|
|
||||||||
Technical Developments |
1 From GE Medical Systems, Milwaukee, Wis (G.S.S., S.N.G., T.K.F.F.); and the Cardiovascular Research Foundation and the Lenox Hill Heart and Vascular Institute, New York, NY (S.D.W.). Received April 3, 2000; revision requested June 5; revision received June 30; accepted July 11. Address correspondence to G.S.S., GE Medical Systems, Johns Hopkins Hospital, 600 N Wolfe St, Rm 110 MRI, Baltimore, MD 21287 (e-mail: glenn.slavin@med.ge.com).
| ABSTRACT |
|---|
|
|
|---|
Index terms: Heart, perfusion, 511.12144 Magnetic resonance (MR), contrast enhancement, 511.12143 Myocardium, blood supply, 511.76, 511.771 Myocardium, MR, 511.121412, 511.12143
| INTRODUCTION |
|---|
|
|
|---|
The feasibility of first-pass contrast materialenhanced perfusion MR imaging has been demonstrated by several groups (26), but the difficulty in optimizing all aspects of the acquisition sequence has resulted in compromised image quality or reduced anatomic coverage. Successful first-pass perfusion MR imaging requires that several issues be addressed, including the following: (a) coverage of the heart from apex to base (which typically requires at least six short-axis views), (b) temporal resolution (ie, time between repeated acquisitions at the same section location) that is sufficient to allow adequate sampling of the first pass of the contrast material bolus, (c) a relationship between signal intensity and contrast material dose (ie, input function) that is quantifiable (preferably linear), and (d) contrast- and signal-to-noise ratios that are sufficiently high to allow discrimination between normal and ischemic regions of the myocardium.
To meet these goals, current perfusion techniques make use of fast gradient-echo MR imaging sequences in combination with various magnetization preparation schemes for T1 weighting, including inversion recovery (25), saturation recovery (710), or partial saturation (11,12). Shortcomings of these approaches include the following: (a) the number of sections acquired is limited owing to long acquisition time or magnetization recovery inversion time (TI), (b) image contrast and signal-to-noise ratio are inadequate owing to use of low preparation flip angles or short TIs, and (c) input functions are not quantifiable owing to sensitivity to arrhythmias.
Regardless of the type of preparation used, the magnetization recovery TI plays a major role in determining contrast, signal-to-noise ratio, and the maximum number of sections that can be acquired. A short recovery time permits the acquisition of more sections per cardiac cycle but sacrifices signal-to-noise ratio and contrast owing to minimal relaxation. A long TI provides higher signal-to-noise ratio and better contrast, but the delay inserted between preparation and acquisition reduces the maximum number of sections that can be acquired.
The purpose of this study was to develop and evaluate a technique for simultaneously achieving long TIs and multisection capability (ie, multiple sections per cardiac cycle) for T1-weighted first-pass myocardial perfusion MR imaging. This technique was developed without inserting explicit delays into the sequence or sacrificing section coverage.
| Materials and Methods |
|---|
|
|
|---|
|
|
The effect of the interleaved notched saturation is demonstrated in Figure 2. The radio-frequency preparation pulse for section n + 1 (t3 in Fig 2), which preceded the acquisition of section n (t4), saturated all tissue in the selected volume except that in section n. Spins in section n were unaffected by this pulse because they fell within the notch. These spins, however, were saturated by the preparation pulse (t1) that immediately preceded the acquisition of section n - 1 (t2). Therefore, the spins in section n recovered for a time spanning the acquisition of section n - 1 and the preparation of section n + 1. This time is the effective TI, which equaled 185 msec for the imaging parameters used in this study.
|
|
The preparation sequence was a 15-msec radio-frequency pulse with a notched frequency response and 2,600-kHz bandwidth that was used in the presence of a section-select gradient. Because seven sections were acquired in an interleaved manner (ie, sections 1, 3, 5, 7, 2, 4, 6), the passband needed to be at least as wide as the largest spacing between any two consecutive sections. In this case, section 2 was saturated when the notch was at section 7. Therefore, the width of each saturation band was designed to be five times the width of the notch; the overall width of the preparation volume was 165 mm. To allow some motion of the section during the TI, the notch width was 50% greater than the section thickness. The total duration of the preparation sequence was 18 msec, including time for the gradient dephasers. The shape of the radio-frequency pulse is shown in Figure 1a, and an actual section profile is shown in Figure 1b.
Patient Study
A feasibility study to evaluate image contrast was conducted with 12 outpatients (eight men and four women; age range, 5476 years; mean age, 66.6 years ± 6.2 [SD]) with known or suspected coronary disease. All patients had undergone cardiac catheterization as part of their clinical evaluation. The only exclusion criteria were contraindications to adenosine or contrast-enhanced MR imaging. All studies were performed after signed informed consent was received from each patient, in accordance with a protocol approved by our institutional review board.
Patients underwent first-pass perfusion MR studies both at rest and with pharmacologic stress. A cardiologist was present in the MR imaging room during all studies. For imaging with stress, adenosine (Adenoscan; Fujisawa Healthcare, Deerfield, Ill) was administered intravenously into an antecubital vein at 140 µg/kg/min, starting 2 minutes before the start of MR imaging. After 2 minutes of adenosine infusion, gadopentetate dimeglumine was administered intravenously into the contralateral arm at 5 mL/sec with use of a power injector (Spectris; Medrad, Pittsburgh, Pa), simultaneous with the start of MR imaging. Patients were instructed to hold their breath and to continue the breath hold as long as possible during imaging. Thirty images (phases) were acquired at each section location during 60 heartbeats. The adenosine infusion was discontinued at the completion of MR imaging. After 15 minutes, the rest study was performed with a second dose of contrast material and identical imaging parameters.
Regions of interest were drawn by one of the authors (S.N.G.) in the posterior or anterior septal wall of a midventricular section for each of the 30 phases. Regions of interest were approximately 150 mm2. In each case, signal intensity as a function of time was measured and averaged for each contrast material dose. Peak enhancement was calculated as the difference between peak and baseline signal intensity in the regions of interest divided by the baseline signal intensity. These results were compared with those of the partial saturation study (45° preparation pulse, 10-msec TI, 12° excitation flip angle, contrast material doses of 0.10 and 0.15 mmol/kg) (11). Ten cases with each contrast material dose in the partial saturation study were selected retrospectively with use of the criterion that no lesions were detected in the left anterior descending coronary artery with conventional coronary angiography. Results were evaluated with a two-sample t test. Differences with a P value less than .05 were considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
The long recovery time and high preparation flip angle also provided better image contrast than did short TI saturation recovery or partial saturation and greater anatomic coverage than did conventional long TI recovery. Furthermore, because blood is saturated on both sides of the imaged section immediately prior to data acquisition, the presence of the notch provides a potential mechanism for blood pool suppression. If saturated blood moves into the imaging plane, it may help discriminate the myocardial wall from the ventricular space and reduce the intensity of flow-related artifacts. The degree of expected in-plane blood suppression is determined by the width of the notch.
Because a given section was prepared before acquisition of the preceding section, the first section represented a special case. The first phase of the first section (ie, the very first image to be acquired) had no preparation. Subsequent phases of the first section, however, were prepared with the saturation pulse that preceded the last section of the previous phase. For example, phase two of the first section was saturated by the preparation pulse preceding phase one of the last section. Consequently, the time between the preparation and acquisition of the first section (ie, the TI) spanned an R wave. The TI for the first section was therefore longer than that for the other sections and was also variable, depending on the R-R interval of the particular heartbeat. Although this effect was not detrimental to image quality or clinical diagnosis, it is possible to apply a custom preparation scheme for the first section.
Imaging was performed through the R wave of the second cardiac cycle for two reasons. First, it maintained the preparation-acquisition timing for all sections except the first, as discussed previously. If half the sections were acquired in each of two cardiac cycles, the middle section would also have a heart-ratedependent TI. Second, imaging through the R wave obviated two trigger windows. This afforded more time for data collection, often allowing the acquisition of an additional section. It should be noted that registration of the sections acquired during the second R-R interval was not appreciably affected by modest changes in heart rate.
In light of the results of this study, we are continuing to optimize the sequence to further improve image quality, by characterizing the blood pool suppression and minimizing any image artifacts due to nonsteady-state effects that can result from imaging with higher flip angles. Bloch simulations of a variable flip angle excitation (17,18) indicate that point-spread-function side lobes can be substantially reduced without loss of overall signal intensity. We confirmed these results in preliminary studies in animal models (unpublished data).
This study evaluated an alternative way of performing magnetization preparation with saturation recovery. The interleaved preparation and acquisition sequences and the overlapping recovery times allowed the most time-efficient implementation of long TI saturation recovery for multisection myocardial perfusion MR imaging. This implementation resulted in substantially better image contrast than can be achieved with short TI saturation recovery or partial saturation and greater section coverage than can be achieved with conventional long TI saturation recovery or inversion recovery.
| FOOTNOTES |
|---|
Author contributions: Guarantors of integrity of entire study, G.S.S., S.D.W., T.K.F.F.; study concepts and design, G.S.S., S.D.W., T.K.F.F.; definition of intellectual content, G.S.S., S.D.W., T.K.F.F.; literature research, G.S.S., T.K.F.F.; clinical studies, S.D.W.; experimental studies, G.S.S.; data acquisition, S.D.W.; data analysis, S.N.G.; statistical analysis, G.S.S., S.N.G.; manuscript preparation, G.S.S., T.K.F.F.; manuscript editing and review, all authors.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. C. Cury, K. Shash, J. T. Nagurney, G. Rosito, M. D. Shapiro, C. H. Nomura, S. Abbara, F. Bamberg, M. Ferencik, E. J. Schmidt, et al. Cardiac Magnetic Resonance With T2-Weighted Imaging Improves Detection of Patients With Acute Coronary Syndrome in the Emergency Department Circulation, August 19, 2008; 118(8): 837 - 844. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schmidt, C. F. Azevedo, A. Cheng, S. N. Gupta, D. A. Bluemke, T. K. Foo, G. Gerstenblith, R. G. Weiss, E. Marban, G. F. Tomaselli, et al. Infarct Tissue Heterogeneity by Magnetic Resonance Imaging Identifies Enhanced Cardiac Arrhythmia Susceptibility in Patients With Left Ventricular Dysfunction Circulation, April 17, 2007; 115(15): 2006 - 2014. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. Cury, C. A. M. Cattani, L. A. G. Gabure, D. J. Racy, J. M. de Gois, U. Siebert, S. S. Lima, and T. J. Brady Diagnostic Performance of Stress Perfusion and Delayed-Enhancement MR Imaging in Patients with Coronary Artery Disease. Radiology, July 1, 2006; 240(1): 39 - 45. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. P. Ingkanisorn, R. Y. Kwong, N. S. Bohme, N. L. Geller, K. L. Rhoads, C. K. Dyke, D. I. Paterson, M. A. Syed, A. H. Aletras, and A. E. Arai Prognosis of Negative Adenosine Stress Magnetic Resonance in Patients Presenting to an Emergency Department With Chest Pain J. Am. Coll. Cardiol., April 4, 2006; 47(7): 1427 - 1432. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Baks, R.-J. van Geuns, E. Biagini, P. Wielopolski, N. R. Mollet, F. Cademartiri, W. J. van der Giessen, G. P. Krestin, P. W. Serruys, D. J. Duncker, et al. Effects of Primary Angioplasty for Acute Myocardial Infarction on Early and Late Infarct Size and Left Ventricular Wall Characteristics J. Am. Coll. Cardiol., January 3, 2006; 47(1): 40 - 44. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Baks, R.-J. van Geuns, E. Biagini, P. Wielopolski, N. R. Mollet, F. Cademartiri, W. J. van der Giessen, G. P. Krestin, P. W. Serruys, D. J. Duncker, et al. Effects of Primary Angioplasty for Acute Myocardial Infarction on Early and Late Infarct Size and Left Ventricular Wall Characteristics J. Am. Coll. Cardiol., December 13, 2005; (2005) j.jacc.2005.09.008v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ishida, H. Sakuma, N. Kato, N. Ishida, K. Kitagawa, T. Shimono, I. Yada, and K. Takeda Contrast-enhanced MR Imaging for Evaluation of Coronary Artery Disease before Elective Repair of Aortic Aneurysm Radiology, November 1, 2005; 237(2): 458 - 464. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Fenchel, U. Helber, U. Kramer, N. I. Stauder, A. Franow, C. D. Claussen, and S. Miller Detection of Regional Myocardial Perfusion Deficit Using Rest and Stress Perfusion MRI: A Feasibility Study Am. J. Roentgenol., September 1, 2005; 185(3): 627 - 635. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Sakuma, N. Suzawa, Y. Ichikawa, K. Makino, T. Hirano, K. Kitagawa, and K. Takeda Diagnostic Accuracy of Stress First-Pass Contrast-Enhanced Myocardial Perfusion MRI Compared with Stress Myocardial Perfusion Scintigraphy Am. J. Roentgenol., July 1, 2005; 185(1): 95 - 102. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. F. Azevedo, L. C. Amado, D. L. Kraitchman, B. L. Gerber, T. Edvardsen, N. F. Osman, C. E. Rochitte, K. C. Wu, and J. A.C. Lima The effect of intra-aortic balloon counterpulsation on left ventricular functional recovery early after acute myocardial infarction: a randomized experimental magnetic resonance imaging study Eur. Heart J., June 2, 2005; 26(12): 1235 - 1241. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Baks, R.-J. van Geuns, E. Biagini, P. Wielopolski, N. R. Mollet, F. Cademartiri, E. Boersma, W. J. van der Giessen, G. P. Krestin, D. J. Duncker, et al. Recovery of left ventricular function after primary angioplasty for acute myocardial infarction Eur. Heart J., June 1, 2005; 26(11): 1070 - 1077. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. K. F. Foo, V. B. Ho, M. Saranathan, L.-q. Cheng, H. Sakuma, D. L. Kraitchman, K. C. Wu, and D. A. Bluemke Feasibility of Integrating High-Spatial-Resolution 3D Breath-hold Coronary MR Angiography with Myocardial Perfusion and Viability Examinations Radiology, June 1, 2005; 235(3): 1025 - 1030. [Abstract] [Full Text] [PDF] |
||||
![]() |
T.H. Giang, D. Nanz, R. Coulden, M. Friedrich, M. Graves, N. Al-Saadi, T.F. Luscher, G.K. von Schulthess, and J. Schwitter Detection of coronary artery disease by magnetic resonance myocardial perfusion imaging with various contrast medium doses: first european multi-centre experience Eur. Heart J., September 2, 2004; 25(18): 1657 - 1665. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.D. Wolff, J. Schwitter, R. Coulden, M.G. Friedrich, D.A. Bluemke, R.W. Biederman, E.T. Martin, A.J. Lansky, F. Kashanian, T.K.F. Foo, et al. Myocardial First-Pass Perfusion Magnetic Resonance Imaging: A Multicenter Dose-Ranging Study Circulation, August 10, 2004; 110(6): 732 - 737. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. F. Azevedo, L. C. Amado, D. L. Kraitchman, B. L. Gerber, N. F. Osman, C. E. Rochitte, T. Edvardsen, and J. A.C. Lima Persistent diastolic dysfunction despite complete systolic functional recovery after reperfused acute myocardial infarction demonstrated by tagged magnetic resonance imaging Eur. Heart J., August 2, 2004; 25(16): 1419 - 1427. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. C. Amado, D. L. Kraitchman, B. L. Gerber, E. Castillo, R. C. Boston, J. Grayzel, and J. A. C. Lima Reduction of "no-reflow" phenomenon by intra-aortic balloon counterpulsation in a randomized magnetic resonance imaging experimental study J. Am. Coll. Cardiol., April 7, 2004; 43(7): 1291 - 1298. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Ishida, H. Sakuma, M. Motoyasu, T. Okinaka, N. Isaka, T. Nakano, and K. Takeda Noninfarcted Myocardium: Correlation between Dynamic First-Pass Contrast-enhanced Myocardial MR Imaging and Quantitative Coronary Angiography Radiology, October 1, 2003; 229(1): 209 - 216. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Y. Kwong, A. E. Schussheim, S. Rekhraj, A. H. Aletras, N. Geller, J. Davis, T. F. Christian, R. S. Balaban, and A. E. Arai Detecting Acute Coronary Syndrome in the Emergency Department With Cardiac Magnetic Resonance Imaging Circulation, February 4, 2003; 107(4): 531 - 537. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L. Gerber, D. A. Bluemke, B. B. Chin, R. C. Boston, A. W. Heldman, J. A. C. Lima, and D. L. Kraitchman Single-Vessel Coronary Artery Stenosis: Myocardial Perfusion Imaging with Gadomer-17 First-Pass MR Imaging in a Swine Model of Comparison with Gadopentetate Dimeglumine Radiology, October 1, 2002; 225(1): 104 - 112. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L. Gerber, J. Garot, D. A. Bluemke, K. C. Wu, and J. A.C. Lima Accuracy of Contrast-Enhanced Magnetic Resonance Imaging in Predicting Improvement of Regional Myocardial Function in Patients After Acute Myocardial Infarction Circulation, August 27, 2002; 106(9): 1083 - 1089. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |