1 From the Departments of Radiology and Biomedical Engineering, Duke University Medical Center, Erwin Rd, Durham, NC 27710. From the 2003 RSNA scientific assembly. Received July 28, 2003; revision requested October 6; revision received January 24, 2004; accepted March 2. Address correspondence to J.A.B. (e-mail: jay.baker@duke.edu).
PURPOSE: To evaluate the variability of true-positive and false-positive cues by using a commercially available computer-aided detection (CAD) system for analysis of 50 malignancies in a screening population.
MATERIALS AND METHODS: Fifty breast cancers detected at screening were analyzed by using a commercially available CAD system. Mean patient age was 62.2 years. Each set of mammograms (craniocaudal and mediolateral oblique views) was digitized and analyzed by the CAD system 10 times. One radiologist compared CAD output with the location of the malignancy at mammography and determined whether each lesion was marked accurately in one mammographic view, both views, or neither. Sensitivity and reproducibility of the CAD system were determined for both case- and image-based analysis.
RESULTS: Overall sensitivity of the CAD system when at least one of the two mammographic views was marked correctly (case-base sensitivity) was 82.4%. Sensitivity when each mammographic view was considered separately (image-based sensitivity) was 61.1%. For case-based analysis, variability in true-positive CAD cues was demonstrated for 14 of 50 (28%) cases. For image-based analysis, inconsistency in CAD output was observed in 33 of 100 (33%) mammographic views that contained malignancies detected at screening. However, the CAD system consistently detected 4043 of the 50 breast cancers in each of the 10 CAD runs. Variability for false-positive marks was significantly greater than that for true-positive marks.
CONCLUSION: Inconsistency was demonstrated for CAD analysis of breast cancers detected at screening. However, the CAD system was reasonably consistent in the overall number of cancers identified from run to run. Greater variability of the CAD system was also demonstrated for false-positive marks, as compared with true-positive marks.
© RSNA, 2004
Index terms: Breast neoplasms, localization, 08.112, 08.115 Breast radiography, quality assurance, 08.112, 08.115 Computers, diagnostic aid
This article has been cited by other articles:
![]() |
S. J. Kim, W. K. Moon, N. Cho, J. H. Cha, S. M. Kim, and J.-G. Im Computer-aided Detection in Full-Field Digital Mammography: Sensitivity and Reproducibility in Serial Examinations Radiology, December 1, 2007; 246(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Yang, W. K. Moon, N. Cho, J. S. Park, J. H. Cha, S. M. Kim, S. J. Kim, and J.-G. Im Screening Mammography-detected Cancers: Sensitivity of a Computer-aided Detection System Applied to Full-Field Digital Mammograms Radiology, July 1, 2007; 244(1): 104 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Kim, W. K. Moon, N. Cho, J. H. Cha, S. M. Kim, and J.-G. Im Computer-aided Detection in Digital Mammography: Comparison of Craniocaudal, Mediolateral Oblique, and Mediolateral Views Radiology, December 1, 2006; 241(3): 695 - 701. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Brenner, M. J. Ulissey, and R. M. Wilt Computer-Aided Detection as Evidence in the Courtroom: Potential Implications of an Appellate Court's Ruling Am. J. Roentgenol., January 1, 2006; 186(1): 48 - 51. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |