|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Breast Imaging |
1 From the Departments of Radiology (C.K.K., P.J., N.M., H.H.S., J.G.) and Gynecology (O.Z.), University of Bonn, Sigmund-Freud-Str 25, 53105 Bonn, Germany, and Philips Medical Systems, Best, the Netherlands (J.G.). From the 2004 RSNA Annual Meeting. Received March 27, 2005; revision requested May 24; revision received May 26; accepted June 21; final version accepted July 26. Address correspondence to C.K.K. (e-mail: kuhl{at}uni-bonn.de).
| ABSTRACT |
|---|
|
|
|---|
Materials and Methods: A prospective intraindividual internal review boardapproved study was performed in 37 women with 53 lesions (25 breast cancers, 28 benign focal lesions) who underwent contrast-enhanced dynamic bilateral subtraction MR imaging twice, once at 1.5 T with a standard technique (voxel size, 1.44 mm3) and once at 3.0 T (voxel size, 0.450.72 mm3) with variable repetition time and flip angle settings. Written informed consent was obtained. Sagittal single breast high-spatial-resolution MR imaging was performed with active fat suppression. Image quality, number and features of enhancing lesions, and Breast Imaging Reporting and Data System categories were compared by using the Wilcoxon matched-pairs signed rank test and Student t test for matched pairs. Diagnostic confidence was compared by using a receiver operating characteristic (ROC) analysis.
Results: With repetition time prolonged to account for longer T1 relaxation times at 3.0 T and a flip angle of 60°, enhancement rates at 3.0 T were substantially below those at 1.5 T. In two patients with benign lesions, enhancement was rated as insufficient to establish diagnosis. When parameter settings were kept equivalent, equivalent enhancement rates were observed with both systems. With these settings, 3.0-T MR imaging yielded homogeneous signal intensity over the entire field of view. No dielectric resonance effects were observed. Overall image quality scores for the dynamic series were slightly higher at 3.0 T (P < .01). A total of 49 lesions were prospectively identified with both systems. Owing to substantial patient motion at 1.5 T, two malignant lesions in one patient were visualized at 3.0 T only. At 3.0 T, differential diagnosis of enhancing lesions was possible with higher diagnostic confidence, as reflected by a larger area under the ROC curve (P < .05).
Conclusion: Initial experiences indicate that contrast-enhanced MR imaging at 3.0 T is nearing readiness for clinical use.
© RSNA, 2006
| INTRODUCTION |
|---|
|
|
|---|
MR systems operating at higher magnetic field strengths (eg, 3.0 T) offer a higher SNRpossibly enough to acquire high imaging matrices with only one signal average or to allow fast acquisition strategies (17,18). Higher magnetic field strengths, however, are associated with physical effects that may not be advantageous for breast imaging; stronger susceptibility effects, longer T1 relaxation times, shorter in-phase echo times, and higher radiofrequency (RF) deposition may impair image quality and complicate or even impede image generation (1923).
Thus, we conducted our study to establish a pulse sequence for dynamic contrast materialenhanced MR imaging of the breast at 3.0 T and to prospectively compare MR imaging at 3.0 T with MR imaging at 1.5 T in the same patients.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Study Design and Patients
The investigation was conducted as a prospective intraindividual comparative study on consecutive patients who underwent contrast-enhanced MR imaging at 1.5 T for clinical reasons and who fulfilled the following inclusion criteria: 18 years of age or older, clinical indications for contrast-enhanced MR imaging of the breast at 1.5 T, no contraindication for MR imaging in general (no contraindications specific to high-field-strength imaging), proof of contrast-enhanced lesion on 1.5-T MR images, and willingness to volunteer for a second contrast-enhanced MR imaging examination performed on a separate day with the 3.0-T system.
A total 37 women (age range, 3369 years; mean age, 50 years ± 9.3 [standard deviation]) met the inclusion criteria and volunteered to participate in the study. Patients were referred for contrast-enhanced MR imaging for preoperative staging of core biopsyproved breast cancer or mammographic BI-RADS category 5 lesions (15 cases). Nine women were referred to clarify equivocal mammographic and/or ultrasonographic (US) findings, six were referred for familial breast cancer screening, five were referred for follow-up after conservation treatment, and two were referred for other reasons. All patients were informed in detail about the scientific purpose of the second imaging examination at high field strength, and written informed consent was obtained.
Care was taken to schedule the second (ie, the high-field-strength) examination after a minimum time period of at least 24 hours to allow for complete contrast agent elimination. After this waiting period had elapsed, the examination was performed as immediately as possible. In premenopausal women, specific care was taken to perform the second examination during the same week of the patient's menstrual cycle. The mean time interval between the examination at 1.5 T and the examination at 3.0 T was 2.4 days, with a median of 1 day. A total 21 patients were examined after 1 day, six after 2 days, three after 3 days, six after 46 days, and one after 9 days.
MR Imaging
Examinations were performed by using 1.5- and 3.0-T whole-body MR imaging systems (Intera 1.5 T and Intera 3.0 T; Philips Medical Systems, Best, the Netherlands). Both systems are equipped with high-performance gradients (master gradients), with a maximum slew rate of 150 (mT · m1)/msec and a maximum gradient strength of 30 mT/m. Dedicated four-element SENSE-compatible breast surface coils (MRI Devices; InVivo Research, Orlando, Fla) were used with both systems. No breast fixation was used with either MR system.
For 1.5-T MR imaging, our standard bilateral dynamic contrast-enhanced protocol, which has been in routine clinical use for many years, was employed (11). This protocol consists of a fast localizer (scout view) gradient-echo sequence, with 25 sections (transverse, coronal, and sagittal) acquired separately for the left and right breasts; this sequence was used to prescribe sections of the subsequent dynamic series to cover the volume of fibroglandular tissue of both breasts exactly. The dynamic series consisted of a T1-weighted two-dimensional (2D) gradient-echo pulse sequence (290/4.6 [repetition time msec/echo time msec], flip angle of 90°), with a total of five dynamic acquisitionsone obtained before and four obtained immediately after a bolus injection of 0.1 mmol per kilogram body weight gadolinium dimeglumine (Magnevist; Schering, Berlin, Germany) at an injection rate of 3 mL/sec. Each dynamic volume consisted of 33 sections (3 mm thick), with an acquisition matrix of 512 x 394 (77% of a 512 matrix) and a field of view of 300330 mm (adjusted to the size of the breasts). With this parameter setting, spatial resolution was 0.6 x 0.8 x 3 mm (noninterpolated), and temporal resolution was 1 minute 50 seconds per dynamic acquisition. To suppress the signal of fat, image subtraction was performed off-line after the actual imaging session.
For 3.0-T MR imaging, the pulse sequence for the dynamic series had to be adjusted to account for the physical effects of the higher magnetic field strength (Table). Our premise was to keep the spatial and temporal resolution equivalent to (or better than) those obtained with the standard 1.5-T setting. The following modifications were made for all pulse sequences at 3.0 T:
|
To identify the appropriate pulse sequence parameters that would best demonstrate enhancing lesions (ie, gadolinium-induced T1-shortening effect) at 3.0 T, we used stepwise variations of repetition time and flip angle in the dynamic series. By this we sought to systematically investigate the influence of these parameters on the enhancement pattern of lesions at 3.0 T, with the enhancement pattern of each lesion at 1.5 T serving as the reference for each patient. The exact pulse sequence parameters regarding repetition time and flip angle are given in the Table.
To account for the fact that T1 relaxation times are prolonged by about 30% at 3.0 T and that RF deposition scales with the degree of flip angle, we began with a longer repetition time at 3.0 T compared with our standard repetition time at 1.5 T and reduced the flip angle to 60°. From there, repetition time was progressively reduced and combined with a flip angle of 60° or 73°, the latter being the highest flip angle that was technically achievable for the given in-phase echo time and SAR limitations. A total of seven different parameter groups with distinct parameter settings were defined as follows: Group 1 (two patients) was investigated with a repetition time of 430 msec and a flip angle of 60°; group 2 (three patients), with a repetition time of 400 msec and a flip angle of 60°; group 3 (three patients), with a repetition time of 300380 msec and a flip angle of 60°; group 4 (three patients), with a repetition time of 300380 msec and a flip angle of 73°; group 5 (six patients), with a repetition time of 260290 msec and a flip angle of 60°; group 6 (13 patients), with a repetition time of 260290 msec and a flip angle of 73°; and group 7 (seven patients), with a repetition time of less than 260 msec and a flip angle of 73°.
Because the parameter setting that was used in the first two patients (group 1) turned out to be inappropriate (see Results), these examinations were not used for further analysis of image quality or for image interpretation. Accordingly, the analysis of image quality and diagnostic confidence was based on data from 35 patients in groups 27 only.
In addition to and after the dynamic series, we acquired sagittal T1-weighted radial-segmented three-dimensional gradient-echo MR images at 3.0 T, with active fat suppression and high spatial resolution (9/2.8, 25° flip angle, 80 sections acquired, 12-mm section thickness, 250-mm field of view, active fat suppression, 0.5 x 0.5 x 1.0-mm pixel size [noninterpolated], and total acquisition time of 90 seconds). These images were not used for the direct comparison of image quality because no such images were obtained at 1.5 T. They were, however, made available for clinical image interpretation.
Data Analysis
Image interpretation and final diagnosis.Images obtained at 3.0 T and those obtained at 1.5 T for the 35 patients in groups 27 were transferred to hardcopy with standardized window settings. Immediately after the examinations, image interpretation was performed prospectively by different radiologists who had substantial expertise in interpreting MR images of the breast (C.K.K. and N.M., with 11 and 6 years of experience, respectively) and who had each read half of the 1.5-T MR images and half of the 3.0-T MR images. The readers noted the number of lesions and assigned a BI-RADS category for each lesion. The same interpretation criteria were used for the 1.5- and the 3.0-T examinations (see below).
Readers were blinded to the findings and final BI-RADS categories given for the same patient at the other field strength. To avoid interreader variability confounding the results, a conjoint reading of images from both examinations was performed for cases in which the lesions were assigned divergent BI-RADS categories. Conventional images, as well as mammograms, US images, and reports, were made available during the reading session just as they would be in the clinical setting. Final management was decided on the basis of results from the 1.5-T examination. For cases in which higher BI-RADS categories were assigned at 3.0 T, the results were carefully discussed with the patient and her surgeon. Final management was decided only after a repeat explanation of the experimental character of the 3.0-T examination was provided.
Masses that had an irregular lesion shape, borders or spicules, and heterogeneous or rim enhancement or non-masslike enhancement, with asymmetric and segmental or ductal configuration, were considered suspicious for malignancy. Rapid enhancement was used to corroborate suspicious findings, except for lesions that appeared morphologically benign and had persistent enhancement in the entire dynamic series. If rim enhancement was identified within a focal mass, the lesion was considered suspicious, irrespective of other findings. If dark septations were present within an oval mass with smooth borders, the lesion was rated as definitely benign (fibroadenoma), irrespective of other findings. Slow, persistent enhancement was considered to corroborate the diagnosis of a benign lesion, except for non-masslike enhancement with a ductal or segmental configuration or for slow, persistent enhancement in masses with morphologic findings suggestive of malignancy (irregular borders or spicules and/or rim enhancement). If washout was observed, the lesion was considered suspicious, except in cases of well-circumscribed lesions with dark septations or small (<5 mm) non-masslike foci of enhancement.
The final diagnosis was established by means of excisional or core biopsy in 20 patients with 31 lesions and by means of follow-up of at least 12 months, including mammography and MR imaging, in 15 patients with 20 lesions.
Image quality.For the prospective analysis of image quality, all images were transferred to a workstation and were displayed with standardized window settings. Images were displayed in a pairwise fashion to allow direct comparison of 1.5- and 3.0-T MR images. No attempt was made to blind the reader to the field strength with which the respective images were acquired because it was obvious from the appearance of the images obtained with SENSE at 3.0 T. A breast radiologist with 11 years of experience in reading MR images (C.K.K.) was asked to rate the image quality by visual assessment. Only the images of the dynamic series were used for comparison; the high-spatial-resolution images obtained before and after the dynamic series at 3.0 T (see above) were considered separately because they were obtained with an entirely different acquisition technique that would not allow fair (or meaningful) comparison between 1.5- and 3.0-T MR images.
Image quality was rated on a scale of 1 (nondiagnostic image quality) to 5 (excellent) on the basis of the following criteria:
A rating of 1 (nondiagnostic image quality) was assigned for cases in which the image quality was considered insufficient for diagnosis because of insufficient signal intensity homogeneity or massive dielectric resonance effects across the field of view, with a complete loss of signal intensity in parts of the field of view owing to severe artifacts (complete image degradation by susceptibility or pulsation) and/or poor visual SNR. A dielectric resonance effect was considered to be present if concentric signal intensity shadowing was noted in the center of the image. A susceptibility artifact was considered to be present if image blurring and image distortions were noted at tissue interfaces.
A rating of 2 (poor image quality) was assigned for cases in which there was insufficient signal intensity homogeneity or dielectric resonance effects, with substantial yet incomplete signal intensity variations across the field of view and/or substantial but incomplete image degradation due to susceptibility or pulsation and/or a low visual SNR.
A rating of 3 (acceptable) was assigned for cases in which there were only slight inhomogeneities in signal intensity or dielectric resonance effects across the entire field of view and/or moderate susceptibility or pulsation artifacts and a high visual SNR.
A rating of 4 (good) was assigned for cases in which there were mild heterogeneity changes in signal intensity across the field of view (eg, slight left-to-right differences in signal intensity), no dielectric resonance effects, mild susceptibility or pulsation artifacts, and high visual SNR.
A rating of 5 (excellent) was assigned for cases in which there were no or hardly perceivable signal intensity variations across the field of view, no dielectric resonance effects, no or only mild susceptibility or pulsation artifacts, and high visual SNR.
Quantitative assessment of enhancement rates.A breast radiologist (C.K.K.) placed a region of interest in all lesions that were identified during hardcopy reading. The size of the region of interest varied with the size of the enhancing lesion and was chosen to selectively include the area with the strongest enhancement, as identified on the first postcontrast subtracted image. The average size of the region of interest was 3 mm2. The enhancement rates (percentage signal intensity increase) of the lesions were calculated by using the following equation: [(SIpost SIpre)/SIpre] · 100, where SIpre is the signal intensity before contrast injection (first dynamic acquisition) and SIpost is the signal intensity after contrast injection (first postcontrast dynamic acquisition). The enhancement rates of each lesion at 3.0 T were compared with those of each lesion at 1.5 T.
Relative enhancement rates (RERs) (ie, the enhancement rate of a lesion at 3.0 T vs the enhancement rate of the same lesion at 1.5 T) were calculated as ER3.0T/ER1.5T, where ER3.0T is the enhancement rate measured at 3.0 T and ER1.5T is the enhancement rate measured at 1.5 T. A RER of 1 would correspond to equivalent enhancement rates at 3.0 and 1.5 T, whereas a RER of 0.5 would indicate that the enhancement rate at 3.0 T was 50% of the enhancement rate at 1.5 T.
RERs were compared between the seven parameter groups to identify the pulse sequence parameters (specifically regarding repetition time and flip angle) that would yield the strongest enhancement at 3.0 T. For this purpose, all the quantitative enhancement rates of lesions that belonged to the same parameter group (ie, those lesions that were imaged with the same parameter settings) were averaged. The resulting mean values were compared to determine the parameter setting that would yield the highest RER (3.0 vs 1.5 T).
Statistical Analysis
The image quality scores of the dynamic contrast-enhanced series were compared for both field strengths; the Wilcoxon matched-pairs signed rank test was used to test for statistical significance. Enhancement rates of lesions at 3.0 T were compared with those of the same lesions at 1.5 T; the Student t test for matched pairs was used to test for statistical significance. To account for several observations in the same patient (ie, in patients with more than one enhancing lesion), a mean was calculated for the different enhancement rates of the different lesions in each patient, and this mean value alone was used for further analysis. The same method was applied for BI-RADS categories in patients with several lesions in one breast. In these patients, lesion classes were defined for the purpose of the receiver operating characteristic (ROC) analysis. A lesion class was used for further analysis if several lesions or if a cluster of lesions with identical appearance were present on the MR image. A P value of less than .05 was considered to indicate a statistically significant difference. Only malignant lesions were evaluated in patients with breast cancers.
To investigate the differential diagnostic power offered by MR imaging at 1.5 and 3.0 T, an ROC analysis was performed (ROCKIT, version 0.9b; C. E. Metz, University of Chicago, Ill), and the areas under the ROC curves were compared for both examinations.
| RESULTS |
|---|
|
|
|---|
The 3.0-T MR images obtained in the two patients from group 1 who had two benign lesions (fibroadenomas) were excluded from analysis because of insufficient enhancement. Accordingly, the analysis of image quality and diagnostic confidence was based on data from 35 patients with a total of 51 enhancing lesions. For the identification of an appropriate imaging protocol, however, data from all 53 lesions in 27 patients were included for analysis.
Lesion Detection and Classification
A total 49 lesions were prospectively identified at 1.5-T MR imaging. These lesions were also prospectively and independently identified at 3.0-T MR imaging. Two additional lesions were diagnosed at 3.0-T MR imaging in a 54-year-old woman who presented for preoperative staging of a biopsy-proved breast cancer in her right breast. During the 1.5-T MR imaging examination, the images of the left breast had been substantially degraded owing to considerable patient motion. At 3.0 T, no motion occurred, and a suspicious mass with adjacent non-masslike linear enhancement was identified in the contralateral left breast. The lesions were scored as BI-RADS category 5 (highly suggestive of breast cancer and ductal carcinoma in situ) at 3.0 T. Preoperative MR-guided localization was performed at 1.5 T; this time, images were obtained without patient motion, and the lesion was clearly visualized. Histologic analysis revealed a ductulolobular invasive cancer (pT1b, TNM classification) with adjacent intraductal cancer.
BI-RADS categories, which were given per lesion and not per breast, were identical in 40 (78%) of 51 lesions. BI-RADS categories differed in 10 patients with a total of 11 different lesions or classes of lesions. In addition to the discrepancies noted in the one patient with two contralateral breast cancers mentioned above, the following discrepancies regarding BI-RADS categories were observed: In one patient (Fig 1) who presented for preoperative staging of a biopsy-proved duct invasive breast cancer, multicentric disease had already been identified at 1.5-T MR imaging. During the 3.0-T MR imaging examination, two additional invasive foci were prospectively identified in a third quadrant and were categorized as BI-RADS category 4 lesions, whereas these foci had been categorized as BI-RADS category 2 lesions at 1.5-T MR imaging. The demonstration of these two additional lesions (which were grouped as a single class of lesions for analysis) did not change management because multicentric disease had already been diagnosed in this patient at 1.5-T MR imaging.
|
|
|
|
Image Quality
Image quality scores for dynamic 2D gradient-echo MR images obtained with both field strengths differed slightly, with a mean score of 4.6 ± 0.5 for MR images obtained at 3.0 T and 4.2 ± 0.5 for those obtained at 1.5 T. This difference proved to be statistically significant (P < .01) by using the Wilcoxon matched-pairs signed rank test. The median image quality score was 5 (excellent) for images obtained at 3.0 T and 4 (good) for those obtained at 1.5 T. In detail, image quality scores for the dynamic series yielded equivalent values in 13 (37%) of 35 patients; a lower score (by one point) was given at 3.0 T in three (9%) of 35 patients, and a higher score (by one point) was given at 3.0 T in 19 (54%) of 35 patients. None of the image quality ratings differed by more than one point. Among the 10 patients in whom the acquisition matrix was increased to 1024 x 680, seven had images of excellent quality that were superior to those obtained at 1.5 T (Fig 3). In three patients, however, blurring and double contours of enhancing lesions on the subtracted images degraded image quality. This was due to the fact that, with very small pixel sizes, the already subtle motion of the patients encompassed a full pixel thickness and therefore produced subtraction artifacts. On none of the 3.0-T MR images were artifacts resulting from the use of parallel imaging with SENSE observed.
|
|
|
|
|
|
|
|
The mean RER for all lesions was 0.89 ± 0.30, indicating that (on average) lesion enhancement at 3.0 T was less than 90% of lesion enhancement at 1.5 T.
Analysis of the influence of variable repetition times and flip angles (Fig 5) showed a correlation between RER, repetition time, and flip angle, with RER decreasing as flip angle decreased and repetition time increased. Enhancement rates at 3.0 T and those at 1.5 T were almost equivalent (ie, RER of about 1) in only those subgroups that were examined with a repetition time at 3.0 T that was at or below the repetition time at 1.5 T, with a flip angle of 73°. In the parameter groups examined with a repetition time of longer than 300 msec (particularly if combined with a flip angle of 60°), enhancement rates obtained at 3.0 T were substantially lower than those obtained at 1.5 T. In the patients imaged with the longest repetition time (460 msec) and a flip angle of 60°, enhancement at 3.0 T was only one-third of the enhancement at 1.5 T (RER of 0.3).
|
| DISCUSSION |
|---|
|
|
|---|
Higher magnetic field strengths (eg, 3.0 T) have been shown to be useful for structural and functional neuroimaging, as well as for MR angiography (3335). The higher SNR afforded by the higher field strength can be successfully translated into improved image quality and higher diagnostic yield. It is clear that high-field-strength imaging for body applications will have to overcome some intrinsic difficulties that are specific to high-field-strength imaging (3639). Homogeneous signal intensity for a large field of view is more difficult to achieve with increasing field strength. For breast imaging, this may interfere with the clinical demand to image both breasts. All types of motion (breathing, pulsation, and considerable or subtle motion) can result in substantially stronger artifacts at higher field strengths, which may prove problematic in any thoracic or abdominal application, including breast imaging.
Susceptibility effects will also be stronger at 3.0 T and may cause image distortions, particularly on images acquired with a gradient-echo pulse sequence, which is the type of sequence that is typically used for breast MR imaging. Fat-water phase cycling is faster with 3.0-T imaging than with 1.5-T imaging such that opposed-phase echo times are reached with only a 1.1-msec shift in in-phase echo time. This means that exact echo times must be maintained for dynamic images that are acquired without fat suppression. T1 relaxation times are longer at higher field strengths, which means that, theoretically, longer repetition times will have to be usedan effect that could compromise temporal resolution if a 2D gradient-echo sequence is used.
Difficulties that are new for 3.0-T MR imaging compared with 1.5-T MR imaging are RF related. Heterogeneous RF distribution (so-called dielectric resonance effects) can lead to extensive areas of signal void on 3.0-T MR images. Second, RF energy deposition in tissue scales exponentially with increasing field strength, which means that SAR limitations are reached sooner at 3.0 T. If no techniques are available to compensate for this effect (such as SENSE), high-field-strength systems will have to be slowed down (ie, the number of RF pulses per time unit will have to be reduced) in order to avoid excessive heating in the patient (17,18,20).
Despite these potential obstacles, our intraindividual comparative study showed that image quality was slightly but significantly higher at 3.0 T than at 1.5 T. Image quality was rated as excellent throughout the high-spatial-resolution images with active fat suppression that were, for SNR reasons, obtained at 3.0 T only.
For bilateral transverse dynamic imaging, the field of view was large enough to accommodate both breasts in all 37 patients. No relevant signal intensity variations were observed across the field of view. Dielectric resonance effects were not observed at all; this finding was expected because the chest and the abdominal wall contribute most to the shielding of RF distribution. Susceptibility effects, particularly blurring, were not observed on the 2D gradient-echo MR images, possibly because the echo time used at 3.0 T was shorter than the echo time used at 1.5 T. Motion artifacts were observed with equivalent prevalence and to an equivalent degree on 3.0- and 1.5-T MR images. Motion was, however, a problem in patients who were examined with high in-plane imaging matrices (1024 x 640) and in whom subtraction artifacts could degrade image quality. This effect was not directly attributable to the higher field strength but was rather due to the smaller pixel size, which at image subtraction could cause subtraction artifacts with even subtle patient motion. We conclude that if very high matrix imaging is performed for a large (bilateral) field of view, some sort of breast fixation will be needed to avoid micromotion, which causes pixel shifts (and thus, artifacts) on subtraction images.
By using SENSE and a reduction factor of two, we were able to reduce the RF deposition by half (compared with non-SENSE imaging). Still, the one 3.0-T-related physical feature that did in fact have a substantial effect on image acquisition (and therefore on image appearance) was RF absorption and/or SAR. Because RF deposition scales exponentially with field strength and with the size of the RF pulse flip angle, we had to reduce the flip angle from 90° (at 1.5 T) to 60° or 73° (at 3.0 T) to adhere to SAR limitations and to fit the flip angle into the shorter in-phase echo time at 3.0 T. Because the size of the flip angle, together with repetition time, is an important determinant of T1 contrast, this may have a substantial effect on the overall enhancement rates observed.
A 30% prolongation of T1 relaxation times has been reported for healthy tissues at 3.0 T (1921). Accordingly, we began with a pulse sequence that had a longer repetition time compared with the standard 1.5-T setting. With this parameter setting, however, the enhancement rates of the same lesions at 3.0 T remained substantially below those at 1.5 T, with a RER on the order of 0.30.4. This was probably at least in part due to the lower flip angle. Even if the same repetition time was used at both 3.0 and 1.5 T and the flip angle was increased to 73°, enhancement rates at 3.0 T would be merely equivalent to those recorded at 1.5 T. Our results indicate that, for 2D gradient-echo dynamic MR imaging with a flip angle limited to 60°73°, it seems prudent not to adapt the repetition time to account for T1 prolongation at 3.0 T but to use the shortest possible repetition time. Moreover, and unlike findings for contrast-enhanced MR imaging of the brain at 3.0 T, the contrast agent dose may not be reduced for breast MR imaging (35,40).
The higher spatial resolution that is attainable at 3.0 T helped improve the classification of 10 of the 51 total lesions in nine of 35 patients. The increase in diagnostic confidence was reflected by a significantly larger area under the curve at ROC analysis. We wish to underscore, however, that this difference regarding diagnostic confidence should not be interpreted as proof of higher sensitivity and/or specificity of MR imaging at 3.0 T. A limitation of our study was a substantial selection bias in our cohort because patients were asked to participate in the study if they had an enhancing lesion observed at 1.5-T MR imaging; accordingly, the cohort is probably not representative of the average group of patients seen in breast MR imaging clinics. So there is reason to assume (but no proof yet) that MR imaging of the breast at 3.0 T may lead to an improved diagnostic accuracy compared with 1.5-T MR imaging.
In our study, we used the extra SNR afforded by 3.0-T MR imaging to improve spatial resolution only, with the intention of improving the analysis of subtle morphologic details. Thus, another study limitation is that we did not investigate the use of the extra SNR for breast imaging with high temporal resolution (ie, with the intention to improve the kinetic modeling of enhancing lesions) (36). Also, we did not investigate the use of the extra SNR for functional breast imaging techniques, such as diffusion-weighted MR imaging or MR spectroscopy. There is evidence to suggest that these techniques may be used to further improve the differential diagnosis of enhancing lesions (4145); at 1.5 T, these techniques have limited SNR and poor spatial resolution. It may be anticipated that the extra signal intensity gained by using 3.0-T systems should also foster the use of these approaches, with another potential increase in specificity.
In conclusion, our data suggest that 3.0-T MR imaging of the breast is nearing readiness for clinical use. Further studies in unselected groups of patients are under way in our department to investigate the effect of high-field-strength imaging on diagnostic accuracy of MR imaging in clinical practice.
| ADVANCES IN KNOWLEDGE |
|---|
|
|
|---|
| FOOTNOTES |
|---|
Abbreviations: BI-RADS = Breast Imaging Reporting and Data System RER = relative enhancement rates RF = radiofrequency ROC = receiver operating characteristic SAR = specific absorption rate SENSE = sensitivity encoding SNR = signal-to-noise ratio 2D = two dimensional
See Materials and Methods for pertinent disclosures.
Author contributions: Guarantor of integrity of entire study, C.K.K.; study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting or manuscript revision for important intellectual content, all authors; manuscript final version approval, all authors; literature research, C.K.K.; clinical studies, C.K.K., P.J., N.M.; statistical analysis, C.K.K.; and manuscript editing, C.K.K.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A M BLAMIRE The technology of MRI -- the next 10 years? Br. J. Radiol., August 1, 2008; 81(968): 601 - 617. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. E. Mountford, P. Stanwell, S. Ramadan, C. K. Kuhl, H. Kooijman, J. Gieseke, and H. H. Schild Breast MR Imaging at 3.0 T Radiology, July 1, 2008; 248(1): 319 - 320. |