|
|
||||||||
Special Reviews |
1 From Siemens Medical Solutions, CT Division, Forchheim, Germany (T.G.F., S.S., K.S., H.B., B.M.O.); Department of Diagnostic Radiology, Tübingen University, Germany (T.G.F.); and Department of Radiology, Medical University of South Carolina, 169 Ashley Ave, Charleston, SC 29425 (U.J.S.). Received January 7, 2004; revision requested March 9; revision received April 26; accepted May 24. Address correspondence to U.J.S. (e-mail: schoepf@musc.edu).
| ABSTRACT |
|---|
|
|
|---|
Supplemental material: radiology.rsnajnls.org/cgi/content/full/2353040037/DC1
© RSNA, 2005
| INTRODUCTION |
|---|
|
|
|---|
A glossary of terms used in this review is available online in Appendix E1 (radiology.rsnajnls.org/cgi/content/full/2353040037/DC1).
| EVOLUTION OF SPIRAL CT: FROM ONE SECTION TO 16 |
|---|
|
|
|---|
Ideally, volume data are of high spatial resolution and are isotropic in nature: Each image data element (voxel) is of equal dimensions in all three spatial axes, and this forms the basis for image display in arbitrarily oriented imaging planes. For most clinical scenarios, however, single-section spiral CT with a 1-second gantry rotation is unable to fulfill these requirements. To prevent motion artifacts and optimally utilize the contrast agent bolus, body spiral CT examinations need to be completed within a certain time frame of, ordinarily, one breath hold (2530 seconds). If a large scan range such as the entire thorax or abdomen (30 cm) has to be covered within a single breath hold, a thick collimation of 58 mm must be used. While the in-plane resolution of a CT image depends on the system geometry and on the reconstruction kernel selected by the user, the longitudinal (z-axis) resolution along the patient axis is determined by the collimated section width and the spiral interpolation algorithm. Use of a thick collimation of 58 mm results in a considerable mismatch between the longitudinal resolution and the in-plane resolution, which is 0.50.7 mm, depending on the reconstruction kernel. Thus, with single-section spiral CT, the ideal of isotropic resolution can only be achieved for very limited scan ranges (5).
Strategies to achieve more substantial volume coverage with improved longitudinal resolution include the simultaneous acquisition of more than one section at a time and a reduction in the gantry rotation time. Interestingly, the first medical CT scanners were two-section systems, such as the EMI (England) head scanner, introduced in 1972, and the Siemens Siretom (Erlangen, Germany), introduced in 1974. With the advent of whole-body fan-beam CT systems for general radiology, two-section acquisition was no longer used. Apart from a dedicated two-section system for cardiac applications, the Imatron C-100 (Imatron, San Francisco, Calif), which was introduced in 1984, the first step toward multisection acquisition in general radiology was a two-section CT scanner introduced in 1993 (Elscint TWIN; Elscint, Haifa, Israel) (6). In 1998, several CT manufacturers introduced multidetector row CT systems, which provided considerable improvement in scanning speed and longitudinal resolution and better utilization of the available x-ray power (710). These systems typically offered simultaneous acquisition of four sections at a gantry rotation time of 0.5 second.
Simultaneous acquisition of m sections results in an m-fold increase in speed if all other parameters (eg, section thickness) are unchanged. This increased performance of multidetector row CT relative to single-section CT allowed the optimization of a variety of clinical protocols. The examination time for standard protocols could be substantially reduced, which proved to be of immediate clinical benefit for the quick and comprehensive assessment of trauma patients and uncooperative patients (11). Alternatively, the scan range that could be covered within a certain time was extended by a factor of m, which is relevant for oncologic staging or for CT angiography with extended coverage (eg, the lower extremities) (12).
The most important clinical benefit, however, proved to be the ability to scan a given anatomic volume within a given scan time with substantially reduced section width at m times increased longitudinal resolution. Because of this, the goal of isotropic resolution was within reach for many clinical applications. Examinations of the entire thorax (13) or abdomen could now be routinely performed with a 1.0- or 1.25-mm collimated section width. Despite these promising advances, clinical challenges and limitations remained for four-section CT systems. True isotropic resolution for routine applications had not yet been achieved, because the longitudinal resolution of about 1 mm does not fully match the in-plane resolution of about 0.50.7 mm in a routine examination of the chest or abdomen. For large volumes, such as for CT angiography of lower extremity vessels (12), thicker (eg, 2.5-mm) collimated sections had to be chosen to complete the scan within a reasonable time frame. Scan times were often too long to allow image acquisition during a purely arterial phase. For CT angiography of the circle of Willis, for instance, a scan range of about 100 mm must be covered (14). With four-section CT at a collimated section width of 1 mm, pitch of 1.5, and gantry rotation time of 0.5 second, this volume can be covered in about 9 seconds, not fast enough to avoid venous overlay, assuming a cerebral circulation time of less than 5 seconds. (Note: The definition of pitch for multidetector row CT is discussed later in this review.)
As a next step, the introduction of an eightdetector row CT system in 2000 enabled shorter scan times but did not yet provide improved longitudinal resolution (thinnest collimation, eight sections at 1.25 mm). The latter was achieved with the introduction of 16detector row CT (15), which made possible the routine acquisition of substantial anatomic volumes with isotropic submillimeter spatial resolution and scan times of less than 10 seconds for 300 mm of coverage (Fig 1). While in-plane spatial resolution is not substantially improved, the two major advantages of fast multidetector row CT are a true isotropic through-plane resolution and a short acquisition time that enable high-quality examinations in severely debilitated and severely dyspneic patients (Fig 1).
|
The limitations of four and eightdetector row CT systems, however, have so far prevented the successful integration of CT coronary angiography into routine clinical algorithms: Stents or severely calcified arteries constitute a diagnostic dilemma, mainly because of partial volume artifacts as a consequence of insufficient longitudinal resolution (22). For patients with a higher heart rate, careful selection of separate reconstruction intervals for different coronary arteries has been mandatory (25). It is almost impossible for patients with manifest heart disease to comply with the breath-hold time of about 40 seconds required to cover the entire heart volume (approximately 12 cm) with four-section CT. The ongoing technical refinement of multidetector row CT, however, holds the promise of gradually overcoming some of these limitations. The most important steps toward this goal are gantry rotation times faster than 0.5 second (26,27) for improved temporal resolution and robustness of use, 16-section submillimeter acquisition for increased longitudinal resolution and shorter breath-hold times, and novel sophisticated approaches for image acquisition and reconstruction.
In this review, ECG-synchronized examinations of the heart and of the cardiothoracic anatomy will be very succinctly discussed, since this topic has been extensively reviewed elsewhere (28). Similarly, advanced 3D postprocessing techniques are omitted. In this article, we will review the general technical principles of multidetector row CT as they apply to the established four and eightdetector row systems, the more recent 16detector row scanners, and generations of CT systems yet to come. On the basis of the technologic description of different scanner types and image-reconstruction approaches, we provide practical "take-home points" to enable better translation into daily clinical practice of the technology and science reviewed here. Useful up-to-date information regarding multidetector row CT is also readily available on the Internet at, for example, the UK Medicines and Healthcare products Regulatory Agency CT Web site (www.medical-devices.gov.uk) or the Advanced Medical Imaging Laboratory site (www.ctisus.org).
| CURRENT TECHNIQUES |
|---|
|
|
|---|
|
For established four-section CT systems, two detector types are commonly used. The fixed-array detector consists of detector elements with equal sizes in the longitudinal direction. A representative example of this scanner type, the Lightspeed scanner (GE Medical Systems, Milwaukee, Wis), has 16 detector rows, each of them defining a 1.25-mm collimated section width in the center of rotation (8,10,29). The total coverage in the longitudinal direction is 20 mm at the isocenter; owing to geometric magnification, the actual detector is about twice as wide. By means of prepatient collimation and combination of the signals of the individual detector rows, the following section widths (measured at the isocenter) can be realized: four sections at 1.25 mm, 2.5 mm, 3.75 mm, and 5.0 mm (Fig 3a). The same detector design is used for the eight-section version of this system and provides eight sections at 1.25- and 2.5-mm collimated section widths.
|
|
|
The selection of the collimated section width determines the intrinsic longitudinal resolution of a scan. In a "step-and-shoot" sequential mode, any multiple of the collimated width of one detector section can be obtained by adding the detector signals during image reconstruction. In a spiral mode, the effective section width, which is usually defined as the full width at half maximum (FWHM) of the spiral section-sensitivity profile (SSP), is adjusted independently in the spiral interpolation process during image reconstruction. Hence, from the same data set, both narrow sections for high-spatial-resolution detail or for 3D postprocessing and wide sections for better contrast resolution or quick review and filming may be derived.
Sixteen-section CT systems usually have adaptive-array detectors. A representative example for this scanner type, the Somatom Sensation 16 scanner (Siemens), uses 24 detector rows (15). The 16 central rows define 0.75-mm collimated section widths at the isocenter, and the four outer rows on both sides define 1.5-mm collimated section widths (Fig 3c). The total coverage in the longitudinal direction is 24 mm at the isocenter. By means of appropriate combination of the signals of the individual detector rows, either 12 or 16 sections with 0.75- or 1.5-mm collimated section width can be acquired simultaneously. The Lightspeed 16 scanner (GE Medical Systems) uses a similar design: It provides 16 sections with either 0.625- or 1.25-mm collimated section width. The total coverage in the longitudinal direction is 20 mm at the isocenter. Yet another design, which is implemented in the Aquilion scanner (Toshiba, Tokyo, Japan), can provide 16 sections with either 0.5-, 1.0-, or 2.0-mm collimated section width, with a total coverage of 32 mm at the isocenter.
Radiation Dose
Radiation dose and dose efficiency.Radiation exposure to the patient at CT and the resulting potential radiation hazard have recently gained considerable attention in both the public and the scientific literature (30,31). Typical values for the effective patient dose for selected CT protocols are 12 mSv for a head CT, 57 mSv for a chest CT, and 811 mSv for abdominal and pelvic CT (32,33). This radiation exposure must be appreciated in the context of the average annual background radiation, which is 25 mSv (3.6 mSv in the United States). Despite the undisputed clinical benefits, multisection CT scanning is often considered to require increased patient dose compared with the dose from single-section CT. Indeed, a certain increase in radiation dose is unavoidable owing to the underlying physical principles.
In the x-ray tube of a CT scanner, a small area on the anode plate, the focal-spot, emits x-rays that penetrate the patient and are registered by the detector. A collimator between the x-ray tube and the patient, the prepatient collimator, is used to shape the beam and to establish the dose profile. In general, the collimated dose profile is a trapezoid in the longitudinal direction. In the umbral region (ie, plateau region of the trapezoid), x-rays emitted from the entire area of the focal spot illuminate the detector. In the penumbral regions, only a part of the focal spot illuminates the detector, while the prepatient collimator blocks off other parts.
With single-section CT, the entire trapezoidal dose profile can contribute to the detector signal, and the collimated section width is determined as the FWHM of this trapezoid. The relative dose utilization of a single-section CT system can therefore be close to 100%. In most cases with multidetector row CT, only the plateau region of the dose profile is used to ensure an equal signal level for all detector elements. The penumbral region is then discarded, either by a postpatient collimator or by the intrinsic self-collimation of the multisection detector, and represents "wasted" dose. The relative contribution of the penumbral region increases with decreasing section width, and it decreases with increasing number of simultaneously acquired images. This is demonstrated in Figure 4, which compares the "minimum width" dose profiles for a four-section CT system and a corresponding 16-section CT system with equal collimated width of one detector section. Correspondingly, the relative dose utilization with four-section 1-mm-collimation CT is 70% or less (10), depending on the scanner type. With 16-section CT systems and submillimeter collimation, dose utilization can be improved to 84%, again depending on scanner type (25). Some multidetector row CT systems offer special implementations of even more dose-efficient modes that use a portion of the penumbral region.
|
Concepts for radiation dose reduction.The most important factor for reducing radiation exposure is an adaptation of the dose to the patients size and weight (3537).
As a general rule for the practicing radiologist, the dose necessary to maintain constant image noise has to be doubled if the patient diameter is increased by 4 cm. Correspondingly, for a patient diameter that is 4 cm smaller than average, half the standard dose is sufficient to maintain adequate image quality. This is of particular importance in pediatric imaging. Dose reduction can be achieved by reductions in the milliampere-seconds and voltage settings. Most CT manufacturers provide dedicated pediatric protocols with, for example, milliampere-seconds and voltage settings adjusted according to the weight of the child.
Another means to reduce radiation dose is to adapt the x-ray tube voltage to the intended application. In contrast agentenhanced studies such as CT angiography, the contrast-to-noise ratio for fixed patient dose increases with decreasing x-ray tube voltage. As a consequence, to obtain the desired contrast-to-noise ratio, the patient dose can be reduced by choosing a lower voltage setting. The potential for dose saving is more substantial for patients with a smaller diameter. This can be demonstrated, for example, by means of phantom measurements of small tubes filled with diluted contrast agent embedded in acrylic plastic phantoms with different diameters (38). The iodine contrast-to-noise ratio at constant radiation dose for various voltage settings is shown in Figure E2 (radiology.rsnajnls.org/cgi/content/full/2353040037/DC1) as a function of the phantom diameter. Compared with a standard scan at 120 kV in a 32-cm-diameter phantom (corresponding to that for an average adult), the same contrast-to-noise ratio is obtained with 0.49 times the dose (1.3 times the milliampere-seconds setting) for 80 kV and 0.69 times the dose (1.1 times the milliampere-seconds) for 100 kV. Thus, ideally, 80 kV should be used for CT angiography in order to reduce patient dose.
Clinical studies (38) have confirmed these findings and demonstrated a potential for dose reduction of about 50% when 80 kV is used for CT angiography instead of 120 kV. In reality, however, the maximum x-ray tube current available at 80 kV is generally not sufficient to scan bigger patients, which limits the routine application of this approach. Therefore, use of 100 kV appears to be a suitable compromise and the method of choice for CT angiography. Figure 5 shows pulmonary CT angiographic images of a patient with pulmonary embolism; the scan was performed on a 16-section scanner at 100 kV and 120 mAs, and the effective patient dose for this scan was 2.3 mSv, 25% less than that for the standard 120-kV protocol. Authors of recent study (39) recommended 100 kV as the standard mode for thoracic and abdominal CT angiography and report dose savings of 30% without loss of diagnostic information.
|
In more sophisticated approaches, tube output is modified according to the patient geometry not only during each rotation but also in the longitudinal direction (automatic exposure control), to maintain adequate dose when moving to different body regions (eg, from thorax to abdomen). In one implementation, the attenuation for each body region of a "standard-sized" patient is stored in the control computer. This attenuation corresponds to the milliampere-seconds setting of the standard protocol. If the actual attenuation of the patient deviates from the "standard" attenuation, the tube output is adapted correspondingly. Figure 6 shows the variation of the milliampere-seconds output for a CT scan of the chest and abdomen in a 6-year-old child. Although the standard protocol with 165 mAs was usedwhich would have resulted in substantially higher radiation dose than necessary in a standard mode of operationthe average milliampere-seconds value throughout the scan was adjusted to 38 mAs by means of automatic exposure control. Automatic adaptation of tube current to patient size prevents both over- and underirradiation, considerably simplifies the clinical workflow for the technician, and eliminates the need for look-up tables of patient weight and size for adjusting the milliampere-seconds settings.
|
| SEQUENTIAL SCANS AND IMAGE-RECONSTRUCTION TECHNIQUES |
|---|
|
|
|---|
The number of images acquired during a sequential scan corresponds to the number of active detector sections. By adding the detector signals of the individual sections during image reconstruction, the number of images per scan can be reduced, and the image section width can be increased. As an example, a scan with four sections at 1.0-mm collimation provides either four images with 1.0-mm section width, two images with 2.0-mm section width, or one image with 4.0-mm section width.
The option to realize a wider section by summing several thin sections is beneficial for examinations that require narrow collimation to prevent partial volume artifacts and low image noise to allow detection of low-contrast details (eg, neurologic examinations of posterior fossa or cervical spine). In the head, partial volume artifacts typically manifest as dark streaks or areas of hypoattenuation and are due to a nonlinear effect that has been described in reference 45. Figure 7 shows an example of a patient who underwent follow-up CT after surgical removal of a pituitary tumor. From the same scan datafour sections at 1.0-mm collimationboth 4.0-mm-thick images with a standard head kernel for soft-tissue evaluation and 1.0-mm-thick images with a bone kernel were reconstructed. For best image quality, the posterior fossa should be scanned with a collimated section width not larger than 1.25 mm, whereas wider collimation can be used in the supratentorial region (46).
|
| SPIRAL SCANS AND IMAGE-RECONSTRUCTION TECHNIQUES |
|---|
|
|
|---|
Definition of Spiral Pitch
An important parameter for characterizing a spiral CT scan is the pitch. According to International Electrotechnical Commission specifications (34), the pitch (p) is given by p = TF/W, where TF is the table feed per rotation, and W is the total width of the collimated beam. This definition holds for both single-section and multidetector row CT. It shows whether data acquisition occurs with gaps (p > 1) or with overlap (p < 1) in the longitudinal direction. With 16 sections at 0.75-mm collimation and a table-feed of 18 mm per rotation, the pitch is p = 18/(16 x 0.75) = 18/12 = 1.5. With four sections at 1.0-mm collimation and a table-feed of 6 mm per rotation, the pitch again is p = 6/(4 x 1) = 6/4 = 1.5. In the early days of four-section CT, the term detector pitch had been additionally introduced, which accounts for the width of a single section in the denominator. For the sake of clarity and uniformity, the detector pitch should no longer be used.
Short Review of Single-Section Spiral CT Reconstruction
Spiral CT requires an interpolation of the acquired measurement data in the longitudinal (through-plane) direction to estimate a complete CT data set at the desired plane of reconstruction. The most commonly used single-section spiral interpolation schemes are the 360° and 180° linear interpolation methods.
The 360° linear interpolation method exploits the 360° periodicity of the projection data (1,2). For each projection angle, a linear interpolation is performed between those two projections on either side of the image plane that are positioned closest to the image plane and are 360° apart (ie, are measured in subsequent rotations). The 180° linear interpolation technique makes use of the fact that for each measurement ray, an interpolation partner is already available after approximately half a rotation (47), when the x-ray tube and detector have exchanged positions. This is the so-called complementary ray. In spiral CT, z-axis resolution is determined not only by the collimated beam width (as in sequential scanning) but also by the effective section width, which is established in the spiral interpolation process. Usually, the effective section width is defined as the FWHM of the SSP. Effective section width increases with increasing pitch for both 360° and 180° linear interpolation, and longitudinal resolution degrades (Fig E3, radiology.rsnajnls.org/cgi/content/full/2353040037/DC1). This is a consequence of the increasing longitudinal distance of the projections used for spiral interpolation. With 180° linear interpolation, the effective sections width equals the collimated section width at a pitch of 1, but effective section width equals 1.27 times the collimated width at a pitch of 2, so that a collimated 5-mm-thick section is an actual 6.4-mm-thick section at a pitch of 2. The image noise in single-section spiral CT is independent of the pitch if the tube current (in milliamperes) is left unchanged, and patient dose decreases with increasing pitch (see Appendix E2, radiology.rsnajnls.org/cgi/content/full/2353040037/DC1).
Single-section spiral CT is based almost exclusively on 180° linear interpolation, owing to the narrower SSP of this algorithm, despite its increased susceptibility to artifacts and increased image noise. For the same milliampere-seconds setting, image noise is about 15% higher than that in sequential CT mode. Spiral artifacts gradually increase as pitch is increased. Spiral artifacts typically manifest as hyper- or hypoattenuating "windmill" structures surrounding z-axis inhomogeneous high-contrast objects (eg, bones), which rotate when scrolling through a stack of images. Spiral artifacts are caused by the spiral interpolation process and can also be observed on multidetector row CT images (see Fig 8). With single-section CT, scanning at a higher pitch is often used to reduce patient dose at the expense of section broadeningif the collimation is kept constantand increased spiral artifacts. For CT angiographic applications in particular, it is more favorable to scan a given volume in a given time by using narrow collimation at a high pitch rather than wider collimation at a low pitch. The motivation for increasing pitch and reducing collimation is to improve longitudinal resolution by narrowing the SSP (48).
|
|
| MULTIDETECTOR ROW SPIRAL CT RECONSTRUCTION APPROACHES THAT NEGLECT CONE-BEAM GEOMETRY |
|---|
|
|
|---|
In general, scanners that rely on 180° or 360° multidetector linear interpolation techniques and extensions thereof provide selected discrete pitch values to the user, such as 0.75 and 1.5 for four-section scanning (29) or 0.5625, 0.9375, 1.375, and 1.75 for 16-section scanning (50). These pitch values are intended to provide optimized sampling schemes in the longitudinal direction and, hence, optimized image quality.
The user has to be aware of pitch-dependent effective section widths. For low-pitch scanning (pitch of 0.75 for four sections and 0.5625 or 0.9375 for 16 sections), the effective section width approximates the collimated section width; for a 1.25-mm collimated section width, the resulting effective section width remains 1.25 mm. The narrow SSP, however, is achieved by using 180° multidetector linear interpolation reconstruction with conjugate interpolation at the price of increased image noise (29,50). For high-pitch scanning (pitch of 1.5 for four sections and 1.375 or 1.75 for 16 sections), the effective section width is approximately 1.27 times the collimated section width, and a 1.25-mm collimated section width results in a 1.51.6-mm effective section width.
When comparing dose and image noise for different pitch values, the widening of the SSP has to be taken into account. To obtain the same image noise as in a sequential scan with the same collimated section width, 0.731.68 times the dose (depending on spiral pitch) is required, with the lowest dose at the highest pitch (see reference 50). Some manufacturers provide a semiautomatic adaptation of the milliampere value to keep the image noise constant if the pitch is changed. In clinical practice, therefore, it is permissible to assume that scanners offering discrete optimized pitch values based on 180° and 360° multidetector linear interpolation techniques are comparable to single-section CT systems in some core aspects: At high pitch, the section widens and the longitudinal resolution degrades; at low pitch, the narrowest possible SSP (comparable to that of 180° single-section linear interpolation at pitch of 1) can be obtained, but a higher dose is necessary to maintain the signal-to-noise ratio. Thus, as a take-home point, when one selects the scan protocol for a particular application, scanning at low pitch optimizes image quality and longitudinal resolution at a given collimation but at the expense of increased patient dose. To reduce patient dose, either milliampere settings should be reduced at low pitch values or high pitch values should be chosen.
Z-Filter Approaches
In a z-filter multidetector row spiral reconstruction (51,52), the spiral interpolation for each projection angle is no longer restricted to the two rays closest to the image plane. Instead, all direct and complementary rays within a selectable distance from the image plane contribute to the image. The weighting function for the rays is selectable, which allows one to adjust both the functional form and the FWHM of the spiral SSP. Still, the cone angle is neglected. A representative example of a z-filter approach is the adaptive axial interpolation algorithm (51) implemented in Siemens CT scanners, which is illustrated in Figure E5 (radiology.rsnajnls.org/cgi/content/full/2353040037/DC1). Another example is the "multislice cone-beam tomography," or MUSCOT, algorithm (52) used by Toshiba. Z filtering allows the system to trade off z-axis resolution (the SSP) with image noise (which directly correlates with required dose).
With adaptive axial interpolation, the spiral pitch is freely selectable in the range 0.52.0, and the same effective section width, which is defined as the FWHM of the spiral SSP, is generated at all pitch values (7,51,53). Therefore, longitudinal resolution is independent of pitch, unlike single-section spiral CT and multidetector row CT that relies on 180° and 360° linear interpolation (51,54). Figure E6 (radiology.rsnajnls.org/cgi/content/full/2353040037/DC1) shows the SSPs of a 2-mm section (for four-section CT at 1-mm collimation) and MPRs of a spiral z-axis resolution phantom for selected pitch values. As a consequence of the pitch-independent spiral section width, the image noise for a fixed tube current (in milliamperes) would decrease as pitch is decreased, owing to the increasingly overlapping spiral acquisition. Instead, the user selects an "effective" milliampere-seconds value, and the tube current is then automatically adapted to the pitch of the spiral scan to compensate for dose accumulation. The dose for fixed effective milliampere-seconds is independent of the spiral pitch and equals the dose of a transverse scan with the same milliampere-seconds setting (see Appendix E2, radiology.rsnajnls.org/cgi/content/full/2353040037/DC1).
Thus, as a take-home point, unlike in single-section spiral CT a change in pitch does not result in a change in dose to the patient. Accordingly, the use of a higher pitch does not result in a dose saving, which is an important practical consideration with CT systems that rely on adaptive axial interpolation.
The intrinsic resolution of a multidetector row spiral scan is determined by the choice of collimation (eg, four sections at 1.0 or 2.5 mm). Z filtering makes it possible to reconstruct images retrospectively with different section widths from the same raw CT data set. Only section widths equal to or larger than the section width of one active detector row can be obtained. In many cases, both thick sections for initial viewing and recording and thin sections for detailed diagnosis or as an input for advanced 3D postprocessing are routinely reconstructed.
The thinnest available section width is the collimated section width (1.0 mm for four sections at 1.0-mm collimation), which is created by using nonlinear spiral weighting functions at the expense of increased image noise and increased susceptibility to artifacts. Thus, as a take-home point, the thinnest available section should only be used for high-contrast applications such as high-spatial-resolution lung imaging. For general purpose scanning, a 1.25-mm section width for four-section CT at 1.0-mm collimation (and 3.0-mm section width for four sections at 2.5-mm collimation) is recommended as the most suitable trade-off between longitudinal resolution, image noise, and artifacts, in particular when thin sections are reconstructed as an input for 3D postprocessing such as for MPR, maximum intensity projection, or volume-rendering techniques. For a 1.25-mm spiral section width reconstructed from four-section CT at 1.0-mm collimation, 0.610.69 times the dose (depending only slightly on spiral pitch) is required to maintain the image noise of a sequential scan at the same collimation (see references 54,55). Unlike 180° and 360° multidetector linear interpolation, image noise is therefore practically independent of pitch at constant dose.
For a given collimation, such as four sections at 2.5 mm, image quality can be optimized with regard to spiral artifacts by lowering the pitch (56). Another means to reduce spiral artifacts is to use narrow collimation: A given section width (eg, 3.0 mm) can be obtained with different collimations, in this case four sections at 1.0 mm and at 2.5 mm. For optimum image quality, collimation that is narrow relative to the desired section width is preferable (51). Furthermore, a more rectangular SSP can be established. Figure 10a shows the SSPs of a 3.0-mm section for four-section CT at both 1.0- and 2.5-mm collimation. Figure 10b shows 3.0-mm transverse sections of a thorax phantom scanned with four-section CT at 2.5- and 1.0-mm collimation. Despite the higher pitch, the 3.0-mm image obtained at 1.0-mm collimation shows fewer artifacts. Similar to single-section spiral CT, narrow collimation at high pitch is preferable to wide collimation at low pitch for artifact reduction.
|
|
In general, more challenging clinical protocols, such as CT of the spine and of the skull base, are reliant on a combination of narrow collimation and low pitch. When multidetector row spiral CT of the head is performed with narrow collimation, low pitch, and z-filter reconstruction of wider sections, the results are equivalent to those of traditional sequential CT. Figure 11 shows an example of a head scan performed with a four-section CT system in which a sequential image (two-section CT at 8 mm) and a spiral image (8-mm section width from four-section CT at 1-mm collimation) are compared in the same patient.
|
| MULTIDETECTOR ROW SPIRAL RECONSTRUCTION APPROACHES THAT ACCOUNT FOR CONE-BEAM GEOMETRY |
|---|
|
|
|---|
AMPR Method
The AMPR approach (67,68) is an extension and generalization of the "advanced single-slice rebinning" (63,64) method. AMPR allows free selection of the spiral pitch with optimized dose utilization, which is beneficial for medical applications. With advanced single-slice rebinning, a partial scan interval (about 240° of scan data) is used for image reconstruction. The image planes are no longer perpendicular to the patient axis; instead, they are tilted to match the spiral path of the focal spot; see Figure 12 for a 16-section scanner at a pitch of 1.5. For every view angle in this partial scan interval, the focal spot is positioned in or near the image planethat is, measurement rays running in or very close to the image plane are available. These conditions need to be fulfilled for a standard two-dimensional reconstruction. In a final z-axis reformation step, the traditional transverse images are calculated by interpolating between the tilted original image planes.
|
|
|
|
Weighted Hyperplane Reconstruction
The weighted hyperplane reconstruction method, which has been described elsewhere (69,70), uses concepts related to AMPR but is derived differently. Similar to AMPR, 3D reconstruction is split into a series of two-dimensional reconstructions. Instead of reconstruction of traditional transverse sections, convex hyperplanes are proposed as the region of reconstruction. The increasing spiral overlap with decreasing pitch is handled by introducing subsets of detector rows, which are sufficient to reconstruct an image at a given pitch value. At pitch of 0.5625 with a 16-section scanner, the data collected by detector rows one to nine form a complete projection data set. Similarly, projections from detector rows two to 10 can be used to reconstruct another image at the same z-axis position. Projections from detector rows three to 11 yield a third image and so on. In a way, these "subimages" are related to the "book pages" of AMPR. The final image is based on a weighted average of the subimages. In the article by Hsieh et al (70), good image quality was demonstrated for a 16-section CT system (Lightspeed 16; GE Medical Systems) with which the weighted hyperplane reconstruction approach was used. By performing parameter optimizations, an optimal balance among various system performance parameters, such as noise, artifacts, and SSPs, can be achieved (72).
| ECG-SYNCHRONIZED SCAN AND IMAGE-RECONSTRUCTION TECHNIQUES |
|---|
|
|
|---|
With retrospective ECG gating, the heart volume is covered continuously by a spiral scan. The basic concepts for ECG-gated spiral imaging, such as single-segment and multisegment reconstruction, had already been developed in 1998 (73). The patients ECG signal is recorded at the same time the CT data are acquired to allow retrospective selection of the data segments used for image reconstruction. Only scan data acquired in a predefined cardiac phase, usually the diastolic phase, are used for image reconstruction (16,17,74,75). The data segments contributing to an image begin with a user-defined offset relative to the onset of the R waves, similar to ECG-triggered sequential scanning. Image reconstruction generally consists of two steps: multidetector row spiral interpolation to compensate for the continuous table movement and to obtain scan data at the desired image z-axis position, followed by a partial scan reconstruction of the transverse data segments. The temporal resolution of an image can be improved up to trot/(2N) by using scan data of N subsequent heart cycles for image formation in a so-called multisegment reconstruction mode (16,7377), where trot is the gantry rotation time of the CT scanner. With increased N, better temporal resolution is achieved but at the expense of slower volume coverage: Increased N and slower patient heart rate require a reduction in spiral pitch.
Multisegment approaches rely on a complete periodicity of the heart motion, and these approaches encounter their limitations in patients with arrhythmia or a heart rate that changes during scan acquisition. Multisegment reconstruction may improve image quality in selected cases, but the reliability of good-quality image acquisitions with N-segment reconstruction is compromised with increases in N.
In general, clinical practice suggests the use of one segment at lower heart rates and two or more (N
2) segments at higher heart rates. Use of single-segment versus multisegment reconstruction is integrated in the data acquisition process in a variety of ways, depending on the scanner type. One approach consists of automatic division of the partial-scan data segment into one or two subsegments, depending on the patients heart rate during acquisition ("adaptive cardio volume" algorithm [74]). With a different approach, single-segment partial-scan images are prospectively reconstructed as baseline images, followed by retrospective two-segment reconstruction for improved temporal resolution in patients with a higher heart rate. Yet another approach is prospective adjustment of the gantry rotation time to the heart rate of the patient to obtain an optimized temporal resolution for a multisegment reconstruction. Again, this approach requires a stable and predictable heart rate during scan acquisition.
Prospective ECG triggering combined with sequential step-and-shoot acquisition of transverse sections has the benefit of smaller patient dose than that of ECG-gated spiral scanning, because scan data are acquired only during the desired heart phases. However, this technique does not provide continuous volume coverage with overlapping sections, and misregistration of anatomic details cannot be avoided. Furthermore, reconstruction of images in different phases of the cardiac cycle for functional evaluation is not possible. Since ECG-triggered sequential scanning depends on a reliable prediction of the patients next R-R interval by using the mean of the preceding R-R intervals, the method encounters its limitations in patients with arrhythmia. To maintain the benefits of ECG-gated spiral CT but reduce patient dose, ECG-controlled dose modulation has been developed (42,43) (see earlier discussion).
The major improvements of 16-section CT, compared with established four-section scanners, include improved temporal resolution due to shorter gantry rotation time, better spatial resolution owing to submillimeter collimation, and considerably reduced scan acquisition times (26,27). The time to cover the entire heart volume (about 12 cm) with four-section CT at 1.0-mm collimation is about 40 seconds, which is at the limit for a scan requiring patient breath holding. ECG-gated CT of the entire thorax or the aorta is not possible within reasonable scan durations. For a 16-section CT system, the time to cover the entire heart volume with submillimeter collimation is about 15 seconds. With 16-section CT, coverage of the entire thorax (30 cm) can be completed in about 38 seconds at 0.75-mm collimation and in about 19 seconds at 1.5-mm collimation. ECG-gated examinations of extended cardiothoracic anatomy became feasible with 16-section CT, which lends itself to a spectrum of applications where suppression of cardiac pulsation is desired. Typical diagnostic pitfalls caused by transmitted cardiac pulsation can be avoided, such as an artifactual intimal flap resembling dissection in the ascending aorta (79). Suppression of cardiac pulsation improves the assessment of paracardiac lung segments and allows confident exclusion of small peripheral pulmonary emboli in segmental and subsegmental arteries (80). In routine thoracic studies, which are not synchronized to the patients ECG signal, cardiac motion usually precludes the assessment of coronary bypass grafts. Figure 16 shows an example of an ECG-gated scan of the entire thorax for a patient with bypass grafts; this scan was acquired with 16-section CT at 0.75-mm collimation and 0.42-second gantry rotation.
|
| APPLICATIONS |
|---|
|
|
|---|
Most protocols even benefit from a combination of all of these advantages. The near isotropic spatial resolution in routine examinations enables 3D renderings of diagnostic quality and oblique MPRs and maximum intensity projections of a resolution similar to that of the transverse images. The availability of multidetector row CT technology has already begun to change the traditional perception of CT imaging. In CT, a distinction is traditionally made between longitudinal and in-plane resolution. This distinction is based mainly on historical reasons. Before the introduction of spiral CT, longitudinal resolution was determined by section collimation alone, while the convolution kernel determined in-plane resolution. With spiral CT, collimation is no longer the only factor used to determine longitudinal resolution; the spiral interpolation function also comes into play. This has been a first step toward decoupling the image section width from the beam width as determined by the collimation. Multidetector row CT now allows reconstruction of arbitrary section widths from a given collimation by using z-filter techniques, as long as the desired section width is not smaller than the collimation. The potential to trade off z-axis resolution and image noise for the same data set is the most important benefit of z-filter reconstruction. In many applications, data acquisition with narrow collimation is recommended independently of the section width desired for primary viewing.
The distinction between longitudinal and in-plane resolution will gradually become a historical curiosity, and the traditional transverse section will loose its clinical importance. In its place, interactive viewing and manipulation of isotropic volume images will become commonplace, with only the key sections or views in arbitrary directions recorded and stored.
Spiral scanning with 16 submillimeter sections, in particular, represents a breakthrough on the way to true isotropic resolution for routine clinical applications. Improved longitudinal resolution is combined with considerably reduced scan times, which facilitate examinations in uncooperative patients and reduce the amount of contrast material needed (although optimized contrast material protocols are also required).
Furthermore, new clinical applications are evolving as a result of the increased speed of volume scanning. CT angiography of the carotid arteries and the circle of Willis with 16 sections at 0.75-mm collimation, 0.5-second rotation time, and pitch of 1.5 requires only 9 seconds for a scan range of about 300 mm (with table feed of 36 mm/sec). For the first time, true arterial phase imaging of the entire carotid artery with high spatial resolution can be performed. Clinical practice indicates the potential of 16-section CT angiography to replace conventional interventional angiography in the evaluation of carotid artery stenosis (81). Evaluation of the supraaortic vessels with 16-section CT is particularly useful in emergency situations, since CT allows a quick diagnosis with optimized patient access.
For patients suspected of having ischemic stroke, both the status of the vessels supplying the brain and the location of the intracranial occlusion can be assessed during the same examination (82). Brain perfusion CT can be performed by using the same modality, with the goal of differentiating irreversibly damaged brain tissue from reversibly impaired tissue at risk. The combined use of nonenhanced CT, perfusion CT, and CT angiography may rapidly provide comprehensive information regarding the extent of ischemic damage in patients with acute stroke (46).
Scan acquisition of the entire thorax (350 mm) with submillimeter collimation can now be performed in approximately 11 seconds. Owing to the short breath-hold time, central and peripheral pulmonary embolism can be reliably and accurately diagnosed even in severely dyspneic patients with limited ability to cooperate (11,83). Meanwhile the use of multidetector row CT for a combined diagnosis of pulmonary embolism and deep venous thrombosis has been clinically established (83). Both a native and a contrast-enhanced scan of the thorax can be obtained within the same breath hold for matching of both image volumes as a basis for investigational applications such as lung perfusion imaging.
Sixteen-section CT enables whole body angiographic studies with submillimeter resolution in a single breath hold. Also, 16-section CT yields the same morphologic information as invasive angiography (84,85). CT angiography of the chest and abdomen with submillimeter collimation can be completed in about 17 seconds for a scan range of 600 mm (Fig 17). When true isotropic resolution is not required, the use of 16-section CT at 1.25- or 1.5-mm collimation enables even shorter examination times or extended scan ranges (eg, for oncologic screening, trauma cases, or CT angiography). Whole-body 16-section CT angiography with 1500-mm scan range, 1.5-mm collimation, 0.5-second rotation time, and pitch of 1.25 (table feed, 60 mm/sec) can be completed in only 26 seconds.
|
|
| FUTURE DIRECTION OF MULTIDETECTOR ROW CT |
|---|
|
|
|---|
Owing to its ease of use and its widespread availability, general-purpose CT continues to evolve into the most widely used diagnostic modality for routine examinations, especially in emergency situations or for oncologic staging. CT primarily provides morphologic information; in combination with other modalities, however, functional and metabolic information can also be obtained (90). Therefore, combined systems for obtaining comprehensive structural and functional diagnoses will gain increasing importance in the near future.
The combination of state-of-the-art multidetector row CT with positron emission tomographic (PET) scanners, for instance, opens a wide spectrum of applications ranging from oncologic staging to comprehensive cardiac examinations. The clinical potential of these scanners is currently being evaluated (91). Reconstruction of the CT images in a sufficient field of view without truncation of anatomic structures (eg, arms) is a prerequisite for adequate attenuation correction of the PET images. An enlarged field of view of up to 70 cm can be realized by extrapolating from the measured CT data. Pertinent algorithms can be found in, for example, reference 92. Figure 19 shows MPRs from CT images in a 46-year-old man with renal cancer who had undergone nephrectomy and chemotherapy, with PET images superimposed. Areas with increased metabolism are enhanced, and a metastatic mediastinal lymph node can be identified, which supports the notion of PET as adding a "new contrast agent" to CT.
|
CT virtual simulation is gaining increasing importance with a more widespread adoption in 3D conformal and intensity-modulated radiation therapy. With general-purpose CT systems that have a gantry opening with a typical diameter of 70 cm, some patients (eg, women with breast cancer) cannot always be scanned in the treatment position. Such applications, along with interventional procedures and trauma protocols, will be facilitated by CT systems with a larger bore (97). Recently, concepts have been introduced for four- and 16-section CT scanners with a bore diameter of up to 85 cm and a reconstruction field of up to 82 cm, owing to image reconstruction based on data extrapolation. These systems will probably gain considerable importance in the near future, in particular with regard to the dramatically increasing number of severely obese patients in the Western countries.
For general purpose CT, we will witness a moderate increase in the number of simultaneously acquired sections in the near future. A new generation of CT systems with 32, 40 andin combination with refined z-axis sampling techniques64 simultaneously acquired sections are currently being introduced. However in contrast to the transition from single-section to four- and 16-section CT, clinical performance will improve only incrementally with further increases in the number of detector rows. The achievable clinical benefit will have to be carefully considered in the light of the necessary technical efforts and the cost. Clinical progress can more likely be expected from further improvements in spatial resolution rather than from an increase in the volume-coverage speed. In clinical reality, the latter has only rarely been a limiting factor since the introduction of 16-section CT. As soon as all relevant examinations can be performed in a comfortable breath hold of not more than 10 seconds, a further increase in the number of sections will not provide a substantial clinical benefit.
At this point, a qualitative enhancement of CT that allows new clinical applications may again bring substantial clinical progress with, for example, the introduction of area detectors large enough to cover entire organs such as the heart, kidneys, or brain in one sequential scan (approximate scan range, 120 mm). With these systems, dynamic volume scanning would become feasible, which would open up a whole spectrum of new applications such as functional or volume perfusion studies.
Area-detector technology is currently under development, but no commercially available system so far fulfills the requirements of medical CT with regard to contrast resolution and fast data readout. A scanner with 256 0.5-mm detector elements has been proposed by one manufacturer and appears to be conceptually promising, but this system is still in the prototype stage. Prototype systems by other vendors use cesium iodideamorphous silicon flat-panel detector technology that was originally used for conventional angiography, which is limited in terms of low contrast resolution and imaging speed. Owing to the intrinsic slow signal decay of flat-panel detectors, rotation times of at least 20 seconds are needed to acquire a sufficient number of projections (
600 projections). The spatial resolution of such systems is excellent, though, because of the small detector pixel size. Excessive dose requirements to date, however, preclude the examination of larger objects. Initial experimental results are thus limited to small high-contrast objects such as joints, the inner ear, or contrast materialfilled vessel specimens (98,99).
Figure 20 shows a prototype set-up, where a flat-panel detector was incorporated into a standard CT gantry (Somatom Sensation 16; Siemens). The detector covers a 25 x 25 x 18 cm scan field of view, and the pixel size is 0.25 x 0.25 mm, both measured at the center of rotation. Figure 21 shows volume renderings of a heart specimen (80 kV, 20 mA, 20-second gantry rotation) that demonstrate excellent spatial resolution, which enables visualization of even very small side branches of the coronary artery tree.
|
|
| ESSENTIALS |
|---|
|
|
|---|
Near-isotropic spatial resolution in routine examinations, which has been achieved with 16-section CT systems, enables 3D renderings of diagnostic quality and oblique MPRs and maximum intensity projections with resolution similar to that of the transverse images.
Scanning at narrow collimation does not markedly increase the radiation dose to the patient, as long as the effective milliampere-seconds level is kept constant.
A key challenge for image reconstruction with multidetector row CT is the cone angle of the measurement rays; this requires novel reconstruction techniques such as 3D back projection, AMPR, or weighted hyperplane reconstruction.
Z filtering makes it possible to reconstruct images retrospectively with different section widths from the same raw CT data set, trading off, in this way, z-axis resolution and image noise.
| FOOTNOTES |
|---|
U.J.S. is a medical consultant to Siemens Medical Solutions, CT Division, Forchheim, Germany.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
U. Metser, S. Ghai, Y. Y. Ong, G. Lockwood, and S. B. Radomski Assessment of Urinary Tract Calculi With 64-MDCT: The Axial Versus Coronal Plane Am. J. Roentgenol., June 1, 2009; 192(6): 1509 - 1513. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. T. Johnson, K. M. Horton, S. Kawamoto, J. Eng, M. J. Bean, S. J. Shan, and E. K. Fishman MDCT for Suspected Appendicitis: Effect of Reconstruction Section Thickness on Diagnostic Accuracy, Rate of Appendiceal Visualization, and Reader Confidence Using Axial Images Am. J. Roentgenol., April 1, 2009; 192(4): 893 - 901. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Vrtiska, R. P. Hartman, J. M. Kofler, M. R. Bruesewitz, B. F. King, and C. H. McCollough Spatial Resolution and Radiation Dose of a 64-MDCT Scanner Compared with Published CT Urography Protocols Am. J. Roentgenol., April 1, 2009; 192(4): 941 - 948. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Marin, C. Catalano, G. De Filippis, M. Di Martino, A. Guerrisi, M. Rossi, and R. Passariello Detection of Hepatocellular Carcinoma in Patients with Cirrhosis: Added Value of Coronal Reformations from Isotropic Voxels with 64-MDCT Am. J. Roentgenol., January 1, 2009; 192(1): 180 - 187. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Funama, K. Awai, K. Taguchi, M. Hatemura, Y. Yanaga, M. Shimamura, and Y. Yamashita Cone-Beam Technique for 64-MDCT of Lung: Image Quality Comparison with Stepwise (Step-and-Shoot) Technique Am. J. Roentgenol., January 1, 2009; 192(1): 273 - 278. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.M. Phal, L.P. Riccelli, P. Wang, G.M. Nesbit, and J.C. Anderson Fracture Detection in the Cervical Spine with Multidetector CT: 1-mm versus 3-mm Axial Images AJNR Am. J. Neuroradiol., September 1, 2008; 29(8): 1446 - 1449. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. W. Goldman Principles of CT: Multislice CT J. Nucl. Med. Technol., June 1, 2008; 36(2): 57 - 68. [Abstract] [Full Text] [PDF] |
||||
![]() |
E J CHOI, Y-W OH, S Y HAM, K Y LEE, and E-Y KANG Comparison between coronal reformatted images and direct coronal CT images of the swine lung specimen: assessment of image quality with 64-detector row CT Br. J. Radiol., June 1, 2008; 81(966): 463 - 467. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Tsampoulas, A. C. Tsili, D. Giannakis, Y. Alamanos, N. Sofikitis, and S. C. Efremidis 16-MDCT Cystoscopy in the Evaluation of Neoplasms of the Urinary Bladder Am. J. Roentgenol., March 1, 2008; 190(3): 729 - 735. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schroeder, S. Achenbach, F. Bengel, C. Burgstahler, F. Cademartiri, P. de Feyter, R. George, P. Kaufmann, A. F. Kopp, J. Knuuti, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: Report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology Eur. Heart J., February 2, 2008; 29(4): 531 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Hashimoto, S Kawashima, S Kameoka, Y Akiyama, T Honjoya, K Ejima, and K Sawada Comparison of image validity between cone beam computed tomography for dental use and multidetector row helical computed tomography Dentomaxillofac. Radiol., December 1, 2007; 36(8): 465 - 471. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Kramer, M. J. Budoff, Z. A. Fayad, V. A. Ferrari, C. Goldman, J. R. Lesser, E. T. Martin, S. Rajagopalan, J. P. Reilly, G. P. Rodgers, et al. ACCF/AHA 2007 Clinical Competence Statement on vascular imaging with computed tomography and magnetic resonance Vascular Medicine, November 1, 2007; 12(4): 359 - 378. [PDF] |
||||
![]() |
C. M. Kramer, M. J. Budoff, Z. A. Fayad, V. A. Ferrari, C. Goldman, J. R. Lesser, E. T. Martin, S. Rajagopalan, J. P. Reilly, G. P. Rodgers, et al. ACCF/AHA 2007 Clinical Competence Statement on Vascular Imaging With Computed Tomography and Magnetic Resonance: A Report of the American College of Cardiology Foundation/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training Developed in Collaboration With the Society of Atherosclerosis Imaging and Prevention, the Society for Cardiovascular Angiography and Interventions, the Society of Cardiovascular Computed Tomography, the Society for Cardiovascular Magnetic Resonance, and the Society for Vascular Medicine and Biology J. Am. Coll. Cardiol., September 11, 2007; 50(11): 1097 - 1114. [Full Text] [PDF] |
||||
![]() |
L. J. M. Kroft, A. de Roos, and J. Geleijns Artifacts in ECG-Synchronized MDCT Coronary Angiography Am. J. Roentgenol., September 1, 2007; 189(3): 581 - 591. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Mahesh and D. D. Cody AAPM/RSNA Physics Tutorial for Residents: Physics of Cardiac Imaging with Multiple-Row Detector CT RadioGraphics, September 1, 2007; 27(5): 1495 - 1509. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Hur, S. W. Hong, S. Y. Kim, Y. H. Kim, Y. J. Hwang, W. R. Lee, and S. J. Cha Uniform Image Quality Achieved by Tube Current Modulation Using SD of Attenuation in Coronary CT Angiography Am. J. Roentgenol., July 1, 2007; 189(1): 188 - 196. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. J. Schoepf, P. L. Zwerner, G. Savino, C. Herzog, J. M. Kerl, and P. Costello Coronary CT Angiography Radiology, July 1, 2007; 244(1): 48 - 63. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Herzog, S. A. Nguyen, G. Savino, P. L. Zwerner, J. Doll, C. D. Nielsen, T. G. Flohr, T. J. Vogl, P. Costello, and U. J. Schoepf Does Two-Segment Image Reconstruction at 64-Section CT Coronary Angiography Improve Image Quality and Diagnostic Accuracy? Radiology, July 1, 2007; 244(1): 121 - 129. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ferencik, D. Ropers, S. Abbara, R. C. Cury, U. Hoffmann, K. Nieman, T. J. Brady, F. Moselewski, W. G. Daniel, and S. Achenbach Diagnostic Accuracy of Image Postprocessing Methods for the Detection of Coronary Artery Stenoses by Using Multidetector CT Radiology, June 1, 2007; 243(3): 696 - 702. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Primak, C. H. McCollough, M. R. Bruesewitz, J. Zhang, and J. G. Fletcher Relationship between Noise, Dose, and Pitch in Cardiac Multi-Detector Row CT RadioGraphics, November 1, 2006; 26(6): 1785 - 1794. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Jagia, K. C. Goswami, S. Sharma, and G. S. Gulati 16-MDCT in the evaluation of coronary cameral fistula. Am. J. Roentgenol., August 1, 2006; 187(2): W227 - W228. [Full Text] [PDF] |
||||
![]() |
T Flohr and B Ohnesorge Developments in CT Imaging, June 1, 2006; 18(2): 45 - 61. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. B. Wolbarst and W. R. Hendee Evolving and Experimental Technologies in Medical Imaging Radiology, January 1, 2006; 238(1): 16 - 39. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |