Experimental Studies |
1 From the Department of Radiology, Division of Pediatric Cardiology, University of California San Francisco, 505 Parnassus Ave, L308, San Francisco, CA 94143-0628. Received May 29, 2001; revision requested July 2; revision received September 7; accepted October 26. Address correspondence to C.B.H. (e-mail: charles.higgins@radiology.ucsf.edu).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: In 11 pigs, six nitinol stents (Memotherm), four platinum stents (NuMed), and one elgiloy stent (Wallstent) were placed in the main pulmonary artery under x-ray fluoroscopic guidance. MR imaging was performed 3 months after stent placement before and after injection of NC100150 contrast medium. Blood flow volumes were assessed with velocity-encoded cine MR imaging through and next to the stent. The signal-to-noise ratio and width of susceptibility artifacts of the stents also were determined. Measurements were analyzed with the paired Student t test and Bland-Altman test, where appropriate.
RESULTS: Blood flow volumes measured through the nitinol and platinum stents disclosed no significant difference between velocity-encoded cine MR imaging measurements through and next to the stent. On cine MR images, small susceptibility artifacts were observed around the nitinol and platinum stents. Signal-to-noise ratio in the stent lumen was reduced in nitinol and platinum stents when compared with that next to the stent. The elgiloy stent produced severe susceptibility artifacts, making measurement of flow volumes impossible. NC100150 injection caused no significant effect on flow volume measurements. It improved the signal-to-noise ratio of the pulmonary arterial lumen outside and, to a lesser extent, inside the stent.
CONCLUSION: Assessment of morphology and flow volumes through nitinol and platinum stents is feasible with MR imaging. Blood-pool contrast media provide persistent signal enhancement in the pulmonary artery and, to a lesser extent, in the lumina of nitinol and platinum stents.
© RSNA, 2002
Index terms: Magnetic resonance (MR), artifact, 944.93 Magnetic resonance (MR), cine studies, 944.121419 Magnetic resonance (MR), contrast media, 944.12143 Magnetic resonance (MR), vascular studies, 944.12144 Stents and prostheses, 944.1268
| INTRODUCTION |
|---|
|
|
|---|
Susceptibility artifacts at the stent wall and radio-frequency (RF) shielding effects inside the stent lumen can be misinterpreted as stent obstruction or occlusion (913). Moreover, quantitative assessment of blood flow through endovascular stents by using VEC MR imaging is considered to be problematic because of phase offset phenomena associated with magnetic susceptibility. Investigators in an in vitro study (14) to determine the accuracy of VEC MR imaging measurements in the presence of a stainless steel coronary stent demonstrated significant decrease in flow velocity when it was measured near the stent. Several recent studies (15,16) have focused on characterization of the effect of a variety of stents at MR imaging and contrast-enhanced MR angiography. In these studies, readable MR images were obtained for some MR imagingcompatible stent materials and designs in conjunction with optimal sequence parameters such as short echo times and high flip angles.
Moreover, administration of extracellular gadolinium chelates was reported to improve MR imaging signal intensity inside the stent lumen (17). On the other hand, the value of blood-pool (intravascular) contrast agents in MR imaging of stents has not, to our knowledge, been evaluated. Blood-pool contrast agents allow prolonged delineation of central and peripheral vasculature and can provide greater signal-to-noise ratio (SNR) than can extracellular gadolinium chelates (18,19). Tello et al (20) demonstrated in eight patients the ability of first-pass gadopentetate dimeglumine in depicting the patency of stents placed in renal and iliac arteries.
The current study was designed to determine the effect of three stent types on VEC MR imaging and fast segmented cine MR imaging in an in vivo animal model. The potential beneficial effect of a blood-pool contrast medium also was evaluated. Specific aims were (a) to determine the accuracy of quantitative flow volume measurements through various types of endovascular stents, (b) to compare the effect of magnetic susceptibility and RF shielding on MR imaging in vivo, and (c) to evaluate effects of blood-pool contrast media on quantitative flow volume measurements, magnetic susceptibility, and RF shielding effects by using steady-state distribution of the contrast media.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Blood-Pool MR Imaging Contrast Medium (NC100150 Injection)
This blood-pool contrast agent (Clariscan; Nycomed Amersham, Oslo, Norway) is composed of single crystals with diameters of 47 nm and is stabilized with a carbohydrate polyethylene glycol. The iron oxide particles (particle concentration, 0.5%1.0% weight/volume) were suspended in an isotonic glucose solution. The final diameter of the NC100150 injection is approximately 20 nm. The longitudinal relaxation rate of the agent is approximately 20 mmol/L/sec, and the transverse relaxation rate is 35 mmol/L/sec at 37°C and 0.5 T. Injection of NC100150 reduces blood T1 to less than 100 msec. Iron oxide particles are considered blood-pool agents, since their plasma half-life is 90 minutes in humans. In clinical trials now being conducted in the United States and Europe, ultrasmall superparamagnetic iron oxide, or USPIO, particles have been well tolerated in humans (21,22). USPIO particles represented by NC100150 injection have been successfully used for MR angiography of peripheral and pulmonary vessels and for three-dimensional cardiac cine MR imaging (2326). The agent was obtained as a septum-sealed vial containing 10 mg iron per milliliter.
Study Design
All procedures were performed in accordance with the National Institute of Health guidelines for care and use of laboratory animals and with approval of the committee of animal research of the University of California, San Francisco. In 11 female farm pigs (mean body weight, 14.1 kg ± 1.5; Pork Power Farms, Turleck, Calif), endovascular stents were placed in the main pulmonary artery by using a transcatheter technique under x-ray fluoroscopic guidance (T.K., K.G., D.T., P.M.). Cine angiography was performed to verify the stent position. Postcatheterization treatment included oral administration of 250 mg/kg per day cephalosporin (Suprax; Wyeth-Ayerst, Philadelphia, Pa) over 1 week and 80 mg per day aspirin over the entire study.
The animals were studied with MR imaging 3 months after stent implantation. At this time, the weight of the animals had increased to 33 kg ± 2.3. Imaging was performed before and after intravenous injection of 0.05 mmol iron per kilogram NC100150. For all procedures, the animals were intramuscularly injected with a mixture of 0.025 mg/kg telazol (Telazol; Wildlife Pharmaceuticals, Fort Collins, Colo), ketamine hydrochloride (Ketaset; Fort Dodge Laboratories, Fort Dodge, Iowa), and xylazine hydrochloride (Anased; Lloyd Laboratories, Shenandoah, Iowa) for induction of and 2% isoflurane (Isoflo; Abbott Laboratories, North Chicago, Ill) inhalation for maintenance of general anesthesia. After completion of the MR imaging study, the animals were euthanized with 200 mg/kg intravenously administered sodium pentobarbital (Nembutal Sodium Solution; Abbott Laboratories). The heart was excised to evaluate the position and morphology of the stent and pulmonary artery. Stent placement was successful in all 11 animals. The stents were well tolerated, without complications or migration, over the 3-month period.
MR Imaging
Flow volume measurements.MR imaging was performed by using a 1.5-T imager (Signa 5; GE Medical Systems, Milwaukee, Wis) with a standard body coil. All MR images were acquired with electrocardiographic gating. The positions of the main pulmonary artery and stent were identified with a set of scout images in sagittal and transverse planes.
VEC MR imaging sequence.This MR imaging sequence was used to quantify pulmonary blood flow volumes through and next to the stent (2729). All VEC MR imaging measurements were made in a double oblique plane perpendicular to the dominant flow direction in the main pulmonary artery. The distance between any visible stent artifact and the site of VEC MR imaging measurements next to the stent was chosen to be greater than 10 mm. Measurement errors due to the stent artifact were not expected at this distance. The effect of surgery on flow volume measurements has been addressed in control animals in our previous publication (30).
For VEC MR imaging measurements next to the stent, velocity maps were obtained by manually tracing the main pulmonary artery cross-sectional area on the magnitude images and then applying it to the phase images (T.K., M.S.). For VEC MR imaging measurements through the stent, the contour of the stent was traced on both VEC magnitude images and VEC phase images, since the phase images showed less susceptibility artifact. Flow volume measurements of both methods for tracing the contour area were compared. There was a 1.0% ± 3.6 difference between the two methods, and these measurements were not significantly different. However, definition of the stent wall was found to be more unequivocal on VEC phase images; therefore, it was used in the calculation of flow volume measurements.
Flow volumes were calculated by multiplying cross-sectional area and spatial average flow velocity within the contour. Instantaneous flow volumes were summed to obtain the total flow volume per cardiac cycle. The following acquisition parameters were used for VEC MR imaging: repetition time msec/echo time msec, 25/7; section thickness, 5 mm; flip angle, 30°; receiver bandwidth, 31.25 kHz; field of view, 24 x 24 cm; matrix, 256 x 192; number of signals acquired, two; velocity-encoding gradient, 200 cm/sec. The average acquisition time was 3.19 minutes.
Cine MR imaging sequence.This sequence was used to determine right ventricular stroke volumes (3133). End-systolic and end-diastolic volume images were measured by manually tracing the area of the endocardial surfaces (T.K., M.S.). Ventricular chamber volumes were computed as the sum of right ventricular volumes of all short axis view sections containing right ventricular chambers. The area of each section was multiplied by section thickness to determine the volume of each level. Stroke volume was calculated as the difference between end-diastolic and end-systolic volumes. The following imaging parameters were used for cine MR imaging: 8/3; section thickness, 10 mm; spacing, 0; flip angle, 20°; receiver bandwidth, 31.25 kHz; field of view, 24 x 24 cm; matrix, 256 x 128; number of phases, 16; acquisition time, 4.17 minutes. Seven to ten sections were acquired to encompass the entire right ventricle.
Magnetic susceptibility and RF shielding.The effects of susceptibility and RF shielding were analyzed quantitatively on VEC MR images acquired perpendicular to and on cine MR images acquired parallel to the long axis of the stent. All VEC and cine MR imaging was performed through the center portion of the stent. Susceptibility artifacts at the stent wall were measured as the number of pixels yielding total signal void. The degree of RF shielding was determined by comparing the MR imaging signal intensity measured in the center portion of the stent lumen with measurements obtained in the center of the pulmonary artery next to the stent (>10 mm) (T.K., M.S.).
In general, the surface area of the examined vessel (without stents) is traced on magnitude images from VEC MR imaging. In this study, however, we measured flow volumes through endovascular stents. It is well known that metallic stents produce different degrees of artifacts on MR images. We found that on magnitude images artifacts were relatively large and superimposing the vessel wall. This made unequivocal delineation of the wall difficult. On phase images, stent artifacts were relatively smaller than on magnitude images. Therefore, unequivocal delineation of the vessel wall was relatively easier. In the current study, we compared the two methods of tracing the vessel wall (once on the magnitude and once on the phase images). We found that the difference in flow volume measurements was 1.0% ± 3.6 (P = not significant; paired Student t test) between the two methods of tracing in the nitinol and platinum stents.
The rationales for examining the susceptibility and RF shielding effects of each stent on flow were: (a) The differences in the alloy of these stents cause different degrees of shielding effects, and (b) the magnitude of susceptibility artifact g is dependent on the material properties of the stent.
Influence of blood-pool contrast medium on cine and VEC MR imaging.To assess the effects of blood-pool contrast media on cine and VEC MR images during steady-state distribution, measurements of pulmonary flow volumes, SNR, and artifact width were repeated after intravenous injection of 0.05 mmol iron per kilogram NC100150 (M.S., T.K.). All measurements were completed within 30 minutes after administration of the contrast agent.
Statistical Analysis
Data are presented as mean ± SD. The unpaired Student t test and analysis of variance were used to determine differences between animals with different stents. The paired Student t test was applied for comparison of measurements within the same animal. A P value of less than .05 was considered to indicate a significant difference. Linear regression analysis with calculation of SDs was used to determine correlation between pulmonary flow volume measurements assessed with VEC MR imaging through and next to the stent. In addition, the Bland-Altman test was performed to determine agreement between pulmonary flow volumes as measured with VEC MR imaging through and next to the stent and with cine MR imaging.
| RESULTS |
|---|
|
|
|---|
Effects of Various Stents on Flow Volume Measurements
The elgiloy stent (Wallstent) caused susceptibility artifacts at the implantation site that prohibited visualization of underlying structures (Fig 1). Therefore, measurement of pulmonary blood flow in or next to this stent was not possible due to susceptibility artifacts.
|
|
Appearance of Various Stents at MR Imaging
Image analysis for the elgiloy stent was not feasible due to severe susceptibility artifacts extending from the stent wall over the entire lumen and beyond the border of the stent on VEC and cine MR images. In the nitinol and platinum stents, circular margins were well defined on VEC phase images (Fig 2). In the nitinol stents, a ring of artifact 1 pixel ± 0.4 wide was noted. For the platinum stents, the artifact width was 1.1 pixels ± 0.8. In contrast, VEC magnitude images showed greater susceptibility artifacts extending over 3.9 pixels ± 0.8 in the nitinol stent and 4.1 pixels ± 0.6 in the platinum stent. The artifacts on the magnitude images prevented accurate identification of the contour and dimensions of the stent.
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
The accuracy of MR imaging to quantify vascular blood flow has been validated in several studies (2729). However, the question of whether flow measurements are feasible after stent placement has not, to our knowledge, been answered. Because magnetic susceptibility results in intravoxel spin dephasing, phase-contrast sequences used for blood flow measurements may be especially prone to phase offset errors. One specific aim of this study was to evaluate the effect of different stent types on VEC MR imaging measurements in situ. Results of the current study show that flow volume measurements are reasonably accurate when obtained inside stents with relatively low magnetic susceptibility and large surface areas. On the other hand, flow measurements obtained through stents with high magnetic susceptibility, such as elgiloy stents, were not feasible. These findings imply that phase errors caused by local field inhomogeneities around nitinol and platinum stents are limited to areas adjacent to the stent wall and do not significantly affect flow measurements through stents with large diameters. In further research, pulse sequences with short echo times or refocusing pulses should be investigated to reduce intravoxel spin dephasing and thus to minimize errors in quantitative flow measurements. By using such pulse sequences, noninvasive MR imaging evaluation of stent patency in small vessels such as the coronary arteries might become feasible.
One major problem in MR imaging of endovascular stents is susceptibility artifacts (34,35). Susceptibility artifacts cause phase and frequency offset phenomena leading to image distortion at the stent wall. The extent of the artifact depends on several factors including but not limited to material properties, gradient field strength, orientation of the stent toward the main magnetic field, and imaging parameters such as echo time and receiver bandwidth. In the current study, MR imaging of elgiloy stents (Wallstents), which are fabricated from a strongly magnetic alloy, showed severe susceptibility artifacts, which prevented appropriate definition of stent wall and lumen patency. In contrast, magnetic susceptibility values of nitinol (Memotherm) and platinum (NuMed) stents are relatively close to those of human tissue. In the current study, both stent types yielded very similar sizes of susceptibility artifacts. On cine MR images, artifact size was relatively small and allowed precise definition of stent morphology and position in the pulmonary artery. On images obtained through the walls of nitinol and platinum stents, the structure of the wire mesh was distinguishable from surrounding tissue. This suggests the possibility that stent obstruction by tissue proliferation might be directly visualized on MR images when using appropriate imaging sequences and stent designs.
Differences in artifact sizes of stents on cine and VEC MR magnitude images are presumably due to differences in the echo times of the two MR imaging sequences (3 and 7 msec, respectively). On VEC phase images, signs of artifacts were even less pronounced than on cine MR images, although longer echo times were used. A possible explanation is that velocity maps compensate for spin frequency offset phenomena, which are caused by local field inhomogeneities due to susceptibility.
Another problem in MR imaging of endovascular stents is the RF (B1-field) shielding effect (36). RF pulses induce electrical eddy currents in closed loops of the stent mesh and thus magnetic flux. The main component of the magnetic flux opposes the originating RF field. Consequently, flip-angle excitation inside the stent lumen is reduced, leading to poor SNR. The amount of shielding depends mainly on the design of the stent mesh, stent geometry (eg, length), orientation of the stent toward the RF pulse direction, and sequence parameters such as flip angle. Stents designed with few or no conducting closed loops would avoid electrical flow of eddy currents (37).
In the current study, differences in SNR were found when measured inside the lumina of platinum and nitinol stents. One obvious difference between the nitinol and platinum stents was length (17 and 25 mm, respectively). Penetration of RF pulses through the open cross-sectional ends of a stent may be one important source for improving intraluminal SNR. Thus, short stents may benefit from this mechanism more than long stents.
There were three reasons for using the blood-pool MR imaging contrast medium: (a) to determine the effect of this blood-pool agent on flow volume measurement, (b) to improve contrast between blood in the left ventricular chamber and in myocardium on cine MR images, and (c) to visualize the patency of the stents after deployment. In the current study, injection of NC100150 improved image quality in the stent lumen slightly but significantly. However, this contrast agent clearly enhanced the contrast between the endovascular stent, vascular blood pool, and surrounding tissue. Lack of substantial enhancement of the stent lumen after shortening of blood T1 is most likely related to the RF shielding effects produced by all of the stents used in the current study. NC100150 injection enabled persistent and excellent delineation of central pulmonary vasculature for at least 30 minutes. In addition, the contrast media did not have any significant effects on VEC MR imaging flow volume measurements.
Bolus administration of the extracellular MR imaging contrast medium gadopentetate dimeglumine has been previously used to assess the patency of endovascular stents placed in the renal and iliac arteries in patients (20). In the current study, in contrast with previous first-pass studies (20), NC100150 injection was used to detect patency of the pulmonary artery by using the steady-state distribution approach of the contrast medium. An advantage of this approach is that MR imaging can be performed in a wide time window (approximately 30 minutes) after injection. It was anticipated that the contrast medium would have no effect on cardiac output.
Authors of recently published reports (3842) have documented the feasibility of MR imaging not only to detect vascular obstruction but also to guide angioplasty and endovascular stent placement for their treatment. High and persistent contrast between the vascular blood pool and endovascular devices, as achieved with NC100150 injection, might be of importance for visualizing stent patency and when using interventional MR imaging techniques for treatment of vascular obstruction or after deployment of endovascular stents. VEC MR imaging flow measurements before and after stent placement could then be used for documentation of the effectiveness of the procedure and for noninvasive follow up.
Practical application: Appropriate imaging sequences and stent types are crucial factors for noninvasive assessment of endovascular stents with MR imaging. Quantitative blood flow measurements with VEC MR imaging and use of blood-pool MR imaging contrast media are promising methods for accurate assessment of stent morphology and patency.
| FOOTNOTES |
|---|
Author contributions: Guarantors of integrity of entire study, T.K., M.S., C.B.H., P.M.; study concepts and design, T.K., M.S., C.B.H., P.M.; literature research, T.K., M.S.; experimental studies, T.K., M.S., K.G., D.T., P.M., G.R.; data acquisition, analysis, and interpretation, T.K., M.S.; statistical analysis, T.K., M.S.; manuscript preparation, T.K., M.S.; manuscript definition of intellectual content, T.K., M.S., C.B.H.; manuscript editing, T.K., M.S., P.M., C.B.H.; manuscript revision/review, T.K., M.S.; manuscript final version approval, all authors.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. Kuehne, M. Saeed, C. B. Higgins, K. Gleason, G. A. Krombach, O. M. Weber, A. J. Martin, D. Turner, D. Teitel, and P. Moore Endovascular Stents in Pulmonary Valve and Artery in Swine: Feasibility Study of MR Imaging-guided Deployment and Postinterventional Assessment Radiology, February 1, 2003; 226(2): 475 - 481. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||