Breast Imaging |
1 From the Departments of Radiology and Biomedical Engineering, Duke University Medical Center, Erwin Rd, Durham, NC 27710. From the 2003 RSNA scientific assembly. Received July 28, 2003; revision requested October 6; revision received January 24, 2004; accepted March 2. Address correspondence to J.A.B. (e-mail: jay.baker@duke.edu).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Fifty breast cancers detected at screening were analyzed by using a commercially available CAD system. Mean patient age was 62.2 years. Each set of mammograms (craniocaudal and mediolateral oblique views) was digitized and analyzed by the CAD system 10 times. One radiologist compared CAD output with the location of the malignancy at mammography and determined whether each lesion was marked accurately in one mammographic view, both views, or neither. Sensitivity and reproducibility of the CAD system were determined for both case- and image-based analysis.
RESULTS: Overall sensitivity of the CAD system when at least one of the two mammographic views was marked correctly (case-base sensitivity) was 82.4%. Sensitivity when each mammographic view was considered separately (image-based sensitivity) was 61.1%. For case-based analysis, variability in true-positive CAD cues was demonstrated for 14 of 50 (28%) cases. For image-based analysis, inconsistency in CAD output was observed in 33 of 100 (33%) mammographic views that contained malignancies detected at screening. However, the CAD system consistently detected 4043 of the 50 breast cancers in each of the 10 CAD runs. Variability for false-positive marks was significantly greater than that for true-positive marks.
CONCLUSION: Inconsistency was demonstrated for CAD analysis of breast cancers detected at screening. However, the CAD system was reasonably consistent in the overall number of cancers identified from run to run. Greater variability of the CAD system was also demonstrated for false-positive marks, as compared with true-positive marks.
© RSNA, 2004
Index terms: Breast neoplasms, localization, 08.112, 08.115 Breast radiography, quality assurance, 08.112, 08.115 Computers, diagnostic aid
| INTRODUCTION |
|---|
|
|
|---|
Because CAD systems use computer algorithms, they are presumed to offer extremely high or virtually perfect reproducibility in their analysis of mammograms (9). Indeed, this reputed reproducibility has been publicized as one of the many benefits of double-reading with a CAD system (9,10). However, a prior study performed in 2000 (11) demonstrated surprising variability in the output of one CAD system. The authors concluded at that time that such systems were insufficiently reproducible for routine clinical use. A recent study (12) also demonstrated inconsistency in a more contemporary version of the same CAD system.
In contrast to the two published studies on CAD system reproducibility (11,12), two manufacturers unpublished studies reported on a marketing Web site (www.r2tech.com/prf/prf001.html#3) and a Food and Drug Administration device labeling submission (iCAD device labeling, 2003) report excellent or near-perfect reproducibility. An important limiting characteristic in all of these prior reports is the patient population studied. In several of the studies in which the patient population can be determined, the study population likely includes symptomatic lesions rather than lesions in a typical screening population.
Further, in three of the four studies, each set of mammographic images was digitized and analyzed only three times, which may result in overestimation of the consistency of the CAD system. Thus, the purpose of our study was to evaluate the variability of true-positive and false-positive cues by using a commercially available CAD system for analysis of 50 malignancies in a screening population.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Between November 1, 2001, and January 31, 2003, 67 biopsy-proved breast malignancies were detected in 12,789 screening mammographic examinations performed at our institution. As in prior similar studies, to simplify statistical analysis, multifocal and multicentric cases were excluded (11). The first 25 consecutive malignant masses and the first 25 consecutive malignant calcification clusters were used to constitute a study population of 50 breast cancers detected at screening. All mammograms were obtained by using standard screen-film technique with one of eight mammography systems qualified according to the Mammography Quality Standards Act (four Mammomat 3000 Nova systems, Siemens, Erlangen, Germany; two Mammomat 3000 systems, Siemens; one Senographe DMR+ system, GE Medical Systems, Milwaukee, Wis; and one MIII system, Lorad, Danbury, Conn).
Patients ranged in age from 40 to 90 years (mean, 62.2 years). Of the 50 malignancies, 17 (34%) represented ductal carcinoma in situ, 13 (26%) represented invasive ductal carcinoma, and 18 (36%) represented invasive ductal carcinoma and ductal carcinoma in situ. Two of the 50 cases (4%) represented invasive lobular carcinoma. Each malignancy was visible in both the craniocaudal and mediolateral views.
Study Design
Each craniocaudal and mediolateral oblique image in the 50 screening cases was digitized and analyzed 10 times by using the hardware and software provided with a commercially available CAD system (ImageChecker M1000, version 3.2; R2 Technology, Sunnyvale, Calif). The CAD system is designed to assist radiologists in the detection of breast cancer by identifying groups of bright specks that are suggestive of calcification clusters and by identifying densities with or without radiating lines that are suggestive of breast masses or foci of architectural distortion (8).
The output of the CAD algorithm was displayed on an 18-inch flat-panel display as a low-resolution mammographic image with small triangles placed to mark locations of possible calcification clusters and small asterisks placed to mark locations of possible breast masses. The CAD system used for this study had been in service for 4 months, and all quality-control tests recommended by the manufacturer were performed daily and weekly, as appropriate.
The output of the CAD system was reviewed by one dedicated breast radiologist (J.A.B.), who was qualified according to the Mammography Quality Standards Act and had 6 years of mammography experience. This radiologist traced the outline of the low-resolution image of each breast on an overlay, and the location of each CAD mark was recorded on the overlay. CAD marks for possible massesdisplayed by the CAD system as asteriskswere differentiated on the overlay from triangle marks, which indicated potential clusters of calcifications.
The radiologist used all available mammographic viewsincluding additional or special viewsto determine the outline of the actual breast cancer (ie, margins of a mass or extent of calcifications) on each screening image. The radiologist then determined whether each CAD mark indicated the location of the malignancy. While there is no absolute rule to determine whether a CAD mark is sufficiently close to a lesion to represent a true-positive mark, previously published guidelines were used to determine true-positive and false-positive marks (13).
A CAD mark was labeled as true-positive if the mark was within the boundary that outlined the mass margin or within the extent of the calcifications. All CAD marks that did not mark the known malignancy were defined as false-positive marks for the purposes of this study.
The radiologist recorded whether each malignancy was marked correctly in the craniocaudal view, medolateral oblique view, both views, or neither view for each of the 10 CAD runs for all 50 cases. Therefore, a total of 500 CAD runs constituting 1000 mammographic images with a visible breast cancer in 50 screening cases were evaluated for this study. The reviewing radiologist recorded how many times and in which views each true-positive and each false-positive location was marked for the 10 CAD runs in each case. By using this system, the number and location of each true-positive and each false-positive mark were determined for the craniocaudal and medolateral oblique views for each of the 10 runs for the 50 cases.
Statistical Analysis
All statistical analysis was performed by using statistical software (SAS, version 8.2; SAS, Cary, NC). For the purpose of this study, each digitization and CAD analysis of the craniocaudal and medolateral oblique views of a single case were termed as one CAD run. Therefore, for this study, 10 CAD runs were performed in each of 50 cases, resulting in a total of 500 CAD analyses performed.
The sensitivity and reproducibility of the CAD system were determined for all 50 cases and were also calculated separately for malignant masses and malignant calcification clusters. In addition, sensitivity and reproducibility were calculated for these three populations by using case-based and image-based analysis. For case-based analysis, a successful mark of the cancer in either the craniocaudal or mediolateral oblique view was considered a true-positive identification of the cancer for that case. For image-based analysis, the craniocaudal and mediolateral oblique views were considered separately, resulting in two CAD analyses considered separately for each malignant case. CAD analysis could be true-positive in one view, both views, or neither view for image-based analysis.
The mean number of false-positive marks per image was calculated for all 50 cases. False-positive marks were defined as marks that the radiologist determined did not require further imaging or biopsy evaluation on the basis of the imaging characteristics. Reproducibility of false-positive marks was determined by using image-based analysis. The craniocaudal and mediolateral oblique views were considered separately (ie, image-based analysis) because false-positive marks cannot always be correlated between the two views. Reproducibility of false-positive marks was compared with the reproducibility of the true-positive marks by using image-based analysis.
Comparison of variance measures of the false-positive and true-positive marks in the craniocaudal and mediolateral oblique views was performed by using the Wilcoxon signed rank test. Comparison of sensitivities of the mediolateral oblique and craniocaudal views was also performed by using the Wilcoxon signed rank test. Comparison of sensitivity of the CAD system for detection of malignant microcalcification clusters and malignant masses was performed by using the Wilcoxon rank sum test. Only locations actually marked as false-positive in at least one view were considered for this comparison. Significance level was defined as P < .05.
| RESULTS |
|---|
|
|
|---|
Sensitivity for the craniocaudal view was compared with sensitivity for the mediolateral oblique view by using the Wilcoxon signed rank test. The sensitivity of all lesions in the 50 craniocaudal views analyzed was 62.0% ± 6.2 (standard error), which was not significantly different than the 60.2% ± 6.1 sensitivity of the same 50 lesions analyzed with the mediolateral oblique view (P > .10).
Sensitivity was also evaluated separately for masses and clusters of calcifications. No cases of masses with associated calcifications were included in this study. The case-based sensitivity for malignant masses detected at screening was 78.4% ± 7.0 (392 of 500 runs), and the case-based sensitivity for screening-detected malignant calcifications was 86.4% ± 5.9 (432 of 500 runs) over all 10 CAD runs.
With the Wilcoxon rank sum test, the sensitivities for detection of masses and calcifications were not significantly different (P > .10). The image-based sensitivity was 52.2% ± 6.0 (261 of 500 images) for screening-detected malignant masses and 70.0% ± 6.6 (350 of 500 images) for malignant calcification clusters. With the Wilcoxon rank sum test, the image-based sensitivity for malignant calcifications was significantly better than that for masses (P = .04).
Variability of CAD Output
While the case-based sensitivity was 82.4% ± 4.6 (412 of 500) for the combined 10 CAD runs in all 50 cases, the range in sensitivity for the 10 separate CAD analyses of the 50 cases was 80%86%. That is, in 50 malignancies detected at screening, the CAD system detected 4043 cancers in at least one routine mammographic view for each of the 10 runs (Table 1). Because there were differences in which cancers were identified in each of the 10 CAD runs, a total of 46 of the 50 cancers (92%) were identified on at least one of the 10 runs.
|
|
|
|
|
|
|
|
|
|
|
A total of 133 false-positive locations were identified from the 10 CAD runs of the 100 mammographic images analyzed. Some of the false-positive locations were identified in only a single CAD run, while others were identified several times. A total of 660 false-positive marks were recorded by the CAD system for the 10 CAD runs of the 100 mammographic images for an average 0.66 (660 of 1000) false-positive marks per image. Only 24 false-positive locations in the 100 mammographic images were identified consistently in all 10 CAD runs.
Comparison of variability of the true-positive marks and false-positive marks by using the Wilcoxon signed rank test indicates that false-positive marks were significantly more variable for both craniocaudal views (P < .001) and mediolateral oblique views (P = .022).
| DISCUSSION |
|---|
|
|
|---|
In the study of Zheng et al (12), in which a more recent version of the same CAD system was used, the authors also reported inconsistency in CAD sensitivity when 100 cases were evaluated three times. In that study, the authors reported that variability had decreased substantially with 78.8% of malignant masses and 93.5% of malignant calcification clusters identified consistently by using case-based analysis in three CAD runs. In the present study, only 52% of malignant masses and 76% (19 of 25) of malignant calcification clusters were identified consistently in all 10 CAD runs. Variability appears substantially higher in our analysis in part because CAD evaluation was performed 10 times in the present study, compared with only three times in the two reports published previously. These additional CAD runs provide substantially greater opportunity to detect CAD variability and, therefore, provide a more exhaustive evaluation of the potential effect of this inconsistency in actual clinical practice.
In addition to the two studies published previously with regard to CAD reproducibility, the manufacturers of the two most widely available commercial CAD systems have reported reproducibility studies that have not been published in the medical literature, including one study listed on a manufacturers Web site (www.r2tech.com/prf/prf001.html#3) and the other reported in the manufacturers submission for Food and Drug Administration device labeling for their CAD system (iCAD device labeling, 2003). Each of these unpublished studies report excellent or virtually perfect reproducibility. However, the design of these unpublished studies, as well as that of the reports published previously, may not precisely reflect the clinical effect of inconsistency in CAD output. Three of four prior studies limit analyses to three CAD runs per examination (11,12). Investigators in at least one prior published study (11) and one unpublished study (www.r2tech.com/prf/prf001.html#3) used older versions of commercial CAD systems. However, we believe the most important limiting characteristic of prior studies is the patient population used.
Because CAD is most commonly used as a "double read" of screening mammograms, a population of screening-detected malignancies should be emphasized in studies on the reproducibility of CAD systems. Although the precise patient population cannot be determined for any of the four prior CAD reproducibility studies, prior studies appear to have included symptomatic patients in the study population. In one manufacturers study (www.r2tech.com/prf/prf001.html#3), only "well-characterized" cancers were evaluated. Given the previously published studies and the data presented here, the near-perfect reproducibility reported in this manufacturers study suggests that the cancers used may have been readily apparent.
The original study on CAD reproducibility (11), as well as the most recent report (12), do not state clearly whether palpableand potentially more apparentcancers were excluded. Just as sensitivity of a CAD system may be overestimated when conspicuous symptomatic cancers are analyzed, consistency of CAD may also be overestimated when the study population includes conspicuous symptomatic cancers.
In the present study, a current, well-maintained CAD system was used to analyze consecutive breast cancer cases detected at screening mammography. Although it is not known whether a CAD system can mark lesions in different ways depending on maintenance, it is possible that debris in the system could affect the digitization process adversely and therefore affect cancer detection. Each mammogram was analyzed by the CAD system 10 times. These factors suggest that the CAD variability presented here most closely reflects the inconsistency that can be encountered in daily clinical practice.
The potential clinical effect of CAD variability is considerable; in one-third of screening mammographic images with a visible breast cancer, the CAD system was inconsistent in its ability to detect and mark the malignant lesion. Indeed, in more than one case in four (28%) in which the cancer was only sporadically marked in at least one view, inconsistency in the output of the CAD system had the potential to directly affect whether a cancer was detected. Further, the cases that were marked with the least consistencysubtle massesare the same cases for which radiologists would be most likely to benefit from consistent CAD assistance.
While inconsistency of the CAD system may directly affect whether an individual breast cancer is detected, less variability was noted for CAD analysis of the entire group of malignant cases. For the entire group, the sensitivity of the CAD system was more reproducible. That is, while individual cases were often marked inconsistently in the 10 CAD runs, the CAD system did consistently detect 4043 of the 50 cancers present. This level of consistency was achieved because different cancers were detected or missed in different CAD runs. While an individual cancer may have been missed in a particular CAD run, a different cancermissed previouslywas detected in that run, maintaining the sensitivity over the study population in a moderately narrow range.
The results of this study indicate that while there is considerable inconsistency in the ability of the CAD system to detect any single cancer, this variability averages out, and the CAD system is reasonably consistent in its sensitivity when an entire population of cases is considered. The source of variability in CAD systems has been explored previously and is likely due to inconsistency in the initial digitization of the mammographic film and concomitant electronic noise caused by the digitization process (11). The present generation of CAD systems requires digitization of the mammographic film as the first step in analysis (14). This initial digitization step is subject to inconsistency, since the film advances into the digitizer in a slightly different lateral position and angle (11). For lesions at the threshold of detectability, this slight difference appears to be sufficient to alter the ability of the CAD system to detect the lesion. Since false-positive regions are likely to be closer to the threshold of detectability, slight differences in film positioning during digitization likely account for the greater variability measured for false-positive cues, as compared with true-positive cues (12).
In the present study, it is interesting to note that the overall sensitivity of the CAD system could have been improved by 10%, from an average of 82.4% to a maximum of 92.0%, by combining the CAD outputs for all 10 CAD runs. This approach to improving CAD sensitivity is clearly not feasible at the present speed of digitization. Further, combining the outputs of all 10 CAD runs would have doubled the false-positive marks from an average of 0.66 marks per film to a maximum of 1.33 marks per film.
There are three noteworthy limitations to the study presented here. First, in this study, we evaluated only one CAD model from only one of several commercial manufacturers. It is not known whether CAD systems developed by other manufacturers with different digitizers and detection algorithms would perform similarly. Second, this study was performed by using only a single example of the CAD model. Although the system was new and well maintained, it is possible that other units of the same model could perform dissimilarly. Finally, this study demonstrates how the CAD system performs independent of the interpreting radiologist. While the CAD system alone may have substantial interrun variability, the skill of the interpreting radiologist may mitigate some of this variability and result in overall greater reproducibility than the CAD system alone.
In summary, the results of this study indicate that state-of-the-art commercially available CAD systems for mammography can suffer from greater inconsistency than that reported previously when used to evaluate a screening population. The performance of the system is more reproducible when a population of cases is considered instead of a single mammographic examination. However, the importance of CAD reproducibility should be considered in conjunction with (a) prior studies that clearly demonstrate the ability of CAD systems to detect more than three-quarters of breast cancers overlooked by radiologists alone (4) and (b) studies that have demonstrated the ability of these systems to substantially improve the sensitivity of interpreting radiologists in actual clinical practice (6). These studies, combined with the data presented here, indicate that CAD systems can indeed help radiologists improve patient care, although CAD systemslike their human counterpartsmay offer different opinions each time a mammogram is reviewed.
| FOOTNOTES |
|---|
Authors stated no financial relationship to disclose.
Author contributions: Guarantor of integrity of entire study, J.A.B.; study concepts, J.A.B., J.Y.L., C.E.F.; study design, all authors; literature research, J.A.B., J.Y.L.; clinical and experimental studies, J.A.B.; data acquisition, J.A.B., J.Y.L.; data analysis/interpretation, all authors; statistical analysis, all authors; manuscript preparation and definition of intellectual content, all authors; manuscript editing, J.A.B., J.Y.L., C.E.F.; manuscript revision/review and final version approval, all authors
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. J. Kim, W. K. Moon, N. Cho, J. H. Cha, S. M. Kim, and J.-G. Im Computer-aided Detection in Full-Field Digital Mammography: Sensitivity and Reproducibility in Serial Examinations Radiology, December 1, 2007; 246(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Yang, W. K. Moon, N. Cho, J. S. Park, J. H. Cha, S. M. Kim, S. J. Kim, and J.-G. Im Screening Mammography-detected Cancers: Sensitivity of a Computer-aided Detection System Applied to Full-Field Digital Mammograms Radiology, July 1, 2007; 244(1): 104 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Kim, W. K. Moon, N. Cho, J. H. Cha, S. M. Kim, and J.-G. Im Computer-aided Detection in Digital Mammography: Comparison of Craniocaudal, Mediolateral Oblique, and Mediolateral Views Radiology, December 1, 2006; 241(3): 695 - 701. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Brenner, M. J. Ulissey, and R. M. Wilt Computer-Aided Detection as Evidence in the Courtroom: Potential Implications of an Appellate Court's Ruling Am. J. Roentgenol., January 1, 2006; 186(1): 48 - 51. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||