|
|
||||||||
Experimental Studies |
1 From the Departments of Cardiology (S.K.H., J.G., T.S., C.V., C.L., S.L., A.E., H.O., H.B., K.T.) and Neurology (C.P.), University of Bonn, Sigmund-Freud-Strasse 25, 53105 Bonn, Germany. Received October 26, 2000; revision requested November 24; revision received February 23, 2001; accepted April 12. S.K.H. supported by a research grant from the German Society of Cardiology, Düsseldorf, Germany. K.T. and C.P. supported by a research grant from BONFOR, Bonn, Germany. Address correspondence to K.T. (e-mail: k-tiemann@uni-bonn.de).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: In vitro studies were performed in a flow phantom. SH U 508A, NC100100, or FS069 was continuously infused at clinically usable doses and infusion rates. To assess agent-specific physical properties, these agents were administered by using a vertically fixed infusion pump and varying infusion start times. The contrast agents were administered by also using a horizontally oriented infusion pump that was either fixed or continuously rotated to homogenize the agent in the syringe.
RESULTS: With SH U 508A and NC100100, constant signals were achieved, regardless of the infusion modality used. Compared with conventional infusion, the continuous homogenization of SH U 508A, although not necessary for signal constancy, increased the agents usefulness (P < .05). With FS069, only continuous homogenization yielded constant signals (P < .001).
CONCLUSION: Continuous infusion of SH U 508A or NC100100 provided constant harmonic power Doppler US signals, regardless of the infusion modality used. Because of the special physical properties of FS069, only homogenization produced constant harmonic power Doppler US signals during continuous infusion of this agent.
Index terms: Contrast media, comparative studies Contrast media, experimental studies Experimental studies Ultrasound (US), contrast media, **.129842, **.12988
| INTRODUCTION |
|---|
|
|
|---|
Continuous infusion of the contrast agent by means of an infusion pump may overcome the limitations of bolus injection. It has been shown that the duration and usefulness of the resulting contrast enhancement can be greatly increased by using continuous infusion (13). This approach allows the clinician to adjust system settings (eg, time gain compensation and gain in terms of pulsed-wave Doppler studies) and to obtain different views of cardiac organs or, with abdominal US, even multiple organs. With continuous infusion, the infusion rate can be customized to individual patients and, thus, the bubble concentration can be titrated to achieve optimal enhancement, with the result of increased usefulness (17). Moreover, there has been considerable skepticism as to whether the greater contrast enhancement obtained with bolus injection obscures possible perfusion defects (18). Finally, most advanced clinical applications, such as the assessment of replenishment kinetics following US-induced destruction of microbubbles, require constant contrast enhancement during the entire study (17,19).
Because current transpulmonary echo contrast materials have specific physical properties, different agents might need different perfusion regimens to produce constant tissue contrast enhancement. Moreover, issues such as agent decay and inhomogeneity in the syringe need to be considered in the development of infusion techniques.
The purpose of this study was twofold: to evaluate the infusion properties of three US contrast agents that are either currently under development or already approved for clinical use and to evaluate different infusion techniques.
| MATERIALS AND METHODS |
|---|
|
|
|---|
US imaging was performed downstream of the mixing chamber by using a Plexiglas tank that contained a polyethylene tube with an inner diameter of 8 mm. For attenuation and for adaptation of the acoustic properties to a physiologic US wave propagation velocity of 1,540 m/sec, the tank was filled with castor oil. In addition, a rough rubber mat was placed in the bottom of the tank. The position of the ultrasound transducer was 8 cm above the tube, with an insonation angle of 85° to avoid reverberation artifacts. To avoid recirculation of the contrast agent, two dialysis cartridges (Hemofilter 14; Gambro Dialysatoren, Hechingen, Germany) were interposed into the flow circuit downstream of the insonation area to filter the blood analog. The efficiency of the filter process with each contrast agent was assessed by using a second ultrasound transducer behind the filter section (Fig 1).
|
Infusion Setup
With each infusion setup, the contrast agents were continuously delivered to the flow phantom by means of a dedicated infusion pump (Braun Perfusor; Braun Melsungen, Melsungen, Germany) that contained a 10-mL syringe (Omnifix; Braun Melsungen). The outlet of the syringe was connected to the infusion port of the flow phantom either directly, with a three-way stopcock and an 18-gauge needle, or by means of special 50-cm connector tubing (Medrad, Indianola, Pa). To evaluate the infusion properties of the three transpulmonary contrast agentsSH U 508A, NC100100, and FS069we compared the following infusion modalities:
Vertical syringe technique.All three contrast agents were administered by using a vertically fixed infusion pump. The outlet of the syringe pointed upward and was directly connected to the infusion port of the flow phantom with a three-way stopcock and an 18-gauge needle, without connection tubing. The infusion was started either immediately after fixation of the syringe at the infusion pump or with a delay of 5 or 10 minutes after filling and fixation. This infusion setup was designed to investigate the possible flotation of the microbubbles in the vertically oriented syringe over time, which would result in variable harmonic power Doppler US signal strengths. We hypothesized that a delay in starting the infusion would result in stronger harmonic power Doppler US signals at the beginning and a rapid decrease in signal strength at the end of infusion, with the final result of bolus-like kinetics of the time-signal curve. Therefore, each infusion was performed, with varying infusion start times, until the syringe was completely emptied.
Horizontal syringe technique.In the horizontal syringe setup, the previously described infusion pump was oriented horizontally. Infusion was started immediately after fixation of the syringe. To evaluate the possible effects of connection tubing (as used in clinical scenarios) on the constancy of contrast enhancement, the syringe was fixed to the infusion port either directly, with a three-way stopcock and an 18-gauge needle, or through special 50-cm connector tubing (Medrad). These measurements were performed only with SH U 508A and NC100100.
We evaluated the effect of homogenization, as a special technique with the horizontally oriented syringe, on the constancy of harmonic power Doppler US signals. To homogenize the contrast agent in the syringe, the infusion pump was manually rotated 180° longitudinally along the vertical axis (about 20 times per minute) during continuous infusion. All three contrast agents were administered by using the previously described infusion pump, which was connected to the infusion port by means of a shortened 6-F, 12-cm-long catheter.
Straight tube technique.We evaluated the straight tube technique, as a special modality for infusion of FS069, to assess the consistency of microsphere delivery over time, as previously described by Miller et al (16). Therefore, vertically oriented 3-mm-diameter extension tubing was filled with a total of 2.5 or 5.0 mL of FS069; the length of the tubing varied from 50 cm for 2.5 mL of FS069 to 100 cm for 5.0 mL of FS069. The outlet of the straight tube pointed upward and was directly connected to the infusion port of the flow phantom with a three-way stopcock and an 18-gauge needle. Another extension line connected the lower part of the straight tube to the infusion pump by way of a three-way stopcock. FS069 (2.5 or 5.0 mL) was introduced into the straight tube at the lowest injection port of the system, which was the distal three-way stopcock. To drive the microbubbles into the flow phantom, normal saline was continuously administered through the infusion pump at varying infusion rates10, 20, and 30 mL/h. This infusion setup has been described by Miller et al (16).
Echo Contrast Agents
Three contrast agents were evaluated in the flow phantom: SH U 508A is a galactose-based US contrast agent composed of air-filled microbubbles, 99% of which are smaller than 7 µm. This agent also contains palmitic acid (0.1%) to stabilize the microbubbles for lung transit (9,12,13). We used a concentration of 400 mg/mL. Therefore, a total of 4 g of SH U 508A granules was suspended in 8 mL of sterile water that was agitated by hand, as recommended by the manufacturer. After preparation, the vial was left for 12 minutes to allow the solution to reach equilibrium. The 10-mL syringe was filled with a total of 8 mL of SH U 508A, and the infusion was started. For each infusion modality, one vial of SH U 508A (8 mL) was infused for 12 minutes at a rate of 40 mL/h.
NC100100 is a second-generation perfluorocarbon-filled contrast agent with a median microbubble diameter of 3.4 µm (20,21). For examinations in humans, the granules are diluted with 2 mL of saccharin solution. For in vitro studies, we used a 1:100 dilutionthat is, 0.1 mL of NC100100 plus 10 mL of saccharin solutionto avoid system saturation and acoustic shadowing. To reduce additional agent decay due to dilution, the infusion was started immediately after preparation of the contrast agent. By using the vertical syringe setup, a total of 10 mL of diluted NC100100 was infused for 12 minutes at a rate of 50 mL/h. For the horizontal syringe (with and without an extension line) and homogenization setups, 10 mL of diluted NC100100 was infused for 10 minutes at a rate of 60 mL/h.
FS069 is a second-generation contrast agent that contains sonicated octafluoropropane-filled albumin microspheres. The microspheres have a mean size of 3.65.4 µm and are supplied in ready-to-use vials that contain 6.39.0 x 108 microspheres per milliliter (11,22). Prior to infusion, we resuspended FS069, as recommended by the manufacturer, by rolling the vial horizontally between the palms. Without further dilution, a total of 2.5 mL of FS069 was infused for 15 minutes at a rate of 10 mL/h during homogenization and for the vertical-syringe setup. As described for the straight tube technique, either 2.5 or 5.0 mL of FS069 was administered to the flow phantom for 20 minutes by means of continuous saline infusion, with the infusion rate varying between 10 and 30 mL/h.
Data Analyses and Statistics
Quantitative offline analyses of the raw data were performed prior to scan conversion by using calibrated software (HDI-Lab; Advanced Technology Laboratory) that takes machine settings into account. Before contrast material administration, a circular 8-mm-diameter region of interest was placed by one observer (J.G.) by using the background B-mode image as a reference for the cross-sectional area of the tube. The suggested regions of interest were routinely cross checked by another author (S.K.H.).
The mean (± SD) strength of the harmonic power Doppler US signals (in decibels) in each region of interest was calculated. Each data point of the resulting time-signal curve represented an average of 10 consecutive measurements at a sampling rate of 1 Hzthat is, one measurement per second was recorded for 10 successive seconds. This type of measurement was performed during three independent contrast agent infusions (average of three passes). Because of the digital acquisition time, we started the measurements at the beginning of every minute of infusion. For clarity, we abstained from additionally displaying SDs on the graphs.
A plateau in signal strength (hereafter referred to as signal plateau) was defined when the differences in signal strength between at least five successive data points did not exceed 1 dB. When a plateau was reached, the mean signal plateau value (in decibels ± SD), plateau duration (in minutes), and systematic increase or decrease in signal plateau (in decibels per minute) were assessed for each contrast agent. If the time-signal curve did not reach a signal plateau, the mean peak signal strength (in decibels ± SD) and systematic increase or decrease in peak signal strength (in decibels per minute) were assessed.
Group comparisons were performed by using the Kruskal-Wallis test. A P value less than .05 was considered to indicate a statistically significant difference. Calculations were performed by using the most recent version of a computer software package (SPSS; SPSS, Chicago, Ill).
| RESULTS |
|---|
|
|
|---|
|
|
|
FS069.Immediate onset of FS069 infusion yielded a bell-shaped time-signal curve, with the highest values after 8 minutes of infusion and a maximum mean value of 17.85 dB ± 0.28. Afterward, the time-signal curve showed a decreasing slope until the infusion stopped after 15 minutes. With delayed infusion starts5 and 10 minutes after filling and fixation of the syringethe time-signal curves revealed a bolus-like kinetic with a steep increase; peak values were reached after 5 minutes of infusion and followed by a slow decrease. The peak signal strength reached a mean value of 22.36 dB ± 0.62 with the 5-minute infusion delay and of 22.08 dB ± 0.31 with the 10-minute delay (Fig 2c). Kruskal-Wallis test results indicated a significant group difference between values obtained at immediate infusion onset and those obtained with the two infusion delays (P < .05). Posthoc Dunn multiple comparisons test results revealed that the peak harmonic power Doppler US signals were significantly stronger with the 5- and 10-minute infusion delays than with immediate infusion (P < .05). There were no significant differences between the peak signal strengths achieved with the two infusion delays (Fig 2c).
Horizontal Syringe Technique
SH U 508A.By infusing SH U 508A directly into the infusion port, we achieved constant harmonic power Doppler US signals at a mean of 3.00 minutes ± 1.00 after infusion onset. The signal strength reached a mean value of 10.43 dB ± 0.29 at the plateau of the time-signal curve and lasted for a mean of 10.33 minutes ± 0.58. Infusing SH U 508A through 50-cm connection tubing resulted in a slight, nonsignificant decay in harmonic power Doppler US signal strength, and a mean value of 9.73 dB ± 0.40 was reached. A signal plateau was achieved after a mean of 3.33 minutes ± 0.58 and lasted for a mean of 10.33 minutes ± 0.58. The time-signal curves for both infusion modalities showed a slight increase in signal plateau: of 0.09 dB/min with direct infusion and of 0.07 dB/min with infusion through the 50-cm connection tubing. During homogenization, constant signals were achieved at a mean of 3.33 minutes ± 0.58 after infusion onset, and a mean value of 11.60 dB ± 0.61 was reached. The signal plateau lasted a mean of 9.67 minutes ± 0.58, and there was a slight increase in strength of 0.07 dB/min (Fig 3a). As indicated by the Kruskal-Wallis test results, there were no significant differences between the signal plateau values obtained at onset and those obtained at delayed infusion among the three infusion modalities. However, we found a significant difference (P = .037) in signal plateau during homogenization, as compared with the signal plateau achieved by using the horizontally fixed syringe, with or without connection tubing (Fig 3a).
|
|
FS069.Owing to results that showed an obvious concentration gradient developing in the vertical syringe and to results of previous studies by Miller and co-workers (16), we abstained from performing FS069 US measurements with the horizontal syringe setup, with or without extension tubing. Only with homogenization were constant signals achieved during continuous infusion of FS069; this finding is concordant with previous clinical experience (18). Constant signals were seen after a mean of 5.33 minutes ± 0.58, and a mean signal strength of 15.99 dB ± 0.31 was reached. The mean plateau duration was 11.33 minutes ± 0.58, with a slight increase in signal strength of 0.06 dB/min (Fig 4).
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Continuous infusion of both SH U 508A and NC100100 with all of the described infusion modalities produced a plateau of constant contrast enhancement. Because SH U 508A has relatively high viscosity and the bubbles do not tend to rise rapidly in the syringe, there was only a slight, nonsignificant increase in signal strength with delayed infusion and the vertical syringe setup. Independently of the infusion modality, constant and prolonged enhancement was achieved 3 minutes after infusion onset and lasted until the end of infusion. Even when extension tubing was connected to the syringe, the conventional infusion setup used in clinical scenarios had a minor effect on signal plateau and duration. Our results are in good agreement with those of Albrecht et al (13), who observed the feasibility of continuous infusion of SH U 508A in a clinical pilot study on Doppler US enhancement. Albrecht et al (13) compared bolus injection with constant infusion of SH U 508A at three different infusion rates for the acquisition of spectral and color Doppler US scans of the femoral and carotid arteries. All infusions yielded a plateau of constant enhancement that started after 12 minutes and lasted until the end of infusion; however, the plateau duration was dependent on the infusion speed. Not only were the saturation artifacts with infusion fewer than those with bolus injection, but also the dose effectiveness was improved with continuous infusion.
Similar results were obtained with continuous infusion of NC100100. We achieved constant signal enhancement with all of the described infusion modalities, and a plateau of the time-signal curve was reached after 3 minutes and lasted until the end of infusion. Although the results of the vertical syringe setup might lead to the assumption that a minor degree of bubble flotation occurs within the syringe, floating bubbles seem to have a minor effect on signal constancy during continuous infusion of NC100100. In several clinical trials involving NC100100 for assessment of myocardial perfusion abnormalities, NC100100 was administered as a bolus injection and followed by a saline flush (2628). Clinical data that indicate whether continuous infusion of NC100100 can provide prolonged contrast enhancement and thus improve the dose effectiveness of this agent are necessary.
In contrast to SH U 508A and NC100100, FS069 cannot be easily infused by using the conventional infusion setup, because the microspheres tend to float rapidly, if not in an agitated manner, and thus create a concentration gradient within the syringe. This concentration gradient has a marked effect on the total amount of microbubbles released over time during the continuous infusion of FS069. To overcome this limitation, several groups have tried to use solutions of this contrast agent, with FS069 diluted in normal saline. Saline bags containing the diluted contrast agent were used frequently. These bags were manually agitated to prevent microbubble flotation. However, it was difficult to maintain consistent myocardial opacification with this dilution infusion method, because the degree of manual manipulation of the diluent affected the microsphere delivery rate (16).
Furthermore, the microbubbles, when diluted in the perfluorocarbon-deficient saline solution, tend to release their encapsulated gas volume rapidly and thereby lose effectiveness in a short time, as demonstrated in a study by Podell et al (22). A decrease of 50% backscatter was observed in a gas-deficient solution (60% air-saturated diluent) within only 90 seconds versus 6 minutes in a 100% air-saturated diluent. This microsphere instability can result in unacceptable variability in the measured acoustic properties, especially when the time after dilution is not controlled. On the other hand, conventional dripping bags act as a bubble trap owing to gravity and therefore do not seem to be useful for infusion. The combination of these limitations makes it impossible to predict the amount of contrast agent infused at any given time.
Therefore, an alternative infusion setup that makes use of the buoyant properties of FS069 microspheres was evaluated by Miller et al (16): The contrast agent was filled into a vertically oriented line and flushed with normal saline. This setup was found to reduce shadowing and the attenuation artifacts caused by bolus injection and thereby provide prolonged contrast enhancement. To evaluate the contrast enhancement constancy achieved by using the straight tube technique, Miller and co-workers (16) introduced a total of 2.5 or 5.0 mL of FS069 into the straight tube, with normal saline used as the vehicle for continuous administration of microbubbles, at three infusion rates.
At low infusion rates, constant signals were observed 5 minutes after infusion onset and lasted for 34 minutes; a slow decrease in signal strength followed. On the other hand, the highest saline infusion rates resulted in the typical wash inwashout enhancement pattern, which resembled a bolus curve with a rapid increase followed by a rapid decrease in signal strength. Miller et al (16) found that at slower saline flow rates, bigger FS069 microbubbles tended to develop faster than the following fluid could push through the straight tube; this resulted in a concentration of larger microbubbles at the head of the flow. Furthermore, as the line was depleted of larger microspheres, the mean diameter and concentration of the microspheres decreased; this resulted in a slow decay in signal at the end of infusion. Because gas exchange in the straight tube has to be considered, we do not believe that this method is ideal for myocardial perfusion assessment modalities, which require a steady state of contrast enhancement.
Because of difficulties in the continuous infusion of FS069, a major part of this study was focused on the evaluation of different setups for FS069 infusion to define the infusion properties of the agent and to design an infusion setup that is suitable for constant microbubble delivery. The flotation properties of FS069 microbubbles were clearly demonstrated in our experiments with the vertical syringe setup. When FS069 was infused immediately after the resuspension of the microbubbles, the resulting time-signal curve was less bolus like and showed relatively constant signals for 5 minutes. On the other hand, a short delay between the preparation and start of infusion resulted in a bolus-like shape of the curve with significantly higher peak values, followed by a fast decay in signal.
We built on our experience with FS069 infusion modalities and evaluated the effect of homogenization on signal constancy. With FS069, homogenization was the only infusion modality that resulted in a constant delivery of the agent during the entire infusion period. In addition, the highest peak signal values and highest signal values in the area under the curve were achieved by using this method. We therefore conclude that with FS069, homogenization during continuous infusion is required to achieve constant signals.
Practical application: The results of this study demonstrate that constant infusion of SH U 508A, NC100100, or FS069 into the vascular system is feasible. Because of the special physical properties of FS069, homogenization is a prerequisite to achieving constant signals during continuous infusion. Although with infusion of SH U 508A or NC100100, homogenization was not necessary to reach a signal plateau, homogenization improved the contrast enhancement achieved with SH U 508A infusion, as compared with the conventional infusion method. Whether the infusion modality used can markedly affect echo enhancement has to be evaluated with other US contrast agents.
We conclude that homogenization might be the superior technique for stable and maximal contrast enhancement with continuous infusion of FS069, whereas conventional infusion of SH U 508A or NC100100 is feasible for constant contrast enhancement.
| FOOTNOTES |
|---|
Author contributions: Guarantors of integrity of entire study, S.K.H., K.T., J.G.; study concepts, K.T., J.G., H.B., S.K.H.; study design, K.T., C.P., J.G.; literature research, S.L., J.G., A.E., C.L.; experimental studies, J.G., C.V., S.L., C.L.; data acquisition, J.G., T.S., C.V., C.L.; data analysis/interpretation, S.K.H., C.P., J.G.; statistical analysis, C.P., J.G., C.L.; manuscript preparation, S.K.H.; manuscript definition of intellectual content, K.T., H.B., S.K.H.; manuscript editing, H.O., S.K.H.; manuscript revision/review, H.B., H.O.; manuscript final version approval, K.T., H.B., H.O. S.K.H. and J.G. contributed equally to this work.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Okada, C. W. Hoffmann, K. J. Wolf, and T. Albrecht Bolus versus Continuous Infusion of Microbubble Contrast Agent for Liver US: Initial Experience Radiology, December 1, 2005; 237(3): 1063 - 1067. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Becher, K. Tiemann, S. Kuntz-Hehner, H. Omran, and T. Schlosser Diagnostic impact of contrast echocardiography for assessment of left ventricular function and myocardial perfusion in patients with coronary artery disease Eur. Heart J. Suppl., March 1, 2002; 4(suppl_C): C12 - C21. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |