Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hennemeyer, C. T.
Right arrow Articles by Derdeyn, C. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hennemeyer, C. T.
Right arrow Articles by Derdeyn, C. P.
(Radiology. 2001;219:732-737.)
© RSNA, 2001


Experimental Studies

In Vitro Evaluation of Platinum Guglielmi Detachable Coils at 3 T with a Porcine Model: Safety Issues and Artifacts1

Charles T. Hennemeyer, BS, Karsten Wicklow, PhD, David A. Feinberg, MD, PhD and Colin P. Derdeyn, MD

1 From the Neuroradiology Section (D.A.F., C.P.D.) and Neuroimaging Laboratory (C.T.H., C.P.D.), Mallinckrodt Institute of Radiology, Washington University School of Medicine, 510 S Kingshighway Blvd, St Louis, MO 63110; and Research and Development Division, Siemens Medical Systems, Iselin, NJ (K.W.). Received August 10, 2000; revision requested October 3; revision received November 16; accepted December 11. Address correspondence to C.P.D. (e-mail: derdeync@mir.wustl.edu).


    ABSTRACT
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
PURPOSE: To evaluate safety-related issues and imaging artifacts of Guglielmi detachable coils in vitro with 3-T magnetic resonance (MR) imaging.

MATERIALS AND METHODS: Two aneurysm models were constructed: one from porcine carotid artery and the other from a pharmaceutical capsule. Both were filled with Guglielmi detachable coils. The models were tested with a 3-T MR imager for heating, deflection, and imaging artifact. Testing for heating and deflection was performed (a) at static points both inside and outside the bore, (b) during movement into the imager, and (c) during clinical imaging sequences.

RESULTS: No change in temperature was measured during movement into the imager bore or at different points within the bore. No differences in heating from radio-frequency energy were found between aneurysm models and controls. Similarly, no evidence of deflection of the coil mass (capsule model) was found. Minor susceptibility artifacts were found in the readout direction during gradient-echo sequences. Magnetic field mapping showed no induced field inhomogeneity.

CONCLUSION: MR imaging at field strengths of 3 T in patients with aneurysms treated with Guglielmi detachable coils is safe. Imaging artifacts are likely to be minimal.

Index terms: Aneurysm, cerebral, 17.73 • Head, MR, 10.121411, 10.121412, 10.12142 • Magnetic resonance (MR), safety, 10.121411, 10.121412, 10.12142 • Radiology and radiologists, iatrogenic injury


    INTRODUCTION
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
With the advent and increasing clinical applications of magnetic resonance (MR) imaging, the effects of strong magnetic fields on implanted metallic objects, such as vascular stents, clips, and filters, have been extensively investigated. Some devices, previously thought to have no ferromagnetic potential, have been found to possess some qualities of ferrous metals (13). Explanations for this susceptibility have included such things as the metallurgical composition of alloys and degree of cold working during manufacture (4), heat treatment (annealing), and undetermined factors such as resterilization causing variability in the product of one maker (5).

Guglielmi detachable coils (GDCs) (Boston Scientific/Target Therapeutics, Boston, Mass), used for the endovascular occlusion of intracranial aneurysms, have been less extensively evaluated than other intravascular devices, particularly at high field strengths (6,7). To our knowledge, researchers in prior studies have not investigated several important factors that might pose dangers to patients with intracranial aneurysms treated with GDCs, such as (a) heating at all points from the imaging center to points outside the portal, (b) heating during movement into and out of the imager, (c) heating during imaging with high radio-frequency (RF) energy commonly used to image the human head (relative to the small amount needed to image a small coil alone), and (d) horizontal and longitudinal deflection forces at these static positions during movement into and out of the imager and during imaging sequences. This study was conducted to evaluate safety-related issues and imaging artifacts of GDCs in vitro with 3-T MR imaging.


    MATERIALS AND METHODS
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
Experimental Models
Two coiled aneurysm models were constructed for these experiments. The first was meant to duplicate as closely as possible an actual coiled aneurysm. This model was constructed from a porcine carotid artery (6-mm outer diameter) obtained after death from an unrelated study, with animal care committee approval. The artery was closed at one end with a 2-0 silk suture, while the lumen at the other end was narrowed to 50% of the luminal diameter with a second 2-0 silk suture, to approximate an aneurysmal neck. The experimental aneurysm was placed in a saline bath. With fluoroscopic guidance, five 0.010-inch-diameter GDCs were placed by means of a 0.014-inch-diameter microcatheter (Prowler; Cordis Endovascular, Miami Lakes, Fla). The first coil was 6 mm x 10 cm, followed by a 4 mm x 6 cm, 3 mm x 6 cm, 2 mm x 4 cm, and 2 mm x 4 cm coils. Each coil was electrolytically detached by using the standard system cables. The placement of the coils and the selection of coil size were performed in a manner to duplicate, as closely as possible, the actual clinical procedure that is used to treat human intracranial aneurysms. This model was kept in a syringe filled with 4 mL of 0.9% saline (Fig 1).



View larger version (91K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1. Fluoroscopic image of both aneurysm models with a 10-mm marker in the center. A, Porcine model. B, Capsule model. The coils are more closely packed in the porcine model compared with the capsule model (as evidenced by the relative lack of lucent pockets in the porcine model). The faint vertical shadow (arrow) above and below the coil mass shows the artery.

 
The second model was constructed from an empty pharmaceutical capsule with the approximate dimensions of 8 x 10 mm (Fig 1). This second model was designed such that it would have minimal mass (other than the coils) to allow the most sensitive measurements of movement. A hole was made in one end of the capsule with a 19-gauge needle. The coil introducer was inserted through the hole, and the coils were placed by hand. The coil was cut just distal to the detachment zone by using a pair of scissors, and a few millimeters of coil left outside the needle hole was pushed into the capsule by hand. Five 0.010-inch-diameter GDCs were placed, beginning with an 8 mm x 20 cm, then a 7 mm x 30 cm, a 5 mm x 10 cm, a 4 mm x 6 cm, and finally a 3 mm x 6 cm coil. The capsule model was kept dry to reduce its mass and heat capacity.

Temperature Probe Placement
Heating measurements were made by using both the capsule and the artery models. Temperatures were recorded with four simultaneous probes in real time by using an MR-compatible in situ fiberoptic temperature-sensing device (model 755 multichannel fluoroptic thermometer; Luxtron, Santa Clara, Calif) sensitive to 0.1°C. Temperature data were recorded by a single investigator (C.T.H.).

MR-compatible Box
To precisely locate the models during both the heating and deflection experiments, a foam core box was constructed to fit the RF head coil of a 3-T superconducting MR imager (Siemens Medical Systems, Iselin, NJ) (Fig 2). This MR-compatible box was 1 m in length, with a square cross-sectional shape and a diagonal measurement of 265 mm to match the inside diameter of the RF head coil. A sliding floor was constructed for the box to allow placement at exact coordinates in the x, y, and z planes (the z plane was along the length of the MR imager bore, including the space outside the portal). The box was then placed in the RF coil on the patient gantry and was driven partially into the bore up to and including the imaging volume (Fig 2). Twenty-five centimeters of the box extended outside of the bore, and 75 cm of the box was located in the bore and included the imaging volume.



View larger version (122K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2. Box apparatus used to hold the models and measurement equipment at precise locations along the x, y, and z axes. A sliding floor (vertical dotted arrow) allowed movement in or out of the imager (horizontal open arrow). The capsule model (horizontal dotted arrow) is suspended by a thread in front of a grid of lines representing deflection angles along the z axis (the model is held slightly away from the grid so as not to contact the surface). The balsa stick with a mirror (vertical open arrow) attached at 45° allows the observer to accurately monitor deflection angles deep in the bore.

 
Assessment of Heating: Static, during Motion, and during Imaging
For the static and dynamic measurements, the probes in the porcine model were placed in a syringe filled with 4 mL of saline, in four locations: (a) in the coil mass, (b) between the mass and the artery wall, (c) directly on the outside surface of the artery, and (d) in the surrounding saline. In the coiled capsule model (suspended in a dry syringe), the temperature probes were placed in four locations: in the dry mass of coils, in the air immediately surrounding the suspended capsule, and in two separate control syringes containing 4 mL of 0.9% saline. Heating experiments were conducted at the imaging center at static points, ranging from 25 cm outside the portal to 75 cm in the bore. After allowing both models to equilibrate to room temperature for 2 hours, six temperature measurements were obtained for each probe at each position. Six measurements were also obtained during movement into and out of the bore at the same points, as the table was driven in and out of the magnet for a total of four complete cycles. The velocity of the table was 0.1 m/sec (similar to that used for a clinical examination).

For the imaging experiment, the four temperature probes were located (a) in the coil mass of the porcine model suspended in 4 mL of saline, (b) in a control artery (without coils) suspended in saline, (c) in a control syringe of the same volume, and (d) in the dry coil mass of the capsule model surrounded by air. Measurements were obtained during two sequences: MR angiographic sequence (29/5 [repetition time msec/echo time msec], 20° flip angle, 0.89-mm section thickness, duration of 6 minutes 33 seconds) and a fast spin-echo sequence (2,510/104, 180° flip angle, 1.0-mm section thickness, duration of 27 minutes). The model and the controls were placed in the center of the imaging volume and were surrounded by five 200-mL intravenous bags of saline to achieve the specific absorption rate limit of 3.3 W per kilogram of body weight for the human head. An RF strength of 119–160 V was achieved, maximizing RF energy deposition. Twenty temperature measurements were recorded for each probe during each imaging sequence.

Assessment of Attractive Forces: Static, during Motion, and during Imaging
Attractive forces were quantitatively measured by using a well-established method (1,8). The dry capsule model was suspended by a thread to allow deflection in the direction of the most rapidly increasing field or highest gradient, while the angle formed from the vertical axis was measured (Fig 1) by two investigators (C.T.H., K.W.). The force F acting on this mass orthogonal to gravity was calculated with the following equation: F = mg tan {theta}, where m is mass (in grams), g is the acceleration of gravity (980 cm/sec2), and {theta} is the angle made with the vertical axis. The attractive force (in dynes) can then be converted to a more intuitive unitless number by using the equation F(g) = F(dyne)/mg (9), where g is the force of gravity.

Attractive forces were measured (C.T.H., K.W.) at numerous static points in the x, y, and z planes, where x is the left-to-right horizontal direction in the bore, y is the vertical distance from the top to the bottom of the bore, and z is the distance from the portal outside the bore or distance from the portal into the bore and the imaging volume. For measurements along the z plane, x was 0 and y was 0; and for measurements along the x plane, y was 0 and x was 0.

Measurements of deflection were obtained at 1- and 10-cm increments along the x and z planes. The z plane was measured not only outside the bore and at the portal but also inside the bore. Since the deflection would be expected along the z direction when x and y are zero, it is impossible to visualize deflection deep in the bore. However, in this experiment, a 2 x 2 cm front-surface mirror mounted at 45° to a 1-m-long balsa rod allowed visualization of even the slightest movement deep within the magnet, preventing parallax error induced by the observer’s eye being placed offline with the string and the protractor (Fig 2).

While longitudinal z and horizontal x forces were measured with this method, vertical y forces were not directly measured, and the reason for this is twofold. First, vertical forces cannot be measured in the same way as the other forces because they are parallel to gravity. Second, horizontal forces, not unlike the vertical forces, are measured at increments approaching the wall of the magnet. Therefore, vertical forces could be estimated from horizontal forces.

The capsule model was observed for deflection for 30 seconds to 1 minute at each position during static measurements and imaging sequences. For the measurements obtained during gantry motion, the inertia of the capsule was accounted for by building a simple mechanism with a fixed-length string attached to a plate that held the capsule as the table accelerated but let the capsule loose once a constant velocity was achieved. These observations were made during four cycles of motion completely into and out of the imager at a rate of 0.1 m/sec.

Assessment of Artifacts during Imaging
Images were acquired with the porcine model suspended in both saline and gadodiamide (Nycomed Amersham, Princeton, NJ) at the appropriate concentration. The saline bags were used again and were placed out of the phase-encoding direction. The selected imaging sequences included the following: (a) field-mapping sequence, 300/100, 90° flip angle, one signal acquired, duration of 1 minute 30 seconds, 5.0-mm section thickness; (b) gradient-echo fast low-angle shot sequence, 250/18, 50° flip angle, two signals acquired, duration of 3 minutes 13 seconds, 2.0-mm section thickness, 40-kHz bandwidth; (c) fast spin-echo T2-weighted sequence, 3,000/96, 180° flip angle, one signal acquired, duration of 3 minutes 40 seconds, 2.0-mm section thickness, 33-kHz bandwidth; (d) spin-echo T1-weighted sequence, 290/14, 90° flip angle, six signals acquired, duration of 11 minutes, 1.0-mm section thickness, 143-kHz bandwidth; and (e) spin-echo T1-weighted sequence, 300/20, 90° flip angle, one signal acquired, duration of 4 minutes 32 seconds; 2.0-mm section thickness, 40-kHz bandwidth.

Higher-spatial-resolution T1-weighted images (sequence d compared with sequence c) were obtained by using thinner sections, higher bandwidths, and more repetitions. Images were visually inspected by all authors for subjective evidence of distortion of length or width, as well as for changes in signal intensity in or near the coil mass.


    RESULTS
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
Heating
In both the porcine and dry capsule models, no heating was encountered at any point in the static magnetic field in or surrounding the MR imager (Fig 3). No heat production in the models or controls occurred during movement into or out of the imager. Temperatures decreased during the first portion of the experiment when the models were moved from the equilibrium point at the edge of the MR suite to positions in the magnetic field along the bore of the magnet. This temperature decrease averaged 0.7°C in the porcine model and control and 0.9°C in the dry capsule model and control (Fig 3).



View larger version (20K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3a. Line graphs show a steady temperature decrease in both the (a) porcine and (b) capsule models, as well as in the control models during sequential measurements along the length of the bore. These positions range from 25 cm outside the portal to 75 cm within the bore. Each line represents a different temperature probe. Six measurements were made at each position. Each point is the average of these six values, and the error bars indicate the 95% confidence limits. In a, {bullet} = probe between the coil and artery wall, {circ} = probe in the coils, {blacktriangledown} = probe outside aneurysm wall, and {triangledown} = probe in the saline bath. In b, {bullet} = probe in the coils, {circ} = probe in air, {blacktriangledown} = probe in control syringe 1, and {triangledown} = probe in control syringe 2. X-axis points for both graphs are as follows: 1, -25 cm; 2, -20 cm; 3, -15 cm; 4, -5 cm; 5, 5 cm; 6, 15 cm; 7, 25 cm; 8, 45 cm; and 9, 75 cm. Locations 10 and 11 represent movement into and out of the imager.

 


View larger version (23K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 3b. Line graphs show a steady temperature decrease in both the (a) porcine and (b) capsule models, as well as in the control models during sequential measurements along the length of the bore. These positions range from 25 cm outside the portal to 75 cm within the bore. Each line represents a different temperature probe. Six measurements were made at each position. Each point is the average of these six values, and the error bars indicate the 95% confidence limits. In a, {bullet} = probe between the coil and artery wall, {circ} = probe in the coils, {blacktriangledown} = probe outside aneurysm wall, and {triangledown} = probe in the saline bath. In b, {bullet} = probe in the coils, {circ} = probe in air, {blacktriangledown} = probe in control syringe 1, and {triangledown} = probe in control syringe 2. X-axis points for both graphs are as follows: 1, -25 cm; 2, -20 cm; 3, -15 cm; 4, -5 cm; 5, 5 cm; 6, 15 cm; 7, 25 cm; 8, 45 cm; and 9, 75 cm. Locations 10 and 11 represent movement into and out of the imager.

 
During imaging with an angiographic sequence, no heating was encountered in the coiled artery, control artery, saline syringe, or dry coiled capsule. However, the second imaging sequence consisting of maximum RF energy deposition at the specific absorption rate limit of 3.3 W/kg, for a duration of 27 minutes, produced a temperature increase of less than 1°C in both the coiled aneurysm and controls (Fig 4).



View larger version (28K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 4. Line graph shows a slight increase in temperature for coiled aneurysm models and controls during prolonged (27 minutes) maximal specific absorption rate (RF energy of 3.3 W/kg) imaging sequence. A similar temperature increase of less than 1°C can be seen in the coiled artery ({bullet}), artery left uncoiled ({circ}), saline-filled syringe ({blacktriangledown}), and dry coiled capsule in air ({triangledown}).

 
Deflection
Deflection of GDCs from the vertical position was not encountered at any position in the x, y, or z direction. Additionally, no deflection was observed during movement into and out of the imager. Subjectively, no visible torque was exerted in the models while they were suspended from a string, at any point in the experiment. No measurable forces could be demonstrated to be acting on the GDCs, indicating lack of ferromagnetic or paramagnetic properties.

Imaging Artifact
Magnetic field mapping sequences failed to demonstrate any visible alterations in field homogeneity caused by GDCs (Fig 5). Minor alterations in field strength were present at the interface between the syringe (containing the gadodiamide-saline solution) and the normal saline–filled cup in which it was suspended. Gradient-echo sequences caused a distortion in size. The coil mass and the surrounding artery wall were enlarged or drawn out (Fig 6) in the readout direction. This artifact was greatest at lower bandwidths. A narrow band of increased signal intensity was noted surrounding the coil mass on spin-echo images (Figs 7 9). This high signal intensity was particularly apparent on T2-weighted fast spin-echo images (Fig 7), as compared with T1-weighted images (Figs 8, 9). It was also less visible on a standard T1-weighted image (Fig 9). Imaging of the dry capsule model was impossible, since no signal intensity from the capsule, coils, or air surrounding either could be visualized.



View larger version (156K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 5. Field-mapping MR image (300/100, 90° flip angle, 5.0-mm section thickness) shows no apparent distortion due to coil mass. The porcine model is suspended in a syringe filled with gadodiamide-saline solution, and the syringe is suspended in a foam cup filled with normal saline. Concentric lines represent magnetic field strength lines, with a difference of 50 Hz between the lines. Notice the moderate distortion (solid arrows) due to the interface of saline and gadodiamide-saline solution; however, also notice the relative lack of field distortion (open arrow) around the GDC mass.

 


View larger version (132K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 6. Gradient-echo fast low-angle shot MR image (250/18, 50° flip angle, two signals acquired, 2.0-mm section thickness) with readout-direction distortion. The width of the coil mass (porcine model suspended in gadodiamide-saline solution) along the readout direction is exaggerated (arrows).

 


View larger version (122K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 7. Ring of high-signal-intensity artifact (arrows) on a T2-weighted fast spin-echo MR image (3,000/96, 180° flip angle, one signal acquired, 2.0-mm section thickness) of the porcine model suspended in gadodiamide-saline solution.

 


View larger version (150K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 8. High-spatial-resolution T1-weighted spin-echo MR image (290/14, 90° flip angle, six signals acquired, 1.0-mm section thickness) of the coiled porcine model suspended in gadodiamide-saline solution. Notice the arterial wall sutured down to a "neck" (left long black arrow) and a suture occluding the arterial lumen (right long black arrow), which together form an aneurysm. Note the GDC mass (white arrow). A saline rim (short black arrows) can be seen between the coil mass and the aneurysm wall, as well as smaller resolvable pockets of saline between the tightly packed coils themselves.

 


View larger version (182K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 9. Conventional T1-weighted spin-echo MR image (300/20, 90° flip angle, one signal acquired, 2.0-mm section thickness) of the porcine model suspended in gadodiamide-saline solution. Band of high signal intensity (arrows) surrounding the coils is less apparent than in

 

    DISCUSSION
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
GDCs are composed of the raw material platinum, which is theoretically neither ferromagnetic nor paramagnetic. However, strong magnetic or RF fields could potentially produce heating or deflection forces for a number of reasons. Various nonmagnetic steel alloys can acquire ferromagnetic domains during their manufacture into complex shapes while they are being bent and cut into their final configuration; this process is termed "cold working" (10). Cold working can alter the crystalline structure from a nonmagnetic, face-centered cubic or austenitic crystal, into a body-centered cubic or martensitic crystal possessing ferromagnetic properties (4). Devices that are cut also develop shear faces that possess magnetic susceptibility (1). GDCs, which are manufactured with extensive cold working into both a microscopic and macroscopic coiled configuration, might acquire substantial ferromagnetic domains. Additionally, there may be a high cut surface–to-mass ratio in a coiled aneurysm composed of multiple individual coils.

The potential for heating of GDCs is founded on the fact that current may be induced in an electric conductor, as the magnetic and RF field gradients are applied or as the conductor moves through the fringe region of the static field outside the imager. Furthermore, a circuit with variable resistance created by overlapping GDCs may lead to heating. Arcing and heat production between various cut ends in the coil mass are also possible. Arcing has been tested in copper and steel rings, without a positive result (8).

To date, no heating has been demonstrated in GDCs; however, to our knowledge, in only one study (6) has the heating in GDCs been tested. In this experiment, measurements were made in a static position (without movement into or out of the imager), and the position in the bore was not stated.

In the present study, we evaluated two additional variables that might lead to induced current and heating (motion in the magnetic field and image acquisition). According to the Faraday law, a voltage is induced in any conducting material placed in a time-varying magnetic field. If a circuit is produced, the current may create localized heating or arcing. MR imaging requires three electromagnetic fields: (a) a static and constant field used to align the protons, (b) a smaller rapidly time-varying second magnetic field for spatial localization, and (c) a time-varying RF field. In modern superconducting MR imagers, the static magnetic field corresponds to the stated type of imager (eg, 1.5 or 3 T). While this field is not time-varying and therefore should not induce a current in a conductor, it becomes time-varying when the patient is moved from outside the bore, where the field is weakest, to the inside of the bore, where it is strongest.

In fact, one of the few reported (11) cases of ferromagnetic complications of MR imaging occurred during the movement of the patient out of the magnet, when an intraocular foreign body caused damage and blindness. To our knowledge, no metallic implants have been tested for movement or heating while being moved into or out of the MR imager (112). The second and third time-varying electromagnetic fields are active during imaging sequences. In these prior studies (112), the effect of imaging sequences on heating was not systematically evaluated.

We observed no evidence of heating at any static location or with movement into or out of the field. Heating was not observed during a standard angiographic sequence. Heating was measured during a 27-minute sequence designed to deposit the maximum allowable RF energy. However, the observed increase in temperature was less than 1°C and was similar between the control saline-filled syringe, the control artery, and the coils (Fig 9). These data were not independent samples, and therefore, a parametric test for differences would be invalid. This experiment was designed to provide the maximum sensitivity to heating: no simulation of blood flow, a small fluid volume for the arterial model, and no fluid volume in the capsule model.

Platinum coils are thought to be nonferromagnetic and are, at most, paramagnetic. Paramagnetic objects have the potential to align along the direction of the lines of magnetic flux and display a force proportional to not only the gradient but also to the field strength (12). Therefore, a 3-T imager could have more effect on platinum coils than a 1.5-T imager, as the magnitude of the field strength is much greater.

An essential part of any attractive force is the size of the gradient. While the static field of an MR imager is uniform in the imaging portion of the magnet, a steep gradient exists outside of the isocenter (9). The point of highest gradient is different for different MR imagers. Kagetsu and Litt (9) have found that in resistive-type imagers, the highest gradient is at the portal, and in superconducting types (as was used in this experiment), the highest point is well within the bore. Despite this, most previous tests (1) have been performed only at the portal of the magnet. In this experiment, forces were measured at all points inside and outside of the bore. Furthermore, the gradient was expected to be steeper, regardless of its location, in a 3-T magnet than in a 1.5-T magnet.

Eddy currents are another potential hazard when moving a conducting material through a magnetic field. When any magnetic field, static or time-varying, is applied to a moving conductor, eddy currents are produced in the conductor. This phenomenon generates a retarding force, which opposes any further movement (9). This may be demonstrated by holding a conducting but nonferromagnetic aluminum plate of any dimension near the bore of an MR imager. As the experimenter attempts to move the plate, a considerable retarding force will be felt to any movement, which becomes more apparent as the plate is moved within the magnet, where the field is strongest. Eddy currents induced in the coils during movement across the fringe regions of a magnetic field were either insufficient or nonexistent, causing neither apparent heating nor deflection forces.

In this study, deflection was not encountered at any point in the length of the magnet outside the homogeneous imaging volume. Time-varying gradients of RF and magnetic fields during imaging were also ineffective in producing any measurable deflection. Neither magnetic nor RF-induced torque (the rotation of an object about an axis) was objectively tested in this experiment. Qualitatively, no torque was observed or felt during the experiment.

With field-mapping sequences (Fig 5), the field homogeneity can be visualized even with slight distortions of that field (each concentric line in Fig 5 represents a small 50-Hz increment). GDCs produced no discernable field inhomogeneity during this sequence. Alterations that did occur arose from the interface of the contrast material–filled syringe and the normal saline–filled cup, rather than the suspended coils themselves. Resolution of detail in the tight coil mass of the porcine model was difficult to achieve with only normal saline. The relatively low signal intensity that arose from the small quantities of normal saline was difficult to distinguish from the signal void of the coils themselves. The addition of gadodiamide increased the signal intensity from the fluid and allowed resolution of fine detail in the tight coil mass (Fig 9).

Two imaging artifacts were identified. Gradient-echo sequences (250/18, 50° flip angle) produced an elongation of the dimensions of the model only along the readout direction (Fig 6). The second artifact was a hyperintense rim of signal intensity that could be seen immediately adjacent to the coil mass (Figs 79). This bright artifact was most apparent on fast spin-echo images and at higher echo time values (Fig 7). This effect was also more pronounced during imaging sequences that used lower bandwidths (33 kHz in Fig 9 vs 143 kHz in Fig 8). This artifact has been previously described (6,1315) with MR imaging of GDC-treated aneurysms at 1.5 T.

Practical application: We found no evidence for heating or deflection of GDCs in a comprehensive series of in vitro experiments at 3 T. We rigorously evaluated several experimental conditions not previously reported, such as eddy currents, phantoms used to increase RF energy, motion, position in the MR imager, and simultaneous temperature monitoring. We conclude that MR imaging at 3 T of patients with GDC-treated intracranial aneurysms is safe. Furthermore, imaging artifacts are minimal, and high-spatial-resolution structural and functional imaging is likely to be feasible.


    ACKNOWLEDGMENTS
 
The authors thank Pete Essex, BA, of Target Therapeutics/Boston Scientific for donation of the GDCs used for these experiments, as well as Jeffrey J. Neil, MD, PhD, for the use of his MR-compatible fiberoptic thermometer.


    FOOTNOTES
 
Abbreviations: GDC = Guglielmi detachable coil, RF = radio frequency

Author contributions: Guarantor of integrity of entire study, C.P.D.; study concepts, C.T.H., C.P.D., D.A.F.; study design, C.T.H., C.P.D.; literature research, C.T.H., C.P.D.; experimental studies, C.T.H., C.P.D.; data acquisition, C.T.H., K.W.; data analysis/interpretation, C.T.H., K.W., C.P.D.; statistical analysis, C.P.D.; manuscript preparation, C.T.H.; manuscript definition of intellectual content, C.P.D.; manuscript editing and revision/review, K.W., C.P.D., D.A.F.; manuscript final version approval, all authors.


    REFERENCES
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 

  1. New PF, Rosen BR, Brady TJ, et al. Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology 1983; 147:139-148.[Abstract/Free Full Text]
  2. Shellock FG, Crues JV. High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces. AJR Am J Roentgenol 1988; 151:389-392.[Abstract/Free Full Text]
  3. Shellock FG, Morisoli S, Kanal E. MR procedures and biomedical implants, materials, and devices: 1993 update. Radiology 1993; 189:587-599.[Abstract/Free Full Text]
  4. Dujovny M, Kossovsky N, Kossovsky R, et al. Aneurysm clip motion during magnetic resonance imaging: in vivo experimental study with metallurgical factor analysis. Neurosurgery 1985; 17:543-548.[Medline]
  5. Kanal E, Shellock FG, Lewin JS. Aneurysm clip testing for ferromagnetic properties: clip variability issues. Radiology 1996; 200:576-578.[Abstract/Free Full Text]
  6. Hartman J, Nguyen T, Larsen D, Teitelbaum GP. MR artifacts, heat production, and ferromagnetism of Guglielmi detachable coils. AJNR Am J Neuroradiol 1996; 18:497-501.[Abstract]
  7. Marshall MW, Teitelbaum GP, Kim HS, Deveikis J. Ferromagnetism and magnetic resonance artifacts of platinum embolization microcoils. Cardiovasc Intervent Radiol 1991; 14:163-166.[Medline]
  8. Davis PL, Crooks L, Arakawa M, et al. Potential hazards in NMR imaging: heating effects of changing magnetic fields and RF fields on small metallic implants. AJR Am J Roentgenol 1981; 137:857-860.[Abstract/Free Full Text]
  9. Kagetsu NJ, Litt AW. Important considerations in measurement of attractive forces on metallic implants in MR imagers. Radiology 1991; 179:505-508.[Abstract/Free Full Text]
  10. Teitelbaum GP, Bradley WG, Jr, Klein BD. MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents, and coils. Radiology 1988; 166:657-664.[Abstract/Free Full Text]
  11. Kelly WM, Pagan PG, Person JA, Santiago AG, Solomon MA. Ferromagnetism of intraocular foreign body causes unilateral blindness after MR study. AJNR Am J Neuroradiol 1986; 7:243-245.[Medline]
  12. Kanal E, Shellock FG. The value of published data on MR compatibility of metallic implants and devices. AJNR Am J Neuroradiol 1993; 15:1394-1396.[Medline]
  13. Derdeyn CP, Graves VB, Turski PA, Masaryk AM, Strother CM. MR angiography of saccular aneurysms after treatment with Guglielmi detachable coils: preliminary experience. AJNR Am J Neuroradiol 1997; 18:279-286.[Abstract]
  14. Gonner F, Heid O, Remona L, et al. MR angiography with ultrashort echo time in cerebral aneurysms treated with Guglielmi detachable coils. AJNR Am J Neuroradiol 1998; 19:1324-1328.[Abstract]
  15. Kahara VJ, Seppanen SK, Ryymin PS, Mattila P, Laasonen EM. MR angiography with three-dimensional time-of-flight and targeted maximum-intensity-projection reconstructions in the follow-up of intracranial aneurysms embolized with Guglielmi detachable coils. AJNR Am J Neuroradiol 1999; 20:1470-1475.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
CirculationHome page
G. N. Levine, A. S. Gomes, A. E. Arai, D. A. Bluemke, S. D. Flamm, E. Kanal, W. J. Manning, E. T. Martin, J. M. Smith, N. Wilke, et al.
Safety of Magnetic Resonance Imaging in Patients With Cardiovascular Devices: An American Heart Association Scientific Statement From the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: Endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance
Circulation, December 11, 2007; 116(24): 2878 - 2891.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
V. Costalat, E. Lebars, L. Sarry, A. Defasque, E. Barbotte, H. Brunel, G. Bourbotte, and A. Bonafe
In Vitro Evaluation of 2D-Digital Subtraction Angiography versus 3D-Time-of-Flight in Assessment of Intracranial Cerebral Aneurysm Filling after Endovascular Therapy
AJNR Am. J. Neuroradiol., January 1, 2006; 27(1): 177 - 184.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
C. B. L. M. Majoie, M. E. Sprengers, W. J. J. van Rooij, C. Lavini, M. Sluzewski, J. C. van Rijn, and G. J. den Heeten
MR Angiography at 3T versus Digital Subtraction Angiography in the Follow-up of Intracranial Aneurysms Treated with Detachable Coils
AJNR Am. J. Neuroradiol., June 1, 2005; 26(6): 1349 - 1356.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
F. G. Shellock, M. Gounis, and A. Wakhloo
Detachable Coil for Cerebral Aneurysms: In Vitro Evaluation of Magnetic Field Interactions, Heating, and Artifacts at 3T
AJNR Am. J. Neuroradiol., February 1, 2005; 26(2): 363 - 366.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
A. Nehra, C. J. Moran, D. T. Cross III, and C. P. Derdeyn
MR Safety and Imaging of Neuroform Stents at 3T
AJNR Am. J. Neuroradiol., October 1, 2004; 25(9): 1476 - 1478.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
F. G. Shellock and J. V. Crues
MR Procedures: Biologic Effects, Safety, and Patient Care
Radiology, September 1, 2004; 232(3): 635 - 652.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
F. G. Shellock, J. A. Tkach, P. M. Ruggieri, T. J. Masaryk, and P. A. Rasmussen
Aneurysm Clips: Evaluation of Magnetic Field Interactions and Translational Attraction by Use of "Long-Bore" and "Short-Bore" 3.0-T MR Imaging Systems
AJNR Am. J. Neuroradiol., March 1, 2003; 24(3): 463 - 471.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hennemeyer, C. T.
Right arrow Articles by Derdeyn, C. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hennemeyer, C. T.
Right arrow Articles by Derdeyn, C. P.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
RADIOLOGY RADIOGRAPHICS RSNA JOURNALS ONLINE