Technical Developments |
1 From the Department of Radiology, Pennsylvania State University Center for NMR Research, Penn State Milton S. Hershey Medical Center, 500 University Dr, Hershey, PA 17033 (J.L.S., T.J.M., Q.X.Y., M.B.S.); Department of Orthopedic Surgery, Thomas Jefferson University Hospital, Philadelphia, Pa (H.E.S.); and Departments of Radiology, Pediatrics, and Biomedical Engineering, University of Cincinnati, College of Medicine, Imaging Research Center, Cincinnati Childrens Hospital Medical Center, Cincinnati, Ohio (B.J.D.). Received November 6, 2003; revision requested January 28, 2004; revision received January 30; accepted February 17. T.J.M. and B.J.D. supported by the Arthritis Foundation and National Institutes of Health/National Institute of Arthritis and Musculoskeletal and Skin Diseases. Address correspondence to T.J.M. (e-mail: tmosher@psu.edu).
| ABSTRACT |
|---|
|
|
|---|
© RSNA, 2004
Index terms: Arthritis, 437.772 Hand, arthritis, 437.772 Hand, MR, 437.121411 Magnetic resonance (MR), arthrography Magnetic resonance (MR), high resolution
| INTRODUCTION |
|---|
|
|
|---|
Thus far, human T2 mapping techniques have been limited to evaluation of articular cartilage of the knee (48). Cartilage damage in large weight-bearing joints reflects a combination of systemic and local biomechanical factors. Extension of MR imaging parametric mapping techniques to evaluation of small nonweight-bearing joints, such as the interphalangeal joints of the hand, could provide a means for comparing and differentiating effects of systemic factors from local biomechanical loading on early cartilage damage.
Application of MR parametric mapping techniques to small joints is technically challenging. Because articular cartilage of the interphalangeal joints is thinner than 1 mm, specialized acquisition techniques are needed to generate high-spatial-resolution images, with pixel resolution less than 100 µm. Such measurements require high signal-to-noise ratio, large gradient amplitudes with rapid rise times, and increased receiver bandwidths to minimize interecho times for accurate determination of T2 relaxation values without compromising spatial resolution. Thus, the purpose of our study was to evaluate the feasibility of MR T2 mapping of the proximal interphalangeal joint of the index finger.
| Materials and Methods |
|---|
|
|
|---|
MR T2 Mapping
Quantitative T2 mapping of the proximal interphalangeal joint was performed with a 3.0-T MR spectrometer (MedSpec S300; Bruker Instruments, Ettlingen, Germany) by using a home-built microimaging gradient coil with a 9.5-cm internal aperture and a maximum gradient strength of 1000 mT/m, combined with a slotted tube transmit-receive coil with a 2.5-cm internal diameter. The frequency-encoding axis was aligned with the long axis of the index finger. Magnitude and quantitative T2 maps were calculated from sagittal data sets obtained through the proximal interphalangeal joint of the index finger by using an 11-echo sequence with a repetition time msec/echo time msec of 1500/9.7106.7, one signal acquired, four sections obtained with 1-mm section thickness, 384 (2) acquisition matrix zero-filled to 512 (2) matrix, and a 1.5-cm2 field of view.
The repetition time and range of echo times were chosen specifically to allow direct comparison with T2 mapping results reported previously for the knee (6). This provided images with an acquired in-plane pixel resolution of 39 µm. Total acquisition time to acquire the source images for quantitative maps was 9.6 minutes. The specific absorption rate was monitored throughout the study and was below the recommended limits of the Food and Drug Administration. None of the subjects experienced discomfort or symptoms of nerve stimulation that could be attributed to high magnetic field per unit time.
Data Analysis
Magnitude images and T2 maps were calculated from 10 spin-echo MR images by means of linear least-squares curve fitting on a pixel-by-pixel basis with custom software (Cincinnati Childrens Hospital Image Processing Software/Interactive Data Language; RSI, Boulder, Colo) (5). The first echo was excluded from the fit to reduce error that resulted from signal produced by stimulated echoes (8). Fitting of the signal intensity for the ith,jth pixel as a function of time, t, can be expressed as follows: SIi,j(t) = SI0i,j · exp(t/T2i,j), where SI0i,j is the pixel intensity at t = 0 and T2i,j is the T2 time constant of pixel i,j. A magnitude image is generated from the pixel SI0i,j data, and a T2 map is generated from the T2i,j data.
To generate regions of interest, segmentation of articular cartilage of the proximal phalynx was performed on each section of the T2 maps by using an interactive subroutine in the custom software. Segmentation of articular cartilage was performed by one investigator (H.E.S.) with 4 years of experience in analysis of cartilage T2 maps of the knee. For the entire region of interest, the software automatically generates multiple T2 profiles by defining perpendicular tangents to the cartilage-bone interface that terminate at the articular surface.
For comparison of cartilage T2 values between volunteers, each profile was normalized for cartilage thickness such that cartilage at the subchondral surface has a normalized distance of 0.0, and cartilage at the articular surface has a normalized distance of 1.0. The T2 profiles for each individual were pooled to determine mean T2 values and standard deviations as a function of normalized distance from bone.
| Results |
|---|
|
|
|---|
|
|
|
| Discussion |
|---|
|
|
|---|
New MR imaging parametric mapping techniques, such as cartilage T2 mapping (4), T1
mapping (10), and delayed gadolinium-enhanced MR imaging of cartilage (11), demonstrate sensitivity to biochemical and structural changes in the extracellular cartilage matrix and thus have the potential to serve as image markers of osteoarthritis. As a clinical research tool, it would be useful to apply image marker techniques to different joints within the same individual to differentiate systemic factors from biomechanical factors that may be localized to a specific joint.
The interphalangeal joints of the hand are a common site for osteoarthritis, with risk factors that differ from those of large weight-bearing joints, such as the hip or knee (1,12,13). Thus far, there has been limited application of MR imaging techniques in the evaluation of articular cartilage in small joints of the hand, and these have been limited to anatomic imaging techniques for the evaluation of tissue morphology. An initial feasibility study by Fry et al (14) demonstrated the potential of MR imaging to depict cartilage of the interphalangeal joints of the fingers. By using a highly optimized 2.4-T MR imaging system, Hodgson and co-workers (15) developed a method for obtaining anatomic images of cartilage of the distal interphalangeal joint of the finger with an in-plane spatial resolution of 75 µm by 150 µm. Peterfy et al (16) were able to obtain accurate reproducible quantification of articular cartilage volume in the metacarpophalangeal joints of the hand by using a clinical 1.5-T MR imager with a home-built receiver coil. When compared with the study of knee cartilage, the relative paucity of studies that involve MR imaging in the evaluation of the small joints of the hand reflects the substantial technical limitations when current techniques are applied with existing MR imaging systems.
Technical Considerations
Technical requirements are even more substantial for application of MR imaging parametric mapping techniques, such as T2 mapping, in the evaluation of articular cartilage in the hand. Thus far, T2 mapping techniques have been limited to evaluation of knee cartilage. The major limitation in the evaluation of phalangeal cartilage of the hand is the small volume of tissue ranging from 34 to 86 µL for phalangeal cartilage of the metacarpophalangeal joint (16). Obtaining reproducible in vivo T2 maps of the hand with a quality similar to that reported previously for knee cartilage requires data acquisition techniques currently used in ultra-high-field-strength MR microscopy studies of excised cartilage plugs.
In vivo cartilage T2 maps with an in-plane resolution of 250 µm have been used in the evaluation of femoral tibial cartilage (6,17). If the average thickness of femoral cartilage is assumed to be 1.5 mm, the T2 map comprises approximately 6 pixels across the cartilage. To obtain similar effective resolution in phalangeal cartilage, pixels of less than 100 µm would be required, which necessitates methods to optimize signal-to-noise ratio. In our study, signal-to-noise ratio was optimized through the use of a 3.0-T imager and a specialized radiofrequency transmit-receive coil with a high filling factor.
Because cartilage has short T2 values that range from 20 to 60 msec, the signal from cartilage will decay rapidly with long TE values. To characterize the T2 decay curve of cartilage accurately, it is necessary to minimize interecho spacing of the multiecho pulse sequence (18). The simultaneous need for high spatial resolution and short interecho spacing necessitates large receiver bandwidths, which further decreases signal-to-noise ratio and requires large magnetic field gradient amplitudes, rapid gradient switching, fast digitizers, and small transmit radiofrequency coils to minimize duration of the 180o refocusing pulses.
For the present study, we used a home-built gradient insert capable of generating 1000 mT/m, with a rise time of less than 50 µsec. With the 2.5-cm transmit-receive coil and a 2.5-kW radiofrequency amplifier, the duration of the refocusing pulse was 2000 µsec. By using a receiver bandwidth of 80 kHz, the interecho spacing could be reduced to approximately 10 msec while an in-plane pixel resolution of approximately 40 µm was obtained.
Cartilage T2 Maps
Prior studies have demonstrated that the T2 of articular cartilage is sensitive to differences in water content (19), collagen content (20), and collagen fiber orientation in the extracellular matrix (21). Early studies by Xia et al (22) involving excised cartilage plugs demonstrated spatial variation in cartilage T2, with longer values observed near the articular surface. This observation was confirmed later with an in vivo study by Dardzinski et al (4) on the evaluation of patellar cartilage.
Despite substantial differences in acquisition methods, magnetic field strength, and hardware, the pattern of spatial dependency of in vivo cartilage T2 of phalangeal cartilage observed in the present study is similar to that reported in prior ex vivo studies of cartilage plugs (21,2325) and in vivo measurements of femoral and/or tibial (6,17) and patellar cartilage (5,7,8). The range of spatial variation in cartilage T2 values of proximal phalangeal cartilage is less than that reported by Smith et al (6) for patellar cartilage (4567 msec) and is similar to that observed in femoral cartilage (4656 msec).
As has been described previously by Smith et al (6), high T2 values are observed near the bone-cartilage interface. By using high-spatial-resolution MR imaging, T2 maps, and histologic findings of bovine cartilage plugs, Nieminen et al (26) found that this zone of high T2 values corresponded to a zone of cartilage that contained an accumulation of chondrocytes. While the elevated T2 value in this zone may be a result of longer T2 values of intracellular water and the higher water content of the pericellular matrix surrounding the chondrocyte, this location is also subject to artifact from volume averaging, with bone and chemical shift artifact making interpretation ambiguous. Similarly, volume averaging with synovial fluid will result in error near the articular surface. This may be one reason why we did not observe low T2 values from the superficial lamina splendens, as has been observed with excised cartilage plugs (26).
Study Limitations
There are inherent limitations in the use of MR imaging techniques for quantitative determinations of T2 (27). Nonuniformity of the B1 field generates stimulated echoes in multiecho pulse sequences. Maier et al (8) showed that T2 is overestimated when a multiecho sequence is used, compared with measurements obtained by using a series of single-echo acquisitions with differing echo times. This error is reduced by excluding the first echo, as was done in the present study (8,27). The use of multisection acquisition produces off-resonance irradiation and incidental magnetization transfer (28) that can lead to an underestimation of T2 (8).
Fortuitously, positive error resulting from signal contribution of stimulated echoes is partially negated by negative error that results from incidental magnetization transfer (8). These errors arise from different factors, however; thus, it cannot be assumed that this effect will improve accuracy. Likewise, these systematic biases will vary with hardware and acquisition parameters and thus may contribute to differences in absolute T2 values acquired from knee cartilage and those acquired from the hand. Careful validation studies and methods for quality control are necessary (a) prior to implementation of these techniques in multicenter trials or (b) following technical modifications at individual sites.
At the bone-cartilage interface and articular surface, partial volume averaging with cortical bone and synovial fluid produces error in cartilage T2 measurements. With a pixel resolution of 39 µm and an average cartilage thickness of 500 µm, approximately 13 pixels are acquired across articular cartilage. With regard to the normalized T2 profiles, the relative resolution is a normalized distance of approximately 0.08. Additional factors can result in error in T2 measurement near the bone-cartilage interface. At the tidemark zone, differences in magnetic susceptibility between calcified bone and cartilage produce local magnetic field gradients that lead to rapid T2 decay, which is not measured accurately with the echo time values used in the acquisition protocol in the present study (18). Chemical shift artifact is increased at 3.0 T; however, this can be reduced with the use of large frequency-encoding gradients. In this case, use of a 80-KHz bandwidth produced a chemical shift of approximately 2 pixels. With regard to T2 profiles, chemical shift artifact from bone would be limited to the region of 0.0 to 0.16 normalized distance.
The results of the present study demonstrate the feasibility of acquiring cartilage MR T2 maps of small joints in the hand with an acquisition time of less than 10 minutes. With clinical implementation of 3.0-T MR imagers, it is likely that these techniques could be applied with relatively minor modifications to commercially available systems. Similar combined gradient and radiofrequency coil inserts have been used to obtain cartilage T2 maps of patellar samples with a 1.5-T clinical MR imager (29).
Cartilage T2 mapping of the hand may have clinical research applications in the study of hand arthritis and may provide novel information that improves understanding of the pathophysiology of generalized osteoarthritis. While biochemical changes in cartilage physiology can be modeled in animal studies, it is difficult to simulate the biomechanical loading conditions of the human joint. The ability to obtain cartilage T2 values in the same individual from both weight-bearing and nonweight-bearing joints provides a potential method to differentiate local biomechanical effects from systemic effects.
This information would be useful in understanding the effect of conditions such as exercise, body habitus, and prior trauma on the composition and organization of the extracellular cartilage matrix.
| FOOTNOTES |
|---|
Author contributions: Guarantors of integrity of entire study, all authors; study concepts and design, J.L.S., H.E.S., Q.X.Y., M.B.S.; literature research, T.J.M.; experimental studies, all authors; data acquisition, J.L.S., B.J.D., H.E.S., Q.X.Y., M.B.S.; data analysis/interpretation, J.L.S., T.J.M., H.E.S.; statistical analysis, J.L.S., T.J.M., H.E.S.; manuscript preparation, T.J.M.; manuscript definition of intellectual content, all authors; manuscript editing, J.L.S., T.J.M., B.J.D., M.B.S.; manuscript revision/review and final version approval, all authors