Neuroradiology |
1 From the Institutes of Clinical Radiology (B.B.E.W., R.T.H., K.H., M.F.R.) and Neuroradiology (R.B.), Klinikum Grosshadern, University of Munich, Marchioninistr 15, D-81377 Munich, Germany; and Center for Statistical Sciences, Brown University, Providence, RI (B.S., J.D.B.). Received April 7, 2003; revision requested June 24; final revision received September 17; accepted October 21. Address correspondence to B.B.E.W. (e-mail: b.ertl-wagner@t-online.de).
PURPOSE: To investigate image quality and vascular delineation of multidetector row computed tomographic (CT) angiography at various kilovoltage settings.
MATERIALS AND METHODS: Thirty patients were investigated with a standardized CT protocol, with three groups of 10 consecutive patients examined at 80, 120, and 140 kV, respectively. Three blinded readers independently evaluated images and graded image quality parameters, diagnostic confidence, and vascular delineation of intracranial arteries and veins. Vascular CT attenuation values, CT dose indices, and dose length products were assessed quantitatively. For data analysis, a Kruskal-Wallis nonparametric rank F test was used to identify trends and variables that required modeling attention. A proportional odds multinomial regression model was then fit with generalized estimating equations to account for the correlated nature of the data.
RESULTS: Image quality was rated higher with higher kilovoltage settings (P < .001). The severity of imaging artifacts was higher with lower kilovoltage settings (P < .001), while the subjectively rated vessel contrast was lower in the 80-kV group than in the 120-kV group and the 140-kV group (P < .05). Diagnostic confidence was higher in the 120-kV group and 140-kV group (P < .005). Vascular delineation was higher with higher kilovoltage settings for most arterial and venous structures. Differences were more significant for structures in close topographic proximity to bone and for subsegmental arteries and were less significant and, in parts, not significant for the main arterial branches and the large venous sinus. Attenuation values were higher with lower kilovoltage settings (P < .05). The mean dose length product could be reduced from 594 mGy · cm in the 140-kV group to 152 mGy · cm in the 80-kV group.
CONCLUSION: This multireader study of image quality and vessel delineation with cranial multidetector row CT angiography at various kilovoltage settings demonstrated a superiority of higher voltages with most pronounced effects for vessels adjacent to bone and subsegmental arteries.
© RSNA, 2004
Index terms: Brain, CT, 17.1211 Computed tomography (CT), image quality, 17.1211 Computed tomography (CT), multidetector row, 17.1211
This article has been cited by other articles:
![]() |
C.-Y. Yang, Y.-F. Chen, C.-W. Lee, A. Huang, Y. Shen, C. Wei, and H.-M. Liu Multiphase CT Angiography versus Single-Phase CT Angiography: Comparison of Image Quality and Radiation Dose AJNR Am. J. Neuroradiol., August 1, 2008; 29(7): 1288 - 1295. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Earls, E. L. Berman, B. A. Urban, C. A. Curry, J. L. Lane, R. S. Jennings, C. C. McCulloch, J. Hsieh, and J. H. Londt Prospectively Gated Transverse Coronary CT Angiography versus Retrospectively Gated Helical Technique: Improved Image Quality and Reduced Radiation Dose Radiology, March 1, 2008; 246(3): 742 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, M. Kiguchi, C. Fujioka, Y. Shen, R. Arie, K. Sunasaka, and K. Ito Radiation Dose, Image Quality, Stenosis Measurement, and CT Densitometry Using ECG-Triggered Coronary 64-MDCT Angiography: A Phantom Study Am. J. Roentgenol., February 1, 2008; 190(2): 315 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Linn, B. Ertl-Wagner, K.C. Seelos, M. Strupp, M. Reiser, H. Bruckmann, and R. Bruning Diagnostic Value of Multidetector-Row CT Angiography in the Evaluation of Thrombosis of the Cerebral Venous Sinuses AJNR Am. J. Neuroradiol., May 1, 2007; 28(5): 946 - 952. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, E. Oechslin, L. Husmann, L. Desbiolles, B. Marincek, M. Genoni, R. Pretre, R. Jenni, S. Wildermuth, and H. Alkadhi Pre- and Postoperative Evaluation of Congenital Heart Disease in Children and Adults with 64-Section CT RadioGraphics, May 1, 2007; 27(3): 829 - 846. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Waaijer, M. Prokop, B. K. Velthuis, C. J. G. Bakker, G. A. P. de Kort, and M. S. van Leeuwen Circle of Willis at CT Angiography: Dose Reduction and Image Quality--Reducing Tube Voltage and Increasing Tube Current Settings Radiology, March 1, 2007; 242(3): 832 - 839. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. V. Sahani, S. P. Kalva, P. F. Hahn, and S. Saini 16-MDCT Angiography in Living Kidney Donors at Various Tube Potentials: Impact on Image Quality and Radiation Dose Am. J. Roentgenol., January 1, 2007; 188(1): 115 - 120. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Nakayama, K. Awai, Y. Funama, D. Liu, T. Nakaura, Y. Tamura, and Y. Yamashita Lower tube voltage reduces contrast material and radiation doses on 16-MDCT aortography. Am. J. Roentgenol., November 1, 2006; 187(5): W490 - W497. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hausleiter, T. Meyer, M. Hadamitzky, E. Huber, M. Zankl, S. Martinoff, A. Kastrati, and A. Schomig Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice: Impact of Different Scanning Protocols on Effective Dose Estimates Circulation, March 14, 2006; 113(10): 1305 - 1310. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |