|
|
||||||||
Medical Physics |
1 From the Department of Radiology, X-ray Imaging Laboratory, University of California, Davis Medical Center, 4701 X St, Sacramento, CA 95817 (J.M.B., K.K.L., J.A.S.); and Department of Radiology, University of California, San Diego (T.R.N.). From the 2000 RSNA scientific assembly. Received January 18, 2001; revision requested March 16; revision received April 10; accepted May 14. Supported in part by grants from the U.S. Army Breast Cancer Research Program (DAMD17-98-1-8176) and the National Cancer Institute (R21 CA82077). Address correspondence to J.M.B. (e-mail: jmboone@ucdavis.edu).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Validated Monte Carlo simulation techniques were used to estimate the average glandular dose (AGD). The calculated photon fluence at the detector for high-quality abdominal CT (120 kVp, 300 mAs, 5-mm section thickness) was the benchmark for assessing the milliampere seconds and corresponding radiation dose necessary for breast CT. Image noise was measured by using a 10-cm-diameter cylinder imaged with a clinical CT scanner at 10300 mAs for 80, 100, and 120 kVp. A cadaveric breast was imaged in the coronal plane to approximate the acquisition geometry of a proposed breast CT scanner.
RESULTS: The AGD for 80-kVp breast CT was comparable to that for two-view mammography of 5-cm breasts (compressed breast thickness). For thicker breasts, the breast CT dose was about one-third less than that for two-view mammography. The maximum dose at mammography assessed in 1-mm3 voxels was far higher (20.0 mGy) than that at breast CT (5.4 mGy) for a typical 5-cm 50% glandular breast. CT images of an 8-cm cadaveric breast (AGD, 6.3 mGy) were subjectively superior to digital mammograms (AGD, 10.1 mGy) of the same specimen.
CONCLUSION: The potential of high signal-to-noise ratio images with low anatomic noise, which are obtainable at dose levels comparable to those for mammography, suggests that dedicated breast CT should be studied further for its potential in breast cancer screening and diagnosis.
Index terms: Breast neoplasms, CT, 00.1211 Breast neoplasms, radiography, 00.11 Breast radiography, radiation dose Computed tomography (CT), radiation exposure, 00.1211
| INTRODUCTION |
|---|
|
|
|---|
Despite its utility, mammography is not without limitations (13). The most widely cited downfall of mammography is its reduced sensitivity in women with dense breasts (1416). Imaging of the dense breast with good sensitivity has become more important as younger women have begun to undergo screening routinely, as the use of hormone replacement therapy has expanded, and as genetic testing has begun to identify younger women at high risk. Digital mammography systems (17,18) that have a wider dynamic range than screen-film mammography have been developed, in part to address the increased challenges of imaging the dense breast. Early results indicate that digital mammography may lead to important incremental improvement in cancer detection in dense breasts.
Although computed tomography (CT) was studied in passing for its utility in breast cancer screening some years ago (1922), this modality has been largely dismissed as having a practical role in breast cancer screening due to concerns about radiation dose and cost-effectiveness. Most earlier studies involved conventional CT scanner technology, in which the images were acquired transversely and thus the x-ray beam had to penetrate the thoracic cavity. With this geometry, not only is a large amount of nonbreast tissue exposed to radiation, leading to substantial radiation dose inefficiency, but also cardiac and respiratory motion have the potential to reduce image quality.
Despite the conventional wisdom that CT is not effective for breast cancer screening, it is generally accepted that CT is far better than projection radiographic techniques in terms of contrast resolution by a factor of about 10 (23). When the complex normal anatomy (structured noise) of the dense breast is factored into the analysis of contrast resolution, the tomographic nature of CT facilitates the ability to eliminate overlapping structures, which are problematic in conventional mammography.
The purpose of this investigation was to evaluate the feasibility of breast CT in terms of radiation dose and image quality. Although a host of other considerations, such as diagnostic accuracy, cost, and interpretation time, remain to be evaluated, dose and image quality are fundamental to the potential of dedicated breast CT. Thus, we believe this investigation represents a necessary first step in assessing the potential of breast CT.
| MATERIALS AND METHODS |
|---|
|
|
|---|
|
The pendulant breast would be approximately cylindrical, with slight tapering of the diameter anteriorly. The dimensions of the cylindrical breast were estimated on the basis of the assumptions illustrated in Figure 2. The breast dimensions of a small cohort of 82 women were evaluated in our breast clinic on the basis of measurements made on the screen-film images. Human use authority (with exemption under category 4) was obtained for this activity. During normal mammographic interpretation, a mammographer (K.K.L.) measured the width of the left breast image at the edge of the film that corresponded to the width of the compressed breast near the chest wall and recorded this dimension on a form. The compressed breast thickness, which was printed onto the film, also was recorded.
|
With the CT gantry in the "parked" position (achieved with assistance from service personnel), the half value layer was determined by using type 1100 aluminum and an ionization chamber (Triad; Keithley, Cleveland, Ohio) for 80, 120, and 140 kVp. The output of the CT x-ray tube (air kerma per milliampere second [mAs] at the isocenter) also was measured. By using a spectral model appropriate for CT (29), we mathematically varied the amount of added aluminum filtration until the modeled spectra matched both the half value layer and the output measurements of the commercial scanner. The filtered spectral model was then used to produce x-ray spectra from 30 to 140 kVp in the Monte Carlo simulations.
With CT, the dose required to obtain a transverse section at a given peak kilovoltage (kVp) is linearly related to the product of the x-ray beam current (in milliampere) and the acquisition time (in seconds); this product is commonly referred to as the mAs. Because the mAs is a part of the technique protocol for any CT scanner, it was used in this study as a descriptor of x-ray beam quantity. Two authors (J.A.S., J.M.B.) determined the relationship between the output of the x-ray tube (in milligray or milliroentgen) and the photon fluence at the isocenter of the commercial CT scanner, as a function of the mAs, at each kVp by using physical exposure measurements combined with spectral modeling techniques.
Radiation dose was computed by using a cylindrical breast geometry with diameters ranging from 6 to 16 cm, which span the range of breast sizes that probably would be encountered clinically. A breast composition of 50% glandular tissue and 50% adipose tissue (ie, "50/50 breast") was simulated by using the data of Hammerstein et al (30). The isocenter is the position in space around which the x-ray tube and detector arrays rotate, and was assumed to be coincident with the center of the breast cylinder. The x-ray sourceto-isocenter distance was assumed to be 54 cm, which is similar to the x-ray sourceto-isocenter distance in a clinical CT scanner at our institution. A fan beam of x rays was incident on the right cylinder, with a 1-mm-thick fan beam positioned orthogonally to the central axis of the cylinder. The x-ray source was rotated in the simulation 360° around the breast cylinder in 3° increments, for a total of 120 different source positions. For each simulated x-ray spectrum (ie, kVp) and breast diameter, a total of 10,000,000 x-ray photons were tracked, and the energy deposition in a grid of 1 x 1 x 20-cm voxels was tallied.
The mean breast radiation dose and SD were computed from 10 Monte Carlo runs of 1,000,000 photons each, and these data were used to compute the coefficient of variation (ie, ratio of the SD to the mean). To evaluate dose homogeneity, the dose distribution was computed by using 1 mm x 1 mm x 20 cm voxels. The out-of-plane scattered radiation dose, regardless of its distance from the collimated CT section, as well as the primary dose deposition, were tallied. This acquisition geometry is equivalent to the measurement of the multisection average dose, which the CT dose index seeks to approximate (31). The CT dose index is the standard measurement used by the CT industry (by Federal statute) and by medical physicists (by convention) to assess CT radiation dose.
The energy deposition of ionizing radiation due to photoelectric interaction and Compton scattering events was tallied in each tissue voxel by a computer program written by one of the authors (J.M.B.). Rayleigh interactions also were tracked, but these do not result in energy deposition in the medium. The dose delivered to the medium studied was corrected to the glandular tissue dose by using the energy-dependent ratio of the mass energy attenuation coefficient of glandular tissue to the mass energy attenuation coefficient of the medium. This correction was performed on an interaction-by-interaction basis. The tallied energy (in joules) in each voxel was normalized by the mass of each voxel (in kilograms) to determine the average glandular dose (AGD, in milligrays). With use of the established relationship between photon fluence and mAs at each kVp, the dose was then normalized to correspond to that delivered by using 100 mAs. The resulting dose values were essentially CT dose index determinations, in milligrays per 100 mAs. The AGD delivered at other mAs settings could then be easily computed from these data.
To compute the radiation dose for a breast CT study, an estimate of the mAs needed at each kVp to produce clinically useful images was required. Since the typical photon fluence levels at the CT detector (which largely determine the signal-to-noise ratio [SNR] on the image) were unknown to us, we used the example of a typical CT technique that is known to produce images with a high SNR. A nonhelical abdominal CT technique performed with 120 kVp, 300 mAs, 5-mm section thickness, and a 32-cm-diameter cylindrical water-equivalent phantom was simulated. This technique was used in a computer simulation to determine the photon fluence striking the center of the CT detector array, integrated over a 360° rotation of the scanner. To maintain the same SNR in breast CT as that in the abdominal CT benchmark, the same photon fluence should be incident on the detector arrays. A simulation was performed for various diameters (616 cm) of cylinders with a 50% glandular tissue composition. The mAs necessary to deliver the same photon fluence to the detector behind the breast for a 1-mm-thick CT image was evaluated for x-ray spectra between 40 and 140 kVp.
The heterogeneity in dose for mammography and breast CT was evaluated (J.M.B.) by using 1-mm3 voxels, and for mammography, a rectangular cross-sectioned breast was simulated. The Monte Carlo techniques described previously were used to assess the dose distribution at CT; however, for mammography, the task was slightly more difficult due to the orthogonal compression used at two-view mammography. To simplify this computation, mammographic exposure to a rectangular breast (in coronal cross section) was simulated. The rectangular shape simplified the application of mathematical compression, as described in the following text.
A 50% glandular 5-cm compressed breast thickness was used, and the median breast width determined from breast size analyses was assumed for the width dimension. In addition, a 4-mm skin layer was assumed (32). Under compression, the cross section of the breast was modeled as a 5.0 x 19.4-cm rectangle, and when uncompressed, the breast cross section was warped by using bilinear interpolation to a 9.85 x 9.85-cm square, which is equal in area to the rectangle. Monte Carlo techniques were used to determine the dose deposition to the rectangular cross section of the breast, and the resulting Monte Carlo depth-dose curve was computer fit by using commercial software (Table Curve 2D; Jandel Scientific, Corte Madera, Calif). After breast exposure in one direction (eg, craniocaudal), the rectangular breast was warped to a square and then warped to a rectangle in the orthogonal direction (19.4 x 5.0 cm) for a second (eg, mediolateral oblique) exposure.
For dose distribution analysis, the breast was warped back to the square orientation. A 26-kVp molybdenum anode x-ray spectrum (33) filtered with 0.030-mm molybdenum and 3-mm poly (methyl methacrylate) (PMMA) (the compression paddle) was used. The entrance skin exposure used in this computation (17.5 mGy air kerma or 2,000 mR) was determined by means of interpolation from technique data measured at our institution for 50% glandular phantoms. For dose comparisons, an 11-cm-diameter breast (equal area) was simulated for exposure by using the breast CT geometry described earlier.
Experimental Studies
A 10-cm-diameter PMMA cylinder, 10 cm in length, was fabricated for this project. The cylinder had two 12.7-mm-diameter holes machined into it to accommodate a CT ionization pencil chamber: One hole was at the center, and one was centered 19 mm from the edge of the cylinder. The cylinder was positioned at the isocenter of the commercial multisection CT scanner (Lightspeed), and a 3-cm3 CT chamber (MDH 1015; MDH, Monrovia, Calif) was placed in the center hole to measure air kerma. At 80, 100, and 120 kVp, CT images were obtained in a series of 14 exposure levels ranging from 10 to 300 mAs. The images were acquired with the scanner by using the detail reconstruction filter (34) with a 10-cm field of view, which corresponded to pixel dimensions of 195 x 195 µm. The section thickness was 1.25 mm at the isocenter of the scanner.
For each image acquired, three regions of interest were evaluated (J.M.B., J.A.S.) by using commercially available image analysis software (EFILM; University of Toronto, Ontario, Canada). The root mean square SD (ie, noise) in each region of interest was recorded, and the average noise (
CT#) of the three regions of interest was computed for each image. The noise was fit as a function of mAs by using the equation
CT# = a x (mAs)b with commercially available software (Freelance 97; Lotus, Cambridge, Mass), where a and b are constants. These data allowed the
CT# to be computed as a function of either mAs or dose for each kVp. The
CT# measurements allowed the denominator of the contrast-to-noise ratio (CNR) to be evaluated as a function of kVp and mAs.
Johns and Yaffe (35) measured the linear attenuation coefficients (LACs) for fibrous (ie, glandular) tissue, fat from the breast (ie, adipose tissue), and infiltrating ductal carcinoma (ie, cancer) for monoenergetic x rays ranging from 18 to 110 keV at eight different energy levels. In addition to these attenuation coefficient data, the elemental compositions of the three tissues (ie, glandular, adipose, and cancer) determined by Hammerstein et al (30) were used, and mass attenuation coefficients were computed by using the mixture rule (36) coupled with published attenuation coefficient data (37). The mass attenuation coefficients of each of the tissue types were multiplied by the physical density to compute the LAC. The computer-generated LACs for breast tissue (30,35) were used to computer fit the measured LACs reported by Johns and Yaffe (35) by using least squares techniques and letting the physical density and the relative glandular fraction vary as free parameters. Once the LAC versus monoenergetic x-ray energy was parameterized for each of the three tissue types, x-ray spectra (29) from 30 to 140 kVp were used to weight the monoenergetic LAC values to produce the effective LAC for each polyenergetic spectrum. This was done for adipose, glandular, and cancer tissues and for water. By using the LACs for each tissue type and for water, we computed the corresponding CT number (CT#), or Hounsfield unit, by using the following formula (23): CT# = 1,000 x [(µt - µw)/µw], where CT# is the CT number of the tissue with an LAC of µt, and µw is the LAC of water.
CNRs were computed by using differences in CT attenuation values as contrast and the noise measurements parameterized from experimental measurements as noise. The CNRs were converted to SNRs for different size objects by using the Rose relationship (38): SNR = CNR x (N)1/2, where N is the number of pixels corresponding to the breast cancer lesion. For lesion diameter d and pixel dimension
, N = (
/4) x (d/
)2.
Cadaveric Breast Imaging
A cadaveric breast was acquired with proper authorization at our institution. The breast was removed from the cadaver with the pectoralis major and minor muscles attached and immediately fixed in 5% formalin. To have the breast in a more natural position during CT scanning, the pectoralis muscle with the accompanying skin flap was sutured onto stiff cardboard by using plastic ties. The breast was placed in the head holder of the clinical multisection CT scanner (Lightspeed), with the long axis of the cylinder of the breast parallel to the table motion. The breast was scanned at 80 kVp, with one acquisition at 50 mAs and without repositioning; another acquisition was performed at 80 mAs. Nonhelical CT images were acquired by using a 1.25-mm section thickness, and a 15.5 x 15.5-cm field of view was reconstructed. This resulted in pixel dimensions of 303 x 303 µm. The CT images were reconstructed by using both standard and detail reconstruction filters.
The CT images were transferred to an imaging workstation for display and analysis. Custom software that enabled the CT images to be loaded into a volume data set was written (by using Microsoft C/C++ 5.0 Compiler; Microsoft, Redmond, Wash), and coronal, transverse, and sagittal views were generated. The viewing software could average any number of adjacent images at any location in the volume to create thicker sections.
For comparison images and dosimetry, the cadaveric breast was placed under compression in a clinical mammography system and imaged by using a prototype computed radiography system (Fuji Medical, Tokyo, Japan) designed for digital mammography. The dedicated mammography imaging plate was read out by using a clinical computed radiography reader (Fuji CR 5000; Fuji Medical) with prototype software that was customized for digital mammography and used 100-µm pixels. The compressed breast thickness averaged 8.0 cm in the compression device. The mammographic radiation dose to the breast was estimated by using an 8-cm-thick phantom designed to emulate a 50% glandular, 50% adipose breast (Computerized Image Reference Systems, Norfolk, Va). This phantom was imaged by using the autofilter mode on a Mammography Quality Standards Actcertified clinical mammography system (Lorad Mark IV; Hologic, Danbury, Conn). This system used 32 kVp and 226 mAs with a molybdenum anode and a rhodium filter. The relationship between mAs and air kerma entrance exposure to the breast was determined, and this value was used to calculate the entrance kerma to the breast. Published tables (32) were used to estimate the AGD for the cadaveric breast.
| RESULTS |
|---|
|
|
|---|
5%). Thus, error toward more conservative higher dose estimates existed. The area of the approximately rectangular cross section of the breast during mammographic compression was assumed to be equivalent to the area of the cylindrical pendulant breast (Fig 2), and the resulting diameters of the breasts were related to breast thickness T by using the following formula: Diameter = 2 x [(19.4 x T)/
]1/2 = 4.97 x (T)1/2.
Monte Carlo Simulations
The graph in Figure 3 illustrates the AGD per 100 mAs as a function of kVp. The data in this figure include the inherent inefficiencies of x-ray production at lower kVp (in terms of air kerma per mAs); these inefficiencies are compounded by the fact that 8 mm of added aluminum filtration was used at all kVps. The glandular dose per 100 mAs, which is essentially a Monte Carlo determination of the CT dose index values for breast CT, is not strongly dependent on the diameter of the breast. Data for breast diameters of 8 cm and 14 cm are shown. Data for intermediate breast diameters were computed and were between the two curves, but they are not plotted in Figure 3 for clarity. For comparison, the solid circles in Figure 3 depict the CT dose index values for a 16-cm head phantom reported by Huda et al (39). The values measured by Huda et al were converted from dose in PMMA to dose in glandular tissue values by using the ratio of mass energy attenuation coefficients. There was excellent agreement with the CT dose index values obtained by Huda et al, considering that the x-ray beam filtration and phantom diameter were slightly different.
|
|
|
|
|
Experimental Results
The CT image noise for a 10-cm PMMA phantom is illustrated in Figure 8. The box inset is a CT image of the phantom with the positions of the three regions of interest that were used to compute the noise. The
CT# was greater at a low mAs setting, as expected, and at the same mAs setting, the
CT# was higher at lower kVp, indicating a reduced number of x-ray photons reaching the detector at lower kVp. The power regression lines (
CT# = a[mAs]b, where a and b are constants) with use of the least squares criterion for fitting demonstrated excellent correlation with the measured data points, with r2 values of 0.998, 0.998, and 0.999, for the 80-, 100-, and 120-kVp curves, respectively. The slopes of these log-log curves were determined to be essentially -
(mean ± SD, -0.508 ± 0.0055), which is consistent with the quantum limited behavior of the CT scanner, where
CT# = a(mAs)-1/2, with a as a constant of proportionality. The data shown in Figure 8 allowed the parameterization of noise versus mAs and of noise versus dose, which was useful for subsequent computation of the CNR versus dose and the SNR versus dose.
|
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Compare the typical mammographic setting of a 5-cm compressed breast imaged with 26 kVp and a molybdenum/molybdenum x-ray beam with the setting of an 11-cm-diameter breast imaged with an 80-kVp x-ray beam, as proposed for breast CT (33,37). For 1,000 input photons to the breast, 14 primary photons emerge at mammography, as compared with 90 primary photons that emerge at breast CTa 6.3-fold increase in photon penetration. In terms of energy fluence, for a spectrum with an entrant energy fluence of 1,000 keV (integral of entire spectrum), 17 keV of primary radiation emerges from the breast at mammography, as compared with 96 keV at breast CTbetter x-ray energy penetration by a factor of 5.8. CT is a high-SNR imaging technique that requires relatively high photon fluence to the detectors; however, the higher beam energy and commensurate increase in penetrability of the proposed breast CT spectrum more than compensate for the high fluence requirements of CT. The results of this research demonstrate that high-quality breast CT can be performed at dose levels that are equivalent to or lower than those used in present-day mammography.
The dose homogeneity of breast CT is far greater than that of mammography. For nearly identical AGD values, the peak dose levels for an appreciable portion of the breast are far higher at mammography than the peak levels at breast CT. If radiation risk is truly a linear no-threshold phenomenon, as assumed by regulatory bodies, then the risks at mammography and breast CT are equivalent. However, if radiation risk is a nonlinear function of dose, as some radiobiologic data suggest (42), then the radiation risk with breast CT is lower than that with mammography due to the greater dose homogeneity.
In addition to using well-validated (27,28) Monte Carlo studies, we performed experimental measurements of the noise properties of breast CT, and published data were used to evaluate the contrast properties of this modality. On the basis of the combined contrast and noise data, the CNRs and corresponding SNRs for small breast cancer lesions are impressive. According to the Rose criterion (43), an object will almost certainly be detected if the SNR exceeds 5, and with use of this criterion, lesions as small as 2 or 3 mm in diameter may be easily detected at breast CT. In comparison, the median lesion diameter detected by using screen-film mammography has been reported to be between 11 and 16 mm (4446). These calculations are for the simple case in which the image background surrounding the lesion is homogeneous, which is not the case with most breast imaging studies. This is where the power of tomography comes to play: Because of the reduced overlapping normal anatomy, the image background of breast CT (Fig 13) is far more homogeneous than that of mammography, in which overlapping breast parenchyma produces a very complicated normal breast background. The problem is worse with dense breasts. For an 11-cm-diameter (median) breast and transversely formatted sections, the acquisition of 1-mm-thick CT sections would reduce the volume of underlying and overlying tissue by a factor of 110.
Whereas the amount of structured noise that the approximately 100 out-of-plane sections contribute to the projection image (but not to the CT section) depends on the distribution of breast density, the combination of the improved SNR and the approximately 10-fold reduction in structured noise (
100-fold reduction in variance) suggests that the performance of breast CT in early breast cancer detection may be impressive.
Digital tomosynthesis is a limited angle tomographic technique that has been studied for use in breast cancer screening (47,48). Although tomosynthesis is extremely promising, its potential for mammography has yet to be fully understood. It is likely that dedicated breast CT will result in substantially less structured noise than will tomosynthesis due to the much thinner tomographic sections that are produced at CT. Tomosynthesis will likely provide better spatial resolution but worse contrast resolution than CT and likely provide lower spatial resolution but better contrast resolution than mammography of the dense breast. What the optimum tradeoff is between these parameters for clinical cancer detection remains to be studied.
Breast compression is a necessity at mammography; however, many women are very apprehensive about the compression and hence the mammographic examination (4951). With breast CT, compression is not needed to produce high-quality images, and, given the rotational acquisition requirements of CT, conventional compression cannot be used to an advantage anyway. Because compression is not required for breast CT, this examination may be better tolerated by some women than is mammography.
The finding of microcalcifications is the sole basis of the diagnosis in a minority (
19%) of breast cancer cases (52), and how well breast cancers with microcalcifications can be detected using breast CT remains to be seen. With flat-panel cone-beam acquisition techniques, it would be possible during CT scanning to produce images with higher spatial resolution for microcalcification detection than the reconstructed CT images. The fundamental importance of microcalcification detection in cancer screening may be overestimated because this is the diagnostic area in which mammography excels. More research is warranted, but it is possible that improvements in contrast resolution, with slight compromises in spatial resolution, may yield better overall cancer detection rates.
According to the Rose criterion (53), objects will be seen with high confidence when their SNR is greater than about 5. The data in Figure 11 demonstrate that a 1-mm breast cancer lesion lying against a glandular tissue background has a SNR of about 5. It is likely, however, that when the added anatomic noise is included, SNR levels will decrease and slightly larger lesions will be required to reach a SNR of 5the point where they would be easily detected. If breast CT was able to achieve a median lesion detection level of 5 mm, for example, this would advance early detection by 0.93 years compared with 11-mm lesions, as at mammography, assuming a doubling time of 100 days (54). A median detectable lesion size of 3 mm would result in a 1.5-year advantage in earlier detection. The potential for much earlier detection has important applications for reducing the morbidity and mortality of breast cancer. For example, a 5-mm-diameter lesion has 9.4% of the cell count of an 11-mm lesion, and a 3-mm-diameter lesion has only 2% of the cells of an 11-mm tumor.
In summary, it has been a general perception for more than 2 decades that the radiation dose at breast CT would prohibit its use for breast screening. In this investigation, the radiation dose delivered in the proposed breast CT scanner design was comparable to or lower than the doses delivered at routine mammography. In addition, analysis of phantom CT images revealed that the CNR and SNR of breast cancer would be high at reasonable dose levels. The coronal CT images of a cadaveric specimen acquired at dose levels lower than those used at mammography were promising at subjective inspection; however, clinical studies are needed to scientifically evaluate the potential of breast CT for breast cancer screening. In our opinion, breast CT technology with modern detector systems should be developed so that such studies can be implemented.
| FOOTNOTES |
|---|
CT# = average noise,
SNR = signal-to-noise ratio Author contributions: Guarantor of integrity of entire study, J.M.B.; study concepts and design, J.M.B., T.R.N.; literature research, J.M.B.; experimental studies, J.M.B., J.A.S.; data acquisition, J.M.B., J.A.S.; data analysis/interpretation, all authors; statistical analysis, J.M.B.; manuscript preparation, J.M.B.; manuscript definition of intellectual content, editing, and revision/review, all authors; manuscript final version approval, J.M.B.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
I. Sechopoulos, S. Vedantham, S. Suryanarayanan, C. J. D'Orsi, and A. Karellas Monte Carlo and Phantom Study of the Radiation Dose to the Body from Dedicated CT of the Breast Radiology, April 1, 2008; 247(1): 98 - 105. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. K. Lindfors, J. M. Boone, T. R. Nelson, K. Yang, A. L. C. Kwan, and D. F. Miller Dedicated Breast CT: Initial Clinical Experience Radiology, March 1, 2008; 246(3): 725 - 733. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. K. Kang, E. J. Kim, H. S. Kim, J. S. Sun, and Y. S. Jung Correlation of Whole-Breast Vascularity with Ipsilateral Breast Cancers Using Contrast-Enhanced MDCT Am. J. Roentgenol., February 1, 2008; 190(2): 496 - 504. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. T. Yang, S. Carkaci, L. Chen, C.-J. Lai, A. Sahin, G. J. Whitman, and C. C. Shaw Dedicated Cone-Beam Breast CT: Feasibility Study with Surgical Mastectomy Specimens Am. J. Roentgenol., December 1, 2007; 189(6): 1312 - 1315. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. G. Harish, S. D. Konda, H. MacMahon, and G. M. Newstead Breast Lesions Incidentally Detected with CT: What the General Radiologist Needs to Know RadioGraphics, October 1, 2007; 27(suppl_1): S37 - S51. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Veit-Haibach, G Antoch, T Beyer, H Stergar, R Schleucher, E A M Hauth, and A Bockisch FDG-PET/CT in restaging of patients with recurrent breast cancer: possible impact on staging and therapy Br. J. Radiol., July 1, 2007; 80(955): 50 |