|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Experimental Studies |
1 From the Departments of Medical Imaging (E.C.) and Radiation Therapy (S.V., M.O.), Cliniques Universitaires St-Luc, Université Catholique de Louvain, Avenue Hippocrate, 10, 1200 Brussels, Belgium. Received April 7, 2005; revision requested June 7; revision received August 12; accepted September 13; final version accepted November 23. Address correspondence to E.C. (e-mail: coche{at}rdgn.ucl.ac.be).
| ABSTRACT |
|---|
|
|
|---|
Materials and Methods: The part of the study involving patients (seven women, four men; mean age, 62 years ± 16 [standard deviation]; range, 4185 years) was approved by the institutional review board. Patients gave written informed consent. Exposure was performed with an anthropomorphic phantom with thermoluminescent dosimeters for fourdetector row CT without the dose-modulation program and 16detector row CT without and with the dose-modulation program with standard protocols for pulmonary CT angiography (120 kV, 144 mAs, four and 16 detector rows with 1.00- and 0.75-mm section thickness, respectively). Digital angiograms were acquired with four standard projections at 80 kV. For digital angiography, radiation dose was calculated according to phantom measurements and adapted to acquisition and fluoroscopy times. Distribution of dose was compared for CT and digital angiography.
Results: During pulmonary CT angiography, mean radiation dose delivered at middle of chest was 21.5, 19.5, and 18.2 mGy for fourdetector row CT and for 16detector row CT without and with dose-modulation program, respectively. At the same level, a mean dose of 91 mGy was delivered with digital angiography. The dose adjusted to clinical conditions was 139.0 mGy for digital angiography and could be reduced after technical adjustment. Ratios of maximum dose to mean dose were 1.15 and 2.96 for CT and digital angiography, respectively. With application of the dose-modulation program at 16detector row CT, radiation dose was reduced 15%20% at the upper chest.
Conclusion: Multidetector row CT delivers a lower radiation dose, with better spatial distribution of dose, than does pulmonary CT angiography. With 16detector row CT and a dose-modulation program, radiation dose is decreased during PE work-up.
© RSNA, 2006
| INTRODUCTION |
|---|
|
|
|---|
Thus, the purpose of our study was to compare the radiation dose delivered at fourdetector row CT and 16detector row CT with a dose-modulation program with that delivered at digital angiography for evaluation of PE.
| MATERIALS AND METHODS |
|---|
|
|
|---|
At the time of experimentation, only the 16detector row CT scanner was equipped with the dose-modulation program (Dose Right, version 2.5; Philips Medical Systems, Best, the Netherlands), a dedicated software that modulates the tube current according to the attenuation of the patient's body (710).
All protocols that were used are the standard CT protocols we used for PE work-up in our Department of Medical Imaging. First, a frontal scout view of the chest was obtained to plan the CT examination. Second, a bolus test was performed (11,12) to determine the circulation time. First, one unenhanced transverse CT image was obtained to define a region of interest at the level of the main pulmonary arteries. Thereafter, repetitive transverse CT images were obtained (Table 1) every second during 20 seconds at the selected level. The number of acquired images for the bolus test was therefore 21. The peak of enhancement in pulmonary arteries was measured by using dedicated software (Bolus Pro; Philips Medical Systems, Cleveland, Ohio).
|
Digital Angiography
A digital angiographic examination, as usually performed in our department, was simulated (Table 2) with a monoplanar vascular x-ray system (Integris V3000; Philips Medical Systems, Best, the Netherlands). Posteroanterior (0°), left posterior oblique (+35°), right posterior oblique (35°), and lateral right (90°) projections were systematically obtained after each selective catheterization. The acquisition was obtained with a delay of 4 seconds, a frame rate of 3 frames per second during 5 seconds and 2 frames per second during 10 seconds, with a 350-mm field of view, 80 kV, and 700 mA (half-value layer of 3.3 mm of aluminum). The values for total accumulated tube currenttime product in our experimental conditions were, on average, 6.4, 29.7, 17.8, and 45.4 mAs for 0°, +35°, 35°, and 90° projections, respectively. Fluoroscopy was performed at 74 kV and 2.5 mA. This standard procedure was used to obtain baseline values of delivered radiation during digital angiography. Sequences performed during clinical work-up were recorded in the last 11 patients (seven women, four men; mean age, 62 years ± 16 [standard deviation]; range, 4185 years) examined in our Department of Medical Imaging for PE work-up from February 2000 to October 2001. This part of our study was approved by the institutional review board of our institution. All patients gave appropriate written informed consent.
|
|
For the bolus test, 20 thermoluminescent dosimeters, with eight at the periphery, were distributed on a section passing throughout the pulmonary arteries (section 15), and six dosimeters were placed on the adjacent section (section 16).
For digital angiography, special care was applied to keep the geometric parametersfocus-phantom distance, source-image distance, and phantom positionconstant. Dose measurements were performed at four standard acquisition projections of 0°, +35°, 35°, and 90° and source-image distances of 100, 106, 106, and 117 cm, respectively (Fig 1). For each projection, the thermoluminescent dosimeters were positioned at 22 different places at the level of section 16. For fluoroscopy, accumulated doses during 4 minutes were measured at 0° and 2.5 mA. Thermoluminescent dosimeters were positioned at the anterior and the posterior part of the chest. Their relative response was in agreement with the ratio of extrapolated 0° digital angiographic data. Doses to the phantom were deduced, with the assumption of the same attenuation level and depth as were used for digital angiographic conditions. The LiF dosimeters were read by using a commercially available reader system with a nitrogen-heating gas flow (model 5500; Harshaw/Bicron, Solon, Ohio) (15) at least 24 hours after irradiation. The annealing cycle consisted of heating the thermoluminescent dosimeters for 1 hour at 400°C, followed by a controlled cooling phase at a temperature reduced to 100°C and held for 3 hours. A set of 100 thermoluminescent dosimeters, individually calibrated in a 6-MV photon therapy beam, has been used. Their response reproducibility expressed as a standard deviation of the mean was 2.6%. The set response was verified at each experiment with eight thermoluminescent dosimeters.
The doses are specified as doses to water at the level of the dosimeters. An increase in the LiF response of 25% and of 39% between the reference beam and the experimental beam is assumed for the measurements at the CT scanner (120 kV, half-value layer of 8.8 mm of aluminum) and at the monoplanar vascular x-ray system (80 kV, half-value layer of 3.3 mm of aluminum), respectively (1618). As a control of the entire dosimetric procedure, a direct comparison of the computer dose index values in polymethylmethacrylate measured by a calibrated-pencil-ionization chamber (Radcal model 2025; Radcal, Monrovia, Calif) and by thermoluminescent dosimeters was performed at 16detector row CT by a physicist.
At fourdetector row CT, measurements were repeated six times; at 16detector row CT with and without the dose-modulation program, measurements were repeated four times and twice, respectively. For pulmonary CT angiography, measurements were repeated three times at each projection. Adjustments of the angiographic suite for maintenance purposes were performed by the manufacturer's engineers after our experiment: For the same protocol and for the same geometric conditions, unchanged tube parameters (tube voltage, filtration, tube current) resulted in a substantial decrease in exposure time. The mean doses before and after adjustment were reported. The distribution of the dose was compared for CT and digital angiography.
| RESULTS |
|---|
|
|
|---|
For the bolus test at fourdetector row CT, the mean radiation dose delivered to section 15 was 120.4 mGy (range, 88.8162 mGy). At 2.5 cm distance, the mean delivered radiation dose was reduced to 18.8 mGy.
For pulmonary CT angiography at fourdetector row CT, the mean radiation dose to sections 12, 16, and 20 was 17.6 mGy ± 3.1, 21.5 mGy ± 1.9, and 21.5 mGy ± 2.8, respectively. The mean dose decreased to 5.7 mGy at the level of the thyroid gland (Table 3).
|
Dose distributions.Dose distribution observed at the middle part of the chest is relatively homogeneous, as can be seen in Figure 2. All the relative doses expressed as percentages of the total mean dose are included between 86% and 108% of the mean measured dose. The standard deviation of the relative dose to the section was 7%. The data also showed that the left side of the patient received a few percentage points less of the dose than did the right side: The greatest measured discrepancy was 10%. For the three selected sections, the maximal dose never exceeded 15% of the mean dose (ratio to the mean dose of 1.15). The same dose distribution was observed for the three acquisitions at the levels of sections 16 and 20. Dose dispersion in section 12 decreased from 18% (standard deviation of the mean) to 8% (standard deviation of the mean) when the dose-modulation program was applied.
|
An average of 6.3 acquisitions (2.3 posteroanterior, 1.1 right posterior oblique, 2.2 left posterior oblique, and 0.7 lateral views) were performed in each patient during digital angiography in the last 11 consecutive patients who were referred for evaluation because they were suspected of having an acute PE; the mean time for fluoroscopy was 10 minutes, and the range was 318 minutes. Therefore, the mean radiation dose delivered to the middle part of the chest in a patient in the clinics was estimated at 139.0 mGy and 83.6 mGy before and after the adjustment performed by the engineers of the manufacturer, respectively (Table 4).
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Angiography versus MultiDetector Row CT
The mean radiation dose delivered to the middle part of the chest at digital angiography is more than four times the mean dose delivered during pulmonary CT angiography at fourdetector row CT in standard conditions and seven times that delivered during this type of imaging in clinical practice. The scout view adds only a small amount of radiation dose that is equal to 0.17 mGy, which is equivalent to an increase of 0.7% of the acquisition dose (21.5 mGy) at the middle part of the chest. The radiation dose delivered during the bolus test was approximately equal to five times the mean radiation dose delivered with pulmonary CT angiography at fourdetector row CT with an acquisition length of approximately 10 mm.
Resten et al (5) obtained total average doses of 28 mGy ± 7.6 and 6.4 mGy ± 1.5 for digital angiography and singledetector row CT, respectively. The effective dose reported by Kuiper et al (6) was 4.2 mSv for fourdetector row CT and 7.1 mSv for digital subtraction angiography. No analysis of delivered radiation dose for the scout view and the bolus test was considered in previous articles. There are some differences between these results and ours, for the CT protocols and the equipment used were not the same. The values we obtained with thermoluminescent dosimeters were controlled with comparison of CT dose index values obtained during CT acquisition by using the same experimental conditions. Variations in patients' doses are large for digital angiography because the number of projections and duration of fluoroscopy vary according to the expertise of the radiologist, the patient's status, and the difficulty of the clinical case. Time of fluoroscopy and number of sequences in digital angiography were systematically recorded by Resten et al (5). The authors recorded four sequences per patient and a mean fluoroscopy time of 9 minutes 47 seconds (range, 1 minute 15 seconds to 26 minutes 20 seconds), and these data agree with our data.
A more homogeneous distribution of delivered radiation dose is observed with multidetector row CT than with digital angiography. At CT, lower doses found on the posterior part of the chest were probably related to CT table absorption, but the higher doses on the right side remain unexplained. The heterogeneity at digital angiography is mainly related to the important attenuation of the x-rays (19,20) with depth and to an unbalanced distribution of the projections.
Effect of MultiDetector Row Factor and Dose-Modulation Program
Analysis of our results shows that a decreased radiation dose is observed when the number of detectors increases. During our study, we observed an overall reduction of 8% of the total amount of delivered radiation dose with the 16detector row CT system compared with that delivered with the fourdetector row CT system. The x-ray tube and its filtration, the geometric features, and the collimator settings were identical for both CT units used in this study. With multidetector row CT, only the plateau region of the dose profile may be used to ensure equal signal level for all detector sections. The penumbral region has to be discarded, with use of either postpatient collimation or intrinsic self-collimation of the multidetector row CT unit and represents a "wasted dose." The relative contribution of the penumbral region decreases with an increasing number of simultaneously acquired sections and increases with a decreasing section width (20,21). This physical property explains the relative lower dose we obtained with 16detector row CT compared with the dose we obtained with fourdetector row CT.
The use of the dose-modulation program with the most recent multidetector row CT unit can decrease radiation dose, particularly at the upper part of the chest. Attenuation-based, online tube-current control has great potential for the reduction of patient dose at CT. The mean value of the radiation dose was decreased by 15%20% at the level of the upper chest and by approximately 5% at the level of the middle and lower parts of the chest with the dose-modulation program in our experiment. Despite that the images are noisier with this software, the diagnosis should not be affected. Tack et al (22) compared values for inter- and intrareader agreement for the diagnosis of PE on CT pulmonary angiographic images acquired with an effective tube currenttime product of 90 mAs and with simulated values of 60, 40, 20, and 10 mAs. The authors showed that values for inter- and intrareader agreement remain stable when the tube currenttime product is reduced from 90 to 10 mAs. A survey of practices and policies for pulmonary CT angiography (23) in pregnant women revealed that the method for reduction of the radiation dose that was most often used consisted of a reduction of the craniocaudal scan coverage.
Limitations
First, the doses were related to the shape of the phantom we used: The anthropomorphic phantom we used is a simulation of a slim patient (body weight, 70 kg). For obese patients, the automatic brightness control on the biplanar angiographic system would increase the tube current, and this increase would result in a higher delivered radiation dose, whereas at CT angiography, radiation dose is relatively much less dependent on patient size (24).
Second, the anthropomorphic phantom that we used is armless, and this structural difference between phantoms and humans probably contributes to an underestimation of the potential effect of online tube-current modulation.
The uncertainty about our determination of the absorbed doses to water is relatively important. First, the phantom (13,14) is not really equivalent to tissue because, within the energy range of 20300 keV, the ratio of mass attenuation coefficients varies by approximately 5% and the ratio of absorption coefficients varies by approximately 7%, both acting in an opposite direction. Second, the thermoluminescent dosimeter response increases by approximately 40% with decreasing energy from 300 to 20 keV. Without the knowledge of the photon spectral distribution at the point of measurement, an "equivalent energy," which characterizes the quality of the photon beam, has to be selected.
As far as the overall uncertainty is concerned, the thermoluminescent dosimeter energy dependence induces a response variation with a depth of the order of 3% for depths that vary between of 0 and 5 cm (17). The beam perturbation caused by the presence of the thermoluminescent dosimeter is assessed to be approximately 1% (16).
Finally, the uncertainty about the dose in the reference beam can be estimated to be 2% at the level of 1 standard deviation (25). In spite of the reproducibility of the thermoluminescent dosimeter, the combined uncertainty about the absorbed dose, predominantly influenced by the estimation of the equivalent energy, could be as high as 15%.
Practical application: Even if bolus tracking concerns a narrow area of the chest, radiologists should be aware that this procedure locally delivers a relatively great amount of radiation. Reduction of the x-raytube current settings will have a major effect on radiation exposure. To date to our knowledge, however, there is no prospective experimental study in which the influence of reduced tube currenttime product and tube voltage on timing values for the bolus test has been tested. Alternatively, reduction of repetition time during the bolus test also could be used to decrease radiation exposure. In our clinical practice, we have empirically decreased tube currenttime product presets for the bolus test from 150 to 30 mAs.
Maintenance procedures are performed three times a year in our angiographic suite by the engineers of the manufacturer of the unit. These procedures consist of measurement of the tube voltage constancy, dose measurements at the level of the monitoring system, and image quality control. The results of our experiment with a phantom demonstrated a technical failure of the detection system of the angiographic unit that was not detected with the routine maintenance procedure. Image quality that resulted from increased tube currenttime product was probably modified but not enough to be reported by clinicians. Therefore, we can suggest that the reproducibility of the relationship between the x-raytube current settings and image quality be systematically verified during the maintenance procedure by using a dedicated phantom.
From our data, we conclude that multidetector row CT delivered a lower radiation dose, with a better spatial dose distribution, than did digital angiography. Multidetector row CT with the higher number of detector rows that we tested is preferable for thin-section examinations, as recommended for PE work-up. Application of the dose-modulation program further helps to decrease the delivered radiation dose. The image quality, however, related to the detection of PE should be evaluated.
| ADVANCES IN KNOWLEDGE |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
Abbreviations: PE = pulmonary embolism
Authors stated no financial relationship to disclose.
Author contributions: Guarantors of integrity of entire study, all authors; study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting or manuscript revision for important intellectual content, all authors; manuscript final version approval, all authors; literature research, all authors; clinical and experimental studies, all authors; statistical analysis, all authors; and manuscript editing, all authors
| References |
|---|
|
|
|---|
Related Article
This article has been cited by other articles:
![]() |
S. Matsuoka, A. R. Hunsaker, R. R. Gill, I. B. Oliva, B. Trotman-Dickenson, F. L. Jacobson, and H. Hatabu Vascular Enhancement and Image Quality of MDCT Pulmonary Angiography in 400 Cases: Comparison of Standard and Low Kilovoltage Settings Am. J. Roentgenol., June 1, 2009; 192(6): 1651 - 1656. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Debnath, E. Coche, and S. Vynckier An Erroneous Conclusion Regarding Pulmonary Embolism Radiology, April 1, 2009; 251(1): 307 - 308. [Full Text] [PDF] |
||||
![]() |
S. Jouveshomme, I. Bohn, and A. Cazaban Diagnosis of pulmonary embolism in hospitalised patients: retrospective survey of an institutional standard Eur. Respir. J., December 1, 2007; 30(6): 1117 - 1123. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Hurwitz, R. E. Reiman, T. T. Yoshizumi, P. C. Goodman, G. Toncheva, G. Nguyen, and C. Lowry Radiation Dose from Contemporary Cardiothoracic Multidetector CT Protocols with an Anthropomorphic Female Phantom: Implications for Cancer Induction Radiology, December 1, 2007; 245(3): 742 - 750. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Heyer, P. S. Mohr, S. P. Lemburg, S. A. Peters, and V. Nicolas Image Quality and Radiation Exposure at Pulmonary CT Angiography with 100- or 120-kVp Protocol: Prospective Randomized Study Radiology, November 1, 2007; 245(2): 577 - 583. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.-C. Pouleur, J.-B. le Polain de Waroux, A. Pasquet, J.-L. J. Vanoverschelde, and B. L. Gerber Aortic Valve Area Assessment: Multidetector CT Compared with Cine MR Imaging and Transthoracic and Transesophageal Echocardiography Radiology, September 1, 2007; 244(3): 745 - 754. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |