of the image. On the old screen-film images, the uppermost level was often overexposed, whereas the two lower levels usually were correctly exposed.
The following aspects of image quality were evaluated: contrast level, noise (ie, quantum mottle), sharpness, visualization of vertebral end plates, subjective suitability for judging skeletal abnormalities, and subjective suitability for angle measurements. Furthermore, the overall quality of the image was assessed.
At each level and aspect, image quality was independently scored on a five-level scale as much worse, worse, equal, better, or much better compared with the reference image quality. The images obtained in the 10 patients who underwent double examinations (a total of 20 images) were graded in pairs, with the digital exposure image in the pair used as the reference. The same grading aspects as those described earlier were used.
This scoring system resulted in nearly 4,000 gradings. All digital images were evaluated at a PACS workstation by using a 1,280 x 1,024-pixel gray-scale monitor (Image Systems, Hopkins, Minn).
Angle Measurements
Angle phantom.The accuracies of Cobb angle measurements obtained with the three methods were evaluated with a phantom made of 13.5 cm of acrylic with 2-mm-thick aluminum sheets simulating two vertebrae in the middle (Fig 4). The aluminum sheets could be rotated between images. Thirty different angles between 0° and 53° were randomly set and imaged with all three methods. Each angle was independently measured once by all three observers. The digital images were measured with the standard PACS software; the film images were marked with a pencil and measured with a protractor. All observers used the same protractor during the entire study. All lines were erased before leaving the image.
|
Statistical Methods
The visual grading material was presented descriptively. Agreement between observers was evaluated with
statistic analysis; verbal interpretations of the
values are shown in Table 2. The angle measurements were evaluated with two-way analysis of variance.
|
| RESULTS |
|---|
|
|
|---|
The results of measurements and calculations in the Alderson phantom are shown in Table 3. Variability between irradiation events was low. The large difference in dose between digital exposure imaging and fluoroscopy should be noted. For comparison, the median kerma area-product value for 20 lumbar spine examinations (four to five images each at 7090 kVp) at our institution is 6.6 Gy · cm2.
|
|
statistic analysis results showed moderate to substantial agreement between observers (
, 0.460.76) on the clinical aspects of visualization of vertebral end plates, suitability for judging skeletal abnormalities, and suitability for angle measurements. For image contrast, noise, and sharpness, agreement was lower, with
values down to -0.03 (poor agreement). However, with regard to the aspects with the lowest
values, the observers agreed in 87 of 90 gradings; this showed the limitation of
statistic analysis.
Digital exposure versus reference image.Compared with the overexposed reference image, at the T3 level, the digital exposure images were judged to be better in all aspects except noise, which was graded as somewhat lower. The two lower levels, T10 and L3, were similar in all aspects, with lower grades on the digital exposure images. The overall quality of the digital exposure images was better.
With regard to visualization of vertebral end plates, suitability for judging skeletal abnormalities, and suitability for angle measurements, there was less agreement between observers at digital exposure image grading; agreement ranged from slight to substantial (
, 0.140.65).
Digital exposure versus screen-film images.Compared with the grades at the T3 level on the screen-film images, those on the digital exposure images were better in all aspects except noise. At the T10 and L3 levels, which had similar grades, contrast, noise, and sharpness were graded somewhat lower on the digital exposure images, but visualization of end plates, suitability for judging skeletal abnormalities, and suitability for angle measurements were graded somewhat higher. The overall quality of the digital exposure images was considerably better. Gradings at the T10 level on screen-film and digital exposure images are illustrated in Figure 6. The results at the L3 level were similar.
|
|
|
|
|
= 0.46), and there was complete agreement on noise among the observers, who all chose the same gradeworse. An intelligible
value could not be calculated with this total agreement. With regard to suitability for judging skeletal abnormalities, agreement was moderate to almost perfect (
, 0.461.00). Agreement on visualization of vertebral end plates and suitability for angle measurements was poor.
Angle Measurements
Angle phantom.Two-way analysis of variance revealed significant results for the two factors of method and observer (method: F = 5.8, P = .005; observer: F = 6.4, P = .003; method and observer: F = 2.3, P = .062). For the post-hoc test, a 95% CI of plus or minus 0.38° was calculated. By using this CI, there were significant differences between observers 1 and 3 and between observers 2 and 3 for digital fluoroscopy. Likewise, there were significant differences between screen-film imaging and fluoroscopy and between digital exposure imaging and fluoroscopy for observer 3. Neither of the other comparisons was significant. Observer 3 and digital fluoroscopy yielded smaller angle measurements than did the other two techniques. Mean angle measurements are shown in Table 5 and Figure 7; the differences between angle measurements obtained with the different methods and the mean angle measurement are shown in Figure 8.
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
The advantages of digital exposure imaging are that all images have a consistent exposure along the entire length of the spine without under- or overexposed areas. Angle measurements are easy to perform with dedicated software at the workstation. This method also works for lateral scoliosis radiographs, but great care has to be taken to get acceptable exposure levels. The resultant images can be stored both with and without annotation on a PACS or printed. Use of the large and unwieldy full-size film used in the conventional method is thus avoided.
Some problems are that the examination is somewhat complicated to perform, and experienced technicians are required for good results. Collimation is a constant problem, because any increase in field size in the craniocaudal direction results in a proportional increase in dose. A fixed collimator might be a solution. Images acquired with the digital exposure method have a large overlap, and each point of the patient is exposed by at least two images. According to the manufacturer, this overlap is necessary for accurate reconstruction of the overview image. The fact that the scanning speed is slower than nominal points to the importance of acceptance testing when installing new equipment. A setup outside the stated specifications might result in a higher-than-anticipated radiation dose.
The imaging geometry with this digital method is unusual, because in the craniocaudal direction the x-ray beam is largely parallel. In the lateral direction, however, it has the traditional fan-beam shape. This causes, at least in theory, the potential for image distortion when objects are at different distances from the receptor (1). This theory was not investigated further in our study.
Radiation Dose
The radiation dose in scoliosis radiography has previously been lowered in various ways (1113). Changes in the screen-film system at conventional imaging can reduce the dose considerably (14). Posteroanterior instead of anteroposterior projection reduces the dose to the breast (15) and other sensitive organs at the cost of a higher bone marrow dose (16). Posteroanterior projection has been reported to be superior in image quality to anteroposterior projection (17), which is now considered obsolete in scoliosis radiography (18). An undercouch tube position, when used with a tabletop, is valuable because the tabletop attenuates considerably. We have measured attenuation as high as 30% with the parameters used in this study. Such an attenuator between the patient and the image receptor should be avoided because this increases the dose to the patient. Increasing x-ray beam filtration is another way to lower the dose (13). With the transition to digital imaging, a further decrease in radiation dose is expected (19,20).
To our knowledge, calculations of the effective dose in scoliosis radiography are scarce in the literature. Values of 0.12 mSv (21) (frontal and lateral views) and 0.05 mSv (22) (frontal view) for conventional screen-film systems have been reported. The dose measurement phantom used in this study was developed for measurements in radiation therapy with higher photon energies (4). The attenuation properties in the diagnostic energy range for this phantom differ somewhat from those for human tissue (23). For the heavily filtered x-ray beam we use, this difference is small, with water having a linear attenuation coefficient that is only 1% lower. The ease of use and general availability of this type of phantom made us choose it despite these potential shortcomings.
Monte Carlo calculations have been used previously in the scoliosis radiography setting, but without any values of effective dose given (15). Our Monte Carlo calculations showed a remarkably good correlation with the measured values of effective dose, even in this complicated setting with multiple images. This indicates that Monte Carlo calculations can be a useful tool in the clinical setting for evaluating the effective dose in various examination types.
Kerma area-product values are easy to obtain and can be a useful tool to compare radiation doses among different institutions and different methods. Kerma area-product values are more sensitive to increases in kilovolt peak level and filtration than effective doses and thus might exaggerate the dose gain.
Image Quality
The digital exposure images were of good quality and had better grades than did the screen-film scoliosis images, despite visible noise, especially when they were magnified. This was also reflected in the visual grading part of the study, in which the digital exposure images were rated as superior to the screen-film images in the subjective aspects of visualization of end plates, suitability for judging skeletal abnormalities, suitability for angle measurements, and overall quality, despite having inferior contrast, noise, and sharpness. With the contrast-detail phantom, digital exposure image quality was lower than screen-film image quality; we interpreted this to be mainly the result of a different imaging chain (image intensifier) and the scanning process with parallax errors. It seems that despite having lower objective image quality, digital exposure images are preferred to screen-film imagesprobably at least in part because of the uneven exposure of the screen-film images (24). The ability to alter the gray scale and contrast dynamically at a PACS workstation also is a great advantage. Digital fluoroscopy images had lower quality and marked noise and were clearly inferior to digital exposure images both at contrast-detail phantom readings and visual grading.
Angle Measurements
There was no significant difference in angle phantom measurements between screen-film and digital exposure images. There was, however, a significant difference between digital fluoroscopy and the other two methods. The mean angles were 21.6° at both screen-film and digital exposure imaging and 21.2° at digital fluoroscopy, with a much lower mean angle SD on the angle phantom, about 0.5°, compared with about 1.0° on the patient images. This indicated a considerably higher precision in the artificial measuring situation, which means that even a small difference between methods turned out to be significant. The difference in mean angles between the three observers was somewhat larger with digital fluoroscopy. It must be stressed, however, that no significant difference between measurements was found on the patient images. The reproducibility, expressed in intraobserver SD, was good with all three methods and better than that reported elsewhere (25). We believe the small difference in angle phantom measurements would not be clinically important in real-life practice, because differences of up to 10° between measurements have been required to indicate a real change in the scoliosis angle (26). The interobserver, intraobserver, and patient differences were much larger. We consider both digital methods to be acceptable for measurements of scoliosis angles, with the caveat that digital fluoroscopy might be slightly less exact. Further studies in this area are needed.
One way to exploit the advantages of the higher image quality of digital exposure imaging and the lower radiation dose with digital fluoroscopy might be to combine the two methods (19). Digital exposure imaging could be used for the initial examination, in which it is important to evaluate underlying abnormalities, whereas digital fluoroscopy could be used for the follow-up examinations, in which the focus is angle measurements.
Statistical Methods
Some authors have set a numeric value to each level (ie, much worse, worse, equal, etc) in the visual grading scale and calculated mean values (9). We disagree with this methodology because the levels are ordinal data with no assumption of equal steps between the different levels and thus should not be treated as continuous data. Instead, we chose to give a descriptive presentation of our material. The maximum
value is 1.00 when agreement is perfect, whereas a value of zero indicates no agreement better than chance.
statistic analysis is of limited value when most of the gradings for all observers accumulate at only one level (eg, are equal). Then, calculations yield low
values despite excellent agreement. In the extreme case when all the grades on all the images for all observers are identical, the
value cannot be calculated owing to a division by zero.
In conclusion, in this study, we evaluated a method for digital scoliosis radiography that enables one to overcome the size restriction of storage phosphor screens. Digital exposure image quality is acceptable, but the radiation dose should be lowered further. A low-dose alternative is to use pulsed digital fluoroscopy, which involves a substantial dose reduction with acceptable accuracy in angle measurements. We are planning future studies in which the described digital exposure examination is optimized further, particularly with lower radiation doses.
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
Author contributions: Guarantor of integrity of entire study, H.G.; study concepts, H.G., B.J.; study design, H.G., T.A., J.P.; definition of intellectual content, H.G., T.A., J.P.; literature research, H.G.; clinical studies, H.G.; experimental studies, T.A., H.G., B.J., K.W.B.; data acquisition, H.G., B.J.; data analysis, H.G., K.W.B.; statistical analysis, H.G., J.P.; manuscript preparation, H.G.; manuscript editing, H.G.; manuscript review and final approval, all authors.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Korner, C. H. Weber, S. Wirth, K.-J. Pfeifer, M. F. Reiser, and M. Treitl Advances in Digital Radiography: Physical Principles and System Overview RadioGraphics, May 1, 2007; 27(3): 675 - 686. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Bath and L G Mansson Visual grading characteristics (VGC) analysis: a non-parametric rank-invariant statistical method for image quality evaluation Br. J. Radiol., March 1, 2007; 80(951): 169 - 176. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |