|
|
||||||||
Technical Developments |
1 From the Division of Abdominal Imaging and Intervention, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, 55 Fruit St, White 270-E, Boston, MA 02114 (T.D., S.M.R.R., M.A.B.); Siemens Medical Solutions, Forchheim, Germany (B.S., C.S., T.F.); and Department of Radiology, Emory University School of Medicine, Atlanta, Ga (S.S., M.K.K.). Supported in part by RSNA R&E Medical Student Departmental Program Grant and University of Vermont, Burlington, Vt. S.M.R.R. supported in part by research fellowship grant from Siemens Medical Solutions. Received September 16, 2004; revision requested November 24; revision received December 6; accepted January 12, 2005. Address correspondence to M.K.K. (e-mail: mkalra{at}emory.edu).
| ABSTRACT |
|---|
|
|
|---|
© RSNA, 2005
| INTRODUCTION |
|---|
|
|
|---|
Since automatic tube current modulation adapts tube current, on the basis of the regional density and attenuation profile, data from some researchers (5) indicate that tube current associated with scanning in patients with a metallic prosthesis is higher in the area of the prosthesis; however, there is an overall reduction in radiation dose for the entire scan length in these patients. Results of prior studies also indicated that an increase in the tube current does not lead to a decrease in streak artifacts from prostheses (6). Therefore, a technique was developed to detect the presence of metallic prostheses from the localizer radiograph before actual scanning of the patient and to avoid an increase in tube current by excluding the contribution of the metallic prosthesis to the overall attenuation profile during the calculation of the appropriate tube current to be used. This technique is a standard component of the current combined modulation technique investigated in this study. Thus, the purpose of our study was to retrospectively evaluate the combined automatic tube current modulation technique in patients with orthopedic metallic prostheses.
| Materials and Methods |
|---|
|
|
|---|
Phantom Study
An elliptical Plexiglas container (32 x 20 cm in transverse and anterior-posterior diameters) was filled with water and was scanned with a 16-section multidetector row CT scanner (Somatom Sensation 16; Siemens Medical Solutions, Forchheim, Germany) with a fixed tube current and a combined-modulation technique (CARE Dose4D; Siemens Medical Solutions). The phantom was first scanned with a fixed effective tube currenttime product of 200 (effective tube currenttime product = [tube current x gantry rotation time]/pitch). Subsequently, the phantom was scanned with the combined-modulation technique with an image quality reference tube currenttime product of 200. The remaining scanning parameters were held constant and included 140 kVp, 0.5-second gantry rotation time, 16 x 1.5 mm detector configuration (16 detector data channels of 1.5-mm section thickness each), 24-mm table feed per gantry rotation, soft-tissue reconstruction algorithm (B31 medium), 5-mm reconstructed section thickness, and 5-mm section intervals.
A cobalt-chrome metallic hip endoprosthesis (Omnifit EON; Stryker Howmedical Osteonics, Mahwah, NJ) with a titanium acetabular cup (Trilogy; Zimmer, Warsaw, Ind), which is used for total hip arthroplasty, was then placed in the phantom to determine the effect of the metallic prosthesis on the combined-modulation technique. The prosthesis was attached to the dependent surface of the phantom with adhesive tape. The phantom was placed in the same position in the gantry isocenter and then rescanned by using identical fixedtube current techniques and combined-modulation techniques. Thus, scanning was performed with fixedtube current techniques and combined-modulation techniques for the phantom without the metallic prosthesis, and then it was repeated for the phantom with the metallic prosthesis.
Two radiologists (M.A.B., with 10 years of experience, and M.K.K., with 5 years of experience), unaware of the scanning technique, independently compared the extent and severity of streak artifacts on images acquired with fixedtube current techniques and combined-modulation techniques. All images were reviewed at a digital picture archiving and communication system diagnostic workstation (Impax RS 3000 1K; Agfa Technical Imaging Systems, Ritchfield Park, NJ) at the same window level and window width (40 and 400 HU, respectively). A three-point scale was used to grade the severity of streak artifacts from the metallic prosthesis (grade 1, minimum streak artifacts; grade 2, moderate streak artifacts; and grade 3, severe streak artifacts).
For each series of images, the CT dose index volume and dose-length product values were recorded from the user interface on the scanner. The CT dose index volume is a descriptor of the average dose within a scan volume (relative to a standardized CT phantom), and it is now required to be displayed on the user interface of the CT scanner. In addition, effective tube currenttime product values used with the scanner were recorded for all images obtained with the combined-modulation technique.
Patient Study
The human research committee of the institutional review board approved the Health Insurance Portability and Accountability Actcompliant study protocol, with waiver of informed consent. Five hundred consecutive contrast materialenhanced abdominal-pelvic CT examinations, performed from April to September 2004 with the same 16-section multidetector row CT scanner as was used for the phantom study with the combined-modulation technique, were retrospectively reviewed to identify patients with metallic prostheses (T.D., S.M.R.R.). All examinations were reviewed at a picture archiving and communication system diagnostic workstation. Nine patients (mean age, 66 years; range, 3586 years; male-female ratio, 5:4) with metallic prostheses served as the study patients. Six patients had a right hip prosthesis, two patients had a left hip prosthesis, and one patient had bilateral hip prostheses. The standard departmental protocol for the combined-modulation technique used for scanning study patients included an image quality reference tube currenttime product of 160. The remaining scanning parameters included 0.5-second gantry rotation time, 140 kVp, 16 x 1.5 mm detector configuration, 24-mm table feed per gantry rotation, soft-tissue reconstruction algorithm (B31 medium), 5-mm reconstructed section thickness, and 5-mm section intervals. The cross-sectional transverse diameter was measured at the upper pole of the right kidney for all study patients (mean, 333 mm; range, 290370 mm).
Nine additional patients (mean age, 56 years; range, 3672 years; male-female ratio, 4:5) with no metallic prosthesis who underwent abdominal-pelvic CT scanning with an identical combined-modulation technique served as control patients. The control group included nine consecutive patients identified in a retrospective review of 77 consecutive contrast-enhanced abdominal-pelvic CT examinations performed in May 2004. All patients without a metallic prosthesis (control patients) were matched for age and cross-sectional transverse dimension at the level of the upper pole of the right kidney with patients who had metallic prostheses (study patients). The average cross-sectional transverse dimension at the level of the upper pole of the right kidney for control patients was 331 mm (range, 281412 mm).
Two radiologists (M.A.B., with 10 years of experience, and M.K.K., with 5 years of experience), who were unaware of the scanning technique, independently evaluated CT images of the patients with metallic prostheses that were performed by using the combined-modulation technique. Images were graded for the severity and extent of streak artifacts from metallic prostheses at a digital picture archiving and communication system diagnostic workstation with the same window level and window width of 40 and 400 HU, respectively. Severity and extent of streak artifacts from the metallic prostheses were graded on a three-point scale (grade 1, streak artifact present but not substantially compromising the evaluation of adjacent structures; grade 2, streak artifact present and slightly compromising the evaluation of adjacent structures; and grade 3, streak artifact present and severely compromising the evaluation of adjacent structures).
To evaluate the effect of metallic prostheses on radiation dose associated with the combined-modulation technique, effective tube currenttime product values used for scanning the patients with metallic prostheses and those without metallic prostheses were recorded at all section positions.
Statistical Analysis
The
2 test, which was calculated with a software program (MedCalc Software; MedCalc, Mariakerke, Belgium), was used to compare CT dose index volume and dose-length product values associated with scanning of the phantom with the metallic prosthesis and of the phantom without the metallic prosthesis. Effective tube currenttime product values for CT images of the phantom with the metallic prosthesis and for those of the phantom without the metallic prosthesis acquired with the combined-modulation technique were compared by using the paired t test with the software program (MedCalc Software; MedCalc). The Student t test also was used to compare the ages and cross-sectional transverse dimensions of patients in the study and control groups with spreadsheet software (Excel; Microsoft, Redmond, Wash). Qualitative scores for the severity of metallic streak artifacts were compared with the Wilcoxon signed rank test, which was determined with a software program (MedCalc Software; MedCalc). The degree of interobserver agreement was determined with the
test by using the software program. The
coefficient values for interobserver agreement were considered as slight (
0.20), fair (0.210.40), moderate (0.410.60), substantial (0.610.80), or almost perfect (0.811.00). The mean values for effective tube currenttime product for study and control patients also were compared by using the Student t test. A P value of less than .05 was considered to indicate a statistically significant difference.
| Results |
|---|
|
|
|---|
coefficient = 1.00, P < .05).
As expected, CT dose index volume and dose-length product values for fixedtube current CT scanning of the phantom with the metallic prosthesis and of the phantom without the metallic prosthesis were identical to each other (20.4 mGy · cm and 325 mGy · cm, respectively). With the combined-modulation technique, CT dose index volume and dose-length product values for the phantom without the metallic prosthesis were 7.8 mGy · cm and 125 mGy · cm, respectively. Likewise, CT dose index volume and dose-length product for CT scanning of the phantom with the metallic prosthesis, performed by using the combined-modulation technique, were 7.6 mGy · cm and 121 mGy · cm, respectively. The
2 test did not reveal a statistically significant difference between CT dose index volume or dose-length product values for scanning of the phantom with the metallic prosthesis and for that of the phantom without the metallic prosthesis performed by using the combined-modulation technique (P = .8). Similarly, there was no statistically significant difference between effective tube currenttime product values for the phantom with the metallic prosthesis and those for the phantom without the metallic prosthesis scanned with the combined-modulation technique (P = .2) (Fig 1).
|
coefficient = 1.00, P < .05).
|
|
| Discussion |
|---|
|
|
|---|
An attenuation profile along the patient's long axis is first estimated from the localizer radiograph in the direction of the projection with a mathematic algorithm. This attenuation profile contains information about the patient's size, anatomic shape, and attenuation at each position in the z-axis. To obtain transverse images, tube-current settings are calculated on the basis of these attenuation profiles. The correlation between attenuation profile and tube current for each image is defined by an analytic function, which adapts the effective tube currenttime product to the patient's size, anatomic shape, and attenuation on the basis of a user-defined image quality reference tube currenttime product, to maintain the desired diagnostic image quality along the scanning direction. The technique performs this tube current modulation online during each tube rotation, with tube current values adapting to the changing angular attenuation profiles of the patient.
The image quality reference tube currenttime product value is selected according to the diagnostic requirements of the study and the individual preference of the radiologist. For a given scanning protocol, this value reflects the mean effective tube currenttime product, which is used for a reference patient defined as a typical adult who weighs 70 to 80 kg (for adult protocols). Since the combined-modulation technique adapts the tube current to the individual patient size on the basis of the image quality reference tube currenttime product values, the image quality reference tube currenttime product is changed only if an adjustment for image quality is required.
Metallic prostheses generate starburst or streak artifacts that can substantially degrade CT image quality and make it hard for the radiologist to evaluate adjacent structures, which is especially crucial when assessment is required for potential surgical revision. These prostheses block incident x-ray beams, which lead to missing projection data and artifacts on the reconstructed CT images (7). The severity of streak artifacts depends on the composition of the metallic prosthesis used. Cobalt-chrome alloy and stainless steel are known to cause more artifacts than do titanium prostheses (8,9). Researchers in prior studies showed that tube current and tube potential (peak voltage) do not affect streak artifacts from metallic prostheses (6). Therefore, it is important to detect and allow for metallic prostheses when scanning with the automatic tube current modulation technique so that patients do not receive higher radiation doses in those regions with a metallic prosthesis. To the best of our knowledge, no investigators in a peer-reviewed study to date have reported automatic detection of a metallic prosthesis from the localizer radiograph and its exclusion from estimation of tube current used for scanning with automatic tube current modulation techniques.
The algorithm was incorporated into the automatic tube current modulation technique assessed in our study. The algorithm is used to estimate any abrupt increase in attenuation numbers caused by metallic structures from the localizer radiograph and neglects these numbers for calculation of the optimal tube current settings. Thus, regions with very high attenuation values, such as those with a metallic prosthesis, do not contribute to the estimation of tube current settings in the combined-modulation technique. The algorithm is important for managing radiation dose associated with automatic tube current modulation techniques, as investigators in previous studies have reported no improvement in image quality or reduction in streak artifacts with an increase in tube current for scanning of body regions with metallic prostheses or devices (6).
Our study shows that use of the algorithm with the combined-modulation technique does not increase the tube current and radiation dose when the area with the metallic prosthesis is scanned. Findings of the phantom study corroborated the results of the patient study and indicated that there was no increase in effective tube currenttime product when the combined-modulation technique was used for scanning regions with a metallic prosthesis. Moreover, there was no difference in the severity and extent of streak artifacts from the metallic prosthesis between images acquired with combined-modulation techniques and those acquired with fixedtube current techniques. Therefore, our results with this algorithm are in contradistinction to those of the prior report of increased patient dose in the region of the metallic prosthesis with a prior version of automatic tube current modulation from a different vendor (5). Successful validation of the technique assessed in our study is important, as investigators in other studies have documented that combined-modulation techniques substantially help in the reduction of radiation dose compared with results with the more commonly used fixedtube current technique (1012). A 20%60% dose reduction, depending on the anatomic region and patient habitus, with improved image quality has been reported with the combined-modulation technique assessed in the present study (10,11). Researchers in another study with the combined-modulation technique from a different vendor (3D Auto mA; GE Yokogowa Medical Systems, Tokyo, Japan) also reported dose reductions of 60% in abdominal-pelvic CT examinations (12). Therefore, incorporation of the technique assessed in our study helps in the maintenance of this desirable radiation dose reduction with the combined-modulation technique in the presence of metallic prostheses.
There were limitations in our study. We did not perform a power analysis to determine how many subjects should have been included in the patient and control groups, since there were no published studies in the medical literature in which the researchers looked at the effect of metallic prostheses on the combined-modulation technique. In addition, we did not examine the effect of the composition of metallic prostheses (titanium, cobalt-chrome, or stainless steel) on radiation dose associated with the combined-modulation technique. Findings in all studies were consecutively reviewed, however, and patients with metallic prostheses were selected without regard for the composition of the metallic prosthesis. The phantom used in our study was a simplified model and did not have surrounding structures, such as those that would represent retroperitoneal fat, visceral structures, and vertebrae, to simulate cross-sectional anatomy. The findings of the phantom experiment, however, were confirmed with those of the patient study, which showed that there was no increase in effective tube currenttime product or radiation dose with the use of this combined-modulation technique in the presence of metallic prostheses and that image quality was not significantly changed.
In summary, results of our study indicate that, with the combined-modulation technique, we successfully avoided an increase in radiation dose caused by the high attenuation of the metallic prosthesis in both the phantom and patients who underwent CT. Moreover, there was no substantial difference in the extent and severity of streak artifacts on images acquired with this type of combined-modulation or fixedtube current technique.
| FOOTNOTES |
|---|
See Materials and Methods for pertinent disclosures
Author contributions: Guarantor of integrity of entire study, M.K.K.; study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting or manuscript revision for important intellectual content, all authors; approval of final version of submitted manuscript, all authors; literature research, M.K.K., S.M.R.R.; clinical studies, T.D., M.K.K., S.M.R.R., S.S.; experimental studies, T.D., M.K.K., S.M.R.R., B.S., C.S., T.F., S.S.; statistical analysis, M.K.K., S.M.R.R.; and manuscript editing, M.K.K., S.M.R.R., B.S., M.A.B., S.S.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. A. Leswick, M. M. Hunt, S. T. Webster, and D. A. Fladeland Thyroid Shields versus z-Axis Automatic Tube Current Modulation for Dose Reduction at Neck CT Radiology, November 1, 2008; 249(2): 572 - 580. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |