|
|
||||||||
Technical Developments |
1 From the Departments of Diagnostic and Interventional Radiology (T.S., P.H., H.K., K.U.W., J.F.D., J.B.) and Cardiology (A.S.), University Hospital, Hufelandstr 55, 45122 Essen, Germany. Received September 11, 2003; revision requested November 24; final revision received March 12, 2004; accepted April 19. Address correspondence to J.B. (e-mail: joerg.barkhausen@uni-essen.de).
| ABSTRACT |
|---|
|
|
|---|
63.1%) even with the most advanced CT scanners. Accurate and reproducible quantification of coronary calcium seems to require analysis of multiple reconstructions. © RSNA, 2004
Index terms: Computed tomography (CT), multidetector row Coronary vessels, calcification, 54.812 Coronary vessels, CT, 54.12115 Coronary vessels, diseases, 54.812
| INTRODUCTION |
|---|
|
|
|---|
Results of several studies have demonstrated a close correlation between the progression of coronary artery disease and increasing calcium scores (8,9). Moreover, lipid-lowering treatment with statins has been shown to reduce coronary calcium progression (10,11). In principle, this provides the opportunity to monitor patients treated with statins and to adjust the dose by quantifying the coronary calcium. This type of follow-up requires accurate and reproducible assessment of coronary calcium.
Electron-beam CT had initially been used for the quantification of coronary artery calcification. Recently, multidetector row CT with fast gantry rotation has been introduced for cardiac CT. For calcium quantification, a good correlation was found between electron-beam and multidetector row CT (12). However, different reconstruction and scoring algorithms, as well as different image reconstruction intervals, affect the calculated scores and result in poor reproducibility of measurements (13,14).
Reproducibility may be improved with the recently introduced 16detector row CT systems with decreased rotation time of 420 msec. Thinner sections decrease partial volume effects, and faster gantry rotations reduce cardiac motion artifacts (15). Thus, the aim of our study was to evaluate the effect of the reconstruction interval on coronary artery calcium scores obtained with use of a 16detector row CT scanner.
| Materials and Methods |
|---|
|
|
|---|
Imaging and Evaluation
CT examinations were performed by using a commercially available 16detector row scanner (Somatom Sensation 16; Siemens, Forchheim, Germany) with a gantry rotation time of 420 msec and a routinely used scanning protocol for coronary artery calcium quantification (section thickness, 3 mm; collimation, 1.5 mm; table feed, 5.7 mm per rotation; reconstruction increment, 1.5 mm). For cardiac protocols, the 12 inner detector rings were applied. Image acquisition was performed by using inspiratory breath holding. Ten separate data sets were reconstructed at different times, which were evenly spaced throughout the cardiac cycle (0%, 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, and 90% of the R-R interval). All reconstructions were transferred to a PC-based workstation (Syngo CaScoring, Wizard; Siemens Medical Solutions, Erlangen, Germany) for quantification of coronary calcifications.
Coronary calcifications were defined on CT images as the presence of more than two contiguous pixels with greater than 130 HU. These lesions were automatically identified and marked in terms of color at the workstation. In a second step, a radiologist (T.S., 2 years of experience in coronary artery CT) and a cardiologist (A.S., 8 years of experience in coronary artery CT) differentiated between calcium and noise by placing regions of interest around the coronary calcifications. All values for the left main, left anterior descending, circumflex, and right coronary arteries were added to calculate the total Agatston (4) and volumetric (7) scores. Both scores were determined for all reconstructions in all patients. The calcium mass was not calculated because no calibration phantom was available.
Data and Statistical Analysis
The patients were assigned to three groups, depending on the mean Agatston score for all reconstructions. The groups were as follows: group 1, score of less than 100; group 2, score of 100400; group 3, score of greater than 400. The total range of Agatston and volumetric scores and the mean scores, standard deviations, and variation coefficients were calculated for all patients. The mean Agatston score and standard deviation of three diastolic reconstructions (50%, 60%, and 70% of the R-R interval) were assessed by using a descriptive statistics tool (version 10.0.7; SPSS, Chicago, Ill) and were compared with the mean Agatston score and standard deviation of three systolic reconstructions (10%, 20%, and 30% of the R-R interval; sign test, P
.05 needed to indicate a significant difference).
The distribution of the minimum and maximum total Agatston scores, as well as the Agatston scores in the left main, left anterior descending, circumflex, and right coronary arteries within the cardiac cycle, was assessed for all patients. Moreover, depending on total Agatston score, age, and sex, patients were assigned to a cardiovascular risk group (percentile 025, low risk group; 2650, moderate risk; 5175, increased risk; 7690, high risk; >90, very high risk) on the basis of the different reconstructions. The percentile values were derived from an American population described by Hoff et al (17).
| Results |
|---|
|
|
|---|
The mean Agatston score among all patients was 324 ± 458. The mean Agatston score was 25 ± 22 for group 1 (n = 10), 218 ± 99 for group 2 (n = 10), and 948 ± 432 for group 3 (n = 10). The mean coefficient of variation for different reconstruction intervals was 63.1% (range, 25.3%174.0%) for group 1, 21.5% (range, 11.8%37.1%) for group 2, and 7.1% (range, 4.3%9.2%) for group 3.
The mean volumetric score among all patients was 298 ± 404. The mean volumetric score for group 1 was 29 ± 22; for group 2, it was 258 ± 124; and for group 3, it amounted to 953 ± 335. The mean coefficient of variation for different reconstruction intervals was 60.2% (range, 12.1%169.4%) for group 1, 19.7% (range, 6.5%33.1%) for group 2, and 9.6% (range, 9.3%10.1%) for group 3. Table 1 lists the variation of the Agatston score and the volumetric score for one patient in each group as a representative example.
|
Table 2 illustrates the distribution of the minimum and maximum total Agatston scores within the cardiac cycle. Most frequently, the minimum scores were found at 90% (n = 5) and 0% (n = 6) reconstruction intervals, whereas the maximum scores were determined mostly at the 80% (n = 5) and 20% (n = 5) reconstruction intervals. A systematic distribution of maximum and minimum scores to certain reconstruction intervals was not observed; this was also true for distribution of the Agatston scores in the left main, left anterior descending, circumflex, and right coronary arteries. However, in 12 (57%) of 21 patients with calcifications in the left anterior descending coronary artery, maximum scores were found with the 70% or 80% reconstruction interval (Table 3).
|
|
|
|
| Discussion |
|---|
|
|
|---|
CT has been shown to be highly sensitive in the depiction of coronary artery calcium. Hence, the technique allows for the early identification of coronary artery disease. For the detection and quantification of coronary calcium, two different CT techniques exist: electron beam CT (4,18) and multidetector row spiral CT (9). Electrocardiographic triggering is a prerequisite for all calcium measurements. For electron-beam CT, it has been demonstrated that optimal electrocardiographic triggering reduces the interscan variation of coronary artery calcium score (19). In contrast to the prospective electrocardiographic triggering used with electron-beam CT, retrospective electrocardiographic gating is used with multidetector row CT. This provides the opportunity to perform multiple reconstructions at different time points throughout the cardiac cycle. Horiguchi et al (20) demonstrated that this technique produces cardiac images with less motion artifacts than does prospective triggering and showed a high correlation with coronary calcium scores determined by using electron-beam CT.
Mahnken et al (13) have recently shown that the total calcium score depends on the image reconstruction interval. They compared the total mean score acquired at different reconstruction intervals by using a fourdetector row CT scanner. The introduction of the latest generation multidetector row CT scanner, with 16 detector rows, shows promise to improve reproducibility by combining retrospective gating with isotropic volumetric imaging. However, our study results demonstrate a high dependence of both the Agatston and volumetric scores on the reconstruction interval, particularly in patients with mild or moderate amounts of coronary calcium. Furthermore, in five patients in whom coronary calcium was detected in one reconstruction interval, one or more reconstructions failed to demonstrate any calcium at all. Table 4 illustrates the effect of the relatively high variation in patients with mild coronary calcifications. The variable Agatston scores of three diastolic reconstructions result in completely different estimations of coronary risk, which range between low risk and increased risk. Overall, 19 (63%) of 30 patients could be assigned to more than one risk group; of these, three patients could be assigned to three risk groups, and one patient could even be assigned to four risk groups.
It appears reasonable that the "real" calcium score in a blood vessel is represented by the highest score, since motion may result in smearing artifacts and an underestimation of the score. However, Horiguchi et al (21) demonstrated in a cardiac phantom study that calcium is deformed and blurred in the systolic phase, which results in increased calcium scores. Our data support these findings. The 60% reconstruction interval, which is commonly used in clinical routine because of minimal cardiac motion, did not show the maximum Agatston scores for the left anterior descending coronary artery in any of the patients. Instead, the minimum and maximum scores were evenly distributed within the cardiac cycle. This indicates that a fixed reconstruction interval is inadequate for the assessment of maximum or minimum calcium scores, even with the newest generation of CT scanners.
To estimate the progression of coronary calcification as a marker of atherosclerosis in an individual patient, accurate and reproducible calcium quantification is an indispensable tool. To assess the accurate score for the follow-up examinations, it seems mandatory to perform reconstructions that cover the entire cardiac cycle. However, it is doubtful whether this time-consuming postprocessing is practicable in a routine clinical setting. Therefore, the mean of a limited number of reconstructions acquired at defined time points within the cardiac cycle may minimize random errors and improve the reproducibility of the Agatston and volumetric scores in clinical routine. Further studies will be needed to show whether the mean values of diastolic or systolic measurements are better suited for follow-up examinations than are reconstructions performed at a fixed time point within the cardiac cycle.
With the use of a 16detector row CT scanner, the Agatston and volumetric scores both proved to be highly dependent on the reconstruction interval used. Minimum and maximum values of the Agatston and volumetric scores were evenly distributed over the entire cardiac cycle. Accurate and reproducible quantification of coronary calcium seems to require the analysis of multiple reconstructions, even with the most advanced CT scanners.
| FOOTNOTES |
|---|
Author contributions: Guarantors of integrity of entire study, T.S., J.B.; study concepts, T.S., P.H., J.B., A.S.; study design, T.S., H.K., K.U.W., J.F.D.; literature research, T.S., A.S., J.B.; clinical studies, T.S., H.K., P.H.; data acquisition, T.S., H.K., K.U.W.; data analysis/interpretation, T.S., P.H., J.B., A.S., J.F.D.; statistical analysis, T.S., J.B.; manuscript preparation and editing, T.S., P.H., A.S., H.K.; manuscript definition of intellectual content, T.S., K.U.W., J.F.D., J.B.; manuscript revision/review and final version approval, all authors
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
N. Matsuura, J. Horiguchi, H. Yamamoto, N. Hirai, T. Tonda, N. Kohno, and K. Ito Optimal Cardiac Phase for Coronary Artery Calcium Scoring on Single-Source 64-MDCT Scanner: Least Interscan Variability and Least Motion Artifacts Am. J. Roentgenol., June 1, 2008; 190(6): 1561 - 1568. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Schlosser, P. Hunold, T. Voigtlander, A. Schmermund, and J. Barkhausen Coronary Artery Calcium Scoring: Influence of Reconstruction Interval and Reconstruction Increment Using 64-MDCT Am. J. Roentgenol., April 1, 2007; 188(4): 1063 - 1068. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, R. S. Blumenthal, J. J. Carr, J. G. Goldin, P. Greenland, A. D. Guerci, J. A.C. Lima, D. J. Rader, G. D. Rubin, et al. Assessment of Coronary Artery Disease by Cardiac Computed Tomography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology Circulation, October 17, 2006; 114(16): 1761 - 1791. [Full Text] [PDF] |
||||
![]() |
S. Tridandapani, J. B. Fowlkes, and J. M. Rubin Echocardiography-Based Selection of Quiescent Heart Phases: Implications for Cardiac Imaging J. Ultrasound Med., November 1, 2005; 24(11): 1519 - 1526. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |