Experimental Studies |
1 From the Center for Molecular Imaging Research, CNY149-5403, Department of Radiology, Massachusetts General Hospital, Bldg 149, 13th St, Charlestown, MA 02129. Received September 26, 2001; revision requested December 3; revision received December 17; accepted January 7, 2002. U.M. supported in part by a research grant from the RSNA Research and Education Foundation. Address correspondence to U.M. (e-mail: mahmood@helix.mgh.harvard.edu).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: The imaging system was composed of modular parts, including a light source, excitation filters, emission filters, and a charged-coupled device for recording images. On the basis of generated excitation and absorption spectra of green fluorescent protein (GFP), tricarbocyanine 5.5 (Cy5.5), and indocyanine green (ICG), filters were selected to allow spectral separation and optimize resultant recorded signal. The system was tested by using a combination of the fluorochromes to confirm spectral separation. In vivo tests were performed in nude mice with tumors that expressed cathepsin B, which could be evaluated by using a Cy5.5-based activatable probe and GFP. For each in vivo tumor type and channel, statistical analysis was performed on the basis of signal intensity in the region of interest.
RESULTS: The different fluorochromes were readily distinguished with the system; characteristics such as power were determined for all wavelengths. The system demonstrated a linear response for GFP, a monotonic response for Cy5.5 over a range of more than three orders of magnitude of concentration, and a more complex response for ICG. In vivo analysis demonstrated the ability to image GFP expression and cathepsin B expression separately in tumors: As expected, marked differences were observed in GFP-expression imaging between tumor types (1,363 arbitrary units [AU] ± 236 [SD] vs 110 AU ± 11 for GFP-positive and GFP-negative tumors, respectively; P < .001), whereas similar cathepsin B expression (1,070 AU ± 285 vs 1,168 AU ± 367; P > .5) was observed. Histologic analysis confirmed in vivo findings.
CONCLUSION: Imaging multiple gene expressions simultaneously in vivo by using optical imaging is feasible.
© RSNA, 2002
Index terms: Experimental study Genes and genetics Neoplasms, experimental studies
| INTRODUCTION |
|---|
|
|
|---|
The goal of this study was to develop and test a multichannel reflectance imaging system for small animals on the basis of a previously developed setup (16). We hypothesized that, analogous to multichannel fluorescence microscopy (which typically uses fluorochromes in the visible region), near-infrared multichannel imaging should be feasible in experimental in vivo settings and ultimately in clinical settings, as well. The major advantages of imaging different biologic targets simultaneously and independently include the ability to (a) colocalize multiple targets, (b) probe for differential expression levels of multiple targets, (c) analyze the combination of expression levels of particular importance in cancer, where one target alone is rarely overexpressed, (d) develop miniarrays for in vivo target assessment, (e) image the temporal and spatial correlation of distinct enzymatic pathways in disease, (f) image the effects on a pathway of therapy targeting another pathway, and (g) provide reference channels for activatable "smart" sensors.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Overall, the imaging system consists of multiple separate functional components encased in a black light-tight box:(a) broadband white-light (halogen) source, (b) selective excitation filters, (c) imaging chamber for the animal, (d) emission filters to select fluorescence, (e) charged-coupled device for recording emission photons, and (f) computer for image acquisition control and storage. The excitation and emission filters are readily exchangeable, with switching times between channels currently approximately a few seconds. To maximize spectral purity, all filters are five-cavity interference filters (Omega Optical, Brattleboro, Vt). All excitation filters are 25 mm in diameter, and all emission filters are 50 mm in diameter.
The absorption and emission spectra for all the fluorochromes of interest were obtained: intracellular enhanced green fluorescent protein (GFP), tricarbocyanine 5.5 (Cy5.5), and indocyanine green (ICG) spectra were determined with the use of a fluorometer (U-4500; Hitachi, Tokyo, Japan) (Fig 1); fluorescein absorption and emission spectra were also obtained (not shown for clarity of Figure). The choice of bandpass filter systems was based on the observed wavelengths of the fluorochromes. The filter sets (especially for GFP and Cy5.5) were chosen to maximize signal received from the fluorescent moleculesthat is, emission filter bandpasses were matched to the emission spectra, whereas excitation bandpasses were shifted to shorter wavelengths than peak excitation frequency to decrease cross-talk between excitation and emission channels. The exact center and bandpass wavelengths for the filter sets are detailed in the Table and are also depicted graphically in Figure 1.
|
|
Fluorochromes
Four fluorochromes were used in this study as primary reporters: fluorescein (Sigma, St Louis, Mo), enhanced GFP, Cy5.5 (Amersham-Pharmacia, Piscataway, NJ), and ICG (Akorn, Buffalo Grove, Ill). ICG and fluorescein were used as examples of nontargeted reporter probes, while GFP was expressed in a gliosarcoma cell line as a fluorescent optical marker gene in tumor cells. Cy5.5 was used as an enzyme-responsive reporter to measure cathepsin B enzyme activity (12). This probe belongs to a group of smart probes in which the fluorochrome is activated (ie, markedly increases in fluorescence) with enzymatic cleavage (17).
GFP-expressing Cell Line
The 9L rodent glioma cell line and a stably transfected clone that overexpresses GFP (18) were cultured in Dulbecco modified Eagle medium supplemented with 10% fetal calf serum and 1% penicillin and streptomycin (all products from Cellgro Mediatech, Washington, DC). The GFP-positive clone was selected by using the antibiotic G418 (Cellgro Mediatech), since the plasmid encoding for GFP also contained a gene for resistance to G418. For in vitro studies, cells were resuspended in Hanks buffered saline without phenol red.
In Vitro and Phantom Evaluation
The imaging system was tested by using phantoms consisting of 96-well cell culture plates (Corning, Corning, NY). For evaluating the effectiveness of spectral separation, wells containing 6 x 106 9L GFP-positive glioma cells, 100 µL of 100 µmol/L (10 nmol) free Cy5.5 dye, or 100 µL of 100 µmol/L (10 nmol) free ICG were used. White-light images were obtained at 0.075 second (minimum system acquisition time). Fluorescence imaging times of 30 seconds per channel were used, although an excellent signal-to-noise ratio can be achieved in a fraction of this time (35 seconds) at these concentrations. Samples were evaluated by using a dilution series from 100 µmol (100 µL x 1 mmol/L) through 0.01 nmol (100 µL x 0.1 µmol/L) for all three fluorochromes (fluorescein, Cy5.5, and ICG) by using acquisition times (for all dilutions) of 15, 24, and 90 seconds for the GFP and/or fluorescein, Cy5.5, and ICG channels, respectively. Fluorescein was used to evaluate the GFP channel for the serial dilution experiments, because unlike intracellular GFP, molar concentration is readily determined. To additionally evaluate the GFP channel more directly, 9L GFP-positive and 9L GFP-negative cells were placed in a 96-well plate with a 16x concentration range evaluated at twofold serial increments in cell number from 62,500 to 1,000,000 cells per well.
Animal Experiments
For in vivo evaluation of the imaging system and demonstration of multichannel imaging of different gene expressions, 105 tumor cells were implanted into the subcutaneous tissues of the lower anterior abdominal walls of four nude mice (nu/nu Taconic, Germantown, NY), with 9L GFP-positive glioma cells injected on the animals right side and 9L GFP-negative glioma cells injected on the animals left side. Animals were imaged 24 hours after intravenous injection of 2 nmol of cathepsin B enzymeactivated probe (containing Cy5.5; see "Fluorochromes," earlier in article) when the tumors were approximately 4 mm in diameter; cathepsin B was expressed in both tumors. In vivo imaging times were 0.075 second for the white-light anatomic image, 2 minutes for the GFP image, and 30 seconds for the cathepsin B enzyme activity image (measured by means of Cy5.5 fluorescence).
Approval was received for this study from our institutions animal care committee. Ketamine (subcutaneous administration of 9 mg per 100 g) and xylazine (subcutaneous administration of 0.9 mg per 100 g) were used as an anesthetic, and pentobarbital (intraperitoneal administration of 20 mg per 100 g) was used for sacrifice. For histologic evaluation, animals were euthanized, and tumors were removed and snap frozen in liquid nitrogen. Sections remained unstained for fluorescence microscopic evaluation. By using standard fluorescent microscopy filter sets (Omega Optical) for GFP and Cy5.5, the same sections were evaluated by using white-light, GFP, and Cy5.5 channels.
Image and Statistical Analysis
All regions of interest (ROIs), both in vitro and in vivo, were placed by one author (U.M.). For GFP-positive and GFP- negative cell phantom experiments and for free fluorochrome experiments, a circular ROI was placed that encompassed the entire well for each cell concentration. Single mean values for signal intensity were obtained for each well, and standard linear regression was performed to correlate signal intensity with cell count per well.
For in vivo evaluation, mean signal intensity values were obtained by using ROI measurements obtained immediately over the region of GFP-positive and GFP-negative tumors for both GFP-channel and Cy5.5-channel images for each tumor. Because of the scattering of light through tissue that resulted in a penumbra surrounding the tumors, a same-sized circular ROI was placed over the area of the tumor with the highest signal intensity. The smaller size of the ROI compared with the tumor size was chosen to result in less variability between measurements of the same tumor and of different tumors. ROI size was 177 pixels. Mean values and SDs of signal intensity for each tumor type and for each channel were then calculated for the four animals collectively on the basis of the mean ROI intensities calculated for each tumor type and channel. Significance was determined by using the Student t test and reported as P values.
| RESULTS |
|---|
|
|
|---|
|
|
|
|
Correlation between Fluorochrome Concentration and Signal Intensity
Figure 3 summarizes the correlation between fluorochrome concentration and normalized fluorescence signal intensity obtained with use of the imaging system. As is evident from the data, there is a monotonic increase in signal intensity with fluorescein and Cy5.5 over a range of more than three orders of magnitude of concentration. At the highest concentration tested (1 mmol/L), quenching of Cy5.5 is observed, as expected. The graph also demonstrates some of the idiosyncrasies of ICG, which is known to demonstrate considerable concentration-dependent shifts and changes in the shape of absorption and emission spectra (19). Hence, dyes other than ICG may be more useful in multichannel optical imaging in combination with Cy5.5.
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
The multichannel paradigm places additional constraints on the system when compared with single-channel systems. In the current study, the filter sets (especially for GFP and Cy5.5) were chosen to maximize signal received from the fluorescent moleculesthat is, emission filter bandpasses were matched with the emission spectra. Because of the relatively small Stokes shifts between excitation and emission (typically on the order of 20 nm), a conscious decision was made to excite the fluorochromes at slightly lower wavelengths than their maximum absorption to decrease the likelihood of excitation photon bleed-through into the final image. The decreased fluorescence yield from exciting at a wavelength that is lower than the maximum can be overcome by increasing excitation photon flux or by increasing the sensitivity of the receiving charged-coupled device. The former method is easier and less costly than the latter and was used in the current configuration. Moreover, relatively wide bandpasses (3040 nm) were chosen to maximize both excitation photon flux and the number of emission photons that were seen with the charged-coupled device. Despite the five-cavity interference mechanism and relatively sharp cutoffs, the bandpass filters have relatively long, albeit small, tails. This is much less of an issue in histologic fluorescence settings, given the markedly smaller volume of background tissue providing a reflective surface for bleed-through when compared with the in vivo setting. Future improvements include the addition of polarizing filters to further separate excitation and emission bands, thus allowing closer frequency spacing of excitation and emission wavelengths. More imaging channels may also be added, including Cy3.5 between GFP and Cy5.5 and channels for dyes further in the near-infrared region of the spectrum beyond ICG.
The current multichannel configuration can be readily applied to tomographic systems. An additional optical switch in series with those typically present to change excitation location can be implemented to change source wavelength. Mechanical, manual, or optical switching of emission filters would also be required. Thus, it is expected that for many applications, more than one biologic parameter will be interrogated. Rational probe design for the multichannel paradigm will be disease specific: Breast cancer, for example, may have activatable probes for extracellular proteases combined with activatable probes at a second wavelength for intracellular kinases. Such combinations, which are essentially targeted in vivo miniarrays, will increase the sensitivity and specificity of disease detection and will allow more specific follow-up of antineoplastic therapy.
Practical applications: A number of the most common diseases that affect people today, including many cancers, cardiovascular diseases, inflammatory diseases, neurologic disorders, and some degenerative disorders, may potentially be better assessed at an earlier stage in the future when optical imaging is incorporated into the clinical evaluationimaging paradigm. Multichannel optical imaging further expands the utility of this modality by allowing the spatial distribution of biologic parameters, including different gene expressions, to be determined nearly simultaneously. Since many pathologic conditions, especially predisease states, fall outside of simple one-gene Mendelian genetics, a multiple-gene imaging approach may help in our quest for earlier detection.
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
Author contributions: Guarantor of integrity of entire study, U.M.; study concepts and design, U.M., R.W.; literature research, U.M.; experimental studies, all authors; data acquisition, U.M., Y.T.; data analysis/interpretation, U.M.; statistical analysis, U.M.; manuscript preparation, U.M.; manuscript definition of intellectual content and editing, U.M., R.W.; manuscript revision/review, all authors; manuscript final version approval, U.M., R.W.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. S. Gee, R. Upadhyay, H. Bergquist, H. Alencar, F. Reynolds, M. Maricevich, R. Weissleder, L. Josephson, and U. Mahmood Human Breast Cancer Tumor Models: Molecular Imaging of Drug Susceptibility and Dosing during HER2/neu-targeted Therapy Radiology, September 1, 2008; 248(3): 925 - 935. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Upadhyay, R. A. Sheth, R. Weissleder, and U. Mahmood Quantitative Real-time Catheter-based Fluorescence Molecular Imaging in Mice Radiology, November 1, 2007; 245(2): 523 - 531. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Alencar, M. A. Funovics, J. Figueiredo, H. Sawaya, R. Weissleder, and U. Mahmood Colonic Adenocarcinomas: Near-Infrared Microcatheter Imaging of Smart Probes for Early Detection--Study in Mice Radiology, July 1, 2007; 244(1): 232 - 238. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Wang, M. Iyer, A. Annala, L. Wu, M. Carey, and S. S. Gambhir Noninvasive indirect imaging of vascular endothelial growth factor gene expression using bioluminescence imaging in living transgenic mice Physiol Genomics, January 12, 2006; 24(2): 173 - 180. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Funovics, R. Weissleder, and U. Mahmood Catheter-based in Vivo Imaging of Enzyme Activity and Gene Expression: Feasibility Study in Mice Radiology, June 1, 2004; 231(3): 659 - 666. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Botnar, A. S. Perez, S. Witte, A. J. Wiethoff, J. Laredo, J. Hamilton, W. Quist, E. C. Parsons Jr, A. Vaidya, A. Kolodziej, et al. In Vivo Molecular Imaging of Acute and Subacute Thrombosis Using a Fibrin-Binding Magnetic Resonance Imaging Contrast Agent Circulation, April 27, 2004; 109(16): 2023 - 2029. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. Thrall Nanotechnology and Medicine Radiology, February 1, 2004; 230(2): 315 - 318. [Full Text] [PDF] |
||||
![]() |
M. F. Kircher, U. Mahmood, R. S. King, R. Weissleder, and L. Josephson A Multimodal Nanoparticle for Preoperative Magnetic Resonance Imaging and Intraoperative Optical Brain Tumor Delineation Cancer Res., December 1, 2003; 63(23): 8122 - 8125. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. R. Herschman Molecular Imaging: Looking at Problems, Seeing Solutions Science, October 24, 2003; 302(5645): 605 - 608. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Matrisian, G. W. Sledge Jr., and S. Mohla Extracellular Proteolysis and Cancer: Meeting Summary and Future Directions Cancer Res., October 1, 2003; 63(19): 6105 - 6109. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Mahmood and R. Weissleder Near-Infrared Optical Imaging of Proteases in Cancer Mol. Cancer Ther., May 1, 2003; 2(5): 489 - 496. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||