Technical Developments |
1 From the Department of Radiology, Erasmus University Medical Center Rotterdam, "Dijkzigt," Dr Molewaterplein 40, 3015 GD Rotterdam, the Netherlands (L.C.v.D., J.v.H., N.A.A.M., P.M.T.P.), and Elephant Dental, Hoorn, the Netherlands (B.P.v.D.). Received March 7, 2000; revision requested April 26; revision received August 1; accepted September 6. Address correspondence to L.C.v.D. (e-mail: lcvandijk@rond.azr.nl).
| ABSTRACT |
|---|
|
|
|---|
Index terms: Magnetic resonance (MR), artifact, 9*.129412, 9*.129422 Magnetic resonance (MR), vascular studies, 9*.129412, 9*.12942 Stents and prostheses, 9*.1268
| INTRODUCTION |
|---|
|
|
|---|
The MR imaging compatibility of a metal alloy depends on the type and quantity of elements used. Iron-containing alloys in particular and, to a lesser extent, alloys that contain cobalt, nickel, or chromium are known to cause MR imaging artifacts (3) and therefore degrade the MR imaging characteristics of vascular stents. We developed ABI alloy, a metal alloy composed of precious metal elements. The purpose of this initial study was to test the MR imaging compatibility of stents made of ABI alloy, especially as compared with those made of stainless steel 316. Because it is known that some MR compatibility is achieved with the use of other metal alloys, such as nitinol (46) and tantalum (7), we also compared the MR imaging of ABI alloy with that of commercially available stents made of nitinol, tantalum, and Elgiloy (Elgiloy, Elgin, Ill).
| Materials and Methods |
|---|
|
|
|---|
We performed three MR imaging examinations. In the first examination, a self-made stent model made of stainless steel 316 was compared with an identical stent model made of ABI alloy. Both models were composed of two 0.4-mm-thick wires made into a helixone clockwise and the other counterclockwise. The two wires were bound together by using silk at each wire crossing to form a basic expanded stent model with a length of 40 mm, diameter of 8 mm, and 12 loops.
In the second examination, we compared the self-made ABI alloy stent model with the following commercially available devices: Memotherm nitinol stent (Bard Interventional Products, Billerica, Mass), Strecker tantalum stent (Boston Scientific/Vascular, Watertown, Mass), and Wallstent elgiloy stent (Schneider, Minneapolis, Minn), all of which have a diameter of 8 mm.
In the third examination, the ABI alloy stent model was compared with the nitinol stent by using a three-dimensional volume gradient sequence, because it is known that nitinol stents have the best MR imaging compatibility.
The MR imaging experiments were carried out with a 1.5-T magnet (ACS-NT; Philips Medical Systems, Best, the Netherlands) and a surface coil. In the first two examinations, images were obtained with a gradient-echo technique (10.0/6.2 [repetition time msec/echo time msec], 30° flip angle, 205 x 256 matrix, 2-mm section thickness). In the third examination, a three-dimensional gradient-echo technique was used (27.0/7.3, 30° flip angle, 512 x 512 matrix, 0.8-mm section thickness). The devices were placed in plastic tubing that was taped to the bottom of a plastic container. The tube and the container were filled with a copper sulfate in water solution (1 g/L).
The presence and degree of artifacts on the resulting MR images were quantified according to the method of Teitelbaum et al (8): no artifact, mild artifact, moderate artifact, severe black-hole artifact, and severe black-hole artifact with marked image distortion (Table). A mild artifact was judged to be smaller than the device causing the artifact (8). A moderate artifact was approximately the same size as the device. A severe black-hole artifact was larger than the device. Distortion was defined as bending, warping, or obliteration of the image contours normally observed.
|
|
|
| Results |
|---|
|
|
|---|
In test 2, the Elgiloy stent caused a severe black-hole artifact, the nitinol and tantalum stents caused moderate artifacts, and the ABI alloy stent model caused a mild artifact (Fig 2).
In test 3, the ABI alloy stent model caused fewer artifacts than did the nitinol stent (Fig 3). On the MR images of the ABI alloy stent model, delineation of the plastic tube from the stent material was possible, but this was not the case with the nitinol stent. Both the nitinol stent and the ABI alloy stent model caused some signal intensity decrease within the lumen.
|
With use of the differential method to determine the radiopacity of both metal alloys, stainless steel 316 was equal to a 5.4-mm aluminum equivalent (±0.5 mm) and ABI alloy was equal to a 12.8-mm aluminum equivalent (±0.5 mm). ABI alloy was therefore approximately 2.4 times more radiopaque than stainless steel 316 (Fig 4).
|
| Discussion |
|---|
|
|
|---|
Balloon-expandable stents are preferred for the coronary arteries by almost all interventional cardiologists and for several peripheral vessels by many interventional radiologists because of their accuracy and ease of placement compared with those of self-expandable stents. Stainless steel 316 has the mechanical properties necessary for construction of balloon-expandable stentsthat is, plastic deformability to allow crimping on a low-profile balloon and expansion to the vessel diameter, combined with good hoop strength at full expansion and resistance to breakage. Because the mechanical properties of ABI alloy are similar to those of stainless steel 316, our main goal was to compare this material with the standard materials used to construct balloon-expandable stents.
Artifacts in stents on MR images are dependent on the material and design of the stents. In the first test, we used self-made stent models with an identical design, which allowed a comparison of the (isolated) effect of the different materials. In the second test, the stent materials and stent designs were different. Therefore, the different MR artifacts observed with the ABI alloy stent model and the nitinol, tantalum, and Elgiloy stents were caused by a combination of material and design influences. Because stent design is dictated by many factors, including the mechanical properties of the material and patent issues, to our knowledge clinically tested stents of identical design but different material are not yet available for in vitro and in vivo comparisons of MR artifacts.
MR artifacts are dependent on the MR sequence. In the described experiments, we used gradient-echo sequences with relatively long echo times to exaggerate the artifact sizes, because longer echo times are associated with larger artifacts. Modern MR systems allow fast imaging sequences for MR angiography and coronary MR angiography with shorter repetition and echo times. It is reasonable to assume that the severity of artifacts will be reduced if these sequences are used, but the differences between the materials will still exist.
MR evaluation of the stent lumen is, to some extent, possible with nitinol and tantalum stents (5,10), but only stenoses of 50% or more can be correctly assessed, because smaller stenoses are overestimated owing to the artifacts (11). The ABI alloy stent model allowed clear delineation of the stent from the plastic tube surrounding the stent. This suggests that with ABI alloy stents, not only MR imaging evaluation of stent patency but also MR analysis of the vessel wall surrounding the stent is possible. The ability to image close to the stent material could mean that MR angiographic evaluation of even mild within-stent recurrent stenosis is possible after ABI alloy stent placement. In the lumina of both the ABI alloy stent model and nitinol stent, some MR signal intensity decrease was noticeable, probably owing to the Faraday cage effect.
The radiograph showed that the radiopacity of ABI alloy is approximately 2.4 times higher than that of stainless steel 316. This implies that fluoroscopic imaging for positioning of stents made of ABI alloy should be at least as good as that for positioning of stents made of stainless steel 316 when stents of identical design are compared. In conclusion, initial examination results showed that ABI alloy causes fewer artifacts at MR imaging than does stainless steel 316.
| FOOTNOTES |
|---|
B.P.v.D. is a research and development employer for Elephant Dental.
Author contributions: Guarantors of integrity of entire study, L.C.v.D., P.M.T.P.; study concepts, all authors; study design, L.C.v.D., J.v.H., N.A.A.M.; literature research, L.C.v.D., J.v.H.; experimental studies, L.C.v.D., J.v.H., B.P.v.D., N.A.A.M.; data acquisition, L.C.v.D., J.v.H.; data analysis, all authors; manuscript preparation, L.C.v.D., J.v.H.; manuscript editing, B.P.v.D., N.A.A.M., P.M.T.P.; manuscript review, N.A.A.M., P.M.T.P; manuscript final version approval, all authors.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. R. Whittaker and M. F. Fillinger The Engineering of Endovascular Stent Technology: A Review Vascular and Endovascular Surgery, March 1, 2006; 40(2): 85 - 94. [Abstract] [PDF] |
||||
![]() |
E. Spuentrup, A. Ruebben, M. Stuber, R. W. Gunther, and A. Buecker Metallic Renal Artery MR Imaging Stent: Artifact-free Lumen Visualization with Projection and Standard Renal MR Angiography Radiology, June 1, 2003; 227(3): 897 - 902. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |