Technical Developments |
1 From the Departments of Medical Physics (T.J.C.) and Radiology (H.A.R., V.M.H.), University of Wisconsin, Madison. Received February 21, 2002; revision requested May 7; final revision received September 13; accepted September 23. Supported by National Institutes of Health grant R01-HL66488-01 A1. Address correspondence to T.J.C., Northwestern University, Departments of Biomedical Engineering and Radiology, 448 E Ontario St, Suite 700, Chicago, IL 60611 (e-mail: t-carroll@northwestern.edu).
| ABSTRACT |
|---|
|
|
|---|
© RSNA, 2003
Index terms: Brain, infarction, 10.4352 Cerebral blood vessels, flow dynamics, 17.76 Magnetic resonance (MR), contrast enhancement, 10.12143 Magnetic resonance (MR), diffusion study, 10.12144 Magnetic resonance, vascular studies, 10.12144
| INTRODUCTION |
|---|
|
|
|---|
Ostergaard and co-workers (5,6) and Wirestam et al (7) have shown that an accurate determination of relative CBF is possible by using MR imaging methods through mathematical deconvolution of a tissue concentration curve with an arterial input function (AIF). In current implementations of the MR imaging perfusion study, the AIF is determined interactively on the basis of an operator-chosen region of interest that is assumed to represent signal intensity changes in a large feeding vessel, such as the middle cerebral artery (6,7). Some groups have proposed methods to calculate an AIF from the time of arrival of the contrast material and the time course of signal intensity changes in the proximal carotid artery (8). In both of these approaches, off-line postprocessing of the MR perfusion data is required, and this postprocessing necessitates additional time, a dedicated workstation, and specially trained operators. These additional steps and the need for advanced training could pose practical barriers to widespread clinical application of dynamic susceptibility contrast MR imaging perfusion techniques.
We have developed a means of determining AIF with dynamic susceptibility contrast MR imaging on the basis of adaptive thresholding. The purpose of this study was to compare the AIFs calculated with an automated method without operator input with those calculated with the conventional interactively determined region of interestbased method and to determine if the automated method provided technically adequate results in clinical practice.
| Materials and Methods |
|---|
|
|
|---|
Automated AIF
The details of the automated AIF algorithm (9) appear as an Appendix to this report. In short, the automated AIF algorithm identifies a single voxel to represent the AIF on the basis of arrival time of contrast medium and integrated signal change of the voxel. Thresholds for the arrival of contrast medium are measured in units of SD of the precontrast signal intensity. Physiologic and hardware-related fluctuations are thereby incorporated into the threshold on an examination-by-examination basis. The mean precontrast signal intensity is calculated from five image sets acquired before the administration of contrast medium, and the SD of the precontrast signal intensity is calculated. The arrival time of the contrast medium is defined as the time at which signal intensity first deviates from the mean by 10 SDs. Voxels with arrival times of more than 2 seconds later than the first arrival times were rejected because they likely represented voxels with predominantly venous circulation. The automated AIF voxel was selected as the time course in the remaining voxels that exhibited the largest integrated change in signal intensity. A time integration window, roughly 8 seconds, containing four consecutive images was used to remove transient noise spikes, as well as to optimize the chosen AIF with respect to relative CBV.
Image Processing
The resulting time-series of images were postprocessed through deconvolution (6,7) of the AIF and indicator dilution techniques (10). To calculate hemodynamic parameters (13), gadolinium concentration as a function of time was determined for each voxel, up to a scale factor, on the basis of the relationship determined with the following equation:
|
|
A stack of two-dimensional images, one image per section, in which signal intensity was proportional to CBF (relative CBF) were produced automatically in less than 1 minute. Additional images of CBV (relative CBV) and MTT were produced (13). All image-processing software used at MR imaging was written in house and installed on the host console computer of the imagers. The calculated relative CBF, relative CBV, and MTT images were immediately reinserted into the image database of the MR imaging unit for near real-time display.
Comparison of AIFs
Detailed comparison and validation of the automated AIF method compared with the standard manual method was undertaken in 20 of 100 cases in which a suspected perfusion abnormality was present. The 20 randomly selected patients included 10 male patients and 10 female patients, with a mean age of 52.05 years ± 27.14 and an age range of 390 years. Manual AIF selection was performed by an operator trained in a previously reported method for choosing an AIF (6,7,11). The operator iteratively placed a region of interest in the location of the proximal middle cerebral artery on anatomic images (mean area = 7.0 mm2 ± 5.4). The operator then viewed a cine loop display of the time course of that voxel on the dynamic T2*-weighted images to determine that it had signal loss characteristic of an artery on the basis of arrival time, bolus shape, and the magnitude of the signal change.
Anatomic Registration of Automated AIF
The anatomic location of the automated AIF was superimposed on the relative CBF images as small crosshairs. These images were presented independently to two board-certified neuroradiologists (H.A.R., V.M.H.), who were instructed to identify the position of the voxel relative to known vascular structures. Discrepant findings were resolved with consensus.
Quantitative Comparison
Signal intensity versus time curves were used to calculate
R2*(t) for both the automated AIF and the user-defined AIF, with
R2*(t) = -ln[S(t)/S0]/TE. The
R2*(t) curves from both AIFs were fitted to a gamma-variate function, that is, C(t) = atb e-ct, by using the Levenburg-Marquart algorithm (12). The effects of recirculation were removed by fitting only the first pass of the bolus of contrast agent. The fitted curves were compared on the basis of three criteria that reflected the partial-volume effects and the extent with which the input function was convolved from the point of injection to the point of measurement. The three criteria were as follows: peak
R2*, width of the
R2* curve, and ratio of the integrated
R2* curves. These are explained further next.
One effect of the convolution is to broaden the width of the AIF. In vivo, when venous injections are performed, the idealized bolus has undergone convolution in transit from the injection site to the arteries in the brain. Further convolution of this bolus occurs as it passes through the capillary bed. We compared the full width half maximum (FWHM) of the two AIFs to determine which AIF represented the least convolved and thus represented a more pristine AIF. Previously, this criterion was used to choose voxels as an AIF (8).
A greater concentration of gadolinium at the peak of the bolus was assumed to indicate less partial-volume averaging, because
R2* and contrast agent concentration are proportional. Therefore, we compared the peak
R2* and the integrated
R2* curves of the user-defined and the automated AIF. With the integrated
R2* curves, we derived a measure of the "relative vascular volume" of the two AIFs by means of defining the user-AIF as a reference and calculated the CBVratio as follows:
|
|
R2* per unit time of the automated AIF and user-AIF(t) is the
R2* per unit time of the user-defined AIF. The ratio exceeded 1 when the vascular volume determined with the automated AIF exceeded that determined with the user-defined AIF.
Statistical Analysis
Statistical analysis was performed for the FWHM, peak
R2*, and CBVratio of the automated and the user-defined AIFs. Significant differences were determined on the basis of a Wilcoxon signed rank test. Significant differences were defined with a P value less than .05. We also compared the integrate
R2* and half width of the raw unfit data.
Technical Validation in Patients
Blood flow maps of relative CBF, relative CBV, and MTT were prepared with the automated method for each case and were interpreted with the picture archiving and communication system together with the anatomic images. In each case, the quality of the images and the presence or absence of an abnormality were assessed by a neuroradiologist. In each case, conventional images were available for interpretation or comparison in the event that the automatically processed images were considered technically inadequate. In all cases, commercially available software was also available to evaluate time courses from selected regions of interest to obtain additional perfusion information. Images were classified as technically inadequate if the relative CBF, relative CBV, and MTT maps did not have marked distinction between gray and white matter, had inadequate definition of the brain contours, or had inadequate definition of a perfusion deficit when one was shown to be present by using any method.
| Results |
|---|
|
|
|---|
|
|
|
R2* fit confirmed that the automated AIF exhibited the characteristic gamma-variate shape. The peak
R2*, the FWHM, and the CBVratio for the two AIFs are included in the Table. The mean differences between the automated and user-defined AIF were -0.16 sec-1 ± 0.99 and 14.87 sec ± 20.66, respectively, for the FWHM and peak
R2* distributions. The peak
R2* was significantly higher (P < .05) with the automated AIF (
R2* = 46.35 sec-1 ± 14.80) than it was with the user-defined AIF (
R2* = 31.48 sec-1 ± 10.78). The CBVratio significantly exceeded 1.0 (CBVratio = 1.97 ± 2.53; P < .05). The FWHM was smaller for the automated AIF (FWHM = 2.47 sec ± 0.68) than it was for the user-defined AIF (FWHM = 2.63 sec ± 1.01); however, these differences were not statistically significant (P = .67).
|
| Discussion |
|---|
|
|
|---|
The automated AIF algorithm can be used to eliminate off-line postprocessing of dynamic susceptibility contrast MR images and the need for specially trained personnel. The automation of the determination of an AIF could potentially remove several steps in the processing of dynamic susceptibility contrast MR images. Streamlining the generation of relative CBF images may prove beneficial for time-critical applications, such as the delineation of ischemic regions in patients who have acute stroke. Elimination of the complexity in image processing could potentially result in broader dissemination of the dynamic susceptibility contrast MR imaging technology in the radiology community.
In this pilot study of the reliability of the automated AIF, perfusion images were successfully produced with no additional off-line postprocessing and were reviewed as part of the nominal patient work-up. The automated AIF curves had a more "arterial" shape than the user-defined AIF curves, in the sense that they were less broadened, as might occur as the bolus of contrast material passes through the vascular bed. Although AIFs calculated with the conventional method are not considered a reference standard, the conventional method represents a reasonable check on the accuracy of the automated method. The automated AIF is a robust means of generating parametric images of relative CBF, relative CBV, and MTT by using dynamic susceptibility contrast MR imaging. It routinely provides technically adequate images in clinical cases and, in particular, in those characterized by severe cerebrovascular disease.
One potential criticism of the automated AIF is that it selects one and only one voxel that is used to deconvolve the entire data set. Classic tracer kinetic modeling assumes each capillary bed is supplied by a unique and different input function. Therefore, some groups have suggested that individual AIFs should be applied to normal and ischemic territories. However, in clinical practice, the perfusion maps are used to identify areas that deviate from normal. Use of the single best internal "control" AIF, therefore, could be viewed as a practical advantage over the use of many individual AIFs, which would actually reduce sensitivity for low-flow lesions.
No failures due to patient motion were identified in the pilot series, but the effect of patient motion on the reliability of the program has not been evaluated. The effect of congestive heart failure, arteriovenous shunting, or other conditions that might affect the automated program were also not systematically evaluated in this study.
The AIF selected by the automated program appeared to localize on or immediately adjacent to a major artery. The program selects the voxel on the basis of a large signal intensity change. A voxel containing an artery with an inferosuperior orientation was most likely chosen by the program, since such an anatomic relationship minimizes partial-volume averaging artifacts. Further processing of this AIF by using iterative techniques may be necessary for quantitative flow in milliliters per 100 g per minute (13). The program rejects veins because of the requirement that voxels selected have the earliest signal intensity changes. Venous structures such as the sagittal sinus appear to be effectively removed from the set of voxels considered as an AIF, and such locations were never chosen for the AIF in the cases analyzed.
The AIF program also effectively discriminated signal intensity changes related to bolus passage from physiologic or imager noise unrelated to the passage of the bolus of contrast agent. In all 20 cases, a voxel with a characteristic gamma-variate shape was identified, which indicated that the chosen voxels were not anomalous noise spikes or other spurious signal changes. By using the proposed method of determining AIF, the potentially confounding effect of transient signal changes due to noise were minimized with integration of the signal intensity changes over several seconds. Selection of the AIF on the basis of minimum width or peak signal change would not likely have produced as reliable a result because of spurious brief fluctuations. An additional benefit of the signal integration is that it correlates well with the determination of relative CBV (ie, rCBV), which may be calculated as follows:
|
|
R2* and the greater integrated contrast than the user-AIF. These results indicate less partial-volume averaging in the automated AIF, which indicates a cleaner measure of the true AIF and, hypothetically, a more accurate determination of flow. The use of the arrival time parameter could potentially be modified and used to identify the time courses of venous efflux voxels. The identification of a venous time course has been shown to be useful in correction of AIF for quantitative determination of CBF (14). The use of this approach to automated quantitative CBF would present some advantages relative to previously reported quantitative CBF measures for which fewer sections were acquired and for which a dedicated imaging section was required to interrogate the distal carotid artery (8). However, in the current investigation, we do not have direct correlative studies with a reference standard. This is a limitation that prevents an evaluation of absolute CBF and the evaluation of the potential improvements afforded by the proposed venous efflux correction.
The automated AIF algorithm, as in the case of the manual method, chooses a single voxel or region of interest to represent the AIF for the whole brain. There is some question as to whether the same AIF is appropriate to deconvolve tissue curves measured in a more proximal section. On the basis of simulations, Ostergaard et al (5) concluded that singular valued deconvolution underestimates flow measured from tissue curves that possess an arrival time delay relative to the AIF. To validate this finding, Wirestam et al (7) compared AIFs recorded in the carotid artery with an AIF manually chosen in a proximal area of the imaging section. They were unable to reproduce the expectations of the simulations and found no difference in absolute cerebral perfusion when these two AIFs were compared. It should also be noted that on the basis of the original work of Ostergaard et al (5), Fourier deconvolution of an AIF will not exhibit the same sensitivity to arrival time. In short, the arrival time issue is one of the chosen deconvolution techniques and not the manner in which the AIF is chosen.
The automated method has limitations. The algorithm depends on six parameters chosen for our specific imaging protocol. The suitability of the algorithm for other protocols in which fewer precontrast points are obtained or smaller amounts of contrast medium are used has not been tested. However, with a suitable choice of parameters, it is anticipated that this algorithm can be suitably adjusted to other acquisitions. For example, in spin-echo acquisitions, the transient signal changes associated with bolus passage are not as great as they are in the gradient-echo acquisitions presented in this work. Therefore, a different criterion for determining the time of departure from baseline may be required with spin-echo acquisitions. It should be noted that we have successfully applied the automated AIF with the current set of parameters to spin-echo acquisitions in a limited number of cases; however, a study in a larger number of patients is needed to establish its utility for these acquisitions. Finally, a more direct optimization of the relative CBV may be possible through the use of the
R2* curve as opposed to the raw signal changes.
In conclusion, we have evaluated an automated rapid method of estimating AIF for the calculation of relative CBF at dynamic susceptibility contrast MR imaging. The automated AIF was found to yield relative CBF parametric images of diagnostic quality in 100 consecutive patients by using a standard MR imager and single-dose gadolinium-based contrast material. In a randomly chosen sample of 20 patients, the voxel that was chosen as the AIF was found to be associated with signal changes in the middle, posterior, or anterior cerebral artery in or other arterial structures. When compared with the user-defined AIF, the automated AIF was found to have significantly higher peak
R2* and similar bolus width as a user-defined AIF.
| APPENDIX |
|---|
|
|
|---|
The precontrast time frames that are used for baseline signal and noise calculations are determined with the whole-brain signal intensity as a function of time. By using these average signal intensities, we estimate the mean precontrast signal and the SD by using frames acquired from 10 to 20 seconds after initiation of imaging. The 1020-second time window is appropriate because our imaging protocol requires intravenous injection of contrast material 13 seconds after initiation of imaging to ensure contrast-free time frames for the determination of baseline signal intensity. The SD was used to determine an upper limit for signal fluctuation to determine precontrast frames (<3
) and a threshold for contrast material arrival (>10
), as shown in Figure A1.
|
(x,y,z), of each individual voxel were calculated and used to determine contrast material arrival time of each voxel. Since the changes of whole-brain signal versus time were dominated by signal in the large blood vessels, we calculated a separate arrival time threshold to determine voxel-by-voxel arrival time. The contrast material arrival time in each voxel is the first time point with a signal depletion of greater than 5 ·
(x,y,z). This arrival time was used to identify any voxels with delayed arrival times. If the arrival time of a voxel was more than 2 seconds after the whole-brain arrival time, it was assumed to reflect flow in large venous structures and, therefore, was removed from further consideration as the AIF voxel. The AIF is defined as the voxel, of the remaining voxels, that exhibits the largest change in signal intensity sustained over four time frames, or roughly 8 seconds, in the current imaging protocol (Fig A1). The AIF was then used to determine the arrival of the contrast material and the recirculation peak of the bolus. This is accomplished with identification of the first time point after peak signal where the derivative of the AIF decreased to less than zero. The arrival time and recirculation points were used to determine the time limits for gamma-variate fits and the deconvolution of tissue concentration curves.
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
R2* = change in R2*,
FWHM = full width half maximum,
MTT = mean transit time,
TE = echo time Author contributions: Guarantor of integrity of entire study, T.J.C.; study concepts and design, T.J.C., H.A.R., V.M.H.; literature research, T.J.C., H.A.R.; clinical studies, T.J.C., H.A.R.; experimental studies, T.J.C., V.M.H.; data acquisition and analysis/interpretation, T.J.C., H.A.R., V.M.H.; statistical analysis, T.J.C.; manuscript preparation, definition of intellectual content, editing, revision/review, and final version approval, T.J.C., H.A.R., V.M.H.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A.M. Saindane, M. Law, Y. Ge, G. Johnson, J.S. Babb, and R.I. Grossman Correlation of Diffusion Tensor and Dynamic Perfusion MR Imaging Metrics in Normal-Appearing Corpus Callosum: Support for Primary Hypoperfusion in Multiple Sclerosis AJNR Am. J. Neuroradiol., April 1, 2007; 28(4): 767 - 772. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Inglese, S.-J. Park, G. Johnson, J. S. Babb, L. Miles, H. Jaggi, J. Herbert, and R. I. Grossman Deep Gray Matter Perfusion in Multiple Sclerosis: Dynamic Susceptibility Contrast Perfusion Magnetic Resonance Imaging at 3 T Arch Neurol, February 1, 2007; 64(2): 196 - 202. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Law, R. Young, J. Babb, M. Rad, T. Sasaki, D. Zagzag, and G. Johnson Comparing Perfusion Metrics Obtained from a Single Compartment Versus Pharmacokinetic Modeling Methods Using Dynamic Susceptibility Contrast-Enhanced Perfusion MR Imaging with Glioma Grade. AJNR Am. J. Neuroradiol., October 1, 2006; 27(9): 1975 - 1982. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Moftakhar, A. S. Turk, D. B. Niemann, S. Hussain, S. Rajpal, T. Cook, M. Geraghty, B. Aagaard-Kienitz, P. A. Turski, and G. C. Newman Effects of Carotid or Vertebrobasilar Stent Placement on Cerebral Perfusion and Cognition AJNR Am. J. Neuroradiol., August 1, 2005; 26(7): 1772 - 1780. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Mlynash, I. Eyngorn, R. Bammer, M. Moseley, and D. C. Tong Automated Method for Generating the Arterial Input Function on Perfusion-Weighted MR Imaging: Validation in Patients with Stroke AJNR Am. J. Neuroradiol., June 1, 2005; 26(6): 1479 - 1486. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Ge, M. Law, G. Johnson, J. Herbert, J. S. Babb, L. J. Mannon, and R. I. Grossman Dynamic Susceptibility Contrast Perfusion MR Imaging of Multiple Sclerosis Lesions: Characterizing Hemodynamic Impairment and Inflammatory Activity AJNR Am. J. Neuroradiol., June 1, 2005; 26(6): 1539 - 1547. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Miller, H. H. Pien, D. Sahani, A. G. Sorensen, and J. H. Thrall Imaging Angiogenesis: Applications and Potential for Drug Development J Natl Cancer Inst, February 2, 2005; 97(3): 172 - 187. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Law, A. M. Saindane, Y. Ge, J. S. Babb, G. Johnson, L. J. Mannon, J. Herbert, and R. I. Grossman Microvascular Abnormality in Relapsing-Remitting Multiple Sclerosis: Perfusion MR Imaging Findings in Normal-appearing White Matter Radiology, June 1, 2004; 231(3): 645 - 652. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||