1 From the Fairfax Radiological Consultants P.C., 2722 Merrilee Dr, Suite 230, Fairfax, VA 22031 (J.P.E., E.L.B., B.A.U., C.A.C., J.L.L., R.S.J.); Applied Statistics Laboratory, GE Global Research Center, Niskayuna, NY (C.C.M.); and Applied Science Laboratory, GE Healthcare Technologies, Waukesha, Wis (J.H., J.H.L.). From the 2006 RSNA Annual Meeting. Received June 11, 2007; revision requested July 27; revision received August 25; final version accepted September 27. Address correspondence to J.P.E. (e-mail: jpearls{at}yahoo.com).
| ABSTRACT |
|---|
|
|
|---|
Materials and Methods: This HIPAA-compliant study received a waiver for approval from the institutional review board, including one for informed consent. Coronary CT angiograms obtained with 64–detector row CT were retrospectively evaluated in 203 clinical patients. A routine RGH technique was evaluated in 82 consecutive patients (44 males, 38 females; mean age, 55.6 years). The PGT technique was then evaluated in 121 additional patients (71 males, 50 females; mean age, 56.7 years). All images were evaluated for image quality, estimated radiation dose, and coronary artery segment assessability. Differences in image quality score were evaluated by using a proportional odds logistic regression model, with main effects for three readers, two techniques, and four arteries.
Results: The mean effective dose for the group with the PGT technique was 2.8 mSv; this represents an 83% reduction as compared with that for the group with the RGH technique (mean, 18.4 mSv; P < .001). The image quality score for each of the arteries, as well as the overall combined score, was significantly greater for images obtained with PGT technique than for images obtained with RGH technique. The combined mean image quality score was 4.791 for images obtained with PGT technique versus 4.514 for images obtained with RGH technique (proportional odds model odds ratio, 2.8; 95% confidence interval: 1.7, 4.8). The percentage of assessable coronary artery segments was 98.6% (1196 of 1213) for images obtained with PGT technique versus 97.9% (1741 of 1778) for images obtained with RGH technique (P = .83).
Conclusion: PGT coronary CT angiography offers improved image quality and substantially reduced effective radiation dose compared with traditional RGH coronary CT angiography.
© RSNA, 2008
| INTRODUCTION |
|---|
|
|
|---|
Radiation dose is a concern for all cardiac imaging studies in which ionizing radiation is used. The reported effective radiation dose in conventional coronary angiography ranges from 3.1 to 9.4 mSv, whereas researchers in most studies report that the coronary CT angiographic dose is even higher (15–20). Reported effective radiation dosage in recent studies with the use of 64–detector row CT have ranged from 9.5 to 21.4 mSv, more than the previously reported doses from 16–detector row CT studies (9,12,13,21) (Table 1). The effective doses reported for coronary CT angiography performed with dual-source CT systems are also relatively high, ranging from 8 to 16.1 mSv (22–24). Previous attempts to develop methods to reduce the radiation dose with coronary CT angiography have primarily involved the use of electrocardiogram (ECG)-controlled tube current modulation or reduction of tube current or tube voltage (25,26).
|
|
| MATERIALS AND METHODS |
|---|
|
|
|---|
Three authors (J.H., J.H.L., and C.C.M.) are employees of GE Healthcare, manufacturer of the CT system used in this study. The other (non–GE Healthcare employee) authors had control of the data and information that might present a conflict of interest for the employee authors.
Patients
Eighty-two consecutive RGH coronary CT angiographic procedures performed in 82 patients (44 males and 38 females; mean age, 55.6 years) were initially obtained during the study period by using a routine RGH coronary CT angiographic protocol. All examinations were performed for clinical reasons. After initially optimizing imaging parameters, we included 121 (92.4%) (71 males and 50 females; mean age, 56.7 years) of the next 131 consecutive clinical patients presenting for coronary CT angiography, as they met our inclusion criteria (heart rate, <70 beats per minute; observed heart rate fluctuation, <10 beats per minute during observation at the scanner prior to performance of the coronary CT angiographic sequence).
Multi–Detector Row CT
All CT examinations were performed with a 64–detector row CT unit (LightSpeed VCT XT, software version 7.1; GE Healthcare, Milwaukee, Wis), with a rotation time of 350 msec, fixed single-sector reformation with a temporal resolution of 175 msec, detector aperture of 0.625 mm, scan field of view of 25 cm, and tube voltage of 120 kV. The scan sequence included a scout scanogram, a low-dose transverse scout or calcium-scoring scan, a test-bolus scan, and a coronary angiogram. A low-dose transverse scout scan is used to determine the position of the left main coronary artery and cardiac apex for prescription of the coronary angiographic sequence, when a calcium-scoring study is not performed. Criteria for performing a calcium-scoring sequence included age older than 45 years, no coronary stent placed, no prior coronary angiography performed, and no calcium-scoring study performed within the past 12 months; 69% of patients underwent a calcium-scoring study.
A dual-headed injector (Optivantage DH; Mallinckrodt, St Louis, Mo) was used for all coronary CT angiographic studies. For both PGT and RGH studies, a test-bolus scan was obtained at the level of the aortic root. Sequential scans were obtained every 2 seconds for 30 seconds after administration of the test bolus of 20 mL iodixanol 320 (Visipaque; GE Healthcare, Princeton, NJ) at 5.5 mL/sec. The diagnostic angiogram included an area 20 mm above the left main orifice to 20 mm below the apex of the heart. All scans were obtained in a cranial-caudal direction. The timing of the main injection of contrast medium (50–80 mL iodixanol 320 at 5.5 mL/sec was determined by the time to peak enhancement from the test bolus plus 8 seconds to allow for coronary artery filling.
A dose of 50–100 mg of a β-adrenergic blocking agent, metoprolol (Mylan Pharmaceuticals, Morgantown, WVa), was administered orally at 45–60 minutes before CT examination if the patient's resting heart rate exceeded 60 beats per minute and there was no contraindication to the use of a β-adrenergic blocking agent. In seven of the patients undergoing RGH CT angiography and eight of the patients undergoing PGT CT angiography, oral medication was supplemented with 5-mg intravenous aliquots of metoprolol administered at 5-minute intervals up to a maximum total intravenous dose of 20 mg.
RGH Technique
A standard RGH technique was performed by three technologists, including one author (J.L.L.), for the RGH technique group (Table 2). The tube current was selected by one of three technologists, each experienced in performing from 300 to 4000 coronary CT angiographic examinations. The selected tube current ranged from 349 to 771 mA, depending on the patient's body mass index (BMI) and chest circumference as determined by the technologist's prior experience. ECG modulation was used in all patients (100% peak tube current during the middle of diastole and 80% reduction during systole). The pitch was determined automatically by the system and is not user defined in the cardiac helical mode; it ranged from approximately 0.2 to 0.3, depending on heart rate. Images were reconstructed throughout the cardiac cycle in 10% increments. A cardiac-specific noise reduction filter (C2) was applied to the images prior to reconstruction.
|
The scan was prescribed by using three to five incremental 64 x 0.625-mm (40-mm volume coverage) image groups, requiring two to four incremental table movements of 35 mm, with a 5-mm overlap. The minimum interscan delay was approximately 1.0 second; this protocol normally requires skipping a cardiac cycle between acquisitions of successive image groups. The tube current was selected by one of three technologists (as with the RGH technique), and it ranged from 300 to 800 mA, depending on the patient's BMI and chest circumference, as determined by the technologist's prior experience. No noise reduction filters were available for use with the PGT studies. The minimum scan time at each transverse location is 230 msec (180° plus a fan angle), which translates to an effective temporal resolution of 175 msec with half-scan weighting.
The prescribed phase for data acquisition was 75% of the cardiac cycle (middle of diastole) for all subjects. Additional padding of the tube-on time was used, depending on the amount of perceived beat-to-beat variability (Fig 2). This technique turns the tube on prior to, and leaves it on after, the minimum required 230 msec. Padding allows the reconstruction to adapt to minor heart rate variations and produce consistent image quality, because the reconstruction window can be modified retrospectively to ensure an identical cardiac phase from scan to scan. For example, when the heart rate fluctuates by 10% from beat to beat (decreases from 60 to 54 beats per minute or increases from 60 to 66 beats per minute), then ±10% of phase (100 msec) padding is used to ensure that images during the prescribed phase location are acquired for all table locations. In this example, if the prescribed phase was 75% of the cardiac cycle, then images are captured during phases of 65%–85% of the cardiac cycle and images from multiple phases can be reconstructed. Actual padding in the study ranged from 0 msec in patients with very stable heart rates to 100 msec in patients with less stable heart rates; this padding value was chosen by the technologist on the basis of observation of the ECG waveform.
|
Data Analysis
Coronary segments were defined according to American Heart Association guidelines (31). All reformatted images were evaluated and classified by three independent readers (J.P.E., B.A.U., and E.L.B., each with experience interpreting images from 500 to 3000 coronary CT angiographic examinations) who were blinded to patient identity. Available images included transverse source images, curved multiplanar reformations, and thin-slab maximum intensity projections. Readers had access to images obtained during all available cardiac phases. Image evaluation and image processing were performed with an independent three-dimensional workstation equipped with software designed for use with the workstation (Advantage Windows with CardIQ software, version 4.3; GE Healthcare, Milwaukee, Wis). Because with the RGH scans more reconstructed phases were available than were available with the PGT scans, the readers could determine the scanning technique used.
For each coronary artery, the readers assessed image quality semiquantitatively by using a previously described five-point ranking scale (32). In this scale, a score of 5 indicated that there were no motion artifacts and there was a clear delineation of the artery; a score of 4, that there were minor artifacts and mild blurring; a score of 3, that there were moderate artifacts and moderate blurring without structure discontinuity; a score of 2, that there were severe artifacts and doubling or discontinuity in the course of the segment; and a score of 1, that the image was not evaluable and blood vessel structures were not differentiable. Examples of each score were provided to the readers prior to assigning a score to the images from the examinations. For coronary artery assessability, a score of 2 or less was an indication that the artery was considered nonassessable. Coronary stenoses were determined by using a 50% diameter reduction threshold value to classify a stenosis as significant; this value was determined by comparing the narrowest diameter with the diameter of the nearest normal lumen.
Statistical Analysis
Differences between imaging parameters for the two patient groups were tested by using unpaired t tests at the 95% confidence level (two tailed). We did not assume equal variances. A P value of .05 was considered to indicate a significant difference.
For analysis of image quality score, we controlled for effects for reader and also for the correlated nature of the data because four arteries were evaluated in each subject. Two analyses with power calculations were performed. First, an ordinal regression was fit to the image quality scores to measure differences in image quality distributions among techniques, readers, and arteries. Second, artery assessability differences between the two techniques were analyzed by using a logistic regression on a binary outcome where an artery was considered assessable when the image quality score was more than 2.
We modeled the image quality score by using a proportional odds logistic regression model (33) with main effects for the three readers, the two techniques, and the four arteries. The proportional odds model used here was specified as a marginal model (34) as follows:
![]() | (1) |
j values are the baseline log odds for each image quality score, and â1r, â2t, and â3a[r] are the log odds ratios (ORs) for the three independent predictors (reader, technique, and artery, respectively). We tested the proportional odds model assumption of a constant OR regardless of the choice of image quality cutoff value by applying separate binary logistic regressions of the form given in Equation (1), where j was varied, and the effects for reader, treatment, and artery were reestimated in each fit. No significant deviations were seen in the ORs from the multiple fits. The generalized estimating equation method was used to estimate the effects for reader, technique, and artery of Equation (1), by using the ordgee function of the geepack library (version 1.0-11) in the R statistical package (version 2.4.1) (34,35). Clusters of size 4 were defined to model the correlation structure among the four artery measurements assigned to a subject by a reader.
In a second analysis, we quantified possible differences in the effect for technique according to artery through the addition of a technique-artery interaction to Equation (1) as follows:
![]() | (2) |
Analysis of artery assessability was performed by using a logistic regression (36), with assessability, defined as an image quality score greater than 3, as the outcome variable. Because of the very small number of arteries called unassessable, we were forced to collapse the data over artery by removing the effect for artery from Equation (1). Main effects were included for reader and technique thus:
![]() | (3) |
A power calculation was performed for both the image quality analysis and vessel assessability analysis. For the image quality analysis, we determined that at least 81 subjects in each technique group would be sufficient to detect a technique OR of 2.6 with 80% power while controlling the type I error rate at 5% in a two-sided hypothesis test (37). On the other hand, we determined that a very large sample size would be required to detect even large technique ORs for vessel assessability (>4) with 80% power because the assessability percentages for both groups were expected to be relatively high, greater than 90%–95%. Collecting such a large number of subjects was outside the scope of this project. Note that the methods given in an article by Walters (37) do not take into account the repeated-measures character of these data. For the purposes of the power calculations, we made the conservative assumption that only one artery was assessed per subject.
| RESULTS |
|---|
|
|
|---|
|
In one of the examinations with the RGH technique, performed in a 52-year-old woman (height, 5 ft 0 inches [150 cm]; weight, 200 lb [90 kg]; BMI, 39.0 kg/m2) with a stent in the right coronary artery and new chest pain, images of the distal right coronary artery were nondiagnostic because of poor image quality. At a later date, the examination was repeated by using the PGT technique, and findings were diagnostic, revealing a distal right coronary artery stenosis (Fig 3).
|
|
|
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
PGT coronary CT angiography offers acquisition of images during a limited number of available reconstruction phases. This was initially a concern prior to having experience with the technique. Depending on the amount of padding used, the range of available cardiac phases for acquisition of images with the PGT technique can be from 10% to 50% of the cardiac cycle. With the RGH technique, images can be obtained during 100% of the cardiac cycle for analysis. In the past it was shown that end-systolic phase images may at times be more optimal for analysis of the right coronary artery; however, Leschka et al (39) and Hong et al (40) recently determined that only at heart rates greater than 85.5 beats per minute does the best reconstruction time shift to end systole. Despite limited available phases in PGT examinations, only 1.4% of coronary segments were nonassessable (image quality score of <3); this percentage was not significantly different as compared with the value of 2.1% for nonassessable segments with the RGH method.
The effective dose for the PGT group (mean, 2.84 mSv) was significantly lower than that for the RGH group (mean, 18.4 mSv) (P < .001). This difference represents an 83% reduction from the effective dose with the RGH technique to that with the PGT technique. Differences inherent in the two techniques, low-pitch helical for the RGH technique and transverse with minimal overlap for the PGT technique, account for most of the effective dose differences. "Overbeaming," extra exposures at the start and end of a helical acquisition, of the RGH technique also contributes significantly to the overall dose increase. The remainder of the difference is likely caused by some variation in other parameters common to both methods, including tube current, and z-axis coverage. Although the tube voltage was held at 120 kV for all patients, on average a higher tube current was used for the RGH group subjects (mean, 647.6 mA; range, 348–771 mA) than for the PGT group subjects (mean, 507.3 mA; range, 300–800 mA). In addition, slightly greater anatomic coverage was prescribed for the RGH group subjects than for the PGT group subjects, also contributing to higher effective dose.
There were seven subjects in the PGT group who received a radiation dose of 1.0 mSv or less. We believe that these are the lowest reported effective doses for coronary CT angiography. Abada et al (26) reported relatively low effective doses of approximately 2 mSv by using a modified RGH technique in a group of 11 slim patients (<60 kg) in which the tube voltage was reduced to 80 kV, ECG modulation was used, and the tube current was 520 mA.
Physicians performing CT in general adhere to the as low as reasonably achievable, or ALARA, principle. However, even with strict adherence to this principle, the radiation dose of coronary CT angiography remains relatively high, and the long-term risk to the patient of developing radiation-induced cancer must be considered (41). Although all patients benefit from decreasing the effective dose of coronary CT angiographic examinations, we believe that a decrease is especially promising for premenopausal women (because of the direct breast exposure) and middle-aged and younger patients because of the longer lag time that exists for a cancer to develop. Additionally, patients who might in the future undergo repeated coronary CT angiographic studies, such as those with coronary stents, bypass grafts, complex congenital anomalies, or known coronary stenoses, may also benefit because of the lower cumulative dose.
Recently, the National Research Council's Committee on the Biological Effects of Ionizing Radiation released their seventh report on the health effects of exposure to low-dose ionizing radiation (42). The committee calculated that a dose of 10 mSv would cause one in 1000 lifetime cancers but acknowledged that this calculation could be off by a factor of two or three (16). The International Commission on Radiological Protection had previously estimated that a dose of 10 mSv would cause one in 2000 lifetime cancers (43). On the basis of these estimates, helical coronary CT angiography with an effective dose of 15 mSv has a range of risk of inducing a fatal cancer of one in 677 (Committee on the Biological Effects of Ionizing Radiation, seventh report [42]) to one in 1333 (International Commission on Radiological Protection [43]). Coronary CT angiography performed with the PGT technique and an estimated dose of 2.8 mSv would yield a theoretical risk of one in 3571 (Committee on Biological Effects of Ionizing Radiation, seventh report [42]) to one in 7143 (International Commission on Radiological Protection [43]).
It may be useful to put the coronary CT angiographic effective doses in context with other cardiac diagnostic tests. Conventional coronary angiography has an approximate dose of 5.6 mSv (15). The effective dose for a rest-stress myocardial single photon emission CT scan obtained by using technetium 99m (99mTc) is 8–17.5 mSv (16,44). The dose for a thallium 201 (201Tl) stress and reinjection scan is 18–25.1 mSv, and a dual-isotope technique study performed by using 201Tl and 99mTc has an effective dose of 27.3 mSv (16,44). The long-term risk of developing a radiation-induced cancer from any single cardiac imaging study is relatively small. However, patients with CAD undergo multiple examinations over their lifetimes. Thus, the risks are likely to be substantially larger. Because PGT coronary CT angiography represents a substantial dose savings as compared with any of the other noninvasive cardiac studies with radiation, widespread use in lieu of these other examinations will likely yield a widespread benefit in relation to reduced overall risk.
We have shown that the quality of images from PGT examinations was significantly better than that of images from RGH examinations. This can be explained by several reasons. First, the PGT examinations had a shorter scan time (4.2 seconds for PGT vs 5.7 seconds for RGH), allowing less time for ectopy or heart rate acceleration or deceleration to occur, all of which can adversely affect image quality. In addition, with the PGT technique, the table remains stationary during the data acquisition, unlike with the RGH technique in which the patient is constantly moving during the scan. With the PGT technique, one can also take full advantage of the predictive gating capability allowed by the stationary table; variation in the heart rate will simply translate to a delayed or earlier data acquisition and reconstruction. With the PGT technique, a newer gated complementary reconstruction algorithm that further improves image quality is used (27). Results of a recent study with a phantom in which PGT and RGH techniques were compared revealed in-plane resolution to be 8% higher and section sensitivity in the z-axis to be 6% better for the PGT technique (45).
There were limitations to our study. Ours was a retrospective analysis of clinical cases. Therefore, many parameters were not strictly controlled, and bias could have been introduced. In the RGH technique group, we identified a higher tube current, which contributed negatively to overall dose but likely helped, or at least did not hinder, overall image quality. There was also a slightly greater amount of z-axis coverage in the RGH technique group, which also increased its effective dose. The current PGT implementation itself has inherent limitations. With its current revision, heart rates are limited to less than 70 beats per minute and administration of β-adrenergic blocking agents is required in most patients because of the lack of multisector reconstruction capability. Although image quality is usually excellent at 60–70 beats per minute, we use β-adrenergic blocking agents in patients in this group to minimize or, hopefully, eliminate heart rate accelerations (or decelerations) during data acquisition because they can contribute negatively to image quality. We found that the limitation of 70 beats per minute allows us to image approximately 92% of clinical patients with the newer technique.
Because of the narrow range of available reconstruction windows, neither global nor regional function can be determined with the PGT technique, although in theory expanded acquisition windows with the PGT technique could offer such capability at the cost of a higher dose. In our clinical practice, most patients presenting for coronary CT angiography have already had either an echocardiogram or a nuclear myocardial perfusion examination prior to the coronary CT angiographic examination, and cardiac function usually is not requested or analyzed with CT. Even in cases in which cardiac function is desired, the additional 10–15 mSv required to change from PGT to a helical technique may not be the optimal approach, as other noninvasive methods for determining function are available with no radiation dose.
In summary, we have shown here that coronary CT angiography performed with a PGT technique, compared with an RGH technique, provides a substantial (>80%) effective radiation dose reduction and significantly improves image quality. We believe this technique has great promise to become a commonly used method for coronary CT angiography.
| ADVANCES IN KNOWLEDGE |
|---|
|
|
|---|
| IMPLICATION FOR PATIENT CARE |
|---|
|
|
|---|
| FOOTNOTES |
|---|
Abbreviations: BMI = body mass index CAD = coronary artery disease ECG = electrocardiogram OR = odds ratio PGT = prospectively gated transverse RGH = retrospectively gated helical
Guarantors of integrity of entire study, J.P.E., J.L.L.; study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting or manuscript revision for important intellectual content, all authors; manuscript final version approval, all authors; literature research, J.P.E., J.L.L., C.C.M., J.H.; clinical studies, J.P.E., E.L.B., B.A.U., C.A.C., J.L.L., R.S.J., J.H.; experimental studies, E.L.B., J.H.; statistical analysis, J.P.E., E.L.B., C.C.M.; and manuscript editing, J.P.E., B.A.U., C.A.C., R.S.J., C.C.M., J.H., J.H.L.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. L. Raff, K. M. Chinnaiyan, D. A. Share, T. Y. Goraya, E. A. Kazerooni, M. Moscucci, R. E. Gentry, A. Abidov, and for the Advanced Cardiovascular Imaging Consortium Radiation Dose From Cardiac Computed Tomography Before and After Implementation of Radiation Dose-Reduction Techniques JAMA, June 10, 2009; 301(22): 2340 - 2348. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach, V. Dilsizian, C. M. Kramer, and W. A. Zoghbi The Year in Coronary Artery Disease J. Am. Coll. Cardiol. Img., June 1, 2009; 2(6): 774 - 786. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. P. Shuman, K. R. Branch, J. M. May, L. M. Mitsumori, J. N. Strote, B. H. Warren, T. J. Dubinsky, D. W. Lockhart, and J. H. Caldwell Whole-Chest 64-MDCT of Emergency Department Patients with Nonspecific Chest Pain: Radiation Dose and Coronary Artery Image Quality with Prospective ECG Triggering Versus Retrospective ECG Gating Am. J. Roentgenol., June 1, 2009; 192(6): 1662 - 1667. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Nagel, J. A.C. Lima, R. T. George, and C. M. Kramer Newer methods for noninvasive assessment of myocardial perfusion cardiac magnetic resonance or cardiac computed tomography? J. Am. Coll. Cardiol. Img., May 1, 2009; 2(5): 656 - 660. [Full Text] [PDF] |
||||
![]() |
L. Husmann, B. A. Herzog, O. Gaemperli, F. Tatsugami, N. Burkhard, I. Valenta, P. Veit-Haibach, C. A. Wyss, U. Landmesser, and P. A. Kaufmann Diagnostic accuracy of computed tomography coronary angiography and evaluation of stress-only single-photon emission computed tomography/computed tomography hybrid imaging: comparison of prospective electrocardiogram-triggering vs. retrospective gating Eur. Heart J., March 1, 2009; 30(5): 600 - 607. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Tatsugami, L. Husmann, B. A. Herzog, N. Burkhard, I. Valenta, O. Gaemperli, and P. A. Kaufmann Evaluation of a Body Mass Index-Adapted Protocol for Low-Dose 64-MDCT Coronary Angiography with Prospective ECG Triggering Am. J. Roentgenol., March 1, 2009; 192(3): 635 - 638. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Mayo and J. A. Leipsic Radiation Dose in Cardiac CT Am. J. Roentgenol., March 1, 2009; 192(3): 646 - 653. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. P. Agarwal, A. Chughtai, F. R. K. Matzinger, and E. A. Kazerooni Multidetector CT of Thoracic Aortic Aneurysms RadioGraphics, March 1, 2009; 29(2): 537 - 552. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Gerber, J. J. Carr, A. E. Arai, R. L. Dixon, V. A. Ferrari, A. S. Gomes, G. V. Heller, C. H. McCollough, M. F. McNitt-Gray, F. A. Mettler, et al. Ionizing Radiation in Cardiac Imaging: A Science Advisory From the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention Circulation, February 24, 2009; 119(7): 1056 - 1065. [Full Text] [PDF] |
||||
![]() |
J. Hausleiter, T. Meyer, F. Hermann, M. Hadamitzky, M. Krebs, T. C. Gerber, C. McCollough, S. Martinoff, A. Kastrati, A. Schomig, et al. Estimated Radiation Dose Associated With Cardiac CT Angiography JAMA, February 4, 2009; 301(5): 500 - 507. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kitagawa, H. Yamamoto, J. Horiguchi, N. Ohhashi, F. Tadehara, T. Shokawa, Y. Dohi, E. Kunita, H. Utsunomiya, N. Kohno, et al. Characterization of noncalcified coronary plaques and identification of culprit lesions in patients with acute coronary syndrome by 64-slice computed tomography. J. Am. Coll. Cardiol. Img., February 1, 2009; 2(2): 153 - 160. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Carrigan, D. Nair, P. Schoenhagen, R. J. Curtin, Z. B. Popovic, S. Halliburton, S. Kuzmiak, R. D. White, S. D. Flamm, and M. Y. Desai Prognostic utility of 64-slice computed tomography in patients with suspected but no documented coronary artery disease Eur. Heart J., February 1, 2009; 30(3): 362 - 371. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Holz, R. Lautamaki, T. Sasano, J. Merrill, S. G. Nekolla, A. C. Lardo, and F. M. Bengel Expanding the Versatility of Cardiac PET/CT: Feasibility of Delayed Contrast Enhancement CT for Infarct Detection in a Porcine Model J. Nucl. Med., February 1, 2009; 50(2): 259 - 265. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J Einstein Radiation risk from coronary artery disease imaging: how do different diagnostic tests compare? Heart, December 1, 2008; 94(12): 1519 - 1521. [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, H. Scheffel, S. Leschka, A. Plass, S. Baumuller, B. Marincek, and H. Alkadhi Influence of Calcifications on Diagnostic Accuracy of Coronary CT Angiography Using Prospective ECG Triggering Am. J. Roentgenol., December 1, 2008; 191(6): 1684 - 1689. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Gibbons Noninvasive Diagnosis and Prognosis Assessment in Chronic Coronary Artery Disease: Stress Testing With and Without Imaging Perspective Circ Cardiovasc Imaging, November 1, 2008; 1(3): 257 - 269. [Full Text] [PDF] |
||||
![]() |
A. Chander, M. Brenner, R. Lautamaki, C. Voicu, J. Merrill, and F. M. Bengel Comparison of Measures of Left Ventricular Function from Electrocardiographically Gated 82Rb PET with Contrast-Enhanced CT Ventriculography: A Hybrid PET/CT Analysis J. Nucl. Med., October 1, 2008; 49(10): 1643 - 1650. [Abstract] [Full Text] [PDF] |
||||
![]() |
H Scheffel, H Alkadhi, S Leschka, A Plass, L Desbiolles, I Guber, T Krauss, J Gruenenfelder, M Genoni, T F Luescher, et al. Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance Heart, September 1, 2008; 94(9): 1132 - 1137. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. P. Shuman, K. R. Branch, J. M. May, L. M. Mitsumori, D. W. Lockhart, T. J. Dubinsky, B. H. Warren, and J. H. Caldwell Prospective versus Retrospective ECG Gating for 64-Detector CT of the Coronary Arteries: Comparison of Image Quality and Patient Radiation Dose Radiology, August 1, 2008; 248(2): 431 - 437. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Kelly, D. Thickman, S. D. Abramson, P. R. Chen, S. F. Smazal, M. J. Fleishman, and S. C. Lingam Coronary CT Angiography Findings in Patients Without Coronary Calcification Am. J. Roentgenol., July 1, 2008; 191(1): 50 - 55. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |