Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


Published online before print February 16, 2006, 10.1148/radiol.2383050262

(Radiology 2006;239:94.)

A more recent version of this article appeared on April 1, 2006
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow All Versions of this Article:
2383050262v1
239/1/94    most recent
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hines-Peralta, A. U.
Right arrow Articles by Goldberg, S. N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hines-Peralta, A. U.
Right arrow Articles by Goldberg, S. N.
© RSNA, 2006

Experimental Studies

Microwave Ablation: Results with a 2.45-GHz Applicator in ex Vivo Bovine and in Vivo Porcine Liver1

Andrew U. Hines-Peralta, MD, Nadeer Pirani, MD, Peter Clegg, PhD, Nigel Cronin, PhD, Thomas P. Ryan, PhD, Zhenjun Liu, MD and S. Nahum Goldberg, MD

1 From the Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, 1 Deaconess Rd, WCC 308B, Boston, MA 02215 (A.U.H., N.P., Z.L., S.N.G.); Department of Physics, University of Bath, Bath, England (P.C., N.C.); and Microsulis Americas, Waltham, Mass (P.C., N.C., T.P.R.). From the 2004 RSNA Annual Meeting. Received February 15, 2005; revision requested April 12; revision received June 2; accepted June 21; final version accepted September 1. Supported by Microsulis. Address correspondence to S.N.G. (e-mail: sgolber{at}caregroup.harvard.edu).

Purpose: To characterize the relationship between applied power and treatment duration in their effect on extent of coagulation produced with a 2.45-GHz microwave applicator in both an ex vivo and a perfused in vivo liver model.

Materials and Methods: All experimentation was approved by the Institute of Animal Care and Use Committee. Multiple tissue ablations were performed in ex vivo bovine liver (120 ablations) and in vivo porcine liver (45 ablations) with a 5.7-mm-diameter 2.45-GHz microwave applicator. The applied power was varied from 50 to 150 W (in 25-W increments ex vivo and 50-W increments in vivo), while treatment duration varied from 2 to 20 minutes (in eight time increments for ex vivo and five for in vivo liver). Three-dimensional contour maps of the resultant short- and long-axis coagulation diameters were constructed to identify the optimal parameters to achieve maximum coagulation in both ex vivo and in vivo models. Multivariate analysis was performed to characterize the relationship between applied power and treatment duration.

Results: Power and treatment duration were both associated with coagulation diameter in a sigmoidal fashion (ex vivo, R2 = 0.78; in vivo, R2 = 0.74). For ex vivo liver, the maximum short-axis coagulation diameter (7.6 cm ± 0.2 [standard deviation] by 12.3 cm ± 0.8) was achieved at greatest power (150 W) and duration (20 minutes). In vivo studies revealed a sigmoidal relationship between duration and coagulation size, with a relative plateau in coagulation size achieved within 8 minutes duration at all power levels. After 8 minutes of treatment at 150 W, the mean short-axis coagulation diameter for in vivo liver was 5.7 cm ± 0.2 by 6.5 cm ± 1.7, which was significantly larger than the corresponding result for ex vivo liver (P < .05).

Conclusion: Large zones of ablation can be achieved with the 2.45-GHz microwave applicator used by the authors. For higher-power ablations, larger zones of coagulation were achieved for in vivo liver than for ex vivo liver with short energy applications, a finding previously not seen with other ablation devices, to the authors' knowledge.

© RSNA, 2006