Experimental Studies |
1 From the Laboratory of Cardiac Energetics, National Heart, Lung and Blood Institute, National Institutes of Health, Department of Health and Human Services, Bldg 10, Rm B1D416, MSC 1061, 10 Center Dr, Bethesda, MD 20892-1061. Received April 11, 2003; revision requested June 30; final revision received January 6, 2004; accepted January 21. Supported by the Intramural Program of the National Heart, Lung and Blood Institute. Address correspondence to A.E.A. (e-mail: araia@nih.gov).
PURPOSE: To compare fluorescent microsphere measurements of myocardial blood flow (MBF) with qualitative, semiquantitative, and fully quantitative measurements of first-pass perfusion at magnetic resonance (MR) imaging.
MATERIALS AND METHODS: Coronary artery occlusion or intracoronary adenosine infusion was successfully performed in 16 beagles; both procedures were performed simultaneously in one animal. MBF was assessed at microsphere analysis. First-pass myocardial perfusion MR imaging was performed during a dual-bolus administration of gadopentetate dimeglumine (0.0025 mmol/kg followed by 0.10 mmol/kg). The absolute myocardial perfusion at MR imaging was calculated by using Fermi function deconvolution methods. Qualitative, semiquantitative, and absolute myocardial perfusion MR imaging measurements were compared with microsphere MBF measurements by using paired t tests, linear correlation, and Bland-Altman analysis.
RESULTS: Fully quantitative (ie, absolute) analysis of MBF at MR imaging correlated with microsphere MBF measurement (r = 0.95, P < .001) across the full range of blood flow rates encountered (from 0 to >5.0 mL/min/g). Similar close correlations were observed in endocardial and epicardial segments (representing approximately 0.85 g of the myocardium). With modest increases in MBF, qualitative measurements plateaued in the hyperemic zones. Semiquantitative measurements did not correlate with MBF as well (r = 0.690.89); they plateaued around 3.0 mL/min/g.
CONCLUSION: Dual-bolus MR imaging enabled accurate measurement of absolute epicardial and endocardial perfusion across a wide range of blood flow rates (0 to >5.0 mL/min/g). Use of qualitative MR imaging measures such as the contrast enhancement ratio led to substantially underestimated hyperemic blood flow measurements.
© RSNA, 2004
Index terms: Animals Experimental study Heart, MR, 511.121412, 511.12143, 511.12144, 511.12145 Heart, perfusion Magnetic resonance (MR), perfusion study, 511.12144
This article has been cited by other articles:
![]() |
R. R. Edelman Can MR Imaging Be Used to Determine the Efficacy of Gene Therapy in the Heart? Radiology, October 1, 2008; 249(1): 1 - 2. [Full Text] [PDF] |
||||
![]() |
N. Beohar, A. K. Erdogan, D. C. Lee, H. N. Sabbah, M. J. Kern, J. Teerlink, R. O. Bonow, and M. Gheorghiade Acute heart failure syndromes and coronary perfusion. J. Am. Coll. Cardiol., July 1, 2008; 52(1): 13 - 16. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. M. Gonzalez, S. Shiva, P. S. Vincent, L. A. Ringwood, L.-Y. Hsu, Y. Y. Hon, A. H. Aletras, R. O. Cannon III, M. T. Gladwin, and A. E. Arai Nitrite Anion Provides Potent Cytoprotective and Antiapoptotic Effects as Adjunctive Therapy to Reperfusion for Acute Myocardial Infarction Circulation, June 10, 2008; 117(23): 2986 - 2994. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Petersen, M. Jerosch-Herold, L. E. Hudsmith, M. D. Robson, J. M. Francis, H. A. Doll, J. B. Selvanayagam, S. Neubauer, and H. Watkins Evidence for Microvascular Dysfunction in Hypertrophic Cardiomyopathy: New Insights From Multiparametric Magnetic Resonance Imaging Circulation, May 8, 2007; 115(18): 2418 - 2425. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. A. Barmeyer, A. Stork, K. Muellerleile, C. Tiburtius, A. K. Schofer, T. A. Heitzer, T. Hofmann, G. Adam, T. Meinertz, and G. K. Lund Contrast-enhanced Cardiac MR Imaging in the Detection of Reduced Coronary Flow Velocity Reserve Radiology, May 1, 2007; 243(2): 377 - 385. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Watkins, K G Oldroyd, and S Frohwein Magnetic resonance myocardial perfusion imaging: a new era in the detection of reversible myocardial ischaemia Heart, January 1, 2007; 93(1): 7 - 10. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |