|
|
||||||||
Cardiac Imaging |
1 From the Institute of Biomedical Engineering and Medical Informatics, University of Zurich and Swiss Federal Institute of Technology, Gloriastrasse 35, 8092 Zurich, Switzerland (S.K., P.B.); and the Department of Cardiothoracic and Vascular Surgery and MR Center, Institute of Experimental Clinical Research, Aarhus University Hospital, Denmark (J.M.H., H.N., P.K.P., E.M.P.). Received January 6, 2000; revision requested February 28; revision received April 12; accepted June 1. Supported by EUREKA grant E!2061 INCA-MRI; Swiss Commission for Technology and Innovation grant 4178.1; Danish Heart Foundation grant 97-2-1-5-22549; Karen Elise Jensen Foundation, Denmark; Desirée and Niels Yde Foundation, Denmark; and Philips Medical Systems, Best, the Netherlands. Address correspondence to P.B. (e-mail: boesiger@biomed.ee.ethz.ch).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Blood velocity was quantified downstream of the valve prosthesis (for replacement after aortic valve stenosis or combined stenosis and regurgitation) in 10 patients by using a magnetic resonance (MR) cine velocity mapping method in which the imaging section position is adapted according to the excursion of the valvular plane of the heart. Two acquisitions were performed to display the blood velocity distributions one-fourth valve diameter and one valve diameter downstream of the valve and to quantify blood volumes and shear rates.
RESULTS: The velocity profiles measured during flow acceleration one-fourth valve diameter downstream were characterized by a distinct pattern of two lateral jets and one central jet of antegrade flow. High shear rates were found along the leaflet tips. The profiles obtained one valve diameter downstream were skewed, with varying velocity patterns among patients. Peak shear rates were found close to the vessel wall. With correction for through-plane motion of the valve, the mean apparent regurgitant fraction (± SD) was 14% ± 6; the mean regurgitant fraction without correction was 9% ± 5.
CONCLUSION: The described noninvasive procedure for velocity mapping enables measurements close to the valve and thus evaluation of blood flow patterns with respect to valve design in humans.
Index terms: Heart, flow dynamics, 51.91, 535.91 Heart, MR, 51.121412, 51.12144 Heart, valves, 535.453 Magnetic resonance (MR), cine study, 51.12144 Magnetic resonance (MR), physics, 51.121412, 51.12144 Phantoms
| INTRODUCTION |
|---|
|
|
|---|
Nygaard et al (68) provided early data on velocity fields and shear stress distal to prosthetic valves in humans. However, the invasive measurement procedure used in that study can be performed only during open-heart surgery.
Cine magnetic resonance (MR) phase-contrast measurements enable noninvasive mapping of velocity distributions. The modality has the potential to provide more detailed information with regard to velocity distributions than does echocardiography, which is currently the reference standard for evaluating heart valves in vivo. Furthermore, MR velocity mapping enables quantification of transvalvular blood volumes as important markers of valve performance.
MR velocity mapping is based on the determination of the acquired phase shift of proton spins moving along a magnetic field gradient. Owing to magnetic field inhomogeneity and susceptibility variations, it is necessary to perform two acquisitions with different velocity sensitivities. The subtraction of the phase-contrast images from the two acquisitions yields a phase-difference map, which is proportional to the velocities.
The nature of blood flow is primarily determined by the ratio of inertial forces to viscous forces. Predominant inertial forces result in turbulence. The locations of turbulence are associated with areas of high-velocity gradients (1,2,6,7). Therefore, mapping of velocity gradients might be used to estimate the complexity of flow fields.
Conventional MR imaging techniques involve a spatially fixed imaging section. Because of cardiac motion, measurements in the immediate vicinity of the heart valve prostheses are not possible. However, the highest flow complexity occurs immediately downstream of prosthetic valves. Measurements further downstream, depending on varying geometric conditions, such as aortic curvature and the valve position relative to the outflow axis, are expected to exhibit alterations in velocity fields. To provide reference data regarding velocity fields downstream of valve prostheses, in vivo data should be acquired as close as possible to the valve.
By using a method for MR velocity mapping and adapting the imaging section position to the cyclic motion of the heart, we tested our hypothesis that motion-corrected velocity and shear rate data can be obtained immediately downstream of valve prostheses in humans.
| MATERIALS AND METHODS |
|---|
|
|
|---|
MR Examination
MR measurements were obtained by using a Gyroscan NT 1.5-T whole-body unit (Philips Medical Systems, Best, the Netherlands) equipped with a gradient system that delivers a maximum gradient amplitude of 21 mT/m with a slew rate of 105 mT/m/msec. The patients were examined while in the supine position. A five-element cardiac phased-array coil was used for receiving signal.
Heart Motion Adaptation
Adaptation of the actual section position to the active motion of the heart was achieved by using a recently proposed three-step strategy (9).
Labeling.A cine imaging sequence with a labeling prepulse was applied within a single breath hold to mark points of the basal plane of the heart. By using a subtraction technique, three points in the basal myocardium, moving during the cardiac cycle, were marked and imaged with a temporal resolution of 30 msec (Fig 1, top). Image processing was used to automatically extract the positions of the three points for each phase of the cardiac cycle. Accordingly, the cyclic motion at the basal plane level of the heart in an individual patient could be quantified.
|
Velocity mapping.Adaptation of the imaging section position according to the basal plane motion was used to map blood velocities in the vicinity of the aortic valve (Fig 1, bottom). The trajectory of heart motion was mapped to the initial section position either one-fourth valve diameter or one valve diameter downstream of the prosthesis.
For velocity mapping, a gradient-echo phase-contrast sequence that interleaved the two velocity-encoding segments was applied. In-plane resolution was 1 x 1 mm2 with a field of view of 256 x 180 mm2, imaging matrix of 256 x 180, and section thickness of 5 mm. Velocity data were recorded in intervals of 30 msec throughout the cardiac cycle, starting 62 msec after the R wave. The initial delay after R-wave detection was required for execution of the navigator pulse, processing of the navigator signal, and the algorithm for motion-adapted gating. The aliasing velocity in the section-selecting direction was set at 160 cm/sec. Accordingly, velocities within a range of -160 to +160 cm/sec could be measured without wraparounds. In the frequency-encoding and phase-encoding directions, flow-compensating gradients were incorporated. To reduce dephasing effects in complex flow fields, 75% of the full echo was acquired to yield an echo time of 3.94.0 msec. Missing samples were zero padded, and an asymmetric weighting function was applied to central k-space data before Fourier transformation.
Accounting for an average gating efficiency of about 75% and a heart rate of 70 beats per minute, the imaging duration was 3 minutes 30 seconds. The measurements one-fourth valve diameter and one valve diameter downstream were performed successively.
Respiratory Motion Compensation
To compensate for the passive motion of the heart due to respiration, a two-dimensional selective excitation pulse (navigator) was applied to detect the actual breathing state of the patient. The pencil-beam navigator pulse was positioned through the dome of the right hemidiaphragm and applied immediately after R-wave detection.
To overcome the shortcoming of low imaging efficiency associated with conventional gating (10), the motion-adapted gating approach (11) was implemented for the velocity-encoding sequence. In real time, the phase-encoding step was calculated according to the actual shift of the diaphragm with respect to the end-expiration level. The phase-encoding number was determined from the actual diaphragm displacement by using a cubic weighting function. Diaphragm displacements within a 10-mm constant gating window were marked as valid for motion-adapted gating. Larger displacements led to rejection of the acquired data.
In addition to navigator gating, adaptive real-time motion correction of the imaging section position was performed (12). The imaging section was shifted according to the current diaphragm displacement within the gating window. A factor of 0.6 was used to scale the diaphragm motion to the craniocaudal motion of the imaging plane (13).
Data Analysis
Initially, phase errors from concomitant gradient fields were corrected by using the method described by Bernstein et al (14). Phase shifts caused by eddy currents were compensated for by subtracting a linear two-dimensional function from the original phase data. The linear function was obtained from a least-squares fit to the phase of static tissue. Static regions were automatically identified as those pixels, on the phase images, that exhibited a low SD during the cardiac cycle (15). Phase pixels with a time SD lower than 4% of the maximum time SD found were considered to be static. Vessel contours were extracted by using active contour-based segmentation (16).
Because the measured velocity data represent a superposition of axial flow velocities through the valve and the through-plane velocity of the section of interest, the section through-plane velocity component was subtracted to obtain the velocity of the blood flow without the influence of aortic root motion. By integrating the pixel velocity over the contour area, the mean blood flow of each heart phase was calculated. The mean forward and reverse flow values were extracted by integrating either the positive or negative blood velocities over the contour area. To match the different heart rates of the patients, the time axis was normalized by assigning the end systole to 100%.
Systolic volume was defined as the volume ejected between 20% and 100% of end systole. Regurgitant volume was obtained by integrating blood flow during diastole (Fig 2). The regurgitant fraction was calculated as the ratio of regurgitant volume to forward volume.
|
Processing for phase correction and vessel segmentation on a DEC-Alpha workstation (Digital Equipment Corporation, division of Compaq, Manalapan, NJ) took 1520 minutes per patient. Values for blood flow, blood volumes, and shear rates were readily available thereafter.
Statistical Analyses
A paired Student t test was performed to determine differences in the following parameters. The regurgitant volume based on velocity data measured one-fourth valve diameter downstream, with correction for through-plane motion, was compared with data in which the through-plane motion correction process was skipped. The ratio of reverse systolic blood volume to forward systolic blood volume measured one-fourth valve diameter downstream was compared with this ratio based on data measured one valve diameter downstream. A P value of less than .05 was considered to prove the statistical significance of the observed differences.
Phantom Validation
To validate the accuracy of MR phase-contrast velocity mapping for measurements downstream of the artificial heart valve, phantom experiments were conducted. A Plexiglas pipe equipped with a bileaflet aortic heart valve was used in a steady flow setup (diameter of valve orifice, 23 mm; pipe diameter downstream of inserted valve, 38 mm; flow rate, 15 L/min). Fluid viscosity and relaxation time constants, T1 and T2, were matched to the values of blood (viscosity, 3.5 mPa · sec; T1, 941 msec; T2, 232 msec). MR velocity mapping was performed 8 mm downstream of the valve, and the data were compared with those measured with laser Doppler anemometry. Linear regression analysis was used for comparison. Laser Doppler anemometry is considered the reference standard for registration of fluid velocity in vitro. The in vitro parameters for MR velocity mapping were identical to those used for the in vivo measurements.
To substantiate the hypothesis of a correlation between high-velocity gradients and regions of highly complex flow spatial velocity, gradients were calculated at MR phase-contrast imaging and compared with the laser Doppler anemometry measurements of fluctuating velocity. Local maximum velocity gradients were extracted from the MR velocity data and compared with local maximums of fluctuating velocity by using regression analysis.
| RESULTS |
|---|
|
|
|---|
Aortic Root Motion
The mean maximum through-plane motion of the aortic root (± SD) was 7 mm ± 2 (range, 411 mm). The peak velocity of through-plane motion of the aortic root was 6 cm/sec ± 1 (range, 47 cm/sec) and observed in early diastole (140%160% of end systole) (Fig 3).
|
With correction for through-plane motion, the mean apparent regurgitant volume (± SD) measured one-fourth valve diameter downstream of the valve was 10 mL ± 4; the mean regurgitant volume without through-plane motion correction was 6 mL ± 3. The corresponding mean regurgitant fraction was 14% ± 6 with correction for through-plane motion; the mean regurgitant fraction without through-plane motion correction was 9% ± 5 (Table 1).
|
|
|
|
Blood Flow
There was good agreement between the stroke volumes calculated independently in the two downstream locations in all 10 patients (Table 2). The mean difference in stroke volume (± 2 SDs) between the two locations was 0 mL ± 6.
To quantify the differences between forward and reverse blood volumes with respect to the distance from the valve, the forward and reverse blood volumes during systolic ejection were compared separately (Fig 6). At the closest distance from the valve, the systolic forward and reverse blood volumes were substantially greater compared with the volumes obtained further downstream (Table 2). The mean regurgitant volume (± SD) for all patients and all measured positions amounted to 10 mL per beat ± 5.
|
|
One valve diameter downstream.The peak shear values one valve diameter downstream were considerably smaller compared with those closest to the valve. The shear boundary was reattached to the right vessel wall, and, accordingly, high shear rate values were found along this site (Fig 7, bottom). The maximum shear rate occurred 242 msec after the R wave.
Phantom Validation
In vitro, the difference between velocities measured with MR and those measured with laser Doppler anemometry downstream of the artificial valve was within 12% for the central 75% of the pipe diameter (Fig 8, A). Linear regression analysis yielded a correlation coefficient (r2) of 0.99; the standard error of the estimate was 4 cm/sec. The locations along the pipe radius that exhibited high shear rates, as derived from MR velocity data, correlated well with the peak values of fluctuating velocity measured with laser Doppler anemometry (r2 = 1.00; standard error of the estimate, 1 mm) (Fig 8, B).
|
| DISCUSSION |
|---|
|
|
|---|
MR velocity mapping is a noninvasive modality that can provide detailed information on velocity distributions. However, because of the cyclic motion in the valvular plane of the heart, conventional MR velocity mapping is limited to measurements downstream of the sino-tubular junction of the aorta. The described method of MR velocity mapping enabled us to assess the blood flow patterns closer to an artificial aortic valve in patients than has been possible previously.
Through-Plane Motion
The considerable through-plane motion of the prosthetic valve was quantified (mean maximum motion, 7 mm) and thus could not be neglected when measurements in the immediate vicinity of the valve were performed. The through-plane motion measured in this study was smaller compared with that in a previous investigation with young healthy volunteers (9). The differences might be attributed to age-related stiffening of the heart and the fact that the patients underwent open-heart surgery and had consequent scar tissue.
The rapid relaxation of the heart and thus rapid through-plane motion of the valve during diastole led to an underestimation of regurgitant flow. However, regurgitant flow was small (mean, 10 mL) compared with systolic blood flow. Therefore, the velocity-to-noise ratio for the diastolic velocity data was approximately 2.5 times lower than that for the peak systolic velocity data. The use of adaptable velocity encoding (18) would be beneficial in future studies. Nevertheless, the estimate of regurgitation in our study compared well with related data in the literature (19).
Blood Velocity Distribution
The velocity data obtained shortly before peak systole one-fourth valve diameter downstream of the valve clearly disclosed the valve design: Three velocity jets emerged from the two major lateral orifices and the small central slit of the valve. This distinct triple-jet pattern has been verified by using laser Doppler anemometry and MR imaging in in vitro steady-flow experiments with models simulating peak systolic flow (20).
Reverse flow within the sinuses of Valsalva, which was indicative of recirculation, occurred during systolic flow acceleration. This finding has been observed in vitro under pulsatile flow conditions (21). Our comparison of systolic forward and reverse flow indicated the existence of recirculation in the cavities of the sinuses during flow acceleration. Flow quantification revealed substantially more antegrade and retrograde flow at the distance closest to the valve.
During early flow acceleration, the two lateral jets were farther apart than they were later during the acceleration phase. This seemed to be caused by flow that was initially accelerated by the opening leaflets toward the vessel wall. Later during systole, flow was directed more in the axial direction of the aorta, which was indicative of fully opened leaflets. The observed symmetry of velocity distribution that corresponded to the two lateral orifices of the valve suggested an optimal utilization of flow area across the valve during systolic flow acceleration. The velocity profiles obtained one valve diameter downstream of the valve exhibited substantial changes compared with the patterns observed close to the valve, with large heterogeneity among the patients. As blood flowed downstream, the skewness of velocity profiles became more accentuated: Either a single velocity jet or two jets migrated toward the right vessel wall. These alterations were considered to be attributed to differences in aortic root geometry and aortic curvature among the patients.
Asymmetry of the velocity jet pattern has been associated with nonoptimal use of the flow area across the valve (22). However, this conclusion was based on data measured one valve diameter downstream of the valve and therefore might be misleading given the large variation in flow profiles among subjects. In this study, measurements one-fourth valve diameter downstream revealed lateral velocity jets of almost equal magnitude.
Shear Rate Distribution
Peak velocity gradients were detected one-fourth valve diameter downstream immediately distal to the leaflet tips during the deceleration phase in all patients. The highest normal Reynolds stresses measured 1.52.0 vessel diameters downstream of the valve in humans typically have been at locations that corresponded to the central slit and the hinges of the valve (7,8). However, in vitro stress measurements performed closer to the valve (one-half diameter downstream) have revealed high turbulent shear stresses also immediately distal to the valvular leaflets (8). As blood flowed further downstream, the peak velocity gradients became considerably smaller, as observed in the present study, and were located along the right posterior vessel wall; this indicated the reattachment of the velocity jet to the vessel wall.
Technical Considerations
MR velocity mapping has limited accuracy when measurements are performed in disturbed or turbulent flow fields (23,24). Turbulence has not been shown to cause distortion of velocity data in straight-tube flow (25). Unlike separated flow, like that found with stenosis, in fully developed turbulent flow the intravoxel phase dispersion can be assumed to be small, because a small random time-dependent velocity component is superimposed on a large mean velocity component. However, flow separation occurs downstream of artificial heart valves owing to the obstruction of flow caused by the hinge mechanisms of the valve and by the leaflets themselves. Total signal voids were not seen; however, slight signal losses were visible on the modulus images, and, thus, phase distortions that affected the accuracy of the velocity encoding were possible.
Nevertheless, in vitro, the differences in velocity profiles downstream of the artificial aortic valve measured with MR and laser Doppler anemometry have been reported to be within 10% (20) with parameter settings (1 x 1-mm2 spatial resolution, 4-msec echo time) similar to those used in the present study. Reports (20,26) indicate that MR velocity mapping with short echo times can be used to assess velocity distributions downstream of artificial heart valves in vivo.
MR measurements closer than one-fourth valve diameter downstream of the valve were not possible because of susceptibility changes around the prosthetic valve and interference of the open-valve leaflets with the imaging plane.
In conclusion, the MR velocity mapping method described in this study enabled the in vivo assessment of blood flow patterns in the near vicinity of the artificial aortic valve and the quantification of regurgitant blood flow through the valve.
The velocity distributions measured in the immediate vicinity of the valve reflected the valve design and provided information about flow distribution through the leaflets. The velocity profiles obtained further downstream of the valve were severely altered and varied among patients owing to effects of the individual aortic geometries.
The described MR velocity mapping method enabled detailed assessment of valve-induced changes in blood flow and quantification of transvalvular blood volumes in humans with an implanted prosthetic valve. Data on in vivo near-valve velocity distributions are an important basis for the evaluation of prosthetic valve designs.
| FOOTNOTES |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. J. Pennell, U. P. Sechtem, C. B. Higgins, W. J. Manning, G. M. Pohost, F. E. Rademakers, A. C. van Rossum, L. J. Shaw, and E. K. Yucel Clinical indications for cardiovascular magnetic resonance (CMR): Consensus Panel report Eur. Heart J., November 1, 2004; 25(21): 1940 - 1965. [Full Text] [PDF] |
||||
![]() |
M. H. Yacoub and L. H. Cohn Novel Approaches to Cardiac Valve Repair: From Structure to Function: Part II Circulation, March 9, 2004; 109(9): 1064 - 1072. [Full Text] [PDF] |
||||
![]() |
S. J. Karlik Visualizing Radiologic Data Am. J. Roentgenol., March 1, 2003; 180(3): 607 - 619. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |