1 From the Departments of Radiology (G.Y.E.) and Orthopaedics and Rehabilitation (K.J.A., H.J.L., M.J.R., T.D.B., C.L.S.), University of Iowa, Carver College of Medicine, 200 Hawkins Dr, Iowa City, IA 52242. From the 2003 RSNA scientific assembly. Received November 26, 2003; revision requested January 21, 2004; revision received February 12; accepted March 23. Supported by the National Institutes of Health Specialized Center for Research on Osteoarthritis (NIAMS) (award number: 5 P50 AR048939). Address correspondence to G.Y.E. (e-mail: george-el-khoury@uiowa.edu).
PURPOSE: To test the accuracy of double-contrast multidetector row computed tomographic (CT) arthrography for measurement of cartilage thickness in cadaveric ankles and to compare this technique with three-dimensional (3D) fat-suppressed spoiled gradient-echo in the steady state (FS-SPGR) magnetic resonance (MR) imaging.
MATERIALS AND METHODS: Five cadaveric ankles were used. In the ankle specimens, five to nine 1.5-mm-diameter holes were drilled across the joint traversing the tibial subchondral bone, tibial articular cartilage, talar dome cartilage, and talar subchondral bone. The ankle specimens were obtained and used according to institutional policies. Each ankle specimen was imaged first by using 3D FS-SPGR MR imaging with a 1.5-T magnet and then by using double-contrast arthrography followed by CT with a fourdetector row scanner (ie, double-contrast multidetector row CT arthrography). The section thickness used for CT scanning was 1.0 mm reconstructed in 0.5-mm intervals. The MR and CT images obtained in the five specimens were then downloaded to a workstation, where a measurement tool was used to measure the cartilage thickness at each hole. The physical measurement of cartilage thickness at each hole was used as the reference standard with which the MR imaging and CT measurements were compared. Linear regression and correlation analyses were performed by using a statistical computer program.
RESULTS: Double-contrast arthrography followed by multidetector row CT, as compared with 3D FS-SPGR MR imaging, enabled more accurate measurement of the physical cartilage thickness in the ankle (P < .001).
CONCLUSION: In this study of five cadaveric ankles, multidetector row CT arthrography was more accurate than 3D FS-SPGR MR imaging for measurement of articular cartilage thickness in the ankle.
© RSNA, 2004
Index terms: Ankle, arthrography, 461.1211, 461.122, 463.1211, 463.122 Ankle, MR, 461.121412, 461.121415, 463.121412, 463.121415 Arthritis, degenerative, 461.77, 463.77 Cartilage, MR, 461.121412, 461.121415, 463.121412, 463.121415
This article has been cited by other articles:
![]() |
K. J. Stevens, R. F. Busse, E. Han, A. C. S. Brau, P. J. Beatty, C. F. Beaulieu, and G. E. Gold Ankle: Isotropic MR Imaging with 3D-FSE-Cube--Initial Experience in Healthy Volunteers Radiology, December 1, 2008; 249(3): 1026 - 1033. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. B. Chou, M. T. Coughlin, S. Hansen Jr, A. Haskell, G. Lundeen, C. L. Saltzman, and R. A. Mann Osteoarthritis of the Ankle: The Role of Arthroplasty J. Am. Acad. Ortho. Surg., May 1, 2008; 16(5): 249 - 259. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Anderson, B. J. Ellis, C. L. Peters, and J. A. Weiss Cartilage Thickness: Factors Influencing Multidetector CT Measurements in a Phantom Study Radiology, January 1, 2008; 246(1): 133 - 141. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Wyler, V. Bousson, C. Bergot, M. Polivka, E. Leveque, E. Vicaut, and J.-D. Laredo Hyaline Cartilage Thickness in Radiographically Normal Cadaveric Hips: Comparison of Spiral CT Arthrographic and Macroscopic Measurements Radiology, February 1, 2007; 242(2): 441 - 449. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |