Internal Barium Shielding to Minimize Fetal Irradiation in Spiral Chest CT: A Phantom Simulation Experiment1
David K. Yousefzadeh, MD,
Matthew B. Ward, MD and
Chester Reft, PhD
1 From the Departments of Radiology (D.K.Y., M.B.W.) and Radiation and Cell Oncology (C.R.), University of Chicago, 5841 S Maryland Ave, MC 2026, Chicago, IL 60637. Received December 27, 2004; revision requested February 28, 2005; revision received June 15; accepted June 27; final version accepted August 2.
Address correspondence to D.K.Y. (e-mail: dyousefz{at}midway.uchicago.edu).
 |
ABSTRACT
|
|---|
Purpose: To use a phantom to prospectively examine the attenuating effect of barium sulfate as an internal shield to protect the fetus.
Materials and Methods: In an adult-size phantom, 1- and 2-cm-thick acrylic slabs containing 315 or 630 mL of water, 2% or 40% barium sulfate suspension, and a 1-mm lead sheet were placed under the diaphragm. In 17 experiments, fetal dose was measured by using thermoluminescent dosimeters that were placed immediately under (near field) and 10 cm below (far field) the water slab (eight experiments), barium sulfate slab (eight experiments), and lead sheet (one experiment). In a pulmonary embolism protocol, the phantom was scanned with single-detector spiral computed tomography (CT) at 130 kVp and 230 mAs.
Results: The control radiation dose was 3.60 mSv ± 0.54 (standard deviation) with the water slab at near field, where the uterus dome is at near term, and 0.507 mSv ± 0.07 with the water slab at far field, the uterus position during early gestation. Scattered radiation was attenuated 13% and 21% with 2% barium sulfate and 87% and 96% with 40% barium sulfate, as calculated in the near and far fields, respectively, and 99% with the 1-mm lead sheet. The extrapolated attenuations for 5%40% barium sulfate suspensions indicated that beyond a 30% suspension, attenuation increased further only slightly.
Conclusion: Study results in the phantom experiment suggest that fetal irradiation during maternal chest CT can be reduced substantially with barium shielding.
© RSNA, 2006
 |
INTRODUCTION
|
|---|
Deep venous thrombosis occurs in 0.53.0 of every 1000 pregnancies in the United States (1,2), accounting for 10% of all hospital deaths of pregnant women and being a contributing factor in another 10% of such deaths (3). Fifty percent of all cases of pulmonary embolism (PE) in the general U.S. population are not diagnosed antemortem (4). During pregnancy, there is a five- to sixfold increased risk of PE (5,6), which continues to be the leading nonobstetric cause of death during pregnancy or the peripartum period (69) (Table 1). PE risk factors include increased lower-extremity venous stasis and relative hypercoagulability of the blood during the hyperestrogenic state (6,7). Although deep venous thrombosis is thought to occur mainly during the third trimester of pregnancy and immediately postpartum, others have observed an almost equal prevalence of the disease during all trimesters (10).
The challenge of diagnosing PE during pregnancy is in balancing the need for examinations that involve the use of ionizing radiation with both the potential risk to the fetus (1116) and the likelihood of childhood cancer caused by pediatric computed tomography (CT), which have been publicized even further during the past 5 years (1719). These concerns are the basis of ongoing debates among experts (2025).
Currently, the following six examinations are available for diagnosing acute PE:
1. D-dimer assay has 97% sensitivity and a 99.6% negative predictive value (26). False-positive results may be caused by pneumonia, myocardial infarct, cancer, sepsis, pregnancy, and/or recent surgery (27).
2. Lower-extremity ultrasonography (US) for assessment of deep venous thrombosis may indicate a risk of PE and justify the use of anticoagulant treatment (26). The sensitivity and specificity of chest wall US in the detection of PE are 94% and 87%, respectively (28,29).
3. Contrast materialenhanced magnetic resonance angiography is not yet practical to perform because it lacks adequate spatial resolution and does not yield detailed information about the lung parenchyma (27).
4. Nuclear medicine examinations for the overall assessment of PE are now used less frequently (30,31) because their results are indeterminate in up to 73% of cases (32). Nonetheless, some still advocate performing perfusion scanning during pregnancy. Performing single photon emission CT lowers the indeterminate result rate (33).
5. Pulmonary angiography is recommended more often than it is performed (34,35). It is invasive and associated with occasional morbidity and mortality (3638) and with interobserver agreement levels of 45%66% (34,39). Therefore, pulmonary angiography is no longer the reference standard imaging tool for the detection of PE (39).
6. Spiral CT angiography, enabling direct visualization of the thrombus (40,41), is now preferred for diagnosing PE (27). Furthermore, up to two-thirds of patients suspected of having PE have disorders that simulate PE, including aortic dissection, pneumonia, lung cancer, and pneumothorax (42). The reported interobserver agreement with CT is 0.72, versus 0.22 with nuclear medicine examinations (32,43), and a diagnostic algorithm for PE that includes CT is more cost effective than other diagnostic schemes (44). Last, CT can correctly depict PE even in the presence of coexisting pneumonia or preexisting lung disease (45).
Unlike single-detector CT, which may not depict emboli in small peripheral vessels (4649), multidetector CT can depict emboli in vessels down to the sixth-order branches (50), with interobserver agreement exceeding that achieved with selective pulmonary angiography (39,51,52).
The fetal radiation dose with single-detector CT has been estimated in various ways and during different trimesters; doses of 0.00330.8621 mSv derived by using low exposure factors (including 100 mAs) in one study (53), of 2 mSv in a second study (54), and of 1.2 mSv in a third involving the use of a phantom model (55) have been reported. Fetal radiation doses also vary according to the trimester during which the experiments are performed. If up to four-section multidetector CT is performed, the dose may increase by 30% (49,56) to 100% (57), and the dose to the uterus will increase by 92%180% (58). Hence, use of multidetector CT technology may lead to a trade-off of radiation safety for improved diagnostic accuracy (5961).
Since to the best of our knowledge, no internal protective shield had been designed before, the purpose of our study was to use a phantom to prospectively examine the attenuating effect of barium sulfate as an internal shield to protect the fetus.
 |
MATERIALS AND METHODS
|
|---|
Phantom
An anthropomorphic phantom (Rando; Alderson Research Laboratories, Stanford, Calif) composed of 2.5-cm-thick tissue-equivalent contoured slabs was chosen as an in vitro model. The phantom had anteroposterior and side-to-side dimensions of 27.5 and 20.5 cm, respectively, and a circumference of 81.0 cm. Two Harshaw lithium fluoride thermoluminescent dosimeters (TLDs) (Landauer, Glenwood, Ill) weighing 0.04 and 0.23 g/cm2 were chosen because (a) they could be easily positioned at various locations within the phantom; (b) a minimum radiation dose of 0.10 mGy could be measured according to the variability of the signal from nonirradiated detectors despite background noise; (c) the measurement precision was higher than 1.5% throughout the study; and (d) the energy response of these TLDs was known, having been previously investigated at various beam qualities. This last characteristic was particularly important because the energy spectrum of the scattered radiation is not well known. The TLDs were calibrated in a 6-MV phantom beam with a calibrated ionization chamber. A correction factor of 1.6 was applied to the 6-MV TLD calibrations to account for the energy response of the detector in the CT beam.
To measure the attenuation of internal radiation scatter in a lung CT examination, we used two volumes of water (315 and 630 mL), two concentrations of barium sulfate (2% and 40%), and one thickness of lead shielding (1 mm) to perform chest CT in the adult-size phantom. Two hollow acrylic slabs, 1.0 and 2.0 cm thick, were made to fit seamlessly between the phantom's slabs. The two slabs had fluid capacities of 315 and 630 mL, respectively; we chose these volumes to compare the attenuation of the scattered radiation between the two volumes. The slabs were filled first with water as a control and then with the 2% and 40% concentrations of barium sulfate. The slabs were then placed between the lowest point of the "thorax" and the upper part of the "abdomen" of the phantom at the L1 level.
TLDs that were 3 x 3 mm and two different thicknesses were housed in 15 5-mm-diameter round slots in the phantom slabs (Fig 1). The TLDs were placed first in the near field corresponding to the L2 plane, immediately below (ie, 0 cm from) the acrylic fluid-filled slabs (Z = 0 cm), and then in the far field corresponding to the L5 level, 10 cm from the slabs (ie, Z = 10 cm, caudal to Z = 0), to represent the approximate positions of the dome of the uterus during the last and first trimesters of pregnancy, respectively. To block the external source of radiation, a 1-mm-thick lead sheet was wrapped around the lower portion of the phantom that contained the TLDs (Fig 2).
Experiments Performed
A total of 17 experiments were performed. Eight experiments were performed with 1- and 2-cm-thick slabs containing water (control), and eight others were conducted with slabs containing the 2% or 40% barium sulfate suspension in the near and far fields. For each experiment, a new set of TLDs was used. Experiment 17 was performed with the 1-mm-thick lead sheet placed between the thorax and the TLDs as the reference shield. A single-detector spiral CT unit (Philips Medical Systems, Cleveland, Ohio), the only system available in our laboratory at the time of these experiments, was used. The following exposure factors were used: 130 kVp, 230 mAs, 3-mm section thickness, a 1.7 pitch, and 30-cm collimation from the lung apices (T1 level) to the lung bases (T12 level). These factors remained constant throughout the 17 experiments. In none of the experiments was a scout CT image acquired.
Attenuation measurements were performed at CT with 0.5- and 1.0-mm lead foils to obtain information about the energy of the scattered radiation, which was needed for TLD calibration and to derive an effective broad-beam attenuation coefficient.
An uncertainty analysis was performed because, when converting the measured TLD signal intensity to a radiation dose, four factors must be taken into account: the calibration of the TLDs at 6 MV, the individual sensitivities of the TLDs, the TLD energy exposure, and the background tissue signal intensity. This is consistent with finding that there was no difference in the dose measurements obtained between the two different-thickness detectors in this study.
 |
RESULTS
|
|---|
The control radiation doses measured in eight different experiments with water slabs ranged from 3.06 to 4.14 mSv (mean, 3.60 mSv ± 0.54 [standard deviation]) in the near field and from 0.437 to 0.577 mSv (mean, 0.507 mSv ± 0.07) in the far field. Table 2 summarizes the attenuation measurements of barium versus water considering four variables: (a) the slab thickness containing water versus barium; (b) the barium solution concentration, 2% versus 40%; (c) the TLD slab placement at Z = 0 versus 10 cm from the water or barium slab (where Z is the distance from the fluid-filled slab to the TLD containing the Rando slab); and (d) the thickness, therefore, the weight of the TLDs, 0.04 g/cm2 versus 0.23 g/cm2. The section thickness was 3 mm on all occasions. These data are provided to give some perspective on the magnitude of scattered radiation that can be attenuated by barium sulfate. The Aeff is equal to 1 minus the ratio of the radiation dose measured with barium sulfate (Db) to the dose measured with an equivalent thicknesstherefore, volumeof water (Dw) replacing the barium sulfate:
 | (1) |
View this table:
[in this window]
[in a new window]
|
Table 2. Summary of Attenuation Measurements of Barium Sulfate Relative to Water Aeff in the Rando Phantom for 0.3-cm Helical CT at 130 kVp and 230 mA
|
|
The results show that the different-thickness detectors yielded the similar results within the range of estimated experimental uncertainties, and the percentages of attenuation were similar at Z = 0 and Z = 10 cm. With the 0.04 g/cm2 TLDs, the attenuation values measured for the 1- and 2-cm 2% barium sulfate slabs were 13% and 17% in the near and far fields, respectively. Corresponding values for the 0.023 g/cm2 TLDs were 11% and 13%. With the 40% barium sulfate concentrationdepending on the slab thickness, field, and TLD weightthe attenuation ranged between 86% and 98%. Assuming that the scattered radiation was attenuated by the barium sulfate as an exponential function of the barium sulfate concentration yielded the following relationship:
 | (2) |
where µeff is an effective linear attenuation coefficient, B is the concentration of barium sulfate, and t is the thickness of the barium sulfate layer. Using data from the attenuation values measured with the 2% and 40% barium sulfate concentrations, we obtained an effective linear attenuation coefficient of 5.9 cm 1. Incorporating Equation (2) into Equation (1) yielded the following equation:
 | (3) |
With this equation, the Aeff rendered by the barium sulfate suspension in different concentrations and different slab thicknesses may be estimated. Aeff values for barium sulfate concentrations of 5%40%, in 5% increments, are listed in Table 3. As an example, with a 20% concentration, we calculated attenuations of 69% and 91%, respectively, for the 1- and 2-cm-thick barium sulfate slabs in the near and far fields. Derived by using the attenuation data for 2% barium sulfate, the composite exponential attenuation profiles for 1- and 2-cm-thick slabs containing concentrations of 2%40% in the near and far fields are shown in Figure 3. The 1-mm lead sheet placed between the thorax and the abdomen attenuated 99% of the scattered photons. From the experiment involving the use of 0.5- and 1.0-mm lead foils, we established an effective energy of 60 keV for the scattered radiation.

View larger version (21K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3: Composite attenuation profiles in near and far fields extrapolated for 1- and 2-cm-thick slabs that contain 2%40% barium sulfate concentrations, in 2% increments, demonstrate that at the 30% concentration, the attenuations reach a plateau, irrespective of the slab thickness and the distance of the TLDs from the barium slab. Further attenuation beyond this concentration point is only slight.
|
|
In our analysis of the attenuation rendered by barium sulfate, we eliminated some of the uncertainties because we evaluated the ratio of the TLD signal intensities. However, four factors were involved in converting the measured TLD signal intensity to a dose measurement: the calibration at 6 MV, the individual sensitivity of the TLD, the TLD energy response, and the signal intensity of the background tissue. For each factor, 1 standard deviation of uncertainty,
, was calculated:
values for calibration at 6 MV, individual TLD sensitivity, TLD energy response, and background signal intensity at Z = 0 cm and Z = 10 cm were ±2.0%, ±1.5%, ±7.0%, ±1.0%, and ±10.0%, respectively. Considering the individual uncertainties in quadrature, the estimated uncertainties in the attenuation measurements at 1 standard deviation, or at the 68% confidence level, were ±8% and ±13% at Z = 0 cm and Z = 10 cm, respectively.
 |
DISCUSSION
|
|---|
Twenty-one percent to 39% of all PEs cannot be detected in subsegmental arteries with single-detector CT, two-detector CT, or even electron-beam CT (46,49). Therefore, multidetector CT is now the scanning method of choicebut at the expense of a 30% (49,56) to 100% (57) higher fetal radiation dose and a 98%180% higher organ dose (58). Because motion artifact, technical factors, and poor opacification may cause 5%10% of CT scans to be nondiagnostic (53), patients may need to undergo either repeat CT or additional examinations such as lung scintigraphy or pulmonary angiography. In fact, the results of a multiinstitutional study indicate that although multidetector CT is the preferred imaging tool, multimodality imaging is still being used frequently (62). The frequent use of multimodality imaging will further increase fetal radiation doses and provoke anxiety in patients, their families, and their physicians. The radiation data given here may not apply to the newest and more efficient multidetector CT technologies (63), which are not available in all centers nation- or worldwide.
The perceptions of physicians regarding the teratogenic risks associated with radiography and CT during pregnancy are not realistic (64)even at academic medical centers in the United States (65). These unrealistic perceptions may lead to the underutilization of new tests and jeopardize the health of mothers and fetuses. On the other hand, some physicians consider the reluctance of some radiologists to use multidetector CT in the work-up of pregnant patients to be an infringement of their right to examine and treat their patients (17). In addition, the estimated potential risk of childhood cancer due to fetal irradiation is not well known. It varies between one case per 1000 radiologic examinations in utero during the first 4 months of pregnancy (66), to 0.03%0.05% per examination (67), to one additional cancer death per 1700 pregnancies, and one case per 10000 radiologic examinations (20). Some suffice to say that "the risk for cancer is not zero at doses of the order of 10 mSv" (14).
In our experiments with water slabs, the control radiation doses were 3.064.14 mSv in the near field, the presumed position of the dome of the uterus in the third trimester, and 0.4370.577 mSv in the far field, simulating the dose in the first trimester. The difference between our projected fetal doses and those reported by Winer-Muram et al (53) and Damilakis et al (55) can be explained in part by the exposure factors that we used, 130 kVp and 230 mAs (vs 120 kVp and 100 mAs), and our imaging of the entire thorax without excluding the lung apices and bases so that simulating disorders could be searched for. Our choice of field size was compatible with those used in daily practice. Other factors may have been differences (a) between the phantom sizes that were available to us and those that were available to the other investigators, (b) in the range of TLD sensitivities, and (c) in the calculation methods used (Monte Carlo technique used by Winer-Muram et al [53] versus our phantom method) and the uncertainties inherent of any analysis similar to ours.
The radiation dose with up to four-section multidetector CT would increase by 30%100% (49,56,57), so our adjusted near-field radiation dose with use of this version of multidetector CT would be 4.688.28 mSv. These doses, without any additional radiation from scout imaging or other imaging examinations involving ionizing radiation, are almost equal to or higher than the 5-mSv limit set by the International Commission on Radiological Protection for the total 9-month occupational dose allowed for pregnant workers and are 1016 times higher than the monthly radiation dose allowed for radiology personnel (62). The targeted gonadal dose in the third trimester, when the head is engaged in the pelvis and the gonadsspecifically the testesare in the near field, is markedly higher.
Our in vitro experiment results show that the protective effects of 40% barium sulfate measured with TLDs of two thicknesses were 90%98% in the near field and 86%94% in the far field. However, a 40% barium sulfate suspension need not be used in every patient. The extrapolated data that we have provided offer the pregnant patient several choices of barium volume and concentration. Both 630 mL of the 20% barium sulfate suspension and 315 mL of the 40% suspension will yield 91% attenuation. Even a 5%10% barium sulfate concentration can reduce fetal radiation exposure by 45%69%. At a 30% concentration, the attenuation curves will plateau, irrespective of the barium volume or TLD positions in the near or far field. Beyond a 30% concentration, further reductions in radiation will be only slight. Therefore, the effectiveness of a 30% or higher barium sulfate suspension is similar to the attenuating effect of a 1-mm-thick lead sheet. It is noteworthy that the attenuating effects of the 40% barium sulfate suspension during the actual experiment with the 1- and 2-cm-thick slabs were almost identical to those measured by using the extrapolation method (90%98% vs 91%99%). These results give credence to the extrapolation technique and to the projected data that we obtained.
In the far field (Z = 10 cm), the presumed position of the uterus during the first trimester, the radiation dose was 0.4370.577 mSv. As low as this dose may seem, it is delivered during the critical phase of organogenesis when the fetus is most sensitive to radiation. The barium shield can attenuate the radiation dose to the fetus by 83%94% during this critical period.
Although efforts to reduce the radiation dose by simply lowering the exposure factors had promising results initially (6870), more recent study data contradict the earlier results (71). Furthermore, none of these studies included patients with PE in the study populations (6871).
In terms of limitations, the specific dose reduction is a complex product that is affected by the maternal body habitus, fetal gestational age, position and orientation of the fetus, quantity and concentration of the barium sulfate suspension, and distribution of the suspension through the gastrointestinal tract. Furthermore, patients who are critically ill or may need to undergo a cesarean section may not be appropriate candidates for oral barium use. Last, a slab of barium sulfate in vitro is not precisely comparable to the anatomy of the gastrointestinal tract in vivo. In vivo protocols are currently being designed to address these differences.
In conclusion, the results of our phantom study show that a 30%40% barium sulfate shield attenuates scattered photons almost as effectively as a 1-mm lead shield. The described radiation attenuation method is not rigid and gives the patient and the radiologist latitude in the choice of volume and concentration of barium without substantially altering the attenuation effect. With a concentration higher than 30%, any further decreases in radiation dose will be only slight. We believe this method is practical and can safeguard the health of the fetus, in accordance with the as low as reasonably possible philosophy. This method can potentially also be used for other imaging examinations that involve the use of ionizing radiation.
Practical applications: The use of oral barium sulfate is safe, effective, and intuitively logical, and it can potentially prevent one to 10 lifetime malignancies per 10000 pregnant patients who undergo chest CT examinations. The use of oral barium sulfate can alleviate the apprehension of acutely ill pregnant patients, their families, and their physiciansat practically no cost. After future in vivo application proves to be practical, the use of this method can be expanded to include pregnant patients who undergo lung scanning with use of nuclear medicine tracers, children who need to undergo repeated chest CT examinations for metastatic surveys, employees at workplaces with radiation hazards, and patients who need to undergo cardiac catheterization and interventional cardiothoracic procedures.
 |
ACKNOWLEDGMENTS
|
|---|
The authors thank Kate A. Feinstein, MD, FACR, Heber MacMahon, MD, Michael Vannier, MD, and Charles Pelizzari, PhD, for their input; Elizabeth Lanzl for editorial assistance; and Sharon Harris and Sheila A. Mitchell for secretarial assistance. The authors dedicate this paper to the memory of their beloved and deceased colleague and scholar, Carl Vyborny, MD, PhD, who was the encouraging force behind this original research.
 |
FOOTNOTES
|
|---|
Abbreviations: Aeff = attenuation measurement of barium sulfate relative to water PE = pulmonary embolism TLD = thermoluminescent dosimeter
Authors stated no financial relationship to disclose.
Author contributions: Guarantor of integrity of entire study, D.K.Y.; study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting or manuscript revision for important intellectual content, all authors; manuscript final version approval, all authors; literature research, all authors; experimental studies, C.R.; and manuscript editing, D.K.Y., C.R.
 |
References
|
|---|
- Atrash HK, Rowley D, Hogue CJ. Maternal and perinatal mortality. Curr Opin Obstet Gynecol 1992;4(1):6171.[Medline]
- Barbour LA, Pickard J. Controversies in thromboembolic disease during pregnancy: a critical review. Obstet Gynecol 1995;86(4 pt 1):621633.[Abstract]
- Lindblad B, Sternby NH, Bergqvist D. Incidence of venous thromboembolism verified by necropsy over 30 years. BMJ 1991;302(6778):709711.[Medline]
- Patriquin L, Khorasani R, Polak JF. Correlation of diagnostic imaging and subsequent autopsy findings in patients with PE. AJR Am J Roentgenol 1998;171(2):347349.[Abstract/Free Full Text]
- Wessler S. Medical management of venous thrombosis. Annu Rev Med 1976;27:313319.[CrossRef][Medline]
- Toglia MR, Nolan TE. Venous thromboembolism during pregnancy: a current review of diagnosis and management. Obstet Gynecol Surv 1997;52(1):6072.[CrossRef][Medline]
- Rosenow EC 3rd. Venous and pulmonary thromboembolism an algorithmic approach to diagnosis and management. Mayo Clin Proc 1995;70(1):4549.[Medline]
- Johnson MS. Current strategies for diagnosis of pulmonary embolus. J Vasc Interv Radiol 2002;13(1):1323.[Medline]
- Goldhaber SZ, Elliott CG. Acute pulmonary embolism. I. Epidemiology, pathophysiology, and diagnosis. Circulation 2003;108(22):27262729.[Free Full Text]
- Toglia MR, Weg JG. Venous thromboembolism during pregnancy. N Engl J Med 1996;335:108114.[Free Full Text]
- Ginsberg JS, Hirsh J, Rainbow AJ, Coates G. Risk of the fetus of radiologic procedures used in diagnosis of maternal venous thromboembolic disease. Thromb Haemost 1989;61(2):189196.[Medline]
- Kal HB, Struikmans H. Pregnancy and medical irradiation: summary and conclusions from the International Commission on Radiological Protection, publication 84 [in Dutch]. Ned Tijdschr Geneeskd 2002;146(7):299303.[Medline]
- Hall EJ. Scientific view of low-level radiation risk. RadioGraphics 1991;11(3):509518.[Abstract]
- Wakeford R, Little MP. Childhood cancer after low-level intrauterine exposure to radiation. J Radiol Prot 2002;22(3A):A123A127.[CrossRef][Medline]
- Wakeford R, Little MP. Risk coefficients for childhood cancer after intrauterine irradiation: a review. Int J Radiat Biol 2003;79(5):293309.[CrossRef][Medline]
- Lowe SA. Diagnostic radiography in pregnancy: risks and reality. Aust N Z J Obstet Gynaecol 2004;44(3):191196.[CrossRef][Medline]
- El-Khoury GY, Madsen MT, Blake ME, Yankowitz J. A new pregnancy policy for a new era. AJR Am J Roentgenol 2003;181(2):335340.[Free Full Text]
- Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 2001;176(2):289296.[Abstract/Free Full Text]
- Hall EJ. Lessons we have learned from our children: cancer risks from diagnostic radiology. Pediatr Radiol 2002;32(10):700706.[CrossRef][Medline]
- Brent RL, Mettler FA. Pregnancy policy [letter]. AJR Am J Roentgenol 2004;182(3):819822.[Free Full Text]
- Brent RL, Mosely RD Jr, Russell LB, Wilson J, Casarrett GW, Tayler LS. Medical radiation exposure of pregnant and potentially pregnant women. Report no. 54. Washington, DC: Ad Hoc Committee on Exposure of Potentially Pregnant Women in Diagnostic Radiology of the National Council on Radiation Protection and Measurement, 1977.
- Brent RL. Utilization of developmental basic science principles in the evaluation of reproductive risks from pre- and postconception environmental radiation exposures. Teratology 1999;59(4):182204.[CrossRef][Medline]
- Schull WJ, Otake M. Cognitive function and prenatal exposure to ionizing radiation. Teratology 1999;59(4):222226.[CrossRef][Medline]
- Boice JD Jr, Miller RW. Childhood and adult cancer after intrauterine exposure to ionizing radiation. Teratology 1999;59(4):227233.[CrossRef][Medline]
- Miller RW. Discussion: severe mental retardation and cancer among atomic bomb survivors exposed in utero. Teratology 1999;59(4):234235.[CrossRef][Medline]
- Schoepf UJ, Costello P. CT angiography for diagnosis of pulmonary embolism: state of the art. Radiology 2004;230(2):329337.[Abstract/Free Full Text]
- Schoepf UJ, Goldhaber SZ, Costello P. Spiral computed tomography for acute pulmonary embolism. Circulation 2004;109(18):21602167.[Abstract/Free Full Text]
- Reissig A, Kroegel C. Transthoracic ultrasound of lung and pleura in the diagnosis of pulmonary embolus: a novel non-invasive bedside approach. Respiration 2003;70(5):441452.[CrossRef][Medline]
- Reissig A, Heyne JP, Kroegel C. Sonography of lung and pleura in pulmonary embolism: sonomorphologic characterization and comparison with spiral CT scanning. Chest 2001;120(6):19771983.[Medline]
- Schibany N, Fleischman D, Thallinger C, et al. Equipment availability and diagnostic strategies for suspected pulmonary embolism in Australia. Eur Radiol 2001;11(11):22872294.[CrossRef][Medline]
- Leveau P. Diagnostic strategy in pulmonary embolism. National French survey [in French]. Presse Med 2002;31:929932.
- Blachere H, Latrabe V, Montaudon M, et al. Pulmonary embolism revealed on helical CT angiography: comparison with ventilation-perfusion radionuclide lung scanning. AJR Am J Roentgenol 2000;174:10411047.[Abstract/Free Full Text]
- Garg K, Welsh CH, Feyerabend AJ, et al. Pulmonary embolism: diagnosis with spiral CT and ventilation-perfusion scanningcorrelation with pulmonary angiographic results or clinical outcome. Radiology 1998;208(1):201208.[Abstract/Free Full Text]
- Crawford T, Yoon C, Wolfson K, et al. The effect of imaging modality on patient management in the evaluation of pulmonary thromboembolism. J Thorac Imaging 2001;16(3):163169.[CrossRef][Medline]
- Prologo JD, Glauser J. Variable diagnostic approach to suspected pulmonary embolism in the ED of a major academic tertiary car center. Am J Emerg Med 2002;20(1):59.[CrossRef][Medline]
- Mills SR, Jackson DC, Older RA, et al. The incidence, etiologies and avoidance of complications of pulmonary angiography in a large series. Radiology 1980;136(2):295299.[Abstract/Free Full Text]
- Stein PD, Athanasoulis C, Alavi A, et al. Complications and validity of pulmonary angiography in acute pulmonary embolism. Circulation 1992;85:462468.[Abstract/Free Full Text]
- Hudson ER, Smith TP, McDermott VG, et al. Pulmonary angiography performed with iopamidol: complications in 1434 patients. Radiology 1996;198(1):6165.[Abstract/Free Full Text]
- Stein PD, Henry JW, Gottschalk A. Reassessment of pulmonary angiography for the diagnosis of pulmonary embolism: relation of interpreter agreement to the order of the involved pulmonary arterial branch. Radiology 1999;210:(3)689691.
- Gurney JW. No fooling around: direct visualization of pulmonary embolism [editorial]. Radiology 1993;188:(3)618619.
- Woodard PK. Pulmonary arteries must be seen before they can be assessed [editorial]. Radiology 1997;204:1112.[Free Full Text]
- van Rossum AB, Pattynama PM, Mallens WM, Hermans J, Heijerman HG. Can helical CT replace scintigraphy in the diagnostic process in suspected pulmonary embolism? a retrolective-prolective cohort study focusing on total diagnostic yield. Eur Radiol 1998;8(1):9096.[CrossRef][Medline]
- van Rossum AB, van Erkel AR, van Persijn van Meerten EL, Ton ER, Rebergen SA, Pattynama PM. Accuracy of helical CT for acute PE: ROC analysis of observer performance related to clinical experience. Eur Radiol 1998;8(7):11601164.[CrossRef][Medline]
- van Erkel AR, van Rossum AB, Bloem JL, Kievit J, Pattynama PM. Spiral CT angiography for suspected pulmonary embolism: a cost-effectiveness analysis. Radiology 1996;201(1):2936.[Abstract/Free Full Text]
- Goodman LR, Lipchick RJ, Kuzo RS, Liu Y, McAuliffe TL, O'Brien DJ. Subsequent pulmonary embolism: risk after a negative helical CT pulmonary angiogramprospective comparison with scintigraphy. Radiology 2000;215(2):535542.[Abstract/Free Full Text]
- Goodman LR, Curtin JJ, Mewissen MW, et al. Detection of pulmonary embolism in patients with unresolved clinical and scintigraphic diagnosis: helical CT versus angiography. AJR Am J Roentgenol 1995;164(6):13691374.[Abstract/Free Full Text]
- Remy-Jardin M, Remy J, Artaud D, et al. Peripheral pulmonary arteries: optimization of the spiral CT acquisition protocol. Radiology 1997;204:157163.[Abstract/Free Full Text]
- Qanadli SD, Hajjam ME, Mesurolle B, et al. Pulmonary embolism detection: prospective evaluation of dual-section row helical CT versus selective pulmonary arteriography in 157 patients. Radiology 2000;217:447455.[Abstract/Free Full Text]
- Schoepf UJ, Helmberger T, Holzknecht N, et al. Segmental and subsegmental pulmonary arteries: evaluation with electron-beam versus spiral CT. Radiology 2000;214:433439.[Abstract/Free Full Text]
- Ghaye B, Szapiro D, Mastora I, et al. Peripheral pulmonary arteries: how far in the lung does multi-detector row spiral CT allow analysis? Radiology 2001;219(3):629636.[Abstract/Free Full Text]
- Diffin DC, Leyendecker JR, Johnson SP, Zucker RJ, Grebe PJ. Effect of anatomic distribution of pulmonary emboli on interobserver agreement in interpretation of pulmonary angiography. AJR Am J Roentgenol 1998;171:(4)10851089.
- Schoepf UJ, Holzknecht N, Helmberger TK, et al. Subsegmental pulmonary emboli: improved detection with thin collimation multi-detector row spiral CT. Radiology 2002;222(2):483490.[Abstract/Free Full Text]
- Winer-Muram HT, Boone JM, Brown HL, Jennings SG, Mabie WC, Lambardo GT. Pulmonary embolism in pregnant patients: fetal radiation dose with helical CT. Radiology 2002;224(2):487492.[Abstract/Free Full Text]
- Classic K. VQ scan and chest CT during pregnancy. Health Physics Society Web site. http://www.hps.org/publicinformation/ate/9411.html. Published October 17, 2000. Updated December 31, 2003. Accessed December 2004.
- Damilakis J, Perisinakis K, Voloudaki A, Gourtsoyiannis N. Estimation of fetal radiation dose from computed tomography scanning in late pregnancy: depth-dose data from routine examinations. Invest Radiol 2000;35(9):527533.[CrossRef][Medline]
- Schoepf UJ, Bruening RD, Hong C, et al. Multislice helical CT of focal and diffuse lung disease: comprehensive diagnosis with reconstruction of contiguous and high-resolution CT sections from a single thin-collimation scan. AJR Am J Roentgenol 2001;177:179184.[Abstract/Free Full Text]
- McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys 1999;26(11):22232230.[CrossRef][Medline]
- Thornton FJ, Paulson EK, Yoshizumi TT, Frush DP, Nelson RC. Single versus multi-detector row CT: comparison of radiation doses and dose profiles. Acad Radiol 2003;10(4):379385.[CrossRef][Medline]
- Rogers LF. Radiation exposure in CT: why so high? [editorial]. AJR Am J Roentgenol 2001;177:277.[Free Full Text]
- Rogers LF. Serious business: radiation safety and radiation protection [editorial]. AJR Am J Roentgenol 2001;177:1.[Free Full Text]
- Golding SJ, Shrimpton PC. Radiation dose in CT: are we meeting the challenge? [commentary]. Br J Radiol 2002;75(889):14.[Free Full Text]
- Brateman L. The AAPN/RSNA physics tutorial for residents: radiation safety considerations for diagnostic radiology personnel. RadioGraphics 1999;19(4):10371055.[Abstract/Free Full Text]
- Mayo JR, Aldrich J, Müller NL. Radiation exposure at chest CT: a statement of the Fleischner Society. Radiology 2003;228(1):1521.[Abstract/Free Full Text]
- Ratnapalan S, Bona N, Chandra K, Koren G. Physicians' perceptions of teratogenic risk associated with radiography and CT during early pregnancy. AJR Am J Roentgenol 2004;182(5):11071109.[Abstract/Free Full Text]
- Lee CI, Haims AH, Monica EP, Brink JA, Forman HP. Diagnostic CT scans: assessment of patient, physician, and radiologist awareness of radiation dose and possible risks. Radiology 2004;231:393398.[Abstract/Free Full Text]
- El-Khoury GY, Madsen MT, Blake ME, Yankowitz JT. Pregnancy policy [reply to letter]. AJR 2004; 182:822.
- Mole RH. Radiation effects on pre-natal development and their radiological significance. Br J Radiol 1979;52(614):89101.[Abstract]
- Lee KS, Primack SL, Staples CA, Mayo JR, Aldrich JE, Muller NL. Chronic infiltrative lung disease: comparison of diagnostic accuracies of radiography and low- and conventional-dose thin-section CT. Radiology 1994;191:669673.[Abstract/Free Full Text]
- Mayo JR, Hartman TE, Lee KS, Primack SL, Vedal S, Muller NL. CT of the chest: minimal tube current required for good image quality with the least radiation dose. AJR Am J Roentgenol 1995;164(3):603607.[Abstract/Free Full Text]
- Ravenel JG, Scalzctli EM, Huda W, Garrisi W. Radiation exposure and image quality in chest CT examinations. AJR Am J Roentgenol 2001;177(2):279284.[Abstract/Free Full Text]
- Mayo JR, Kim KI, MacDonald SL, Johkoh T, Kavanagh P. Reduced radiation dose helical chest CT: effect on reader evaluation of structures and lung findings. Radiology 2004;232(3):749756.[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
S. J. Patel, D. L. Reede, D. S. Katz, R. Subramaniam, and J. K. Amorosa
Imaging the Pregnant Patient for Nonobstetric Conditions: Algorithms and Radiation Dose Considerations
RadioGraphics,
November 1, 2007;
27(6):
1705 - 1722.
[Abstract]
[Full Text]
[PDF]
|
 |
|