Gastrointestinal Imaging |
1 From the Department of Diagnostic Radiology, Mallinckrodt Institute of Radiology, 510 S Kingshighway Blvd, St Louis, MO 63110 (A.K.H.), and the Departments of Diagnostic Radiology (C.D.J., R.L.M., T.J.W., C.H.M.) and Biostatistics (W.S.H.), Mayo Clinic, Rochester, Minn. Received June 12, 2000; revision requested July 26; revision received September 22; accepted November 1. Address correspondence to A.K.H. (e-mail: haraa2@yahoo.com).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: A total of 237 consecutive patients received subcutaneously administered glucagon and underwent prone and supine CT colonography with singledetector row CT (n = 77) and multidetector row CT (n = 160), followed by colonoscopy. Examination results were graded for colonic distention, respiratory artifacts, and polyp depiction by two radiologists working independently.
RESULTS: Suboptimal colonic distention was significantly more common with singledetector row CT and was present in at least one segment in 52% (40 of 77 patients) of examinations versus only 19% (30 of 160 patients) with multidetector row CT (P < .001). Mild respiratory artifacts were present in 61% (47 of 77 patients) of singledetector row CT examinations versus only 16% (26 of 160 patients) of multidetector row CT examinations (P < .001). Depiction of polyps larger than 10 mm was 89% (eight of nine polyps) for singledetector row CT and 80% (eight of 10 polyps) for multidetector row CT (P > .05).
CONCLUSION: CT colonography performed with multidetector row CT significantly improved the demonstration of colonic distention and depicted fewer respiratory artifacts compared with singledetector row CT. No significant differences in the depiction of polyps larger than 10 mm were demonstrated between single and multidetector row CT for a small number of polyps. Studies with a larger prevalence of clinically important polyps are needed for further evaluation of differences in polyp detection.
Index terms: Colon, CT, 75.12115 Colon neoplasms, CT, 75.12115 Colon neoplasms, diagnosis, 75.311, 75.32 Colonoscopy, 75.1289 Computed tomography (CT), comparative studies Computed tomography (CT), helical, 75.12115
| INTRODUCTION |
|---|
|
|
|---|
| MATERIALS AND METHODS |
|---|
|
|
|---|
Only patients (n = 237) who received subcutaneously administered glucagon were included in this study to eliminate bias related to colonic distention. Forty-six patients refused glucagon. Of the 237 patients who received glucagon, 77 patients underwent CT colonography with the use of a singledetector row CT scanner (HiSpeed Advantage; GE Medical Systems, Milwaukee, Wis), and 160 patients underwent CT colonography with use of a multidetector row CT scanner (LightSpeed; GE Medical Systems). The singledetector row CT study group consisted of 50 men and 27 women (age range, 4674 years; mean age, 63.3 years ± 6.4 [SD]). A similar age and sex distribution was present in the multidetector row CT study group, which consisted of 102 men and 58 women (age range, 4175 years; mean age, 63.5 years ± 6.5).
All patients underwent standard oral colonoscopic preparation (polyethylene glycol electrolyte solution and bisacodyl tablets) in the 24 hours preceding the CT colonographic examinations. Glucagon, 1.0 mL, was subcutaneously administered to all patients 10 minutes preceding the examination. Colonic insufflation to maximal patient tolerance without pain was performed by the same radiology nurse for 213 (90%) of 237 examinations, with the remaining 24 (10%) performed by a second experienced radiology nurse. Approximately 2 L of CO2 was manually administered through a rectal tube.
The scanning protocols for single and multidetector row CT used in this study are summarized in Table 1. The scanning protocol for singledetector row CT examinations has been previously reported (7,8) and includes the acquisition of an anteroposterior localization image followed by the acquisition of transverse images with the following helical CT protocol: 5-mm collimation, 6.5 mm/sec table speed (pitch of 1.3), 3-mm reconstruction interval, 512 x 512 matrix, field of view to fit, 70 mA, 1-second exposure time, 120 kVp, and 22-second breath holds. Three to four 22-second breath holds were needed to scan 40 cm by using singledetector row CT with a 3-cm overlap between acquisitions.
|
All images were obtained with the patient first in the supine position and then in the prone position, with reinsufflation as needed based on the scout image. The effective radiation dose was approximately 450 mrem (4.5 mSv) in men or 670 mrem (6.7 mSv) in women for both the single and multidetector row CT examinations.
Image processing was performed with a computer workstation (Ultra2; Sun Microsystems, San Jose, Calif) by using customized software developed at our institution. Each prone and supine image was graded for overall colonic distention as follows: 1, excellent throughout; 2, suboptimal in one to two segments; 3, suboptimal in more than two segments. Each image was graded for respiratory artifacts as follows: 0, none; 1, mild; 2, moderate; 3, severe and nondiagnostic. For grading purposes, the colon was divided into eight segments: cecum, ascending colon, hepatic flexure, transverse colon, splenic flexure, descending colon, sigmoid colon, and rectum.
The Wilcoxon rank sum test for ordered categories was used for evaluation of colonic distention and respiratory artifacts (10). For all tests, a P value of .05 or less indicated a statistically significant difference.
All colonoscopic examinations were performed immediately after CT colonography and were videotaped. Three radiologists (C.D.J., R.L.M., T.J.W.) evaluated the CT colonographic findings. Findings of each examination were independently evaluated by two of the three radiologists by using magnified axial CT images with lung window settings (width of 1,500 HU, level of -500 HU) to first screen the colon for abnormalities. Any suspicious areas were indicated with a cursor on the axial image, and fully interactive two- and three-dimensional endoluminal CT colonographic images were then immediately available at the specified location for problem solving. Polyp size and location were independently identified. Both observers were instructed to ignore polyps smaller than 5 mm, since these require more time to evaluate and are of little clinical importance (11).
CT colonographic results were compared with colonoscopic findings, which served as the standard. The Fisher exact test was used to compare the sensitivity and specificity of CT colonographic polyp detection with the use of each type of scanner. Single and multidetector row CT results were compared in two ways: by requiring only one observer to correctly identify the positive examination findings and then by requiring both observers to call the examination findings positive. CT colonographic results were also analyzed by randomly choosing one of the two observers and using his or her CT colonographic result, as well as by using a weighted mean among all of the radiologists. Interobserver results were also compared to determine statistically significant differences. The results were compared by a research nurse familiar with CT colonographic examinations. Any questions concerning the findings were discussed with a radiologist (C.D.J.), and the CT colonographic findings were reviewed. If the location of the polyp at colonoscopy was within one segment of the location at CT colonography and if the sizes were similar, the videotape was not reviewed for confirmation. If both CT colonographic reviewers reported a false-positive finding of 10 mm or larger, the videotape was sent to a gastroenterologist for review.
All CT colonographic examinations with findings of polyps 10 mm or larger that were identified at colonoscopy and missed at CT colonography were retrospectively reviewed to determine the cause of the false-negative finding. In addition, all false-positive CT colonographic findings of polyps 10 mm or larger were also reviewed to determine the cause of the error.
| RESULTS |
|---|
|
|
|---|
|
|
|
|
The majority of false-positive results of polyps larger than 10 mm were caused by retained stool. Stool was responsible for five of seven false-positive findings at singledetector row CT and six of 11 false-positive findings at multidetector row CT. One false-positive finding at singledetector row CT was attributed to a haustral fold that was mistaken for a polyp, and another cause remained indeterminate (ie, not definitely a fold or stool at CT colonography or videotaped colonoscopy). There were two cases at singledetector row CT that had both true- and false-positive findings of polyps larger than 10 mm, and these findings were ultimately classified as true-positive. The false-positive findings in these cases were attributed to stool and a haustral fold. The other false-positive findings at multidetector row CT were attributed to haustral folds (n = 2), an appendiceal stump (n = 1), and incomplete distention (n = 1). The remaining false-positive finding at multidetector row CT remained indeterminate.
| DISCUSSION |
|---|
|
|
|---|
Although in the optimal comparison of multi and singledetector row CT colonography the same patients would undergo both techniques, this was not performed because of radiation dose and patient compliance. Instead, this study involved two separate groups of patients who underwent either multi or singledetector row CT colonographic examination.
A comparison of these two groups demonstrated that respiratory artifacts were significantly reduced with multidetector row CT. Most (84%) multidetector row CT images in this study had no identifiable respiratory artifacts, while less than half (39%) of the singledetector row CT images had the same results. The decrease in respiratory artifacts at multidetector row CT may be due to the patients ability to perform a single breath hold well (21 seconds with multidetector row CT) rather than to perform the multiple long breath holds required with singledetector row CT. In addition, faster scanning with multidetector row CT eliminates image misregistration between overlapping acquisitions, since all the images are acquired in a single breath hold. Although respiratory artifacts were recognized by the experienced observers in this study and did not affect diagnostic results for polyps larger than 10 mm, they can prolong image evaluation time and cause diagnostic errors (8,12,14). For example, image misregistration artifacts with partially imaged folds can mimic polyps. Minimizing these types of artifacts and associated diagnostic errors would be most beneficial in reducing evaluation time and diagnostic errors for less-experienced observers.
Another potential advantage of multidetector row CT is improved colonic distention, as the patients are not required to retain the air for as long as with singledetector row CT. Faster scanning minimizes the risk of gas being expelled via the rectum or refluxing into the small bowel through an incompetent ileocecal valve. Results of a previous study (15) in 23 patients indicate that only eight (35%) of 23 examinations were graded as having optimal distention by using a singledetector row CT overlapping breath-hold acquisition in the supine and prone positions. In another study (16), patients were instructed to suspend respiration for as long as possible and then to resume shallow respirations while the last pelvic images were acquired. With this faster combined single breath hold and shallow respiratory technique lasting less than 60 seconds, optimal colonic distention was reported in all 60 examinations.
Similar results were found in our study, in which only 44% of singledetector row CT examinations performed with multiple overlapping breath-hold acquisitions were graded as having optimal distention. Colonic distention was improved by using a multidetector row CT breath-hold acquisition, with 80% of examinations having optimal distention. Other factors such as the experience of the person inflating the colon, variable patient tolerance levels, and anatomic differences (redundant colonic segments) may also have contributed to improved colonic distention. It is unlikely that variable insufflation of the colon contributed to the differences between results at single and multidetector row CT in this study, since 90% of the examinations were performed by the same nurse. It is possible, however, that with more experience the use of the colonic insufflation technique improved during the later multidetector row CT examinations, contributing to the improved results.
Multiple technical factors are also improved with multidetector row CT. For example, at multidetector row CT, faster imaging of narrow sections is allowed, compared with imaging at singledetector row CT. Also, at multidetector row CT, a wider x-ray beam is used so that the x-ray tube is used more efficiently, and cooling restrictions are reduced (13,9). Image acquisition time is also reduced to a single 21-second breath hold, allowing faster patient throughput.
Potential disadvantages of multidetector row CT include possible increased radiation doses for some protocols in which thinner sections are used, as well as the introduction of additional artifacts. The pitch used in this study was less than 1; that is, the table moved 15 mm per acquisition with a 20-mm x-ray beam. Therefore, each tube rotation overlapped by 25%, increasing the radiation dose if all other parameters were unchanged (4).
Prior to this study, multiple in vitro and in vivo multidetector row CT studies (9) were performed at our institution to maintain detection of 5-mm polyps while matching patient dose and minimizing image artifacts and breath holds. The use of 40 mAs for multidetector row CT (50 mA with a 0.8-second rotation speed) compared with 70 mAs for singledetector row CT (70 mA with a 1.0-second rotation speed) increased the center CT dose index by 10% (multidetector row CT, 0.37 rad [3.7 mGy], vs singledetector row CT, 0.35 rad [3.5 mGy]) (4,9). The effective dose, however, was equal between the two techniques (450670 mrem [4.56.7 mSv]) since the 3-cm overlap between breath-hold acquisitions with singledetector row CT was not required with singlebreath-hold multidetector row CT (9). No previously undescribed artifacts were identified in the in vitro or in vivo multidetector row CT studies. Therefore, by reducing the milliampere seconds and eliminating overlapping breath-hold acquisitions, the effective dose of multidetector row CT equals that of singledetector row CT, without a reduction in polyp depiction.
Theoretically, improved polyp detection is expected with the use of multidetector row CT, since collapsed colonic segments and respiratory artifacts have been identified as important causes of false-negative findings. An adequate assessment of polyp detection with the use of multidetector row CT compared with singledetector row CT, however, was difficult in this study, which included a high-risk screening population with a low prevalence of polyps. Only five and nine single and multidetector row CT examinations, respectively, revealed polyps larger than 10 mm. Although detection of polyps larger than 10 mm was slightly better at singledetector row CT (eight [89%] of nine polyps) compared with that at multidetector row CT (eight [80%] of 10 polyps), the numbers are too small to indicate a statistically significant difference. Improved polyp depiction with the use of multidetector row CT may be indicated in future studies, since in multidetector row CT, sections as thin as 1.25 mm may be obtained in the same time required for a current singledetector row CT examination. Although these thin sections would likely improve detection of smaller polyps, the larger number of images would increase image storage requirements and image processing times. In addition, the thinner section mode has an increased (by approximately 100%) radiation dose, compared with the mode used in this study (4). The detection of polyps smaller than 10 mm is of debatable clinical benefit (11) and will likely be addressed in future studies.
In conclusion, optimization of the CT colonographic examination continues with the implementation of multidetector row CT. CT colonography with the use of multidetector row scanners compared with singledetector row scanners demonstrates decreased respiratory artifacts, improved colonic distention, and faster scanning times. Detection of polyps larger than 10 mm was not significantly different between single and multidetector row scanners. Future studies may demonstrate improved polyp depiction not only by decreasing respiratory artifacts and improving colonic distention with faster scanning but by allowing rapid acquisition of sections that are even thinner than those currently used.
| FOOTNOTES |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. Sosna, J. Bar-Ziv, E. Libson, M. Eligulashvili, and A. Blachar CT Colonography: Positioning Order and Intracolonic Pressure Am. J. Roentgenol., October 1, 2008; 191(4): W175 - W180. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Kim, J. M. Lee, C.-I. Shin, H. C. Kim, J.-G. Lee, J. H. Kim, J. Y. Choi, H. W. Eun, J. K. Han, J. Y. Lee, et al. Effects of Spatial Resolution and Tube Current on Computer-aided Detection of Polyps on CT Colonographic Images: Phantom Study Radiology, August 1, 2008; 248(2): 492 - 503. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Sosna, T. Sella, O. Sy, P. T. Lavin, R. Eliahou, S. Fraifeld, and E. Libson Critical Analysis of the Performance of Double-Contrast Barium Enema for Detecting Colorectal Polyps >= 6 mm in the Era of CT Colonography Am. J. Roentgenol., February 1, 2008; 190(2): 374 - 385. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Hashimoto, S Kawashima, S Kameoka, Y Akiyama, T Honjoya, K Ejima, and K Sawada Comparison of image validity between cone beam computed tomography for dental use and multidetector row helical computed tomography Dentomaxillofac. Radiol., December 1, 2007; 36(8): 465 - 471. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Mang, A. Maier, C. Plank, C. Mueller-Mang, C. Herold, and W. Schima Pitfalls in Multi-Detector Row CT Colonography: A Systematic Approach RadioGraphics, March 1, 2007; 27(2): 431 - 454. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Veit-Haibach, C. A. Kuehle, T. Beyer, H. Stergar, H. Kuehl, J. Schmidt, G. Borsch, G. Dahmen, J. Barkhausen, A. Bockisch, et al. Diagnostic Accuracy of Colorectal Cancer Staging With Whole-Body PET/CT Colonography JAMA, December 6, 2006; 296(21): 2590 - 2600. [Abstract] [Full Text] [PDF] |
||||
![]() |
T Xiong, K McEvoy, D G Morton, S Halligan, and R J Lilford Resources and costs associated with incidental extracolonic findings from CT colonogaphy: a study in a symptomatic population Br. J. Radiol., December 1, 2006; 79(948): 948 - 961. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wessling, R. Fischbach, A. Borchert, H. Kugel, T. Allkemper, N. Osada, and W. Heindel Detection of Colorectal Polyps: Comparison of Multi-Detector Row CT and MR Colonography in a Colon Phantom Radiology, October 1, 2006; 241(1): 125 - 131. [Abstract] [Full Text] [PDF] |
||||
![]() |
A G Schreyer and R Kikinis Combined PET/CT colonography: is this the way forward? Gut, January 1, 2006; 55(1): 10 - 12. [Full Text] [PDF] |
||||
![]() |
P Veit, C Kuhle, T Beyer, H Kuehl, C U Herborn, G Borsch, H Stergar, J Barkhausen, A Bockisch, and G Antoch Whole body positron emission tomography/computed tomography (PET/CT) tumour staging with integrated PET/CT colonography: technical feasibility and first experiences in patients with colorectal cancer Gut, January 1, 2006; 55(1): 68 - 73. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Yasumoto, T. Murakami, H. Yamamoto, M. Hori, R. Iannaccone, T. Kim, H. Abe, M. Kuwabara, K. Yamasaki, N. Kikkawa, et al. Assessment of Two 3D MDCT Colonography Protocols for Observation of Colorectal Polyps Am. J. Roentgenol., January 1, 2006; 186(1): 85 - 89. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Burling, S. A. Taylor, S. Halligan, L. Gartner, M. Paliwalla, C. Peiris, L. Singh, P. Bassett, and C. Bartram Automated Insufflation of Carbon Dioxide for MDCT Colonography: Distension and Patient Experience Compared with Manual Insufflation Am. J. Roentgenol., January 1, 2006; 186(1): 96 - 103. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Halligan, D. G. Altman, S. A. Taylor, S. Mallett, J. J. Deeks, C. I. Bartram, and W. Atkin CT Colonography in the Detection of Colorectal Polyps and Cancer: Systematic Review, Meta-Analysis, and Proposed Minimum Data Set for Study Level Reporting Radiology, December 1, 2005; 237(3): 893 - 904. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. G. Mang, C. Schaefer-Prokop, A. Maier, E. Schober, G. Lechner, and M. Prokop Detectability of Small and Flat Polyps in MDCT Colonography Using 2D and 3D Imaging Tools: Results from a Phantom Study Am. J. Roentgenol., December 1, 2005; 185(6): 1582 - 1589. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Macari and E. J. Bini CT Colonography: Where Have We Been and Where Are We Going? Radiology, December 1, 2005; 237(3): 819 - 833. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Pickhardt, A. D. Lee, E. G. McFarland, and A. J. Taylor Linear Polyp Measurement at CT Colonography: In Vitro and in Vivo Comparison of Two-dimensional and Three-dimensional Displays Radiology, September 1, 2005; 236(3): 872 - 878. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Yee, N. N. Kumar, S. Godara, J. A. Casamina, R. Hom, G. Galdino, P. Dell, and D. Liu Extracolonic Abnormalities Discovered Incidentally at CT Colonography in a Male Population Radiology, August 1, 2005; 236(2): 519 - 526. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Park, H. K. Ha, M.-J. Kim, K. W. Kim, A. Y. Kim, D. H. Yang, M.-G. Lee, P. N. Kim, Y. M. Shin, S.-K. Yang, et al. False-Negative Results at Multi-Detector Row CT Colonography: Multivariate Analysis of Causes for Missed Lesions Radiology, May 1, 2005; 235(2): 495 - 502. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. P. Mulhall, G. R. Veerappan, and J. L. Jackson Meta-Analysis: Computed Tomographic Colonography Ann Intern Med, April 19, 2005; 142(8): 635 - 650. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ozgun, E. Rollven, L. Blomqvist, S. Bremmer, R. Odh, and A. Fransson Polyp Detection with MDCT: A Phantom-Based Evaluation of the Impact of Dose and Spatial Resolution Am. J. Roentgenol., April 1, 2005; 184(4): 1181 - 1188. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Chung, K. C. Huh, W. J. Choi, and J. K. Kim CT Colonography Using 16-MDCT in the Evaluation of Colorectal Cancer Am. J. Roentgenol., January 1, 2005; 184(1): 98 - 103. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Cohnen, C. Vogt, A. Beck, K. Andersen, W. Heinen, S. vom Dahl, V. Aurich, D. Haeussinger, and U. Moedder Feasibility of MDCT Colonography in Ultra-Low-Dose Technique in the Detection of Colorectal Lesions: Comparison with High-Resolution Video Colonoscopy Am. J. Roentgenol., November 1, 2004; 183(5): 1355 - 1359. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Macari, E. J. Bini, S. L. Jacobs, Y. W. Lui, S. Laks, A. Milano, and J. Babb Significance of Missed Polyps at CT Colonography Am. J. Roentgenol., July 1, 2004; 183(1): 127 - 134. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Gluecker, J. G. Fletcher, T. J. Welch, R. L. MacCarty, W. S. Harmsen, J. R. Harrington, D. Ilstrup, L. A. Wilson, K. E. Corcoran, and C. D. Johnson Characterization of Lesions Missed on Interpretation of CT Colonography Using a 2D Search Method Am. J. Roentgenol., April 1, 2004; 182(4): 881 - 889. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. F. Geenen, S. M. Hussain, F. Cademartiri, J.-W. Poley, P. D. Siersema, and G. P. Krestin CT and MR Colonography: Scanning Techniques, Postprocessing, and Emphasis on Polyp Detection RadioGraphics, January 1, 2004; 24(1): e18 - e18. [Abstract] [Full Text] |
||||
![]() |
R. Iannaccone, A. Laghi, C. Catalano, J. A. Brink, F. Mangiapane, S. Trenna, F. Piacentini, and R. Passariello Detection of Colorectal Lesions: Lower-Dose Multi-Detector Row Helical CT Colonography Compared with Conventional Colonoscopy Radiology, December 1, 2003; 229(3): 775 - 781. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Sosna, M. M. Morrin, J. B. Kruskal, P. T. Lavin, M. P. Rosen, and V. Raptopoulos CT Colonography of Colorectal Polyps: A Metaanalysis Am. J. Roentgenol., December 1, 2003; 181(6): 1593 - 1598. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Pickhardt Three-Dimensional Endoluminal CT Colonography (Virtual Colonoscopy): Comparison of Three Commercially Available Systems Am. J. Roentgenol., December 1, 2003; 181(6): 1599 - 1606. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. W. Lui, M. Macari, G. Israel, E. J. Bini, H. Wang, and J. Babb CT Colonography Data Interpretation: Effect of Different Section Thicknesses--Preliminary Observations Radiology, December 1, 2003; 229(3): 791 - 797. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Taylor, S. Halligan, C. I. Bartram, P. R. Morgan, I. C. Talbot, N. Fry, B. P. Saunders, K. Khosraviani, and W. Atkin Multi-Detector Row CT Colonography: Effect of Collimation, Pitch, and Orientation on Polyp Detection in a Human Colectomy Specimen Radiology, October 1, 2003; 229(1): 109 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wessling, R. Fischbach, N. Meier, T. Allkemper, J. Klusmeier, K. Ludwig, and W. Heindel CT Colonography: Protocol Optimization with Multi-Detector Row CT--Study in an Anthropomorphic Colon Phantom Radiology, September 1, 2003; 228(3): 753 - 759. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Macari, E. J. Bini, S. L. Jacobs, N. Lange, and Y. W. Lui Filling Defects at CT Colonography: Pseudo- and Diminutive Lesions (The Good), Polyps (The Bad), Flat Lesions, Masses, and Carcinomas (The Ugly) RadioGraphics, September 1, 2003; 23(5): 1073 - 1091. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Sosna, M. M. Morrin, J. B. Kruskal, R. J. Farrell, I. Nasser, and V. Raptopoulos Colorectal Neoplasms: Role of Intravenous Contrast-enhanced CT Colonography Radiology, July 1, 2003; 228(1): 152 - 156. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Gluecker, C. D. Johnson, W. S. Harmsen, K. P. Offord, A. M. Harris, L. A. Wilson, and D. A. Ahlquist Colorectal Cancer Screening with CT Colonography, Colonoscopy, and Double-Contrast Barium Enema Examination: Prospective Assessment of Patient Perceptions and Preferences Radiology, May 1, 2003; 227(2): 378 - 384. [Abstract] [Full Text] |