|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Technical Developments |
1 From the Division of Image Processing (B.P.F.L., R.J.v.d.G., J.H.C.R.), Department of Radiology (H.J.L.), Leiden University Medical Center, Bldg 1 C2-S, Albinusdreef 2, PO Box 9600, 2300 RC Leiden, the Netherlands; and Interuniversity Cardiology Institute of the Netherlands, Utrecht, the Netherlands (H.W.M.K., H.J.C.R.). Received December 22, 2000; revision requested February 5, 2001; revision received March 29; accepted May 14. Address correspondence to J.H.C.R. (e-mail: j.h.c.reiber@lumc.nl).
| ABSTRACT |
|---|
|
|
|---|
Index terms: Heart, function, 524.91 Heart, MR, 524.121416 Heart, ventricles, 524.121416, 524.91 Magnetic resonance (MR), cine study, 524.121416 Magnetic resonance (MR), volume measurement, 524.121416
| INTRODUCTION |
|---|
|
|
|---|
Recently, image analysis tools were introduced to help automated planning of the short-axis image and reduce radiologist interaction to a minimum. This approach is based on the matching of a digital atlas of the thorax to scout or localizer images in a particular patient, and technical details are described elsewhere (1213). This procedure can be applied to calculate the two parameters required to plan a short-axis volume: the position and the orientation of the LV long axis. The purpose of our study was to assess whether the accuracy and reproducibility of the automated image planning method are equal to those achieved with conventional manual image planning.
| Materials and Methods |
|---|
|
|
|---|
By matching this thorax model to scout images, the geometric transformations that map the model to the scout images can be applied to estimate the position and orientation of the LV long axis. On the basis of these estimates, the geometry of the short-axis image volume is automatically defined directly from the scout images. The three-dimensional angulation of the LV long axis, which is defined in this study as the geometric principal axis of the left ventricle, is used as the orientation of the image. Figure 1 is an example of a matching result and the corresponding automatically defined short-axis image geometry.
|
1. Cardiac gated scout image (nine coronal, nine sagittal, and nine transverse images) acquired with a multistack protocol during free breathing (turbo field-echo pulse sequence: repetition time of 8 msec, echo time of 4 msec, flip angle of 20°, section thickness of 10 mm, section gap of 1 mm, field of view of 450 x 450 mm).
2. Cardiac gated vertical long-axis image, which was planned manually on a transverse section in the scout images. The image was acquired in a plane parallel to the imager bore axis and the LV long axis, through the LV apex (fast field-echo echo-planar pulse sequence: repetition time equal to the R-R interval, echo time of 11 msec, prospective triggering, end expiration, section thickness of 8 mm, reduced field of view of 400 x 237 mm).
3. Four-chamber image, which was planned on the vertical long-axis image. The image was acquired in a plane through the apex that was oriented perpendicular to the vertical long-axis image and parallel to the LV long-axis image, as seen in the vertical long-axis image (fast field-echo echo-planar pulse sequence: repetition time equal to the R-R interval, echo time of 11 msec, prospective triggering, end expiration, section thickness of 8 mm, reduced field of view of 400 x 237 mm).
4. Multisection multiphase gated short-axis cardiac MR image, which was acquired perpendicular to the four-chamber image through the apex and perpendicular to the LV long axis (fast field-echo echo-planar pulse sequence: repetition time equal to the R-R interval, echo time of 11 msec, prospective gating, end expiration, section thickness of 10 mm, section gap of 1 mm, reduced field of view of 400 x 237 mm).
Experimental Setup
Two experimental studies, both approved by our institutional review board, were performed. Manual and automated short-axis image planning were compared on the basis of two criteria: (a) prospectively, the accuracy and reproducibility of clinical parameters, such as LV EF, end-diastolic (ED) volume, end-systolic (ES) volume, LV wall mass, and ED and ES LV wall thicknesses; and (b) retrospectively, image geometry.
All MR studies were performed with a standard 1.5-T MR imaging system (Gyroscan NT; Philips, Best, the Netherlands).
Clinical parameters.To compare the accuracy and reproducibility of clinically relevant parameters calculated with the automated planning protocol and with the manual planning protocol, five healthy adult volunteers (three men and two women; age range, 2238 years; mean age, 29 years) without history of cardiovascular disease underwent manually and automatically planned short-axis MR imaging. Our institutional review board approved the study. Informed consent was obtained from all subjects. To investigate the reproducibility, the same procedure was repeated after the subjects were removed from the imager bore and subsequently repositioned. The number of sections in the automatically defined short-axis grid was fixed at 13 to ensure complete coverage of the LV cavity in the image volume. Computations were performed with a standard workstation (Ultrasparc 10; Sun Microsystems, Mountain View, Calif).
Owing to the developmental stage of the automated image planning procedure, the automatically computed imaging parameters were manually transferred to the imager console. No subjective user interaction was allowed during automated image planning. The time required for automated image planning was 57 minutes, including acquisition of the scout views (2 minutes) and calculation of the acquisition parameters for the imaging volume (35 minutes). On all image pairs acquired (2 x 5 manual images; 2 x 5 automatic images; total, 20 short-axis images), the contours of the LV endo- and epicardium were manually drawn by an independent expert (R.J.v.d.G.) by using cardiac MR postprocessing software (MRI-MASS; Medis Medical Imaging Systems, Leiden, the Netherlands). Contours were outlined on the basis of the conventions for endo- and epicardial contour drawing (10,14), that is, papillary muscles and epicardial fat were excluded from the contours. LV mass, ED volume, ES volume, and EF were calculated. In addition, the mean LV wall thickness was calculated in end diastole and end systole by averaging results in several contiguous sections of the left ventricle. The most apical and the most basal sections were excluded from these wall thickness analyses because, in some cases, no endocardium was present in systole in these outer sections as a result of LV longitudinal contraction.
Image geometry.To compare the accuracy of image geometry obtained with the automated planning method with that achieved in routine clinical practice, 20 patient examinations were collected in a retrospective study. Our institutional review board approved this component of our study and did not require patient informed consent. No gender or age selection was performed. Patients were randomly selected. Their diagnoses included various pathologic cardiac conditions: myocardial infarction, n = 14; cardiomyopathy, n = 3; congenital anomaly, n = 1; arrhythmia, n = 1; and LV aneurysm, n = 1. All examinations were acquired with the manual planning protocol. Automated image planning was applied retrospectively to the 20 scout views to determine the position of the center of gravity of the image volume and the orientation of the imaging planes.
Comparison between manual and automated image geometry.To compare the geometry of the automatically and manually planned imaging volumes, the orientation of the geometric LV long axis at end diastole in the manually imaged data sets was selected as the reference standard for the orientation of the LV long axis. To calculate this orientation, the contours of the endocardium were manually drawn by an expert (R.J.v.d.G.) on all short-axis cardiac MR images at end diastole. The orientation of the reference standard LV long axis was calculated by fitting a straight line through the contour centroids of the ED contours by means of a least squares distance criterion. Figure 2 depicts an example of a reference standard LV long axis.
|
C between the ED and ES geometric LV long axes as determined from the contours on the short-axis images in end diastole and end systole, respectively.
The variability introduced in the manual (M) image planning process was assessed by calculating the angle
M between the reference standard LV long axis and the orientation (normal vector) of the manually planned short-axis images (Fig 3). This angle expresses the error introduced by the radiologist that was accumulated during all the stages of the manual planning procedure, and it provides an indicator of the accuracy with which the LV long axis is determined in routine clinical practice with use of a standard clinical MR system.
|
A was calculated. The angle
A represents the differences in orientation between the automatically defined and reference standard imaging volumes.
Statistical Analysis
Values of the functional parameters EF, ED and ES volumes, and ED and ES wall thicknesses were obtained for the automated and manual image planning procedures. A two-tailed paired-samples t test was used to compare these measurements. The level of significance was a P value of .05.
To evaluate the image geometry, the difference between
M and
A was investigated with a two-tailed paired-samples t test. The difference in SDs in
M and
A was tested with a one-tailed F test. The level of significance was a P value of .05.
| Results |
|---|
|
|
|---|
|
|
|
|
C, was a mean 6.1° (SD, 5.0°; range, 0.2°22.2°). Angle
M was a mean 9.7° (SD, 5.8°; range, 2.2°28.9°), while the angular offset angle
A was a mean 12.2° (SD, 6.8°; range, 1.9°25.2°). The difference between
M and
A was not significant in a two-tailed paired samples t test (P = .23). Figure 5 shows two examples of automatically defined image grids to illustrate the geometric accuracy of automated image planning.
|
| Discussion |
|---|
|
|
|---|
Regarding the reproducibility of clinical parameters assessed with both methods, no statistically significant difference between repeated study measurement pairs was present. Again, variations were well within the range of inter- and intra-observer variabilities associated with manual contour drawing.
The manually introduced angle between the short-axis image orientation and the reference LV long axis (
M) together with the change in direction of the LV long axis during the cardiac cycle (
C) resulted in differences from the reference standard orientation in an order of magnitude between 2° and 29°. This illustrates that short-axis views as planned manually in routine clinical practice may deviate substantially from the optimal image perpendicular to the geometric LV long axis. The error in the automated planning method,
A, however, varied in a slightly smaller range (between 2° and 25°) as compared with the error in the manual planning procedure. The mean angular offset in the automatically planned short-axis volumes was 12.2°, which is comparable to the inaccuracies in the manual image planning procedure (mean, 9.7°). This difference was not statistically significant (P = .23).
The automated method also demonstrated an SD comparable to that with the manual method (6.8° vs 5.8°) in the angular offset from the reference standard LV long axis. This difference was not significant (one-tailed F test, P = .49), and it is in the same order of magnitude as the SD in
C, 5.0°, which represents the variation in orientation of the LV long axis during the cardiac cycle.
On the basis of these findings, we conclude that the automated and manual image planning procedures are equally accurate and reproducible for the assessment of LV function in terms of EF, ED LV mass, and ED and ES volumes. This conclusion is supported by the fact that the measurements in ED and ES wall thicknesses and the orientation of the LV long axis did not differ significantly between the methods.
The automated image planning method was intended for application in a supervised manner. In this study, we eliminated any user interaction to investigate the accuracy and reliability of the method. The automated planning procedure performed with accuracy and reproducibility comparable to those with manual image planning in routine clinical practice. However, we recommend implementation of user interaction by means of visualization of the calculated image volume on the scout image sets and by allowing the radiologist to intervene. On the basis of the results from this study, we expect that such user interaction was not required in 90% of the cases.
The number of sections in the short-axis image volume was fixed in the present study. As a result, a few redundant sections that did not contain the left ventricle were included in some cases. In practice, this parameter can be adjusted manually if required. Future investigations should focus on the automated assessment of this and a number of other imaging parameters required for different cardiac function protocols, such as the navigator position for nonbreath-hold protocols and the location of the aortic arch for the acquisition of velocity-encoded aortic-flow images.
The automated image planning method was developed to work with current standard MR imaging hardware. A recent advance in MR acquisition technology was the development of real-time MR imagers, which allow interactive image planning of the short-axis imaging plane. This may substantially reduce the time required to plan the short-axis image planes. However, the main advantage of automatic image planning over real-time planning is that it is not subjective and does not require specialized expert knowledge to determine the optimal short-axis views.
The time required for the current automated image planning procedure is 57 minutes. In general, this is only slightly shorter than the time required for the manual image planning procedure. However, we foresee two possibilities that will speed up the automated process. First, the rapid evolution of computer hardware will enable us to further shorten the automated matching procedure. Second, we performed the initial experiments by applying the automated model matching procedure to only a subset of the scout image volume. In this way, the planning procedure could be shortened by a factor of 24 with comparable accuracy. These two developments may potentially reduce the automated planning time to 3 minutes, including scout imaging. Comparative studies to evaluate the time efficiency of the manual and automated short-axis planning protocols are needed.
In conclusion, this study investigated the feasibility of an observer-independent image planning method for short-axis cardiac MR images; results with the automated planning procedure showed accuracy and reproducibility comparable to those with the manual planning used in routine clinical practice. Currently, the time for the automated planning protocol is slightly shorter than that for manual planning, mainly because no additional long-axis support images are required. The main advantage of the automated method during the manual planning procedure, however, is that it provided an observer-independent uniform planning approach that required no expert knowledge. The method yielded images with quality comparable to that with manual planning. No statistically significant differences were found between the methods in terms of accuracy and reproducibility of measurements of EF, ED LV and ES LV volumes, ED LV wall mass, and ED and ES wall thicknesses.
| FOOTNOTES |
|---|
Author contributions: Guarantor of integrity of entire study, J.H.C.R.; study concepts and design, B.P.F.L.; literature research, B.P.F.L.; clinical and experimental studies, R.J.v.d.G.; data acquisition, H.J.L., H.W.M.K.; data analysis/interpretation, H.J.L.; statistical analysis, R.J.v.d.G.; manuscript preparation and definition of intellectual content, R.J.v.d.G., B.P.F.L.; manuscript editing, J.H.C.R.; manuscript revision/review, B.P.F.L.; manuscript final version approval, J.H.C.R.
| REFERENCES |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |