Head and Neck Imaging |
1 From the Department of Radiology, Cardiovascular Institute and Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing (S.Z.); Departments of Radiology (S.Z., J.C., C.H., C.Y.) and Electrical Engineering (Y.L.) and Division of Vascular Surgery (T.S.H.), University of Washington, Seattle; The Mountain-Whisper-Light Statistical Consulting, Seattle, Wash (N.L.P.); and the Department of Radiology, PLA General Hospital, Beijing, China (J.C.). From the 2001 RSNA scientific assembly. Received April 24, 2002; revision requested June 21; final revision received November 5; accepted November 19. Supported in part by NIH grants HL56874, HL60213, and HL61851. Address correspondence to S.Z., Department of Radiology, MRI, University Hospitals of Cleveland, 11100 Euclid Ave, Cleveland, OH 44106 (e-mail: zhang.shaoxiong@uhrad.com).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Eleven consecutive patients scheduled to undergo carotid endarterectomy were recruited for the study. A 3D fast gradient-recalled-echo sequence was applied to acquire both precontrast and postcontrast images of the carotid artery wall. The same sequence was used to image the ex vivo excised plaque as a reference for measurement of carotid wall volume and maximum wall area.
RESULTS: The mean difference in maximum wall area between the precontrast in vivo measurements and the ex vivo measurements (mean ± SD, 18.22 mm2 ± 15.61) was significantly larger than that between the postcontrast in vivo measurements and the ex vivo measurements (12.33 mm2 ± 14.49) (P = .02). The difference in wall volume between the precontrast in vivo measurements and the ex vivo measurements (41.81 mm3 ± 36.51) was larger than that between the postcontrast in vivo measurements and the ex vivo measurements (32.73 mm3 ± 35.00) (P = .004). Postcontrast images yielded better correlation with ex vivo images than did precontrast images, in both carotid luminal area (R = 0.88 for postcontrast images, R = 0.80 for precontrast images) and outer wall boundary area (R = 0.79 for postcontrast images, R = 0.71 for precontrast images) measurements.
CONCLUSION: Postcontrast 3D MR imaging may be useful in the measurement of carotid artery plaque burden.
© RSNA, 2003
Index terms: Carotid arteries, MR, 1721.121412, 1721.12143 Carotid arteries, stenosis or obstruction, 1721.721
| INTRODUCTION |
|---|
|
|
|---|
Magnetic resonance (MR) imaging is a noninvasive method that demonstrates excellent soft-tissue contrast and sensitivity to flow. It has advantages over other modalities in depicting the arterial wall and may play a unique role in monitoring atherosclerotic lesion progression and regression. Previous work has demonstrated the ability of MR imaging to characterize the composition of human atherosclerotic plaques, both ex vivo and in vivo (46). Two-dimensional black-blood high-spatial-resolution MR imaging techniques with multiple contrast weightings have been shown to be useful in the characterization of atherosclerotic tissue (4,6,7). Black-blood techniques have also been used in plaque area measurement (813). While bright-blood MR techniques have been widely used in identifying vessel lumen stenosis, they have only recently been applied to plaque characterization (6,14), especially the identification of fibrous caps (14).
Among bright-blood MR imaging techniques, three-dimensional (3D) time-of-flight and 3D gradient-recalled-echo (GRE) techniques may be of particular use because of their inherent high-spatial-resolution data acquisition and their usefulness in both lumen and wall depiction (15,16). Study findings have shown that gadolinium-based contrast agents can penetrate into the arterial wall, including the adventitia layer of the artery, which marks the outer wall boundary (17,18). Thus, the combination of contrast enhancement and 3D data acquisition may provide a very useful tool for evaluating plaque area and volume. The purpose of this study was to investigate whether 3D MR imaging performed after contrast material administration would yield more accurate measurement of carotid wall volume and MWA (both are measures of plaque burden) than 3D MR imaging performed before contrast material administration.
| MATERIALS AND METHODS |
|---|
|
|
|---|
MR Examination
All MR imaging examinations were conducted with a 1.5-T whole-body imager (Signa Horizon Echospeed, version 5.8; GE Medical Systems, Milwaukee, Wis). Four-element custom-built phased-array coils (two elements on the right side of the neck and two on the left) were used to improve the signal-to-noise performance of the imager (19). Studies of the performance of these phased-array coils, which have overall dimensions of 6.4 x 10.8 cm, demonstrated a 37% improvement in the signal-to-noise ratio compared with the signal-to-noise ratio of commercially available 3-inch-diameter (7.6 cm) surface coils (19).
A 3D fast GRE sequence was applied to acquire both precontrast and postcontrast images of the carotid artery centered at the bifurcation of the side of surgery (within 510 minutes after injection of contrast agent). Imaging parameters were 9.8/2.4 (repetition time msec/echo time msec) with a 13-cm field of view, 2-mm section thickness, matrix of 256 x 256, two signals acquired, and 20° flip angle.
A total dose of 20 mL of gadodiamide (Omniscan; Nycomed Amersham, Oslo, Norway) was injected at a rate of 2 mL/sec. To reduce the signal from subcutaneous fat and to avoid chemical shift artifacts, a fat-saturation technique was applied in 3D fast GRE sequences. A zero-filled Fourier transform was used to create voxels of 0.25 x 0.25 x 1 mm for 3D fast GRE imaging. All in vivo MR imaging examinations were performed less than 3 days before surgery. During surgery, the atherosclerotic plaque was removed en bloc and placed in saline at 4°C.
A 3D fast GRE imaging study was conducted ex vivo on the excised plaque as a reference for carotid wall volume measurement. The imaging parameters were identical to those used in vivo with the exception of a section thickness of 1 mm and a field of view of 8 cm. MR imaging of this surgical specimen was performed at body temperature (37°C) within 4 hours of the surgery while the specimen was submerged in saline.
Measurement of Wall Volume and MWA
Carotid luminal and outer wall boundaries were measured by using a custom-designed program, the Quantitative Vascular Analysis Tool, or QVAT (9,12), which was developed with Interactive Data Language. For in vivo images, the outer vessel wall boundary was defined as the vessel wallsoft-tissue interface; for ex vivo images, the outer vessel wall boundary was defined as the vessel wallbackground saline interface.
To measure the cross-sectional area of the carotid artery, the operator manually placed initial points near the lumen and outer wall boundaries. Boundaries were then searched by the Quantitative Vascular Analysis Tool program by using the Snake algorithm (9,12), and the area was automatically calculated. Wall area was defined as the area encircled by subtracting the luminal area from the outer wall boundary area. Two reviewers conducted a blinded review of in vivo and ex vivo images. One author measured in vivo images, and the other author measured ex vivo images. In vivo and ex vivo images were evaluated independently to avoid potential bias.
Definition of Wall Volume and MWA
Wall volume was calculated by multiplying the sum of cross-sectional areas and the section thickness. In each case, 10 locations centered at carotid bifurcation were used for wall volume measurement. MWA was defined as the largest area among the common carotid, bifurcation, and internal carotid artery wall.
Data Analysis
The location of the carotid bifurcation was determined with both in vivo and ex vivo images and was defined as the location of conversion from a single lumen to dual lumina. Once this location was defined, all other image locations were referenced to it in millimeters: Distal locations were given positive values, and proximal locations were given negative values. The in vivo and ex vivo images were then matched according to their locations.
Precontrast and postcontrast wall volume and MWA measurements were compared by using ex vivo images as reference. The mean absolute differences in wall volume and MWA of precontrast or postcontrast and ex vivo values were calculated and assessed by using the Student paired t test. A value of P < .05 was considered to indicate a statistically significant difference. Cross-sectional areas of the carotid artery lumen and outer wall boundary on precontrast and postcontrast in vivo images were compared with those on ex vivo images. The Pearson correlation coefficient was used as a descriptive measure of the strength of association between measurements of in vivo and ex vivo images.
| RESULTS |
|---|
|
|
|---|
As shown in Figure 1, the carotid artery outer wall and jugular vein boundaries are better delineated on postcontrast images than on precontrast images because of the substantial enhancement of the adventitia layer of the artery. On the basis of visual review, nine of the ll patients had improved delineation of the outer wall boundary on postcontrast images and two showed no difference. On postcontrast 3D fast GRE images, the enhancement of the outer wall boundary of the carotid artery appeared as full or partial circumference. The boundary between the carotid artery and the jugular vein was better delineated in comparison to precontrast 3D fast GRE images (Fig 1). In addition, seven patients had better luminal delineation on postcontrast images than on precontrast images. In all patients with clear postcontrast delineation of outer wall and luminal boundaries, measurements obtained from postcontrast images yielded closer wall volume and MWA results compared with ex vivo images than did measurement obtained from corresponding precontrast images compared with ex vivo images.
|
|
|
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Currently, carotid area is usually measured on precontrast MR images (813). Our study findings show that postcontrast 3D fast GRE imaging yields better measurements for both plaque volume and MWA than does precontrast 3D fast GRE imaging. This result suggests that 3D bright-blood technique may be useful in the quantitation of plaque volume.
There have been reports that the bright rim of the vessel wall can be identified with postcontrast MR imaging (17,21,22). Aoki et al (17,21) and Lin et al (22) reported that the outer wall can have atherosclerosis-like enhancement. As we found, on postcontrast MR images the outer wall boundary is more clearly delineated, which directly contributes to better measurement of the area of the carotid artery. In our study, visible enhancement of the outer wall boundary on postcontrast images was identified in nine (82%) of 11 patients, while two (18%) of 11 patients showed no visible changes. Such minimal enhancement may be due to images that were not acquired at the peak perfusion time of the contrast agent and to substantial plaque enhancement that made the enhancement of the outer wall boundary less obvious.
The main limitation of precontrast 3D bright-blood MR imaging is signal loss caused by (a) saturation of slow flow in normal and disease conditions and (b) dephasing due to complex or turbulent flow, which is particularly problematic in patients with severe stenosis (23). If a T1-shortening contrast agent is injected, however, postcontrast 3D MR imaging could eliminate the signal loss in lumen that is caused by saturation of slow flow and turbulent flow, especially at the carotid bifurcation. Thus, better luminal delineation would be achieved on postcontrast images. Without contrast enhancement, the difficulty in recognizing the boundary between the carotid artery and jugular vein could introduce error in measurement. In general, postcontrast 3D MR imaging provides better definition than precontrast imaging of carotid artery lumen and outer wall boundary, as well as the carotid-jugular vein boundaries, all of which affect the accuracy of carotid artery wall volume and MWA measurement.
Previous studies in which MWA was measured on two-dimensional black-blood images demonstrated a mean difference (value from in vivo image minus value from ex vivo image ± SD) of 13.1 mm2 ± 6.5 for T1-weighted images and of 14.1 mm2 ± 11.7 for intermediate-weighted images (11). MWA measurement on postcontrast 3D fast GRE images, in comparison to two-dimensional black-blood images, yielded comparable or slightly better results with a mean difference of 12.33 mm2 ± 14.49, while the difference of the measurement was increased on precontrast 3D fast GRE images with a mean difference of 18.22 mm2 ± 15.61 (Table 2). These results clearly demonstrate the importance of accurate demarcation of the outer wall boundary on 3D fast GRE images as a result of contrast enhancement. A recent publication demonstrated the usefulness of postcontrast high-spatial-resolution MR imaging in the characterization of atherosclerotic tissue (18). Thus, postcontrast 3D fast GRE imaging may be particularly useful in providing information on both the evaluation of plaque burden and the characterization of atherosclerotic tissue. This may aid in the identification of neovascularization and in the differentiation of necrotic core from fibrous tissue.
Vessel stenosis and MWA measurements from ex vivo imaging, which is used as the reference standard, have been reported previously (11,24). The technique, which was described by Pan and co-workers (24), showed that there was little change in luminal size between in vivo and ex vivo studies at the site of atherosclerosis. In our study, the ex vivo measurements were conducted within 4 hours of surgery while the specimen was at body temperature. We expect there was little shrinkage or other alteration to the plaque specimen in this time. It is expected that wall volume and MWA measurements from in vivo images are larger than those from ex vivo images in this study. During surgery, only plaque is removed from the vessel wall, so specimens are truly parts of the vessel wall. Carotid wall in vivo measurements, however, include media and adventitia, as well. To our knowledge, there has been no report on the ability of MR imaging to allow differentiation of the three layers of vessel wallintima, media, and adventitiain vivo. If this ability is achieved in the future, direct comparison will be possible between in vivo and ex vivo specimens of plaque burden. A study on comparison of presurgical, postsurgical, and ex vivo specimens may give us better understanding on this issue.
In conclusion, our results demonstrate that postcontrast 3D fast GRE MR imaging is superior to precontrast MR imaging in quantifying plaque burden, as assessed with MWA and plaque volume. Contrast enhancement provides better delineation of the lumen and outer artery wall boundaries and less flow artifact from slow flow, which results in closer agreement between in vivo measurements and the standard of reference, ex vivo measurements. Measurement of plaque burden with MR imaging will have important applications in studies in which the relationship between lesion size and clinical outcome is examined and in clinical trials in which plaque regression in response to therapeutic intervention is examined.
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
Author contributions: Guarantor of integrity of entire study, S.Z.; study concepts, S.Z., J.C., C.H.; study design, S.Z., J.C., Y.L.; literature research, S.Z.; clinical studies, T.S.H.; experimental studies, S.Z., J.C.; data acquisition, S.Z., J.C.; data analysis/interpretation, S.Z., Y.L., N.L.P.; statistical analysis, S.Z., N.L.P.; manuscript preparation, S.Z., C.H.; manuscript definition of intellectual content, S.Z., C.H., C,Y.; manuscript editing, S.Z.; manuscript revision/review, N.L.P., C.Y.; manuscript final version approval, T.S.H., C.Y.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. A. Wasserman, A. R. Sharrett, S. Lai, A. S. Gomes, M. Cushman, A. R. Folsom, D. E. Bild, R. A. Kronmal, S. Sinha, and D. A. Bluemke Risk Factor Associations With the Presence of a Lipid Core in Carotid Plaque of Asymptomatic Individuals Using High-Resolution MRI: The Multi-Ethnic Study of Atherosclerosis (MESA) Stroke, February 1, 2008; 39(2): 329 - 335. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Touze, J.-F. Toussaint, J. Coste, E. Schmitt, F. Bonneville, P. Vandermarcq, J.-Y. Gauvrit, F. Douvrin, J.-F. Meder, J.-L. Mas, et al. Reproducibility of High-Resolution MRI for the Identification and the Quantification of Carotid Atherosclerotic Plaque Components: Consequences for Prognosis Studies and Therapeutic Trials Stroke, June 1, 2007; 38(6): 1812 - 1819. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Takaya, C. Yuan, B. Chu, T. Saam, N. L. Polissar, G. P. Jarvik, C. Isaac, J. McDonough, C. Natiello, R. Small, et al. Presence of Intraplaque Hemorrhage Stimulates Progression of Carotid Atherosclerotic Plaques: A High-Resolution Magnetic Resonance Imaging Study Circulation, May 31, 2005; 111(21): 2768 - 2775. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Sirol, V. V. Itskovich, V. Mani, J. G. S. Aguinaldo, J. T. Fallon, B. Misselwitz, H.-J. Weinmann, V. Fuster, J.-F. Toussaint, and Z. A. Fayad Lipid-Rich Atherosclerotic Plaques Detected by Gadofluorine-Enhanced In Vivo Magnetic Resonance Imaging Circulation, June 15, 2004; 109(23): 2890 - 2896. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |