Breast Imaging |
1 From the Dept of Radiology, Stanford Univ Medical Center, Room S-068A, Rte 1, 300 Pasteur Dr, Stanford, CA 94305-5105 (D.M.I., R.L.B.); R2 Technology, Sunnyvale, Calif (K.F.O.); Dept of Radiology, Univ of California, San Francisco (E.A.S.); Eisenberg Keefer Breast Center, John Wayne Cancer Institute, St Johns Health Center, Santa Monica, Calif (R.J.B.); and Dept of Radiology, Geffen School of Medicine, Univ of California, Los Angeles (R.J.B.). Received February 13, 2003; revision requested May 2; revision received July 15; accepted August 22. Address correspondence to D.M.I.
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Four hundred ninety-three mammogram pairsan initial negative screening mammogram and a subsequently obtained screening mammogram showing cancerwere collected. The mean interval between examinations was 14.6 months. In 169 cases, in which 172 cancers were later depicted, findings on the initial mammogram were subtle enough that either none or only one or two of five blinded radiologists recommended screening recall. On the initial negative mammograms, of the 172 areas where cancer later developed, 137 (80%) had subtle nonspecific findings and were retrospectively judged as having a benign or normal appearance. The mammograms with these subtle findings were evaluated with a commercially available CAD program, and the numbers of CAD marks on these nonspecific findings were analyzed.
RESULTS: Of the 172 cancers, 129 (75%) were invasive and 43 (25%) were ductal carcinoma in situ. The CAD program marked 72 (42%) of the 172 findings that subsequently developed into cancer: 24 (29%) of 82 findings recalled by none, 25 (49%) of 51 findings recalled by one, and 23 (59%) of 39 findings recalled by two of the five radiologists. Among the 137 areas with nonspecific normal or benign findings, 41 (30%) areas where cancer subsequently developed were marked by the CAD program.
CONCLUSION: A subset of cancers have perceptible but nonspecific mammographic findings that may be marked by a CAD program, even when the findings do not warrant recall as judged at blinded and unblinded radiologist review. The authors believe failure to act on such nonspecific but CAD-marked findings prospectively does not constitute interpretation below a reasonable standard of care.
© RSNA, 2004
Index terms: Breast neoplasms, 00.31, 00.32 Cancer screening Computers, diagnostic aid Radiography, digital
| INTRODUCTION |
|---|
|
|
|---|
We recently reported on the nonspecific findings seen on the initial normal screening mammograms obtained in women in whom breast cancer was later detected at follow-up screening (4). In our report, we described perceptible but normal benign findings that would not be recalled for further evaluation, even in retrospect. In all cases, cancer was evident on follow-up screening mammograms obtained 914 months later. Thus, the purpose of our study was to evaluate, by using CAD, the nonspecific findings on normal screening mammograms obtained in women in whom breast cancer was later detected on follow-up screening mammograms.
| MATERIALS AND METHODS |
|---|
|
|
|---|
The previously obtained negative screening mammograms (also referred to as prior or initial mammograms) from 493 cases were available for review. The mean time between the initial and follow-up screening examinations was 14.6 months (range, 924 months). Sixty-two of the 493 cases were excluded because of prior breast surgery that resulted in scars or findings marked by metallic skin markers. Four other cases were excluded because the original film hard copies were needed at the facility site before the end of the study. Thus, a total of 427 cases comprised the study cohort.
One of three board-certified radiologists, who were not among the facility radiologists, reviewed the 427 cases to determine if the cancers were visible in retrospect on the prior mammograms. One radiologist reviewed 242, one radiologist reviewed 103, and one radiologist reviewed 82 mammograms. Each radiologist used the previously created film overlays to locate the cancer on the prior mammograms. If a perceptible finding was deemed visible on the prior negative mammogram, the radiologist marked the location of the retrospectively visible finding by using a second set of transparent film overlays, thus establishing a reference-standard location of the subsequently detected cancer on the prior mammograms.
In 286 (67%) of the 427 cases, there were findings that were judged to be visible on the prior negative mammograms in locations where cancer later developed. The 286 prior negative mammograms were divided into four sets of approximately 75 mammograms each. Forty-five additional mammograms were added to each case set: five mammograms on which no abnormalities could be seen, 20 mammograms with small subtle cancers, and 20 negative mammograms, as confirmed on the basis of at least one subsequent mammographic examination with negative results during a 2-year follow-up period.
Blinded and Unblinded Radiologist Case Review
To determine if the findings on the prior negative mammograms should have been further evaluated, four panels of radiologists, each consisting of five members, performed a blinded review of the respective four case sets. The radiologists reviewed the original film hard copies and were unaware of the study purpose or the case mix. These panel radiologists (half of whom had a primary work focus in mammographic interpretation) were Mammography Quality Standards Act certified, had practiced radiology for a mean time of 17 years (range, 335 years), and had read a mean of 300 screening mammograms per month (range, 401,000 mammograms per month).
Each panel member independently assessed approximately 120 cases and categorized them according to American College of Radiology Breast Imaging Reporting and Data System (BI-RADS) assessment codes (5). Lesions assigned BI-RADS codes 1 and 2 were considered normal or benign, and those assigned BI-RADS codes 0, 4, or 5 were considered abnormal. The use of BI-RADS code 3 was discouraged; however, the data showed that there were 16 cases with BI-RADS code 3 classifications, which for the purposes of this study were grouped with the BI-RADS codes 1 and 2 cases.
For clarity, we will refer to the cases that were assessed by the majority of five radiologists as BI-RADS codes 0, 4, or 5 as abnormal, meaning that the finding required immediate action. We will refer to mammograms showing findings that led the majority of radiologists to judge the case as BI-RADS code 1, 2, or 3 as normal, meaning that the finding was negative, benign, or not requiring immediate action. The panel radiologists were given the patients ages, they were shown only the prior negative mammograms (ie, mammograms obtained 924 months before the cancer was diagnosed at screening mammography), and no earlier obtained mammograms were presented. The panel radiologists had a mean sensitivity of 84% (mean of 16.8 of 20 cases) for cancer detection in the 20 cases of subtle cancer findings added to each case set and a mean specificity of 81% (mean of 16.2 of 20 cases) for diagnosing the 20 normal cases added to the case set.
Mammograms that were judged by three or more of the five blinded panel radiologists to show abnormal findings at the reference location were considered missed cancers, the rationale being that if the majority of radiologists at blinded review interpreted the mammogram as requiring immediate work-up, then the finding had been prospectively missed (1,2). One hundred twelve cases were judged to be abnormal by using these criteria (ie, by the majority of the panel) and thus were excluded from this study.
Three or more of the five blinded panel radiologists judged the remaining 174 cases to be negative, benign, or requiring no immediate work-up. We classified these cases as having nonspecific findings by using the rationale that if the majority of radiologists at blinded review interpreted a mammogram as normal, then the finding was very subtle, normal, or benign in appearance. Five of the 174 nonspecific finding cases were excluded: four cases in which the cancer location indicated on the mammogram was judged to be inconsistent with the pathologic diagnosis or the cancer location and one symptomatic case that should have been excluded in the original study. The remaining 169 cases, in which 172 cancers were depicted at subsequent follow-up screening, comprised our final study group or case set.
All mammograms were digitized at 50-µm spatial resolution capability by using a Lumisys LS85 digitizer (Lumisys, Sunnyvale, Calif) and downloaded to an Imation HQ969 laser printer (Imation Enterprises, St Paul, Minn) at 12 bits per pixel and 100-µm spatial resolution at the time of the initial study. These digitized images were then available on screen and printed to film hard copies for subsequent case review for this study. The digitized images were also archived for later CAD analysis.
The purpose of the unblinded review was to have the findings independently assessed by breast imaging specialists who knew the reference location of the subsequently detected cancer, to retrospectively reconfirm the assigned BI-RADS categories, to categorize each finding appearance, and to determine the reasons why the findings were nonspecific to the extent that they were categorized prospectively as BI-RADS category 1, 2, or 3 by the majority of five blinded radiologists. Two radiologists who specialize in breast imaging and have 14 (D.M.I.) and 19 (R.L.B.) years of experience in interpreting mammograms jointly reviewed the 169 cases in an unblinded review to categorize the findings and assess the possible reasons for the nondetection of and/or the nonaction on these findings. The digital-copy mammograms printed on film were used for this part of the study.
To ensure that the digital-copy mammograms were of sufficient quality for analysis, 20 original mammograms that included both masses (n = 12) and calcifications (n = 8) were recalled from the sites and compared side by side with the corresponding digital copies on dedicated mammography alternators. The two radiologists rated the image quality of the original and digital-copy mammograms by using a scale from 1 to 5with 1 meaning unable to read, 3 meaning acceptable quality, and 5 meaning good qualityand a narrative description of mass or calcification visibility. The average quality ratings for the original (4.5) and digital-copy (4.4) mammograms were similar. The narrative descriptions revealed no cases in which the quality of the digital copy compromised the detection or characterization of a mass or calcification, further supporting the acceptability of using digital-copy mammograms for our study.
To assess the mammographic characteristics of the visible findings, the 169 prior negative digital-copy mammograms, the subsequently obtained follow-up mammograms on which cancer was detected, and the corresponding reference-location clear overlays were reviewed on a two-tiered dedicated motorized mammography alternator (RADX MS-604A; S&S X-Ray Products, Houston, Tex), with bright lights and magnifying lenses available for use. The four-view prior negative mammogram and its corresponding reference-location overlay were displayed on the top row, and the mammogram obtained 924 months later showing the cancer and its corresponding overlay were displayed on the bottom row. At the time of case review, although the locations of the subsequent cancers were evident from the follow-up mammograms, no patient information, examination date, or pathologic data were available to the unblinded reviewers.
The perceptible finding identified by using the reference-location overlay on the prior negative mammogram (reviewed on digital copies of the films) was analyzed according to finding type, size, location, and depth in the breast. The visibility of the lesion on each appropriate view, as well as the breast density, was also recorded. Each finding that was visible in retrospect was categorized by using the BI-RADS lexicon for masses and calcifications and BI-RADS categories 05, with category 3 excluded (5). Because we endeavored to fully describe all findings, we used several nonBI-RADS terms to describe the normal and benign findings that are not included in the lexicon. Nonspecific finding terms included focal islands of normal-appearing tissue, benign-appearing calcifications, few benign calcifications, and densities. Otherwise, the findings were characterized by using the BI-RADS lexicon. The term that best described the major characteristic of the perceived finding was considered the finding type.
After mammogram analysis, the pathology reports were reviewed (by D.M.I.) for each cancer that was subsequently detected on the follow-up screening mammograms and each finding on the prior negative mammogram was compared with the subsequently detected cancer according to type and grade.
The patient population associated with and the cancer types and descriptors of the findings in this study have been fully described previously (4). The following is a summary of these data: The average age of the 169 patients in whom the 169 mammograms were obtained was 62.3 years. Sixteen (10%) patients were aged 4049 years; 50 (30%) patients, 5059 years; 58 (34%) patients, 6069 years; 40 (24%) patients, 7079 years; and five (3%) patients, 80 years or older. The breast tissue was judged to be fatty in 16 (10%) of the 169 cases, was of scattered fibroglandular density in 78 (46%) cases, was heterogeneously dense in 54 (32%) cases, and was extremely dense in 21 (12%) cases.
One hundred seventy-two cancers were detected at follow-up screening mammography performed in the 169 patients. At the time of diagnosis, 43 (25%) of the 172 cancers were ductal carcinoma in situ (DCIS) and the remaining 129 (75%) were invasive cancers. The median DCIS lesion size was 10 mm (range, 275 mm). The sizes of 24 (56%) of the 43 DCIS lesions were determined from the pathology report, and the sizes of the remaining DCIS lesions were based on the maximal diameter measurements of abnormal calcifications on the mammogram. The median lesion size of the invasive cancers was 10 mm (range, 155 mm). The sizes of 119 (92%) of the 129 invasive cancers were determined from the pathology report, and the sizes of the remaining invasive cancers were based on measurements of the abnormal findings on the mammograms. Of the invasive cancers, 112 were T1 tumors and 17 were T2 or higher-stage tumors. Of the 104 women with invasive cancer and known axillary node status, 22 (21%) had lymph nodes that were positive for metastatic disease.
All five of the blinded panel radiologists rated nearly half (n = 80 [47%]) of the 169 mammograms as normal at review. Fifty-one (30%) of the 169 mammograms were rated as normal by four of the five blinded panel radiologists, and 38 (22%) were rated as normal by three of these radiologists.
At unblinded review, with knowledge of the subsequent cancer location, the two breast imaging specialists would have recalled 35 (20%) of the 172 findings, rating them as BI-RADS 0 or 4 tumors. At unblinded review, the remaining 137 (80%) of the 172 findings were considered to be nonspecific, even in retrospect, by the two unblinded radiologists, and all of these findings were judged to be BI-RADS 1 and 2 tumors with the radiologists having knowledge of the subsequent cancer location.
CAD Evaluation
The 169 original mammograms (ie, the high-spatial-resolution digital images copied from the original mammograms) showing the 172 findings were processed through a CAD system, and the number of CAD marks on each image was recorded. The CAD system (V2.3; R2 Technology, Los Altos, Calif) that was used consists of a laser digitizer, a computer that uses proprietary signal-processing algorithms, and a customized motorized viewer with video display monitors. The CAD algorithm involves the use of a neural network that identifies features of microcalcifications (ie, clusters of bright spots) and marks them with triangles. The program identifies masses or architectural distortions as regions of high density with or without radiating lines and marks them with asterisks. This software version has been found to make an average of two marks per four-film normal case. In this study, there were an average of 3.9 (range, 012) CAD marks per case.
In general, cases with findings have more marks per case owing to the necessary addition of one or two marks per finding, depending on whether the finding is marked on one or on both standard mammographic views. A low-spatial-resolution (640 x 480 pixels) version of the marked digital mammogram is displayed on a small monitor on a motorized viewer directly below the digital mammograms when the unit is prompted. In normal clinical use, the marked images would be displayed only after the original film hard-copy images had been reviewed.
The 169 original mammograms were digitized and analyzed by the CAD program, and the resulting low-spatial-resolution images with marked areas were reviewed (by D.M.I., R.L.B., K.F.O.). The CAD-marked areas were directly compared with the findings on the digital-copy prior negative mammograms, and the corresponding reference-location overlays, described by the two unblinded breast imaging specialists. Each finding was judged to be either unmarked or marked by the CAD system. Marked findings included marks of the correct type on any part of the finding on either view, marks of either type on calcified masses, and calcification marks on any part of a calcification cluster. Unmarked findings were marks of the incorrect type on the finding or no marks on the findings. All other marks were recorded and counted as marks that were unrelated to the findings in this study.
Marks were recorded as "marks" or as "marks unrelated to the findings under study" and were compared with the pathology results, with each finding, and with the ratings assigned the findings at review by the five blinded panel radiologists and the two unblinded radiologists.
| RESULTS |
|---|
|
|
|---|
|
|
|
|
|
|
|
|
|
|
The data in Table 3 show the numbers of views on which a finding was seen and the numbers of views on which the finding was marked, with both groups of data stratified according to the BI-RADS categories assigned by the two unblinded radiologists. Six findings were judged to have the appearance of normal fibroglandular tissue, which was essentially undetectable on both views. None of these six findings were marked by the CAD system. Of the 64 findings that were seen on only one mammographic view, 18 (28%) were marked by the CAD system on that view. Of the 102 findings that were seen on two views, 23 were marked by the CAD system on both views (22%). The CAD system marked 31 (30%) of the 102 findings on only one of the two views.
|
| DISCUSSION |
|---|
|
|
|---|
Our finding that the CAD system marks normal or benign findings in locations where cancer later develops prompts several questions: When CAD is used on screening mammograms in a normal clinical setting, what is the clinical importance of CAD marks on nonspecific mammographic findings? Do all CAD marks on nonspecific findings warrant recall? Do all CAD marks on areas where cancer later develops warrant recall prospectively, even when the finding is nonspecific and cannot be distinguished from other nonspecific findings?
The CAD system marks a percentage of suspicious findings that might represent cancerusually spiculated or irregular masses or pleomorphic calcifications. The benefit of CAD manifests when the radiologists attention is drawn to a finding that is interpreted as abnormal, the patient is recalled at the screening, and the work-up results in a diagnosis of breast cancer that is treated adequately, and, thus, an improved prognosis for the woman is expected. With this scenario, one assumes that the radiologist interprets the marked finding correctly. However, mammographic interpretation is based on the mammographic features of the finding and the experience of the radiologist and not specifically on the CAD marks.
In clinical practice, nonspecific findings are often seen on mammograms. We have shown in prior work (4) that these nonspecific findings do not necessarily warrant recall, and it is our opinion that failure to act on these findings does not deviate from the appropriate standard of care. On the basis of standard medical practice, nonspecific findings do not require recall at screening (6,7) because these findings are ordinarily not reported. Thus, the CAD system may mark nonspecific findings, but these marks should be subordinated to standard interpretation. If nonspecific findings are correctly interpreted as benign and returned to screening in routine clinical practice, then it is reasonable that these findings will be returned to screening even when they are marked by a CAD system.
CAD algorithms are specifically designed to have a high stand-alone sensitivity for cancer detection, with a corresponding high number of false marks per case. Radiologists who purchase CAD systems are informed of the average number of marks per negative case and that the majority of marks do not indicate breast cancer. It is neither necessary nor reasonable for a radiologist to call back all findings marked by a CAD system. The CAD system used in this study made approximately two false-positive marks per mammographic examination. To calculate how many benign findings might be marked in this scenario, let us assume that there was a prevalence of six cancers per 1,000 screening mammograms and a (high) rate of possible missed cancer of one cancer per 1,000 screenings and that the CAD system marked the missed cancer on two views. If we also assume that there were five radiologist-detected cancers per 1,000 screenings and that the CAD system marked these cancers on two views, then there would be 10 marks on the radiologist-detected cancers. This would result in 1,988 marks on the benign findings, 10 marks on the five radiologist-detected cancers, and two marks on the missed cancer. With these assumptions, the CAD system would mark more than 1,000 benign findings for every cancer that would otherwise be missed by a single radiologist reader.
In clinical practice, it is expected and recognized that many benign and normal findings will be marked by the CAD system and that the threshold to mark findings emphasizes the systems increased sensitivity. The radiologist interprets CAD-marked findings on the basis of the findings mammographic features and the radiologists experience and knowledge of mammographic appearances of breast cancer. Thus, the most important element of image interpretation in the setting of CAD marks is how the radiologist uses his or her knowledge of mammography and breast cancer. Recalling all findings marked by a CAD system is neither prudent nor reasonable, especially given the known programming thresholds of this tool. Ultimately, it is the radiologists knowledge of breast cancer imaging and diagnostic acumen that influences the choice to recall a finding, not the marking of a finding by a CAD system.
Our study findings are relevant not only to prospective medical management outcomes but also to medicolegal issues. Medicolegal issues are not trivial: According to the results of a recent national study (8), delayed diagnoses of breast cancer remain the most common reason that physicians are sued, and radiologists remain the most commonly named defendants. In a Physicians Insurers Association of America study (9,10), follow-up mammography depicted a nonpalpable lesion in 48 patients, with an average delay in diagnosis of 20.1 months after the initial mammographic examination.
The presence of subthreshold or nonspecific features seen retrospectively on mammograms obtained in patients who developed breast cancer has been previously reported (4,1013). What effect, then, should CAD marks have on the potential legal liability of the radiologist who decides not to recall the patient for immediate evaluation? The law has repeatedly held physicians to the standard of care of a reasonable and prudent physician under similar circumstances (14). The CAD mark is intended to draw the radiologists attention to a given area and is used to mitigate the possibility of detection error in case the anatomic site of concern has not been sufficiently evaluated (15). Once an area is identified, either with or without a CAD mark, the decision to recall the patient is based on the radiologists training, familiarity with signs of breast cancer, and experience. This decision is subject to the legal test of reasonablenessnot of accuracy or certaintyso that a mistake is not tantamount to breaching the requisite standard of care (16).
Our study results indicate that CAD systems mark a finite percentage of nonspecific mammographic findings in areas where cancer subsequently develops. However, that percentage is neither sufficiently high nor designed for specificity to the extent that it can be an independent variable in deciding the reasonableness of recall. Similarly, if a finding that would be reasonable to recall is not recalled, then the presence (or absence) of a CAD mark is subordinate to the analysis of the specific mammographic features of that area. Although research of ways to reduce the variability of diagnoses by using CAD is underway (17), CAD technology has not been advocated with respect to recall rates, the determinants of which have been reported in other studies (18,19). Our study results indicate that when a nonspecific area is identified by using CAD in a detection setting, the clinical importanceand especially the legal importanceof the mark with respect to the decision to recall are moot.
There were several limitations in our study. The first limitation was our use of digital-copy mammograms, which never have the spatial resolution or lesion conspicuity that completely matches those of the original films. Standard parameters were used to print the digital-copy films. However, we performed a matched evaluation of a subset of original and digital-copy mammograms, which revealed no substantial difference between the two images, validating our results.
A second possible limitation was our use of digitized mammograms from analog films. The fact that most commercially available CAD systems are used on digitized film, however, justifies the technique used in this study despite the noise and image degradation produced by the digitization process. In the future, other studies might be performed by using CAD on directly acquired digital mammograms.
Another limitation was the use of the five-radiologist panel that performed experimental blinded readings, as compared with truly prospective blinded readings, to define the set of nonspecific findings. It is well known that retrospective readings may yield more positive findings than prospective readings (13). The unblinded two-radiologist readings used to characterize and assess the nonspecific findings were another potential but unavoidable limitation of our study, because complete case knowledge was necessary to truly know whether a finding was located where the cancer later developed. Also, since this was a retrospective study, whether the CAD marks would have prompted a change in prospective assessments was not evaluated, because CAD was not available at the time of the original interpretations. Although determining radiologists responses to CAD marks would have been interesting from a psychological aspect, it was beyond the scope of this study.
Our study results show that a subset of cancers have perceptible but nonspecific mammographic findings that do not warrant recall, as judged by a majority of blinded radiologists and by two unblinded radiologists, and that these findings may be marked by a CAD program. We believe failure to act on the nonspecific findings marked by a CAD program does not necessarily constitute mammographic interpretation below a reasonable standard of care.
| FOOTNOTES |
|---|
Abbreviations: BI-RADS = Breast Imaging Reporting and Data System, CAD = computer-aided detection, DCIS = ductal carcinoma in situ
Author contributions: Guarantor of integrity of entire study, D.M.I.; study concepts and design, D.M.I., R.L.B.; literature research, D.M.I., R.L.B., E.A.S., R.J.B.; clinical and experimental studies, D.M.I., R.L.B., K.F.O.; data acquisition, D.M.I., R.L.B., K.F.O.; data analysis/interpretation, D.M.I., R.L.B., E.A.S., K.F.O.; statistical analysis, D.M.I., K.F.O.; manuscript preparation, definition of intellectual content, editing, revision/review, and final version approval, all authors
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. Li, R. Engelmann, C. E. Metz, K. Doi, and H. MacMahon Lung Cancers Missed on Chest Radiographs: Results Obtained with a Commercial Computer-aided Detection Program Radiology, January 1, 2008; 246(1): 273 - 280. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Kim, W. K. Moon, N. Cho, J. H. Cha, S. M. Kim, and J.-G. Im Computer-aided Detection in Full-Field Digital Mammography: Sensitivity and Reproducibility in Serial Examinations Radiology, December 1, 2007; 246(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Georgian-Smith, R. H. Moore, E. Halpern, E. D. Yeh, E. A. Rafferty, H. A. D'Alessandro, M. Staffa, D. A. Hall, K. A. McCarthy, and D. B. Kopans Blinded Comparison of Computer-Aided Detection with Human Second Reading in Screening Mammography Am. J. Roentgenol., November 1, 2007; 189(5): 1135 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Yang, W. K. Moon, N. Cho, J. S. Park, J. H. Cha, S. M. Kim, S. J. Kim, and J.-G. Im Screening Mammography-detected Cancers: Sensitivity of a Computer-aided Detection System Applied to Full-Field Digital Mammograms Radiology, July 1, 2007; 244(1): 104 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Skaane, A. Kshirsagar, S. Stapleton, K. Young, and R. A. Castellino Effect of Computer-Aided Detection on Independent Double Reading of Paired Screen-Film and Full-Field Digital Screening Mammograms Am. J. Roentgenol., February 1, 2007; 188(2): 377 - 384. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. R. Pai, N. E. Gregory, A. E. Swinford, and M. Rebner Ductal Carcinoma in Situ: Computer-aided Detection in Screening Mammography Radiology, December 1, 2006; 241(3): 689 - 694. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Dean and C. C. Ilvento Improved cancer detection using computer-aided detection with diagnostic and screening mammography: prospective study of 104 cancers. Am. J. Roentgenol., July 1, 2006; 187(1): 20 - 28. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. Sumkin, D. Gur, R. L. Birdwell, and D. M. Ikeda Computer-aided Detection with Screening Mammography: Improving Performance or Simply Shifting the Operating Point? Radiology, June 1, 2006; 239(3): 916 - 918. [Full Text] [PDF] |
||||
![]() |
R. J. Brenner, M. J. Ulissey, and R. M. Wilt Computer-Aided Detection as Evidence in the Courtroom: Potential Implications of an Appellate Court's Ruling Am. J. Roentgenol., January 1, 2006; 186(1): 48 - 51. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. L. Khoo, P. Taylor, and R. M. Given-Wilson Computer-aided Detection in the United Kingdom National Breast Screening Programme: Prospective Study Radiology, November 1, 2005; 237(2): 444 - 449. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Doi Current status and future potential of computer-aided diagnosis in medical imaging Br. J. Radiol., January 1, 2005; 78(suppl_1): S3 - s19. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||