1 From the Cardiovascular Magnetic Resonance Unit, Royal Brompton Hospital, Sydney St, London SW3 6NP, England (N.H.B., A.S.J., R.H.M., D.J.P.); Siemens Medical Solutions, Erlangen, Germany (C.H.L.); and Minneapolis Heart Institute, Minneapolis, Minn (J.R.L.). Received November 29, 2001; revision requested February 13, 2002; final revision received August 30; accepted October 17. Supported by CORDA, the Heart Charity, and the Welcome Trust. Address correspondence to D.J.P. (e-mail: d.pennell@ic.ac.uk).
PURPOSE: To compare the accuracy of multisection true fast imaging with steady-state precession (FISP) with gadolinium-enhanced magnetic resonance (MR) angiography for the detection of coronary artery bypass graft patency.
MATERIALS AND METHODS: Twenty-five patients with coronary artery bypass grafts who had recently undergone conventional coronary angiography underwent MR angiography with a 1.5-T system. True FISP angiographic images were acquired in transverse and coronal planes. Coronal cardiac-gated MR angiography was performed with 0.2 mL per kilogram of body weight of gadopentetate dimeglumine injected at a rate of 2 mL/sec. With conventional angiography as the reference standard, the sensitivity, specificity, and accuracy of each technique for the detection of graft patency were determined. Image quality and duration of analysis were determined by two experienced radiologists.
RESULTS: In 25 patients, 46 of 56 venous grafts were patent and 22 of 23 arterial grafts were patent. In all grafts at true FISP angiography, sensitivity for patency was 84% (57 of 68 grafts), specificity was 45% (five of 11 grafts), and accuracy was 78% (62 of 79 grafts). At MR angiography, sensitivity was 85% (58 of 68 grafts), specificity was 73% (eight of 11 grafts), and accuracy was 84% (66 of 79 grafts) (difference not significant). Image quality scores were similar with both techniques, but duration of analysis was significantly longer with MR angiography than with true FISP angiography (29 minutes 24 seconds vs 14 minutes 6 seconds, P < .001).
CONCLUSION: Accuracy for detection of coronary artery bypass graft patency was similar with gadolinium-enhanced MR angiography and true FISP angiography, with a trend toward more false-positive findings for occlusion and reduced visualization of arterial grafts with true FISP angiography.
© RSNA, 2003
Index terms: Coronary angiography, comparative studies, 54.11, 54.121412, 54.12142 Coronary vessels, bypass graft, 54.457 Coronary vessels, MR, 54.121412, 54.12142 Coronary vessels, stenosis or obstruction, 54.11, 54.121412, 54.12142, 54.457 Magnetic resonance (MR), vascular studies, 54.121412, 54.12142
This article has been cited by other articles:
![]() |
D. A. Bluemke, S. Achenbach, M. Budoff, T. C. Gerber, B. Gersh, L. D. Hillis, W. G. Hundley, W. J. Manning, B. F. Printz, M. Stuber, et al. Noninvasive Coronary Artery Imaging: Magnetic Resonance Angiography and Multidetector Computed Tomography Angiography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention, and the Councils on Clinical Cardiology and Cardiovascular Disease in the Young Circulation, July 29, 2008; 118(5): 586 - 606. [Full Text] [PDF] |
||||
![]() |
M. Jemmali, D. Marini, G. Calcagni, F. Brunelle, D. Sidi, D. Bonnet, and P. Ou Pediatric Coronary Artery Bypass After Arterial Switch Operation: Noninvasive Evaluation With ECG-Gated 64-Slice CT in Routine Practice Ann. Thorac. Surg., October 1, 2007; 84(4): 1398 - 1399. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Malagutti, K. Nieman, W. B. Meijboom, C. A.G. van Mieghem, F. Pugliese, F. Cademartiri, N. R. Mollet, E. Boersma, P. P. de Jaegere, and P. J. de Feyter Use of 64-slice CT in symptomatic patients after coronary bypass surgery: evaluation of grafts and coronary arteries Eur. Heart J., August 1, 2007; 28(15): 1879 - 1885. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yamamoto, F. Kimura, H. Niinami, Y. Suda, E. Ueno, and Y. Takeuchi Noninvasive Assessment of Off-Pump Coronary Artery Bypass Surgery by 16-Channel Multidetector-Row Computed Tomography. Ann. Thorac. Surg., March 1, 2006; 81(3): 820 - 827. [Abstract] [Full Text] [PDF] |
||||
![]() |
N I Stauder, A M Scheule, U Hahn, M Fenchel, F S Eckstein, U Kramer, C D Claussen, and S Miller Perioperative monitoring of flow and patency in native and grafted internal mammary arteries using a combined MR protocol Br. J. Radiol., April 1, 2005; 78(928): 292 - 298. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. R. Moreno and V. Fuster The year in atherothrombosis J. Am. Coll. Cardiol., December 7, 2004; 44(11): 2099 - 2110. [Full Text] [PDF] |
||||
![]() |
T. Schlosser, T. Konorza, P. Hunold, H. Kuhl, A. Schmermund, and J.o. Barkhausen Noninvasive visualization of coronary artery bypass grafts using 16-detector row computed tomography J. Am. Coll. Cardiol., September 15, 2004; 44(6): 1224 - 1229. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Edelman Contrast-enhanced MR Imaging of the Heart: Overview of the Literature Radiology, September 1, 2004; 232(3): 653 - 668. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |